STERILIZATION HEALTHCARE Steam Sterilizers Operation and Preventative Maintenance.

Similar presentations


Presentation on theme: "STERILIZATION HEALTHCARE Steam Sterilizers Operation and Preventative Maintenance."— Presentation transcript:

1 STERILIZATION HEALTHCARE Steam Sterilizers Operation and Preventative Maintenance

2 Sterilization: A process by which all forms of microbial life including bacteria, viruses, spores, and fungi are destroyed  Note: In a sterilization process, the nature of microbial death is described by an exponential function. Therefore, the presence of microorganisms on any individual item can be expressed in terms of probability. While this probability can be reduced to a very low number, it can never be reduced to zero. (ANSI/AAMI ST46) Sterilization Fundamentals

3 Sterility Testing: Product is inoculated directly into growth medium. Process Monitoring: Use of mechanical, biological and chemical indicators to monitor critical sterilization parameters. Parametric release: Declaring product sterile based on records demonstrating that the process parameters were delivered within specified tolerances. This method does not include the use of Biological Indicators. General Terminology Review

4 Bio Burden –# of microorganisms on a contaminated item Reduction of Bio Burden: –Hand washing; soap and water –Chemical; high level disinfection –Sanitization –Decontamination –Disinfection

5 What is a Steam Sterilizer ?

6 Steam sterilizers for healthcare have one primary purpose: Technology Fundamentals Render instruments sterile for re-use in a surgical setting

7 Steam Sterilizers –Temperatures ranging from 121-134  C at pressures of 15-30 psi are generally recommended to sterilize wrapped or unwrapped surgical instruments. –Why Steam? Sterilization Fundamentals Highly Effective Non Toxic Readily Available Easy to Control Inexpensive

8 Two Types of Steam Sterilizers: Vacuum: used for large volumes of wrapped surgical instruments Gravity: used for individual unwrapped surgical instruments Sterilization Fundamentals

9 Alternate Sterilization Technologies: ETO (Ethylene oxide) Gas Plasma Ozone Dry Heat Sterilizers Low temp. Ultra High temp.

10 Healthcare Steam Sterilizers –Typical Dimensions: 16“ wide x 16" high x 16" long (gravity or vacuum) 20“ wide x 20" high x 38" long (gravity or vacuum) 24“ wide x 36" high x 36" long (usually vacuum) 24“ wide x 36" high x 48" long (usually vacuum) 24“ wide x 36" high x 60" long (usually vacuum) Floor loader: 26.5 x 62 x up to 76” Sterilization Fundamentals small medium large

11 Steam Sterilizers –Smaller sterilizers: instruments inserted individually for quick sterilization –Larger sterilizers (36" or larger): can hold entire “carriage” of instruments for sterilizing –Floor loader: rises to floor level where entire cart and carriage of instruments are inserted for sterilization Sterilization Fundamentals

12 Factors Affecting Sterilization Time –Sterility Assurance Level (SAL) 10 -6 –One survivor out of a million (infectious organisms) Temperature –250F/270F/275F Moisture –Required to coagulate protein –However, steam quality requires a maximum vapor content of 3% Sterilization Fundamentals

13 Steam Quality Through RO Water Reverse-Osmosis (RO): the process of squeezing water through a semi-permeable membrane. RO will separate pure water molecules from dissolved solids such as Calcium, Sodium and Iron Chlorine, Nitrates and Nitrites are removed in the pre-filtration process. Sterilization Fundamentals

14 Benefits of RO in Steam Generation ( Clean Steam) Balances pH Reduces downtime Reduces costly repairs Extends useful equipment life Removes hardness Improves equipment reliability Sterilization Fundamentals

15 Proper sterilizer design and use to achieve time and temperature Sterilizer area design and quality utilities feeding unit Bio Burden reduced prior to sterilization Adequate contact of the sterilant for all surfaces Other Factors Affecting Sterilization Sterilization Fundamentals

16  Air and steam do not mix  Air inhibits steam contact  Air is 1.6 times denser than steam  Steam condenses on cool goods  Loading and load size  Packs must be dry Why is air removal important? Sterilization Fundamentals

17 Decimal Reduction Value (D- Value) The exposure to sterilant required to reduce the population of test organism by 90% (one logarithm) under stated conditions. Sterility Assurance Level (SAL) Probability of a non-sterile unit. As a minimum sterilization cycles are designed to provide a 1 in a million chance of a non-sterile unit. SAL = 10 -6 SAL Sterilization Fundamentals

18 More about D-Value and SAL... The SAL is extrapolated from the survivor curve based on the D-value of the Biological Indicator and exposure time. Traditional cycle development is driven by determining the time for complete kill of a 10 6 BI and double the cycle time to achieve SAL of 10 -6 Sterilization Fundamentals

19 3 Primary Factors of Sterility Assurance 1.Process Control - Reproducible and controlled conditions 2.Process Monitoring - Biological indicators, chemical indicators, mechanical gauges, cycle charts or printouts, etc. 3.Good Manufacturing Practices (GMP) - Includes record keeping, inspection of materials, process validation and equipment maintenance and calibration Sterilization Fundamentals Complete sterility assurance can only be achieved by testing each item processed - a destructive and impractical procedure.

20 Sterilization Cycle Talk: Getinge Vacuum Cycle Sterilization Fundamentals The actual sterilizer cycle as logged in our digital software solution, T-Doc

21 Sterilizer Maintenance Why should we have preventative maintenance on a sterilizer? What happens in the sterilizer during a cycle? What should I look for the start of my shift? What can effect the sterilization process that I can control? What can effect the sterilization process that I can not control directly?

22 Sterilizer Maintenance Why should we have preventative maintenance on a sterilizer? 1.Keeps the machine running to manufactures recommended specifications 2.Minimizes the amount of downtime 3.Provides assurance that the unit is running optimally 4.Extends the life of the sterilizer 5.Ensures the unit is safe for operating- no chamber cracks, faulty valves etc 6.Ensures the appropriate sterility is met by the system as it was intended to provide at time of manufacturing

23 Sterilizer Maintenance Sterilizer Video - What Happens during a sterilization cycle

24 Sterilizer Maintenance What should I look for when start of my shift? 1. Ask the person who you are replacing if there have been any autoclave problems during their shift. Keep a log book of events for of date, time, unit number, copy of print out and what occurred. 2.Check gasket for wear and tear on the machines you are using 3.Check chamber and drains to ensure no paper or items are obstructing the drain 4.Check the printer paper to ensure you have enough 5.Verify you have appropriate materials for logging, biological tests and bowie dick tests if required for your shift 6.Check to see if you’ll be required to process stat loads etc during your work period to allow you to manage the flow

25 Sterilizer Maintenance What can effect the sterilization process that I can control? 1.The proper loading and spacing of the sterilizer cart/trolley 2.Using the proper packaging material 3.Checking to ensure drain is not plugged and is cleaned at regular intervals 4.Cleaning of the chamber and carts at regular intervals 5.Proper cool down area away from air conditioning 6.Items being sterilized are dry, containers, blue silicone liners dry as well as plastic bowls etc 7.Being able to properly read the tests – BD, Biological, Indicators, Integrators etc 8.Prequalification of the units after a major repair of the sterilizer

26 Sterilizer Maintenance What can effect the sterilization process that I can not control directly? 1.The sterilizer malfunctioning 2.Steam supply staturation drops below 97% 3.Water pressure drops 4.Electrical power failure 5.Failure of BD Test – redo test and check results – 2 nd failure contact service for repair 6.Consistently wet steam or dirty steam – (tea effect) – You can perform a warm-up prior to the bowie dick test and this will help remove as much moister from the lines as possible. 7.Dirty steam – you can have filters placed at the feed of the units to filter out debris and some of the chemicals used by maintenance to clean the steam lines

27 Who establishes the proper sterilization parameters for reusable medical devices? Who establishes the proper sterilization parameters for reusable medical devices? Answer: The responsibility for safe and effective sterilization of reusable medical devices is the shared responsibility of both manufacturers, and users of the devices. Reichert & Young, p. 132 Sterilization Fundamentals

28 Summary - Key Learnings Reviewed General Terminology Reviewed the Sterilization Fundamentals Reviewed Sterilizer Maintenance and areas that can be controlled by MDRD employee

29 Sterilizer Maintenance Questions? Thank you for your Time Documents%20and%20Settings/Rdemedeiros/My%20 Documents/cyberclean.html Documents%20and%20Settings/Rdemedeiros/My%20 Documents/cyberclean.html