MULTISLICE CT
3
Neusoft ProVis 8 Product details Extremely easy to operate
Specifications GANTRY Multi-slice 2, 4, 7, 8 Detectors: 8 X 896, solid state Slice Thickness (mm): 0.5, 1, 1.5, 2, 2.5, 3 Gantry Tilt: -30 to +30 degrees XRAY TUBE Heat Storage: 7,500,000 HU Heat Dissipation Rate (HU/min.): 1,386,000 Focal Spot Dim. (mm): 0.8 X 0.9, 1.4 X 1.6 X-RAY GENERATOR kW Output: 60 kW kVp Range: 80 to 140 in 1 kVp step mA Range: 0 to 500 in 1 mA step SPIRAL SCANNING Max Scan Time (sec.): 100 Max Scan Volume (cm): 170 Reconstruction Time Per Image (sec.): 0.5 PATIENT COUCH Horizontal Travel (cm): 173 Scan able Range (cm): 170 Maximum Load Capacity With Accuracy (Kg): 205 OTHER FEATURES Auto mA, Contrast Tracking, Auto Voice, Auto Film, Auto Send, Auto Archive, CTA, CT Fluoro, Cardiac Scoring, Dental package, Perfusion, Bone Mineral Analysis, Cardiac Imaging Product details Extremely easy to operate Excellent low and high contrast resolution 0.5 second multi-slice scan and reconstruction 7.5 MHU X-ray tube with 1.38 MHU per minute heat dissipation rate 60 kW X-ray generator Powerful and user friendly physician workstation Fully compliant DICOM 3.0
4
SINGLE vs MULTI
6
CT EVOLUTION
7
1998 MULTISLICE CT SCANNERS INTRODUCED AT RSNA MEETING
9
SINGLE SLICE CT LIMITATIONS
SPEED PERFORMANCE 3D,MPR,MIP Z-AXIS RESOLUTION MPR LOW ACCURACY 360 LIA ( IMAGE QUALITY & ARTIFACTS) CT ANGIO LIMITATIONS MORE INTRAVENOUS CONTRAST
29
SLICE THICKNESS IN SINGLE SLICE CT DETERMINED BY:
PITCH PRE-PATIENT COLLIMATION
30
MULTISLICE CT COLLIMATION
BEAM FALLS ON THE ENTIRE MULTIROW DETECTOR ARRAY 8 MM WIDE BEAM FALLING ON FOUR DCETECTORS CAN PRODUCE: 4 –2MM SLICES
31
BEAM GEOMETRY SINGLE SLICE SPIRAL MULTISLICE FAN CONE
32
BEAM GEOMETRY AS THE NUMBER OF DETECTORS IN A MULTIROW DETECTOR ARRAY INCREASES THE BEAM BECOMES WIDER TO COVER THE 2D DETECTOr ARRAY. LARGER # OF ROWS IN THE DETECTOR ARRAY WILL RESULT IN A WIDER BEAM IN Z-AXIS DIRECTION ( CONE BEAM)
33
POST DETECTOR COLLIMATION IN MULTISLICE CT
THE COLLIMATORS ARE POSITIONED BETWEEN DETECTOR COLUMNS.
36
IMAGE RECON IN MULTI SLICE CT
SPECIAL FAN BEAM RECON ALGORITHMS HAVE BEEN DEVELOPED
37
INSTRUMENTATION
38
DAS ALMOST ALL MULTISLICE CT SCANNERS ARE BASED ON III GENERATION SYSTEM DESIGN ROTATE-ROTATE
39
GENERATOR HIGH FREQUENCY GENERATOR 60 KW OUTPUT.
40
X-RAY TUBE ROTATING ANODE TUBE WITH VERY HIGH HEAT CAPACITY