1 Autoclave Training November 16, 2005 University of Kentucky Department of Biosafety

2 Learning Objectives What materials can be autoclaved? What Personal Protective Equipment (PPE) is necessary for autoclave use? How does one properly load the autoclave? What steps assist in efficient decontamination of biohazardous waste? How does one verify autoclave performance for decontamination of biohazardous waste?

3 Importance of Proper Autoclave Use Tremendous pressure from steam in chamber provides explosive potential High temperatures and presence of extremely hot water creates potential for burns and scalding Inadequate decontamination allows for the potential of biological hazards and personnel and environmental contamination

4 Autoclave Explosion

5 What can be autoclaved Surgical Instruments Glassware Plastic tubes and pipette tips Solutions and water Animal food and bedding Waste

6 Types of waste Waste TypeContainerWhat You Need To Do Biological Infectious WasteOrange or clear autoclave bagAutoclave Medical Waste Biological Infectious Waste which cannot be autoclaved Red incineration bag in red waste container Medical Center- close and place by trash in lab for pick-up, Outside Medical Center- contact Environmental Management for pick-up Sharps- All needles, syringes, scalpels, blades, and all other sharp medical waste Sharps containerMedical Center- close and place by trash in lab for pick-up, Outside Medical Center- contact Environmental Management for pick-up Animal Carcasses- no biological, chemical, or radiological hazard Opaque bagReturn to appropriate DLAR location Animal Carcasses- intentionally infected Red incineration bagIncinerate (Medical Center or call Environmental Management pick- up) Used Biological Culture MediaGlass or plasticDisinfect with bleach or autoclave

7 Steam penetrates objects in the autoclave Condensation creates negative pressure and draws in additional steam Moist heat kills microorganisms via coagulation of proteins Two types of autoclaves Gravity Displacement Vacuum Assisted Principles of Autoclave Operation

8 Gravity Displacement

9 Vacuum Assisted

10 Autoclave Use Basics Personal Protective Equipment (PPE) Packaging Loading Operating User logs Maintenance logs Unloading Improper Autoclave Practices

11 PPE for Autoclave Users Eye Protection Lab Coat, Buttoned Closed-toed Shoes Heat-resistant Gloves

12 Packaging for Autoclaving Be sure the material should be autoclaved No volatile chemicals, sharps, red bag waste, radioactivity Utilize containers and autoclave bags appropriate for autoclaving Clear or orange bags UK Stores 6532-0046 Do not overfill containers

13 Carefully add water to autoclave bags Add approximately 1 liter of water to 24” x 30” bag Aids in reaching appropriate temperature quickly Creates steam in the autoclave bag to replace cold air Prepare packaging to allow steam penetration Check that all containers including bags are vented Loosely close autoclave bags Vent lids on bottles containing solutions Place packaged material in a secondary container Stainless steel tray or autoclavable polypropylene bin Packaging for Autoclaving

14 Loading the Autoclave Load material to allow efficient steam penetration Autoclave clean items and waste separately Do not allow material to be autoclaved to touch the sides or top of the chamber

15 Operating the Autoclave Be sure the autoclave is functioning properly before use Record information in User Log Close door properly and securely Choose the correct conditions for your material For decontamination: Temperatures between 121-124°C Total processing time 60-120 minutes Exposure time greater than 20 minutes Pressure minimum 15 PSI Select liquid cycle or slow exhaust cycle

16 Autoclave User Log Maintain User Logs for Two Years Complete User Log with Every Use

17 Autoclave Maintenance Log Maintain Maintenance Logs for Five Years Complete Maintenance Log for Every Repair Maintain records of Yearly Calibration

18 Unloading the Autoclave Put on Personal Protective Equipment Allow the autoclave to completely finish cycle Pressure gauge must read zero Verify cycle conditions were met Open door slightly to allow steam to escape Allow contents to cool before removal Carefully remove items Be especially careful with fluids and plastic bins

19 Improper Autoclave Practices

20 Results of Improper Autoclave Use

21 Program will ensure that all potentially hazardous biological waste is properly decontaminated prior to deposition in landfill Program requires testing of the ability of campus autoclaves to kill microorganisms There are over 50 autoclaves on campus which require this verification program Autoclave Verification Program

22 Autoclave Verification Each Load of Biohazardous Waste Autoclaves with automatic documentation Review the printed report If conditions were met, initial, date, and place User Log number on printed report If conditions were not met, do not remove the load, repeat autoclave cycle. If conditions are not met again, autoclave must be labeled out of order. Contact responsible individual posted on autoclave. Responsible individual contacts service company to schedule repairs.

23 Each Load of Biohazardous Waste Autoclaves without automatic documentation A Chemical Integrator Test Pack (CITP) must be incorporated in each load. Based on color change the performance of the autoclave will be evaluated. Tape the autoclaved CITP strip in the appropriate section of the User Log for your load. Autoclave Verification

24 Monthly Verification for Biological Waste Autoclaves Biological Indicator Test Checks all conditions of autoclave cycle, time, temperature, and pressure Verifies ability of autoclave to kill a microorganism Utilize the 3M Attest Biological Indicator Monitoring Kit Record results in Biological Indicator Results Log Autoclave Verification

25 Biological Indicator Results Log Maintain Results Logs for Five Years Complete Results Log with Every Monthly Test

26 Biological Indicator Test Procedure Label Attest package containing biological indicator vial with date, autoclave ID, and load number from user log Place the indicator vial package in the autoclave with the biohazard waste load Run the autoclave cycle as usual Using proper unloading procedures, remove the indicator package from the autoclave

27 Biological Indicator Test Procedure Label the indicator vial with all information from package Check indicator vial label for color change from rose to brown Place the biological indicator vial into the 56°C incubator

28 Place a non-autoclaved, labeled control tube in the incubator Examine the indicator tube at 8, 12, 24, and 48 hours for any color change Yellow color change indicates bacterial growth Compare to non-autoclaved control at each time point Record 48 hour test results in Results Log Biological Indicator Test Procedure

29 Biological Indicator Test Results Unprocessed Positive Control Processed

30 Learning Objectives Review What materials can be autoclaved? Surgical Instruments Glassware Plastic tubes and pipette tips Solutions and water Animal food and bedding Waste

31 What Personal Protective Equipment (PPE) is necessary for autoclave use? Learning Objectives Review

32 How does one properly load the autoclave? Learning Objectives Review Load material to allow efficient steam penetration Autoclave clean items and waste separately Do not allow material to be autoclaved to touch the sides or top of the chamber

33 What steps assist in efficient decontamination of biohazardous waste? Add approximately 1 liter of water to 24” x 30” bag Loosely close autoclave bags Place packaged material in an approved 2° container Learning Objectives Review

34 How does one verify autoclave performance for decontamination of biohazardous waste? Each Load Verify cycle conditions via autoclave printout CITP indicator use Monthly Utilize Attest Biological Indicator Vial

35 References and Links http://ehs.uky.edu/biosafety/autoclave.html http://www.uottawa.ca/services/ehss/autoclaves.htm http://www.3m.com/ http://www2.umdnj.edu/eohssweb/aiha/technical/labequipment. htm#Autoclave http://www2.umdnj.edu/eohssweb/aiha/technical/labequipment. htm#Autoclave http://ehs.uky.edu/em/ Environmental Management 323-6280

36 Questions or Additional Information Brandy Nelson Sr. Biological Safety Specialist Department of Biological Safety 252 East Maxwell Street Lexington, KY 40506-0314 Phone: 859-257-3753 Cell: 859-699-6082 bnels3@email.uky.edu http://ehs.uky.edu