1 © 2002-2003, James H Philip, all rights reserved
New Anesthesia Delivery Systems at BWH Ohmeda Aestiva Draeger Fabius GS Draft C © , James H Philip, all rights reserved

2 Ready

3 James H Philip MD CCE Draft C
New Anesthesia Delivery Systems at BWH Ohmeda Aestiva Draeger Fabius GS James H Philip MD CCE Draft C

4 Short Clinical Conference first Surprise !!

5 Case 1 Ohmeda Aestiva 65 yo M for CABG on a Tuesday morning.
You induce anesthesia, intubate the trachea and confirm correct tube placement. You turn on the ventilator on your Ohmeda Aestiva Anesthesia Machine Ventilator alarms “Cannot Deliver Correct Volume” Soon, it alarms “Cannot Ventilate” If you know what is wrong and how to correct it Raise your hand now If you do not know what is wrong and how to correct it Raise your hand now

6 Case 2 Draeger Fabius GS 45 yo M for Thoracotomy on a Saturday
You intubate trachea and confirm tube placement. You turn on the ventilator on your Draeger Fabius GS Anesthesia Machine Vent Monitor shows alarm “Vent Fail” There is no mechanical ventilation If you know what is wrong and how to correct it Raise your hand now If you do not know what is wrong and how to correct it Raise your hand now Draeger Fabius GS

7 Ventilators do fail 1) Use reservoir bag and manually ventilate
2) If all else fails, use the Manual Resuscitator (Ambu Bag) on the back of the Anes Mach. 3) Determine ventilator problem and eliminate it Get help Repair your self

8 Ohmeda SmartVent Design
SmartVent is used in these Ohmeda ADSs SmartVent 7900 Aestiva Compare expired volume to set volume Adjust the next breath to correct this by adjusting movement of the bellows Flow sensors are the key element They are required for ventilation Measurement is more difficult than Control

9 Key Element = Flow Sensors

10 Differential Pressure Flow Sensors
DP Differential Pressure Flow Sensors DP

11 Differential Pressure Flow Sensor
Pressure Transducer Variable Orifice Flow

12 When Aestiva or any Ohmeda SmartVent fails
Usually a wet flow sensor User Action Ventilate by hand Change or shake dry the flow sensors Call for help if required

13 Fabius GS Introduction
We will soon have > 15 of these machines We all must learn about them We each must document that we have learned Dept could not buy “old machines” They are no longer for sale Dept did not buy more Ohmeda Aestivas Early failures led some clinicians to be wary

14 Gas flow interface Mechanical Needle Valves and Electronic Flow Measurement and Display
Pipeline pressures Total flow Mechanical rotameter Digital flow display O2 supply low LED Flow control valves Wall pressure gauge

15 High-resolution display at BWH
Pipeline pressures Total flow Mechanical rotameter Finer needle Valves Are coming Digital flow display O2 supply low LED Flow control valves Wall pressure gauge

16 User Control Many Keys, One Knob
Alarm Lamps Warn Caution Advisory Alarm limits Ventilation mode Alarm Silence Volume Setup menu Pressure Future 1 Future 2 AC power LED Standby Reading lamp Home screen Function Rotary Control Knob And, Press to Confirm

17 Ventilation Monitor Screen
Alarm Thresholds Time Alarm messages Status Inspired Oxygen Fresh Gas Virtual flowmeters Ventilation Volume Ventilation Pressure Pressure threshold Pressure waveform Function key descriptions

18 COSY COmpact SYstem for CO2 Absorption

19 COSY Breathing System – Top
Expiratory valve Expiratory port APL valve Manual breathing bag connection PEEP / Pmax valve Inspiratory port APL bypass valve Inspiratory valve

20 COSY Breathing Circuit - Bottom
Flow sensor Scavenging connection Fresh gas hose connection Pressure sensor connection Ventilator hose connection Absorber canister

21 Ventilator (eVent) Electrically driven = consumes no drive gas
Ventilation Modes Volume controlled Pressure controlled + future options Easy access to ventilator by opening ventilator door but, can’t see much Ventilator movement visible through window Ventilation Monitor displays all information

22 Ventilator Piston

23 Flow sensor

24 Hot wire is cooled by flow
Anemometer

25 Cause of Ventilation Failure
Piston and its membrane are mis-aligned because Membrane has been drawn off piston surface This occurs from negative pressure in circuit Immediate Solution Cycle ventilator off and on times Vacuum draws membrane to piston. Problem is solved. Ventilator ventilates. Long term Solution Draeger believes that this no longer occurs. Piston software prevents this.

26 Let’s look at Fabius GS Breathing Circuit

27 Constant Volume FGF Decoupling COSY-1

28 Puff Leak from high flow through CO2 Absorber caused re-design

29 Constant Volume FGF Decoupling COSY-2
Prev Bag location Moved

30 Constant Volume FGF Decoupling
Fresh Gas is Stored until the next breath

31 Virtual Anesthesia Machine
Operation Simulation Virtual Anesthesia Machine Virtual Fabius GS University of Florida Sem Lampotang

32 VAS

33 Check Out and Use

34 Power Switch ON Hidden on Back of Machine OFF

35 Fabius GS Startup Standby when not in use No need to power off
Startup simple and straightforward No automatic self test Manual leak test Manual calibration of sensors

36 Gases Connect pipeline gas supplies Connect reserve cylinders
Check that the supply pressures are in the specified range (green zone) Connect reserve cylinders Check cylinder pressure (open and re-close cylinder valve Verify pressure (O2 > 725PSI, N2O > 435 PSI)

37 Electric Verify electrical connection Verify Power Switch is on
The green LED on the user interface illuminates whenever AC power is connected Verify Power Switch is on Power switch is located on the rear next to the AC power cord

38 Internal Diagnostics are automatic
Result FUNCTIONAL CONDITIONALLY FUNCTIONAL - Use Non-critical fault Use machine now - Press Rotary Knob Call for repair BB to continue operation. NON-FUNCTIONAL - Don’t use Serious fault was detected Can’t use it

39 1 2 3 User Tests Defaults to Standby
User Tests components Simply follow directions System Test (done) Calibrate Flow Sensor Calibrate O2 Sensor Run Leak Test Access alarm log Restore site defaults Done 1 2 3 Show all

40 Calibrate Flow Sensor Hot-wire flow sensor (Spirolog®) Julian and Evita products. Calibration time <15 seconds

41 Standby Mode Run Leak Test
The leak test function key opens the support page for the leak test Leak test instructions Airway pressure numeric Oxygen flow numeric For machines equipped with the auxiliary pressure gauge, the leak test can be performed by building up a pressure of 30 cmH2O and observing the pressure drop on the gauge. The leak test includes the vaporizers. There is no need for a negative pressure test.

42 Calibrate Oxygen Sensor
Defaults to Standby User Tests components System Test (done) Calibrate flow sensor Calibrate O2 sensor Run leak test Access alarm log Restore site defaults

43 Calibrate Oxygen Sensor
Calibration time is < 15 seconds Room air 2 minutes first !

44 [Access Alarm Log] Access Alarm Log Alarm History up to 99 messages
The site defaults volume control settings, pressure control settings, alarm limits and alarm volume can be restored by pressing the “Restore Site Defaults” function key.

45 Sleep Mode Do not power off, put to sleep
Happens after 5 minutes No Buttons pressed No Fresh Gas Flow Pressing any button for Standby. Emergency wakeup Turn of FGF and awakens Man/Spont mode No confirmation required

46 Common Errors

47 Everything requires confirmation with Control Knob
Rotary Control and Push to Accept Knob

48 Flow Sensor can get wet Remove it, shake it dry, re-insert it
If replaced, calibrate before next case Reported expired volume will be wrong Delivered volume measured will be right

49 Compensate for Desflurane high conc
High concentration of desflurane increases specific heat of the gas combo cools hot wire anemometer more than right Des Compensation On corrects this Fabius GS does not measure agent so does not know when to correct

50 Leak and Functionality Check
You have done them, automatically And, the ADS knows it was checked compensates for compliance

51 Further Fabius GS Education
Duty Room today Patrick Parisi, Draeger representative Mike Bradish Anes Engineering Elective Res Support Info ftp://jphilip.bwh.harvard.edu/technologyblock Philip Grand Rounds Lecture Bradish Check Out Directions for Use PDF file BWH Check Out for older machines Anesthesia Written Board Exam Questions Simulation at [University of Florida]

52 Thank you End

53 Anesthesia Machines November 5, 2003
Study Guide Anesthesia Written Board Exam Questions ftp://jphilip.bwh.harvard.edu/technologyblock My answers available to all who answer the questions Grand Rounds New Anesthesia Delivery Systems at BWH Ohmeda Aestiva Draeger Fabius GS Clinical Conference Anesthesia Ventilator Failures Support Info ftp://jphilip.bwh.harvard.edu/technologyblock Simulation at [University of Florida]

54

55 Clinical Conference

56 Case 1 Ohmeda Aestiva 65 yo M for CABG on a Tuesday morning.
You induce anesthesia, intubate the trachea and confirm correct tube placement. You turn on the ventilator on your Ohmeda Aestiva Anesthesia Machine Ventilator alarms “Cannot Deliver Correct Volume” Soon, it alarms “Cannot Ventilate” If you know what is wrong and how to correct it Raise your hand now If you do not know what is wrong and how to correct it Raise your hand now Answer Dry the flow sensor tubes Use an HME to keep circuit dry

57 Case 2 Draeger Fabius GS 45 yo M for Thoracotomy on a Saturday
You intubate trachea and confirm tube placement. You turn on the ventilator on your Draeger Fabius GS Anesthesia Machine Vent Monitor shows alarm “Vent Fail” There is no mechanical ventilation If you know what is wrong and how to correct it Raise your hand now If you do not know what is wrong and how to correct it Raise your hand now Answer Cycle between Vent and Bag/Man Ask why software not upgraded Draeger Fabius GS


Download ppt "© , James H Philip, all rights reserved"