The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Search for Scintillation in Doped Lead Fluoride Crystals for the HHCAL Detector Concept Rihua Mao, Liyuan.

Slides:



Advertisements
Similar presentations
Silvia Franchino Universita' Pavia IFAE09 Bari - 15/04/09 1 Nuovi cristalli per la calorimetria a doppio readout DREAM collaboration (INFN Italy & US)
Advertisements

Exploiting VERITAS Timing Information J. Holder a for the VERITAS Collaboration b a) School of Physics and Astronomy, University of Leeds, UK b) For full.
NLC – The Next Linear Collider Project Colorado Univ. - Boulder Calorimetry Cornell-ALCPG Calorimetry Detector Study Plans at Colorado Uriel Nauenberg.
Materials for Homogenous Hadron Calorimetry Marcel Demarteau Fermilab Internal Review August 2, 2010, Fermilab.
DREAM Collaboration: Recent Results on Dual Readout Calorimetry. F.Lacava for the DREAM Collaboration Cagliari – Cosenza – Iowa State – Pavia – Pisa –
Dense Scintillating Crystals and Glasses for HEP Dual Readout Calorimeter Tianchi Zhao University of Washington Dec. 4, 2007 Dual Readout Calorimeter Biweekly.
17-May-15FCAL collaboration workshop, Status of sensor R&D in Minsk K. Afanaciev, M. Baturitsky, I. Emeliantchik, A. Ignatenko, A. Litomin, V. Shevtsov.
Time-of-Flight at CDF Matthew Jones August 19, 2004.
1 Thermally induced 4f – 5d transitions in LuAlO 3 :Ce (LuAP) A.J. Wojtowicz, S. Janus Institute of Physics, N. Copernicus Univ. Toruń, POLAND IEEE 9th.
1 VUV spectroscopy of BaF 2 :Er A.J. Wojtowicz Institute of Physics, N. Copernicus Univ. Toruń, POLAND FPS 2007 French-Polish Symposium on Spectroscopy.
1 Scintillation light yield of Ce-doped LuAP and LuYAP pixel crystals A.J. Wojtowicz 1), W. Drozdowski 1), M. Ptaszyk 1) Z. Galazka 2), J.L. Lefaucheur.
October-November 2003China - ALICE meeting1 PHOS in ALICE A PHOton Spectrometer with unique capabilities for the detection/identification of photons and.
Characterization of CF 4 primary scintillation Andrey Morozov.
1 Tianchi Zhao University of Washington Concept of an Active Absorber Calorimeter A Summary of LCRD 2006 Proposal A Calorimeter Based on Scintillator and.
PPARC forum on developments in scintillator technology R M Brown - RAL 1 Scintillating Crystals for Particle Physics Outline  Performance considerations.
Heavy Scitillating Crystals and Glasses for a Combined EM and HCal at ILC Tianchi Zhao University of Washigton Sept. 25, 2007.
Heavy Scintillating Glasses for Future High Energy Particle Physics Experiments Chun Jiang School of Electronic Information and Electrical Engineering.
Report on SiPM Tests SiPM as a alternative photo detector to replace PMT. Qauntify basic characteristics Measure Energy, Timing resolution Develop simulation.
Cherenkov radiation and Cherenkov detectors in the Cherenkov laboratory A.S.Belousov Lebedev Physical Instirute Russian Academy of sciences Report on the.
Optical Characterization of GaN-based Nanowires : From Nanometric Scale to Light Emitting Devices A-L. Bavencove*, E. Pougeoise, J. Garcia, P. Gilet, F.
Adam Para, Fermilab, April 26, Total Absorption Dual Readout Calorimetry R&D Fermilab, Caltech, University of Iowa, Argonne National Laboratory,
Photophysical Properties of CdSe/ZnS Quantum Dots Embedded in Polymer Films and Solubilized in Toluene Final Presentation Jamie Golden CHEM /30/10.
Experimental set-up Abstract Modeling of processes in the MCP PMT Timing and Cross-Talk Properties of BURLE Multi-Channel MCP PMTs S.Korpar a,b, R.Dolenec.
PWO/APD/CSP Development at Hiroshima U. Hiroshima ALICE-PHOS Group (T.Sugitate, K.Shigaki, et al.) 17 August, 2004, at CERN.
Scintillators Hoda Bagheri Spring  Light emission  Scintillators  Scintillation mechanism  Photomultiplayer.
Heavy Scintillating Crystal Fibers for calorimetry
Setup for hypernuclear gamma-ray spectroscopy at J-PARC K.Shirotori Tohoku Univ. Japan for the Hyperball-J collaboration J-PARC E13 hypernuclear  -ray.
Recent status of dark matter search with ULE-HPGe detector Tsinghua University Qian Yue nd Korea-China Joint Seminar on Dark Matter Search.
Valery Dormenev Institute for Nuclear Problems, Minsk
Halide Scintillator Growth & Characterization for Rare Decay Search Presented by Gul Rooh Kyungpook National University Republic of Korea.
P. Lecoq CERN 1 March 2014 Fast timing workshop, Clermont Fd, March 12-14, 2014 Ultimate Time Resolution in Scintillator-based detectors for Calorimetry.
T. Sugitate / Hiroshima / PHX031 / Nov.01 The Photon Spectrometer for RHIC and beyond PbWO 4 Crystal Density 8.29 g/cm 3 Radiation length 0.89 cm Moliere.
SEM- Schematic Overview. Electron Detection Tungsten Filament Electron Source.
Luminescent detectors of ionising radiation. L. Grigorjeva, P. Kulis, D. Millers, S. Chernov, M. Springis, I. Tale IWORDI Sept. Amsterdamm Institute.
Quenching of Fluorescence and Broadband Emission in Yb 3+ :Y 2 O 3 and Yb 3+ :Lu 2 O 3 3rd Laser Ceramics Symposium : International Symposium on Transparent.
The Case for the Dual Readout Calorimetry for SiD Adam Para, December 4, 2007.
1 Quantum phosphors Observation of the photon cascade emission process for Pr 3+ - doped phosphors under vacuum ultraviolet (VUV) and X-ray excitation.
Timing Studies of Hamamatsu MPPCs and MEPhI SiPM Samples Bob Wagner, Gary Drake, Patrick DeLurgio Argonne National Laboratory Qingguo Xie Department of.
November 4, 2004Carl Bromberg, FNAL LAr Exp. Workshop Nov. 4-6, Liquid argon as an active medium Carl Bromberg Michigan State University & Fermilab.
Photon Detector with PbWO 4 Crystals and APD Readout APS “April” Meeting in Denver, CO on May 4, 2004 presented by Kenta Shigaki (Hiroshima University,
Characterization of the QUartz Photon Intensifying Detector (QUPID) Artin Teymourian UCLA Dark Matter Group Dept. of Physics and Astronomy.
1 Plannar Active Absorber Calorimeter Adam Para, Niki Saoulidou, Hans Wenzel, Shin-Shan Yu Fermialb Tianchi Zhao University of Washington ACFA Meeting.
Gamma Detector of Plastic Scintillator Oct. 2, 2009 IP-BSM Group 1.
SrCl 2 crystal for EC/  + search Presented by J.H. So (KNU)
International Workshop on Linear Collider, Jeju, Korea, Ren-yuan Zhu, Caltech Aug. 27, Comments on Linear Collider Calorimetry Case for Precision.
Simulation studies of total absorption calorimeter Development of heavy crystals for scintillation and cherenkov readout Dual readout in the 4 th concept.
Presented by Jianghua Kyungpook National University 5 th Seminar on Dark Matter Search and Double beta decay, Sept 22-23, 2011, Yangyang, Korea.
Scintillating Bolometers – Rejection of background due to standard two-neutrino double beta decay D.M. Chernyak 1,2, F.A. Danevich 2, A. Giuliani 1, M.
University of Milano-Bicocca - UNIMIB DEPARTMENT OF MATERIALS SCIENCE The Department of Materials Science of the University of Milano-Bicocca has been.
R&D on Scintillation Crystals and Special Glass at BGRI Mingrong ZHANG, Yuncheng GE (Beijing Glass Reseach Institute) 2nd HHCAL workshop, Beijing, May.
09/04/2006NDM061 CANDLES for the study of 48 Ca double beta decay OGAWA Izumi ( 小川 泉 ) Osaka Univ. ( 大阪大学 )
Gamma Spectrometry beyond Chateau Crystal J. Gerl, GSI SPIRAL 2 workshop October 5, 2005 Ideas and suggestions for a calorimeter with spectroscopy capability.
WIMPs Direct Search with Dual Light-emitting Crystals Xilei Sun IHEP International Symposium on Neutrino Physics and Beyond
Commissioning of the Crystal evaluation setup
Ultra fast SF57 based SAC M. Raggi Sapienza Università di Roma
Properties of Recent SIC Crystals
Denys Solodovnikov Marc H. Weber, Kelvin G. Lynn
G. Tamulaitis, A. Augulis, V. Gulbinas, S. Nargelas, E. Songaila, A
Where are we? What do we want to do next? Some thoughts
superB Detector Workshop I University of Perugia and INFN
G. Tamulaitis, S. Nargelas, A. Vaitkevicius, Vilnius University,
Irina Bavykina, MPI f. Physik
Sr-84 0n EC/b+ decay search with SrCl2 crystal
Studies on Heavy Scintillating Glasses
Yifang Wang Institute of High Energy Physics Elba, May 22, 2012
Alternative CEPC Calorimeter
Comments on Linear Collider Calorimetry Case for Precision ECAL?
LSO/LYSO Crystals for Future High Energy Physics Experiments
A CERN contribution to the dual readout calorimeter concept
Laboratory of High-Power & Semiconductor Lasers
Presentation transcript:

The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Search for Scintillation in Doped Lead Fluoride Crystals for the HHCAL Detector Concept Rihua Mao, Liyuan Zhang, Ren-yuan Zhu California Institute of Technology

2 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Introduction A Fermilab team (A. Para et al.) has proposed a total absorption homogeneous HCAL detector concept for the International Linear Collider to achieve good jet mass resolution, where both Cherenkov and scintillation lights are measured for good hadronic energy resolution. Requirements of Crystals:  High density.  Good UV transmittance.  Some scintillation light, not necessary bright and fast.  Cost-effective material (70 to 100 m 3 !).

3 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 HHCAL Conceptual Design A. Para, ILCWS08, Chicago: GEANT simulation shows jet energy resolution of about 22%/  E after corrections. This is much better than what has been achieved with PFA. R.-Y. Zhu, ILCWS-8, Chicago: a HHCAL cell with pointing geometry

4 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 ParametersBi 4 Ge 3 O 12 (BGO) Bi 4 Si 3 O 12 (BSO) PbF 2 (PbF) PbWO 4 (PWO) NaBi(WO 4 ) 2 (NBW) ρ (g/cm 3 ) ? λ I (cm) λ max τ decay (ns)300100?10-30 /10-200? λ max (nm)480470?420/ Cut-off λ (nm) / Light Output (%)10020?2<1 Melting point (°C) Raw Material Cost (%) Comparison of Crystals for HHCAL

5 The 2 nd HHCAL Workshop, Beijing, China5/9/ Candidates for HHCAL Following 2/19/08 workshop at SICCAS, 5 x 5 x 5 cm 3 samples evaluated

6 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Cosmic Setup with Dual Readout UG11 and GG 400 used to select Cherenkov & scintillation Agilent 6052A (500 MHZ) DSO with rise time 0.7 ns Hamamatsu R2059 PMT (2500 V) with rise time 1.3 ns

7 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Cherenkov Needs UV Transparency Cherenkov figure of merit Using UG11 optical filter Cherenkov light can be effectively selected with negligible contamination from scintillation

8 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Scintillation Selected with Filter GG400 optical filter effectively selects scintillation light with very small contamination from Cherenkov

9 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Ratio of Cherenkov/Scintillation 1.6% for BGO and 22% for PWO with UG11/GG400 filter and R2059 PMT BGOPWO

10 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 No Discrimination in Front Edge S Consistent timing and rise time for all Cherenkov and scintillation light pulses observed. SS C CC

The 2 nd HHCAL Workshop, Beijing, China 11 5/9/2010 Luminescence of PbF 2 at Low Temperature 1.Luminescence Kinetics of PbF 2 Single Crystals, M. Nikl, K. Polak, Phys. Status Solidi, A117, pK89, Luminescence of orthorhombic PbF 2, D. L. Alov, S. I. Rybchenko, J. Phys.: Condens. Matter, vol 7, p1475, Photoluminescence of orthorhombic and cubic PbF 2 single crystal, M. Itoh, H. Nakagawa, M. Kitaura, M. Fujita, D. Alov, J. Phys.: Condens. Matter, vol 11, p3003, Figs. of ref. 3 Othrorhombic Cubic

The 2 nd HHCAL Workshop, Beijing, China 12 5/9/2010 Luminescence / Scintillation of PbF 2 at Room T 1.Deformation and thermal treatment application to heavy scintillators production, S.N.Baliakin et. Al., proceedings of the Crystal 2000 Int. Conf. on Heavy scintillators for scientific and industrial applications, Chamonix, France, Sept , 1992, p A search for scintillation in doped and othrorhombic lead fluoride, D.F. Anderson, J.A. Kierstead, P. Lecoq, S. Stoll, C.L. Woody, NIM, A342, p473, Observation of fast scintillation light in a PbF 2 :Gd crystal, C.L. Woody, S.P. Stoll, J.A. Kierstead, IEEE TNS, vol.43, p1303, 1996 Figs. of ref. 3 Radioluminescence of PbF 2 (Gd)PbF 2 (Gd) Response to MIP of 1 GeV/c

13 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 PbF 2 Samples A total of 116 samples with various rare earth doping were grown by vertical Bridgman method at SIC and Scintibow. SIC samples are of 1.5 X 0 (14 mm) cube, while most of the Scintibow samples are of Φ 22 x 15 mm. Photo: SIC sample/Scintibow sample.

14 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Photo- and X-luminescence Photo luminescence was measured by using Hitachi F fluorescence spectrophotometer. An AMTPEK portable X-ray tube (ECLIPSE-III) was used for the X-luminescence measurement. Monochromator and PMT UV lamp and monochromator X-ray tube Hitachi F-4500

15 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Luminescence: Er & Eu Doped PbF 2 Consistent Photo- and X-luminescence observed

16 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Luminescence: Gd & Ho doped PbF 2 Consistent Photo- and X-luminescence observed

17 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Luminescence: Pr & Sm Doped PbF 2 Consistent Photo- and X-luminescence observed

18 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Luminescence: Tb doped PbF 2 While total 7 rare earth dopings show consistent Photo- and X- luminescence, no broadband emission of d-f transition was observed.

19 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Decay Time Measurement

The 2 nd HHCAL Workshop, Beijing, China 20 5/9/2010 Verified with BGO & CsI(Tl) 1280 ns290 ns Decay time consists with well known values

The 2 nd HHCAL Workshop, Beijing, China 21 5/9/2010 Decay Time: Er & Eu Doped PbF ms 8.5 ms

22 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Decay Time: Ho and Sm Doped PbF ms 1.3ms

23 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Decay Time: Tb Doped PbF ms Under UV plused laser excitation, the luminescence decay time of Gd and Pr doped samples is too weak to be measured.

24 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Comparison with the Ref-2 DopantCaltechRef-2 Smf-f Euf-f Tbf-f Tmno Ceno Ndno Hof-fno Erf-fno Dynof-f Ybnof-f? Prf-fN/A Gdf-f ?N/A No fast luminescence of d-f transition was observed.

25 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Anode Current Measurement Distance between source and sample: 2 cm

26 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Anode Current: PWO & Un-doped PbF 2 PWO: L.O. = 20 p.e./MeV, anode current = 240 nA

27 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Anode Current: All Samples

28 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Summary of Anode Current IDAnode current (nA) Size (mm)Doping Scintibow x12 x10Eu Scintibow-18 52Φ22X15Eu/Gd Scintibow-27 53Φ20X15Eu/Tb Scintibow-B19 56Φ20X15Eu/Tb/Na Scintibow-B21 83Φ22X15Eu/Bi/Na Scintibow-B23 73Φ20X15Eu/Bi/Na Undoped 4214 x 14 x14--

29 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 ɣ -ray Excited Pulse Height Spectrum

30 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 ɣ -ray Excited PHS: PWO

31 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 ɣ -ray Excited PHS: Doped PbF 2 No evident photopeak was observed with up to 4000 ns gate. PbF 2 : EuPbF 2 : Eu/Bi/Na

32 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Summary Lead fluoride crystal samples doped with various rare earth dopant were grown by Bridgman method. Consistent photo and x-ray luminescence found in samples with Er, Eu, Gd, Ho, Pr, Sm and Tb doping. The decay time of doped samples was found to be very long at ms scale as expected from the f-f transition of the rare earth elements. While some doped samples show anode current larger than the un-doped samples, their ɣ -ray excited pulse height spectra were found identical to un-doped sample, indicating no scintillation light.

33 The 2 nd HHCAL Workshop, Beijing, China5/9/2010 Future Plan Investigation will continue to search for scintillation in doped lead fluoride. Rescan the dopants showing consistent photo and x- ray luminescence with samples of improved optical quality, i.e. grown in platinum crucibles. Rescan these samples at low temperature to understand the quenching mechanism at the RT. A spectrophotometer capable of measuring luminescence in a cryostat system is being procured. Looking into the other phase: othrorhombic. Can we get large size and good quality α-PbF 2 crystal? Evaluate other potential candidate materials for the HHCAL detector concept, such as BSO, glasses and ceramics etc.