Dual-Energy Computed Tomography

Slides:



Advertisements
Similar presentations
PRINCIPLES OF CT.
Advertisements

Image Reconstruction.
Dual-Energy Multidetector CT:
Improved Conspicuity of Abdominal Lesions with Single-Source Dual-Energy MDCT Hadassah Hebrew University Medical Center Jerusalem, Israel Ruth Eliahou.
Wednesday night and Thursday Lab Mega Group.  1967: The first Computer Tomography (CT) theory was developed  1972: The CT scan was invented by Godfrey.
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 6 Basic physical principles of.
Computed Tomography II
Seeram Chapter 13: Single Slice Spiral - Helical CT
SCANCOMEDICAL Computed Tomography SCANCO User Meeting 2005 Dr. Bruno Koller SCANCO Medical AG
BMME 560 & BME 590I Medical Imaging: X-ray, CT, and Nuclear Methods
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods X-ray Instrumentation Part 1.
Types of Radiation Interactions All or Nothing Many Small There is a finite probability per unit length that the radiation is absorbed. If not, there is.
Computed Tomography RAD309
tomos = slice, graphein = to write
Chapter 2 Stewart C. Bushong
5.4.1 X-Rays. (a) describe the nature of X-rays Stowmarket Physics X-rays - nature Forms of electromagnetic radiation Short wavelength High frequency.
COMPUTED TOMOGRAPHY I – RAD 365 CT - Scan
X-Ray Medical Imaging Physics – IB Objectives
Medical Image Analysis Medical Imaging Modalities: X-Ray Imaging Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan,
Computed Tomography Physics, Instrumentation, and Imaging
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
Resident Physics Lectures
PRINCIPLES OF CT Dr Mohamed El Safwany, MD. Intended learning outcome The student should learn at the end of this lecture principles of CT.
X - RAYS IN DIAGNOSTICS D. Krilov
CT Instrumentation and X-ray system
The production of X-Rays X-Rays are produced whenever charged particles are slowed down or stopped. X-rays are generated via interactions of the accelerated.
X-ray microanalysis in the electron microscope
Part No...., Module No....Lesson No
Quality Assurance.
Introducing cross- sectional imaging Dr.Ahmed S. Tawfeek FIBMS(Rad.)
Clinical and Pre-clinical applications spectral x-ray detectors Anthony Butler MBChB GradDipSc FRANZCR PhD.
Date of download: 5/29/2016 Copyright © 2016 SPIE. All rights reserved. (a) Photograph of a single detector module. X-rays are incident from the top of.
Introduction to Medical Imaging Week 3: Introduction to Medical Imaging Week 3: CT – Reconstruction and uses Guy Gilboa Course
Introduction to Medical Imaging Week 2: X-ray and CT
Date of download: 6/23/2016 Copyright © 2016 SPIE. All rights reserved. Pixel layouts for (a) a standard 3-T APS containing reset (R), row select (RS)
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
J Cho, G Ibbott, M Kerr, R Amos, and O Mawlawi
-42 Impact of signal non-repeatability on spectral CT images
CT Multi-Slice CT.
Computed Tomography Basics
Computed Tomography Data Acquisition
Learning Objectives By the end of this lesson you should…
CT PRINCIPLES AYMAN OSAMA.
Resident Physics Lectures (year 1)
Andrew D. Hardie, MD Carlo N. De Cecco, MD
Part No...., Module No....Lesson No
Optical Coherence Tomography
Stacy Kopso, M.Ed., RT(R)(M)
Strategies for Reducing Radiation Exposure From Multidetector Computed Tomography in the Acute Care Setting  Aaron Sodickson, MD, PhD  Canadian Association.
Characteristic Radiation in Tungsten Targets Shell # of electrons Binding energy L Char M N O P Eff X-ray Energy K
بسم الله الرحمن الرحیم ..
Applications of Dual-Energy Computed Tomography for the Evaluation of Head and Neck Squamous Cell Carcinoma  Reza Forghani, MD, PhD, Hillary R. Kelly,
Resident Physics Lectures (Year 1)
X Rays Medical Physics Notes.
Volume 56, Issue 2, Pages (August 2009)
Spectral CT Overview Different materials have different absorption profiles depending on the incident x-ray energy Spectral CT* uses this property to separate.
Basic principles Geometry and historical development
Resident Physics Lectures (year 1)
David J. Brenner, Maria A. Georgsson  Gastroenterology 
Role of dual energy spectral computed tomography in characterization of hepatocellular carcinoma: Initial experience from a tertiary liver care institute 
Computed Tomography (CT)
Resident Physics Lectures (Year 1)
Dual-Energy Computed Tomography
Thermal Photonics and Energy Applications
The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT.
Volume 27, Issue 3, Pages xvii-xviii (August 2017)
Generation of X-rays Q&A
Computed Tomography (C.T)
Computed Tomography (C.T)
Computed Tomography (C.T)
Presentation transcript:

Dual-Energy Computed Tomography Reza Forghani, MD, PhD, Bruno De Man, PhD, Rajiv Gupta, MD, PhD  Neuroimaging Clinics  Volume 27, Issue 3, Pages 371-384 (August 2017) DOI: 10.1016/j.nic.2017.03.002 Copyright © 2017 The Authors Terms and Conditions

Fig. 1 Typical X-ray spectrum for X-ray tube voltages of 80 kVp and 140 kVp, including inherent X-ray tube filtration (generated with the XSpect simulator). The peaks represent characteristic lines of a tungsten anode, and the continuous spectrum is the result of bremsstrahlung. This example illustrates the polychromatic nature of the X-ray beams used for DECT scanning. For DECT scanning, it is desirable to achieve as much separation as possible between the low-energy and high-energy spectra. In addition, the tube voltage used must not be too low or too high. If it is too low, it will be excessively absorbed by the body, yielding little information; if it is too high, it will not yield useful information because of poor tissue contrast. In routine DECT, 80-100 kVp (low-energy spectrum) and 140-150 kVp (high-energy spectrum) are commonly used settings for this purpose, although there may be variations depending on the scanner model, filters used, or the specific clinical application at hand. (Courtesy of Reza Forghani, MD, PhD, Montreal, Quebec, Canada and Bruno De Man, PhD, Niskayuna, NY.) Neuroimaging Clinics 2017 27, 371-384DOI: (10.1016/j.nic.2017.03.002) Copyright © 2017 The Authors Terms and Conditions

Fig. 2 Example of tissues with weak and strong spectral characteristics based on their elemental composition. Axial non-contrast-enhanced CT image of the neck acquired in dual-energy mode using a fast kVp switching scanner is shown (A). Region of interest analysis was performed comparing the spectral Hounsfield unit attenuation curves of muscle (green) to that of the thyroid gland (blue) (B). Most of the soft tissue in the human body, including muscle, is composed of low-Z materials such as oxygen (Z = 8), carbon (Z = 6), and hydrogen (Z = 1). As a result, there is little energy dependency of measured attenuation of muscle on an uninfused study (B; green curves). The thyroid gland, on the other hand, contains iodine (Z = 53) with strong energy dependence due to photoelectric effect. Note the marked energy-dependent increase in its attenuation at low energies approaching the K-edge of iodine (33.2 keV) (B; blue curves). Because most clinically used CT contrast agents are iodine based, iodine’s strong energy dependence can be exploited in a variety of clinical settings. Neuroimaging Clinics 2017 27, 371-384DOI: (10.1016/j.nic.2017.03.002) Copyright © 2017 The Authors Terms and Conditions

Fig. 3 Dual-source DECT (Siemens AG). Schematic illustration of dual-source detector combination scanners with the 2 imaging chains in a nearly orthogonal configuration, allowing the same slice to be scanned simultaneously at the 2 energies. Yellow is used to illustrate the low-energy spectrum, and blue, the high-energy spectrum. Typically, 80-100 kVp (low energy spectrum) and 140-150 kVp (high energy spectrum) are used depending on the model, but other combinations may be used for specific applications. Because there are 2 separate source-detector pairs, a filter can be placed to harden the high-energy spectrum. For some higher generation dual-source scanners, it is possible to use a higher kVp for the lower spectrum (90 or 100 instead of 80 kVp) in conjunction with a filter. Because of the limited space in the CT gantry, there is only sufficient space for a smaller second detector, which in turns places restrictions on the usable field of view of the dual-energy CT mode, as shown in the illustration. (Courtesy of Reza Forghani, MD, PhD, Montreal, Quebec, Canada and Bruno De Man, PhD, Niskayuna, NY.) Neuroimaging Clinics 2017 27, 371-384DOI: (10.1016/j.nic.2017.03.002) Copyright © 2017 The Authors Terms and Conditions

Fig. 4 Single-source DECT with rapid kVp switching: GSI (GE Healthcare). Schematic illustration of this type of a single source-detector combination system. Yellow is used to illustrate the low-energy spectrum, and blue, the high energy spectrum, typically 80/140 kVp. DECT projection data are acquired by very fast switching between low- and high-energy spectra combined with fast sampling capabilities of a proprietary, garnet-based scintillator detector with low afterglow for spectral separation at each successive axial or spiral view. (Courtesy of Reza Forghani, MD, PhD, Montreal, Quebec, Canada and Bruno De Man, PhD, Niskayuna, NY.) Neuroimaging Clinics 2017 27, 371-384DOI: (10.1016/j.nic.2017.03.002) Copyright © 2017 The Authors Terms and Conditions

Fig. 5 Layered or sandwich detector DECT (Philips Healthcare). Schematic illustration of this type of a single source-detector combination system in which spectral separation is achieved at the level of the detector. This system takes advantage of the polychromatic nature of the beam produced at the source and highly specialized detectors that consist of 2 layers having maximal sensitivity for different energies. The first layer (yellow) preferentially absorbs the low-energy photons, by design approximately 50% of the total incident photon flux. The second layer (blue) absorbs the remaining high-energy photons. (Courtesy of Reza Forghani, MD, PhD, Montreal, Quebec, Canada and Bruno De Man, PhD, Niskayuna, NY.) Neuroimaging Clinics 2017 27, 371-384DOI: (10.1016/j.nic.2017.03.002) Copyright © 2017 The Authors Terms and Conditions

Fig. 6 Single-source DECT with beam filtration at the source: TwinBeam DECT (Siemens AG). Schematic illustration of this type of a single source-detector combination system in which a split filter consisting of gold and tin is placed at the output of the tube, resulting in separation of the beam into low- and high-energy spectra. The corresponding halves of the detector are then used for detection of the low- and high-energy spectra. (Courtesy of Reza Forghani, MD, PhD, Montreal, Quebec, Canada and Bruno De Man, PhD, Niskayuna, NY.) Neuroimaging Clinics 2017 27, 371-384DOI: (10.1016/j.nic.2017.03.002) Copyright © 2017 The Authors Terms and Conditions

Fig. 7 Sequential scanning approaches to DECT scanning. This is one of the earliest and technologically most straightforward ways to obtain DECT scans. With this approach, the spectral data at 2 different energies are acquired sequentially at the same table position, or a range of table positions using different tube voltages. Although simpler to implement, this approach has significant limitations because of the delay between the acquisition of low- and high-energy data, as discussed in the text. One way to minimize the delay between the acquisition of low- and high-energy data is by alternating scanning of high and low kVp data for each gantry rotation, instead of scanning the entire volume with multiple rotations at one energy followed by the other. (Courtesy of Reza Forghani, MD, PhD, Montreal, Quebec, Canada and Bruno De Man, PhD, Niskayuna, NY.) Neuroimaging Clinics 2017 27, 371-384DOI: (10.1016/j.nic.2017.03.002) Copyright © 2017 The Authors Terms and Conditions

Fig. 8 Schematic illustration of a photon counting scanner, one of the most advanced spectral CT systems currently under development. These scanners use photon counting detectors to resolve the energy of individual photons or photon bins. Theoretically, these highly specialized and efficient detectors would count each individual incident X-ray photon and measure its energy. Narrow selectable subranges (or bins) of the spectrum can then be used to detect and classify materials based on their spectral response, enabling robust multienergy material characterization. (Courtesy of Reza Forghani, MD, PhD, Montreal, Quebec, Canada and Bruno De Man, PhD, Niskayuna, NY.) Neuroimaging Clinics 2017 27, 371-384DOI: (10.1016/j.nic.2017.03.002) Copyright © 2017 The Authors Terms and Conditions

Fig. 9 Examples of abdominal images obtained with a full field-of-view photon-counting CT scanner prototype developed by GE Healthcare and installed at Rabin Medical Center, Israel in 2008. Two different slices are shown in 2 different representations: (A, C) monochromatic images at 70 keV and (B, D) effective Z images. Different Z numbers are mapped to different colors. The images were obtained with a 32-slice helical scan, 1-second gantry rotation, 140 kVp and 140 mA. (Courtesy of Dr Ofer Benjaminov, Rabin Medical Center, Israel; with permission.) Neuroimaging Clinics 2017 27, 371-384DOI: (10.1016/j.nic.2017.03.002) Copyright © 2017 The Authors Terms and Conditions