Artifacts and Errors.

Slides:



Advertisements
Similar presentations
Technical Aspects: the machine the image
Advertisements

IMAGE ARTIFACTS.
Artifact analysis.
X-ray films are generally mad of an emulsion-gelatin containing radiation sensitive silver halide. with a flexible, blue tented base X-ray films are sensitive.
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Quality Control in Diagnostic Radiology
Components of Radiographic Image Quality
Radiographic Quality Chapter 5.
Radiology Artifacts.
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Radiographic Grids II By Professor Stelmark.
Resident Physics Lectures
The Grid Kyle Thornton DMI 50B.
Analyzing the Image Density. Density Overall blackening of the image.
EXPOSURE AND TECHNIQUE ERRORS
Dr. Mustafa Zuhair Mahmoud Mr.Ali B Alhailiy بسم الله الرحمن الرحيم 1.
DARKROOM ,PROCESSING & Artifacts REVIEW
Week 10 a Artifacts.
Technical Artifacts and Errors
FILM ARTIFACTS.
RADIOGRAPHIC PROCESSING
Basic quality control tests in medical radiography
Dental X-ray Film Processing
Lecture 4 Beam restricting Devices, their Use and Maintenance
Quality Assurance/Quality Control Chapter 10. Quality Assurance Definition A system of activities whose purpose is to provide assurance that the overall.
Unit IV Analyzing the Image. Unit IV Analyzing the Image.
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
Conventional and Computed Tomography
Quality Control Rad T 110.
Control of Scatter Radiation
1 Components of Image Quality & Radiographic Artifacts Radiologic Technology A SPRING 2012.
Quality Control.
IMAGE QUALITY REVIEW RT
Dr Mohamed El Safwany, MD.
Components of Image Quality & Radiographic Artifacts Radiologic Technology A Spring 2010 Final Final.
Components of Image Quality & Radiographic Artifacts
1 Possible Radiographic Errors Radiology. 2 During Exposure Film is bent during set up and exposure Causing distortion Film is bent during set up and.
Computed Radiography (CR), & Digital Radiography (DR) Images
IMAGE QUALITY AND ARTIFACTS 1 2 WHY YOU SEE WHAT YOU SEE… The films or images have different levels of density – different shades of gray X-rays show.
1 ARTIFACTS: AN UNWANTED DENSITY ON THE FILM
RADIOGRAPHIC ARTIFACTS
Radiographic Quality Visibility and Sharpness
Quality Assurance in the Dental Office
Grids George David Associate Professor Department of Radiology
Copyright © 2012, 2006, 2000, 1996 by Saunders, an imprint of Elsevier Inc. Chapter 20 Exposure and Technique Errors.
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
RAD 354 Chapt. 16 Image QUALITY
TECHNICAL ARTIFACTS & ERRORS Lavin: pp CTVT: pp
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Lavin: Chapter 7 CTVT: pp
Quality Assurance in Dental Radiography
Optimizing The Image RVT: Chapter 7
Film Processing ChemicalsSafety The Enlarger What’s Wrong?
Image Evaluation Guidelines for Image Analysis Chapter 1.
AJADI ADETOLA DEPARTMENT OF VETERINARY MEDICINE AND SURGERY.
RADIATION PROTECTION 04/12/2016.
Chapters 21 & 22.
X-Ray Generators and Ancillary Radiographic Equipment
Factors Effecting the Production
Exposure Factors Chapter 4
RADIOGRAPHIC ERRORS AND ARTIFACTS
Quality Control Testing of Screen Speed
Processing Radiographs
What makes a Quality Image
Image Production and Evaluation
Moderator : DR.UMA MAHESWARI presentor : dr.a.p.jaganathan
Creating a Veterinary Radiographic Technique Chart
GRIDS PRESENTER-Dr.Vishwanath patil PG resident
Presentation transcript:

Artifacts and Errors

Artifact Any opacity that doesn’t correspond to an actual anatomic structure

Film storage and handling artifacts Pressure/crescents Scratches Fingerprints Static Material in cassette Fog

Pressure/crescents Finger crescent Appear black and focal crease mark on film Improper handling of film Prevention Take time Handle film by edges only

Scratches Scratch in the emulsion Can appear black or white White

Fingerprints Can be either black or white Prevention White more common Hands clean and dry Touch only film edges

Static Static electricity Prevention Remove film from box slowly Don’t drag film across surface as placing in cassette Moderate humidity Anti-static cassette cleaners

Material in cassette/dirty screen White and focal Visible light from screen can’t reach and expose film emulsion

Film fog Any unwanted film exposure turn part of film black Before or after exposure

Light leak Type of film fog Looks black, usually focal

Positioning artifacts Magnification Distortion Grid cutoff Patient rotation Cassette upside down

Magnification Patient closer to film Patient farther away from film

Distortion Patient is not parallel to film

Grid cutoff Purpose of a grid If not centered the more of the primary beam is absorbed

Patient rotation Form of distortion

Cassette upside down Obvious how this happens

Exposure artifacts Motion Double exposure Overexposure Underexposure

Motion Blurry image Prevention Sedation or better restraint Closing mouth to prevent panting Faster time

Double exposure Appears as blurred image 2 identical images superimposed Pushed exposure button twice Forgot to change cassette between exposures Prevention Push button only once Change cassettes

Other Types of Artifacts Wet haircoat

Other Types of Artifacts Folded film in cassette

Other Types of Artifacts Contrast media on table, patient, cassette etc.

Other Types of Artifacts Portion of Film Unexposed Causes Collimation Lead shield Central x-ray beam not perpendicular to cassette Cassette not directly under beam (buckey tray not pushed in)

Other Types of Artifacts White Dots

Quality Control

Quality Control Program Preventative maintenance Equipment calibration In-service education Evaluation of new products

Purpose Provides way to minimize dose of radiation Creates the production of quality radiographs Its use leads to a decreased number of repeated films

QC in Veterinary Radiology Want to minimize exposure of assistants, not as concerned about the patients Vet Tech major part of QC program Perform tests Interpret the data Keep records The data collected is objective, not based on opinion or personal preference for image contrast and quality

Equipment Folder for records Measuring tape Thermometer Nine pennies Carpenters level Screen-film contact mesh

SID-source image distance Measure from focal spot on tube housing to tabletop. Measure from table top to top of cassette in the bucky tray Add measurements together Sometimes referred to as FFD – film focal distance Source image distance If no mark on the tube housing, divide the tube housing end cap into fourths. Use the bottom fourth as the focal spot

Perpendicularity To ensure the x-ray beam is properly centered, be sure the tube stand, collimator and x-ray tube are perpendicular and properly aligned Use the level to confirm that the tube is level and parallel to the table Stand at the end of the table and look at the tube, collimator and tube stand. Verify they are perpendicular Stand alongside the table and verify the same info. Adjust as needed

X-ray Apparatus Tube/Table Locks X-ray field light

X-ray Apparatus Cont. To ensure the light field determined by the collimator dials is accurate

X-ray Apparatus Cont. Angulation indictor Place carpenters level on the tabletop (it should be level Place the carpenters level on the bottom of the collimator (also should be level) Rotate x-ray tube 15 degrees Using protractor measure degree of angulation *should also be 15 degrees Repeat at 30 and 45 degrees

Viewbox Uniformity Clean view box If super dirty the spots could mislead interpretation

Light Field/X-ray Field Alignment Center x-ray tube over table Set SID to 40 inches Put cassette in bucky tray Set collimator field indicators at a field that is approximately 6 x 8 inches Turn collimator light on 9 penny collimation test

Screen-Film Contact

Calibration Should be at least annually kVp mA Timer Filtration

Darkroom QC Needs to be clean Check for light leaks Check safe light for distance and bulb wattage and proper filter If doing manual processing temperature checks