1 „Cryogen-free“ Dilution Refrigerators
50 Years of Dilution Refrigeration, Manchester, 16 September 2015 „Cryogen-free“ Dilution Refrigerators Kurt Uhlig Walther Meissner Institute Garching, Germany

2 1990: Patrick Pari, Adv Cryog Eng 35: 1079 first cryogen-free DR
precooled by GM cryocooler 1999: delivery of PT4-05 (#2 in CRYOMECH production line) 2002: Cryogenics 42: 73 „ 3He/4He DR with pulse tube precooling “ 2007: Prouvé et al., JLTP 148: 909 „Pulse-Tube Dilution Refrigeration Below 10 mK“

3 „Cryogen-free“ dilution refrigerators
Cryomech PT405 T(mc2) = 4.3 mK trap PT1 heat leak: Q(tot) = 70 nW GM-precool: Q(GM) = 170 nW PT2 shield cfhx flow restriction IVC Cryogenics 44 (2004) 53 hx‘s double mc 3He melting curve therm.

4 „Cryogen-free“ dilution refrigerators
VeriCold Inc.: no inner vacuum can (IVC) big mixing chamber plate separate pulse tubes from rotary valve all electrical 300K precool circuit (or gas heat switches, BlueFors) High refrigeration powers ( mK) (1.75 mK) Superconducting magnets (up to 18 T) Clear shots Top loading Bottom loading (cold sample exchange) Automatic cooldown ~ 75% of newly sold DRs are CF

5 CF-DR with 1K-stage 4He-1K-stage: PT Q = 100 mW @ 1.3 K
(80 m3/h pump) T(min) = 1.05 K CF-DR with 1K-stage (exp. on sc quantum circuits) PT 405 Cryomech PT405 RM: 1st stage: 32 K 40 75 K 2nd stage: 2.5 K 0.5 K DR-stage: mK T(min) = 4.0 mK (CMN + fixed point device) CMN sc FPD

6 Vibrational heat leak:
run fridge at base temperature fill 1K-stage with 4He(liq) operate 1K-pot in single cycle mode turn off pulse tube cryocooler see if mixing chamber temperature drops (and make sure, still temperature and inlet pressure remain unchanged) Cryogenics !

7 J. Hoehne, Cryocoolers 18 (2014) 319
piston cylinder oil 90 mm gear pump ν = 1 Hz el. motor pV - efficiency: η ~ 70% J. Hoehne, Cryocoolers 18 (2014) 319 J. Hoehne: arXiv: (CEC/ICMC 2015)

8 Thanks for your attention!

9

10

11 AUTOMATION! Characteristics of dry fridge:
„Cryogen-free“ dilution refrigerators Characteristics of dry fridge: Wide geometry, no He, no 1K-pot, automatic cooldown and warmup AUTOMATION! 2004: helium: 5 €/l power 9c / kWh PTR: 25,000 € compare with He-cryostat 20 l (liq) / day ) 2015: helium: 7.50 €/l power €/kWh PTR: 33,500 € amortization: 280 days 5kW-PTR (320 days 10 kW-PTR) amortization: 266 days 5kW-PTR (330 days 10kW-PTR)

12 „Cryogen-free“ dilution refrigerators
J. Hoehne: arXiv: (CEC/ICMC 2015)