NIR Sensor Technology-Spectroscopy

Similar presentations


Presentation on theme: "NIR Sensor Technology-Spectroscopy"— Presentation transcript:

1 NIR Sensor Technology-Spectroscopy
Moving Grating-Scanning Monochromator AOTF MEMS FTIR Filter Photometer Fixed Grating-Diode Array

2 Acquisition Time FTNIR ~ 8 seconds AOTF~10 seconds
NIR Diode Array~1 sec or less Scanning Monochromator~30 seconds 50 X 20 msecs = 1 sec In presentation by Martin Warman, Pfizer, Inc. at IFPAC 2004

3 Types of NIR Spectrometers
Scanning Devices Moving Grating-Scanning Monochromator AOTF MEMS FTIR Non-Scanning Devices Filter Photometer Fixed Grating-Diode Array

4 An avenue to avoid interference and ambiguity
Multiple-sensor detection (CCD’s, diode arrays) An avenue to avoid interference and ambiguity

5 Sample The Sample is Moving! For the following unit operations:
Blending Coating Drying Granulation Roller Compaction On-Line Content Uniformity The Sample is Moving!

6 Resolution & Spectral Range
How much resolution do you need? What are spectral signatures of the materials you are looking at typically in the NIR spectral range? What spectral range is adequate?

7 NIR Diode Array Specifications
Standard InGaAs – 1.7 micron No of Pixels Spectral Range nm Resolution nm 128 6.5 256* 3.25 512 1.6125 1024 0.8 * Most popular choice

8 NIR Diode Array Specifications
Extended InGaAs – 2.2 micron No of Pixels Spectral Range nm Resolution nm 128 8.8 256* 4.4 512 2.2 1024** 1.1 * Most popular choice **Not readily available

9

10

11 Other Desirable Features
Ruggedness Diode Array spectrometers have no moving parts Ease of use Low Maintenance

12 Unique Features of CDI NIRS
Automated built in references for wavelength and amplitude, can add customer specific references Do not have to remove from vessel to do OQ, can be done on the fly, even during experiment Meets USP 1119 specifications-generates report Dual redundant light source All in one package, Class I Div II, EP available

13 Instrument Rating Pentagon Score: 0 = low; 5 = high
Cost Robustness Index Ease of use Richness of data Speed of Acquisition fti 1998