PTR-TOF-MS: A New Instrument For Real-Time Analysis Of Multi-Component Systems: Applications To Food Analysis Rob Linforth, Annie Blisset, Andy Taylor.

Slides:



Advertisements
Similar presentations
Chemistry of Life.
Advertisements

Chemistry Bellwork.
Chapter 20 Molecular Mass Spectrometry Mass spectrometry is capable of providing information about (1) the elemental composition of samples of matter.
Real-time Monitoring technology of Mass spectrometry
FC-MS from Teledyne Isco CombiFlash ® a Name You Can Rely On.
Improvements in Mass Spectrometry for Life Science Research – Does Agilent Have the Answer? Ashley Sage PhD.
Module 2: Analytical methods and separation techniques.
Chemistry Review Unit 3, Lesson 1 © 2007, CSCOPE Elements 90 occur naturally 25 essential to living organisms C,H,O,N make up 96% of a human body C HOPKINS.
Chapter 1 Section 1.
Unit F321 Module Electron Structure. Atomic Structure Protons, neutrons, electrons How to make ions Relative atomic mass.
Environmental Chemistry
Empirical and Molecular Formulas
Understanding mass spectroscopy. Mass spectroscopy is a very powerful analytical tool that can provide information on the molecular mass of a compound,
The Chemistry of Life pg Why is the study of cells so important? Effects of medicine Diverse Careers Benefits of Foods Ethical testing Improved.
Ionic Bonds During ionic bonding, 1 or more valence electrons are transferred from one atom to another. Atoms are usually neutral because they have the.
Molecular mass spectrometry Chapter 20 The study of “molecular ions” M + e -  M e -
Mass Spectrometry Mass spectrometry involves ionizing a compound, separating the ions that result on the basis of mass to charge ratio (m/z) and obtaining.
Chapter 2~The Chemistry of Life
The Chemistry of Life. Section 2.1 Atoms The basic unit of matter are atoms. 100 million atoms would make a row only about 1 cm long! Consists of subatomic.
Chemistry of Life A Brief … Overview. Matter Matter occupies space and has weight. It can exist as a solid, liquid, or gas. It may be possible to break.
Regents Biology Ch. 6The Chemical Basis of Life Atoms, Elements, & Molecules Why are we studying Chemistry?
 Chemistry is the science of matter, especially its properties, structure, composition, behavior, reactions, interactions and the changes it undergoes.
Chemistry is the study of the composition, structure, and properties of matter, the processes that matter undergoes, and the energy changes that accompany.
Chapter 2 The Chemistry of Life © 2005 Jones and Bartlett Publishers.
CHM 312 Fall 2008 Special Topics in MS Dr. Ralph Mead.
Chemistry 101 Answers.
CHAPTER 1 THE CHEMISTRY OF LIVING THINGS. VOCABULARY TERMS 1. ELEMENTS MADE UP OF ONE KIND OF ATOM MAKE UP THE PERIODIC TABLE 97% OF ORGANISM COMPOSITION.
The Chemistry of Living Cells. What are ATOMS? Atoms are the basic building blocks of matter that make up everyday objects. A desk, the air, even you.
Chapter 4 Atomic Structure. Atom Atom – smallest part of an element that retains the properties of that element. Atomic Theory – proposed by John Dalton.
Background Aerosols are studied for –Environment impact Direct climate effect Indirect climate effect –Biofuels –Human health impact Medicinal Cigarette.
Chapter 2. The Chemical Context of Life. Why are we studying chemistry? Biology has chemistry at its foundation.
Chemistry Review Unit 3, Lesson 1 © 2007, CSCOPE Elements 90 occur naturally 25 essential to living organisms C,H,O,N make up 96% of a human body C HOPKINS.
Chapter 2 Chemistry of Life Section 1: Nature of Matter Section 2: Water and Solutions Section 3: Chemistry of Cells Section 4: Energy and Chemical Reactions.
Chapter 8 Earth Chemistry Section – Atomic Number ► An atom of a specific element is different from other elements by the number of protons it.
Quantitative Chemistry Calculations in Chemistry.
3.3 The Atom. Periodic Table of Elements Atomic Number: Number of protons in an atom; usually appears directly above the chemical symbol. Number of protons.
Objective  To develop methods for analysis of compounds in organic aerosol particles Why is this important?  Environmental impact  Alternative fuels.
Section 2Solutions Water: A Common Solvent 〉 Why is water called the universal solvent? 〉 Water is called the universal solvent because many substances.
The Nature of Molecules Chapter 2. 2 Do Now Complete Vocab pre-quiz for Ch 2.
Basic Chemistry. Background Element:Element: a substance that cannot be broken down into simpler chemical substances. 96%96% of the human body is composed.
Part 1: Download a Specification AS and A level combined specification
Warm Up #5 Look at Carbon on the Periodic Table. Why is Carbon’s atomic mass number and not a whole number? I am a mystery element. I have 32 protons.
ACIDS & BASES An Overview. Acids A compound that produces hydrogen ions (H+) when dissolved in water. Properties include sour taste, reactivity with metals,
Unit F321 Module Electron Structure. Atomic Structure Protons, neutrons, electrons How to make ions Relative atomic mass.
Biochemistry, Chemistry of Life
Atoms, Elements, Compounds, and Ions
CHEMISTRY, MATTER, AND LIFE
Chapter 1 Objectives Define chemistry.
What is Matter?.
Glencoe: Chapter 11 Sections 11.1 & 11.2
Introduction to Chemistry
Mass Spectrometry Mass spectrometry involves ionizing a compound, separating the ions that result on the basis of mass to charge ratio (m/z) and obtaining.
Chemistry Part 1 : Download a Specification
Science ideas I need to know
The Basics of Chemistry Text pages 20 – 24.
Instrumental Chemistry
9/1/2017 the Chemistry of Life.
Section 1 Chemistry Is a Physical Science
Chemistry-Part 1 Inside the Atom
Properties of Matter (1.5)
Properties of Matter (1.5)
9/1/2017 THE CHEMISTRY OF LIFE.
Chemistry Brief Intro. Module 2.
Chapter 1 Objectives Define chemistry.
Properties of Matter (1.5)
What is Chemistry & Matter?
Dr. Ramón González-Méndez MRSC
Biochemistry, Chemistry of Life
Chapter 1 Section 1.
Mass spectrometry (MS) is an analytical technique that can be used to determine the mass, elemental composition or chemical structure of molecules. Mass.
Presentation transcript:

PTR-TOF-MS: A New Instrument For Real-Time Analysis Of Multi-Component Systems: Applications To Food Analysis Rob Linforth, Annie Blisset, Andy Taylor Dept. of Food Science University of Nottingham Caroline Lamont-Smith, Steve Mullock & Fraser Reich Kore Technology Ltd. Ely, Cambs

Talk: Outline Overview of the Proton Transfer Reactor method Why a PTR-TOF-MS? The Instrument Some Example Data Summary First a confession: This talk is heavy on the instrument description, its properties and possibilities of the instrument, because the speaker is one of the team that designed the instrument.

PTR: Proton Transfer Reactions Essentially, PTR is a subset of Chemical Ionisation. It was defined and refined by Werner Lindinger in the 1990’s, and resulted in the first commercially available instrument (Ionicon) Aim is to achieve ‘Soft’ ionisation so as not to fragment the molecule(s) of interest: less fragments = less mass spectral ‘noise’. In positive ion mode, most frequently used reaction is: H 3 O + + R H 2 O + RH + For this protonation reaction to occur, the proton affinity of R must be greater than that of water The ionisation probability is almost the same as the collision cross-section, so the aim is to induce sufficient collisions to ensure efficient analyte ionisation

PTR: Proton Transfer Reactions With H 3 O + CompoundPA (kJ/mol) Oxygen421 Nitrogen493 Carbon Dioxide541 Sulphur Dioxide672 Water691 Hydrogen Sulphide709 Benzene750 Methanol754 Toluene784 Naphthalene803 Ethyl Acetate836 Tri ethyl phosphate909 Compounds above water in the proton affinity table will not be ionised, whereas compounds below will The good news is that most VOCs have higher proton affinities than water and will receive a proton from H 3 O + Generally, larger molecules have larger proton affinities

Linearity of Response Provided that [RH+]<< [H 3 O+], the H 3 O+ signal does not change with analyte concentration, and the detected analyte intensity is linear with concentration: [RH+] = [H 3 O+] 0 (1-e -k(R)t )  [H 3 O+] 0 [R] kt 0-50ppm H 2 S calibration: Raw data plot (k is rate constant for the reaction, and t is time to travel through the drift tube) H 2 S proton affinity= 709kJ/mol

Soft Ionisation vs. Electron Impact Ionisation 70eV EI spectrum of Ethyl Butyrate Molecular Ion appears at mass 116 m/z Significant fragmentation to lower masses

Soft Ionisation vs. Electron Impact Ionisation PTR-TOF-MS Spectrum of 5ppm Ethyl Butyrate Molecular Ion at 117 m/z (molecular ion + proton) Very little fragmentation

Charge Transfer In The PTR Reactor Need to induce sufficient collisions: H 3 O + and analyte mixed in a reactor tube (drift cell) Voltage gradient across reactor Pressure ~ 1 mbar, approximately 2000 collisions down 10cm length reactor Aim is to have dilute analyte, so that analyte molecules do not collide with anything except H 3 O +. This keeps the ionisation scheme and calibration simple, unlike in Ion Mobility Spectrometry (complex inter-analyte charge transfer reactions)

The PTR Reactor and the E/n Ratio E is the voltage gradient down the reactor in V/cm n is the gas density in molecules / cc (1mbar in reactor, typically) E/n in Townsend; 1 Townsend = cm 2 / Vs Lindinger et al. (1998) showed that T is ideal for non- fragmentation of organic molecules If E/n too high, fragmentation of molecule occurs If E/n too low, clustering of water molecules occurs and can present a problem for certain species

The PTR Reactor and the E/n Ratio Tri-ethyl phosphate, protonated mass Townsend 195 Townsend – loss of the ethyl groups (28 mass units) 260 Townsend PO 4 + proton

Why Use A TOF-MS? Parallel detection means no analytical price to pay for monitoring many species, unlike a sequential analyser such as a quadrupole or magnetic sector TOF Cycle frequency typically kHz = high data rate Full mass spectra to several hundred masses with 4-5 orders dynamic range in one second Possibility of ‘real time’ data analysis with < 100ms resolution, multiple species and sensible counting statistics Ability to interrogate a data set ‘retrospectively’: intensity of any species as a function of time, mass spectra for a variety of ‘time slices’ Parallel detection and full mass spectra more suitable to software data reduction techniques, e.g. principle component analysis

Why Use A TOF-MS? This mass spectral plot, acquired with a PTR-Quad-MS (a sequential scanning device) took 40 seconds A full mass spectrum to several hundred m/z can be acquired in less than a second with a PTR-TOF-MS Mass spectrum from breath (U. Nijmegan)

TOF-MS: Accurate Mass Capability m/z NO m/z O m/z CH 3 OH m/z (H 2 O) 2 +

Downside Of A TOF-MS? Main drawback is analysis of a continuous analyte stream: When the TOF cycle has started, no further ions can be injected without resulting in multiple overlapping spectra ‘Duty cycle’ describes the % of the analyte stream that is sampled In an orthogonal TOF (analyte stream enters TOF source at 90°), the duty cycle can reach 10%: when the Kore TOF source pulses, it ‘empties out’ ~5-10 microseconds worth of ions. Usage is therefore 5µs in 50µs, I.e. 10% Currently we are observing up to 500kc/s of Hydronium at the detector, approximately half that reported by the PTR-Quad (single ion mode). Prospect of improving duty cycle using ‘Hadamard’ methods – high frequency random pulsing of source with mathematical deconvolution of spectra - still not really proven as a workable solution

TOF MS: Orthogonal Pulsing TOF Source Continuous Sample Detector Ions Extracted Light Ions Heavy Ions Ions of different masses within a single ‘cycle’ arrive at the detector at different times according to the relation: K.E. = mv 2 /2 Ions with m/z = 1000 has flight time ~20-50  s, therefore ‘cycle time’ =  s, so typical pulsing frequency = kHz VoltsSource Off t Source On 0 -2kV Extractor pulses to –400v TOF Source Variant on ‘Wiley – Maclaren’ source. Compensation for ion position within source

TOF MS: Orthogonal Pulsing TOF Source Overview schematic of the instrument: TOF source to detector

TOF MS: Orthogonal Pulsing TOF Source Overview schematic of the instrument GD Source 2mbar PTR Reactor 1mbar Transfer Optics mbar Mass spectrometer and detector mbar

Mass Spectral Analysis: Applications to Food Analysis University of Nottingham Department of Food Science Professor Andy Taylor Dr. Rob Linforth Annie Blisset “Breath-by-breath” Research Main technique in lab: APCI Wanted to add PTR-TOF-MS Instrument recently delivered; has both PTR and APCI functionality Data acquired at Nottingham by one of us (FR) with Rob Linforth and Annie Blisset

Breath-by-Breath Analysis Of Juicy Fruit Gum The principle compound released during chewing of Juicy Fruit gum is ethyl butyrate, molecular mass 116 Soft tube inserted lightly into the nostril of subject Breathing normally, the mass spectrometer begins acquiring data. Most of the exhaled breath passes out into open space, but a side-mounted capillary pipe samples the breath into the proton transfer reactor.

Breath-by-Breath Analysis: Release of Ethyl Butyrate Idealised appearance of breath markers, such as acetone, as exhalation occurs Release of flavours during mastication Signal intensity Time First data set: take ‘raw data set’ (all ion flight times recorded as a function of elapsed experiment time) Integrate the ethyl butyrate protonated ion (mass 117) every second

Breath-by-Breath Analysis At Different Data ‘Granularity’ Reconstruction of ethyl butyrate signal with 1 second integration time Reconstruction of Acetone signal with 1 second integration time Breathing rate ~5/min

Breath-by-Breath Analysis At Different Data ‘Granularity’ Reconstruction of ethyl butyrate signal with 0.25 second integration time- further structure emerging still Reconstruction of ethyl butyrate signal with 0.5 second integration time- further structure emerging

Breath-by-Breath Analysis At Different Data ‘Granularity’ Reconstruction of ethyl butyrate signal with second integration time Reconstruction of ethyl butyrate signal with second integration time: no further structure emerging

Breath-by-Breath Analysis: Different Person Reconstruction of Acetone signal with second integration time Note different breathing rate ~10/min for different person Overlay of ethyl butyrate signal with acetone

Full Mass Spectra From 125ms Time Slices – seconds – seconds Logscale data Acetone and Ethyl butyrate

Mass Spectrum From Headspace of Freshly Macerated Tomatoes 3 hexenol 3 hexenal

3-Hexenol vs. Hexanal, Both C 6 H 12 O Hexanal: protonated masses at 101, 83, 55 3-Hexenol: protonated masses at 101, 83, 59 and 55 In previous slide, masses 101 and 83 were identified as 3- Hexenol, but in truth there is the possibility that 101 and 83 can be due to Hexanal.

3-Hexenol vs. Hexanal, Both C 6 H 12 O Is it possible to find an E/n value that will give a mass 101 for one compound but not the other? Hexanal3 Hexenol No ‘threshold phenomenon’ observed Also, no E/n found for 3-Hexenol at which intensity of 101>83 Possibility of modulating PTR voltage and using software tools for compound differentiation? More work required, clearly!

Summary PTR-MS instruments based on quadrupole mass spectrometers are now widely used in studies of environmental and atmospheric chemistry as well as food and medical applications A TOF-MS increases the possibilities of real-time analysis down to <100ms “data granularity”, with full mass spectral acquisition in each time slice A TOF-MS permits any ion chromatogram to be re-constructed from the data set A TOF-MS can be operated at higher mass resolutions than a quadrupole mass spectrometer. Even at relatively low mass resolution the mass accuracy is better than 30 millimass units Greater mass range capability, with no discrimination against high masses up to several hundred Daltons