1 Conventional UT B-scan Software Option
EPOCH 1000 Series Conventional UT B-scan Software Option

2 Presentation Contents
Overview and Key Features Required Parts Option Activation Creating an Encoded B-scan Creating a Timed B-scan Cross Sectional B-scan Reviewing Acquired B-scan Data Encoder Calibration

3 Overview and Key Features
3

4 Conventional UT B-scan Option
B-scan software option now available for use in UT mode! Encoded or time based New cable for connection to B-scan buggy NEW Single Value or Cross Sectional B-scan

5 Conventional UT B-scan Option
Available beginning with software version Once option is purchased it is remotely activated on the EPOCH 1000

6 Key Features Three B-Scan Inspection Modes Display Modes
Bi-Directional encoded Encoder is active. Scan moves in the same direction as encoder. Uni-Directional encoded Encoder is active. Scan moves forward when encoder moves forward or backward. Manual non-encoded (Timed) The encoder is not active. B-Scan updates according to a time rate that is set with the PRF control. Display Modes B-Scan only (Single Value or Cross-Sectional) B-Scan and A-Scan simultaneously

7 Key Features (cont.) Ability to store ALL A-scans during B-scan collection EPOCH XT only allowed for A-scan storage on alarm trigger or minimum depth Maximum Scan Rate has been increased to up to 1200 Hz With A-Scan Storage active the maximum scan rate is 800 Hz NOTE: The scan rate is equal to the PRF for values less than or equal to the maximum scan rate.

8 Key Features (cont.) B-Scan is compatible with Echo-to-Echo for coated surfaces or when there is a varying water path. Gate 1 width must be set wide enough to completely cover Gate 2 to avoid loss of signal. Cross-Sectional B-scan draws the B-scan image based on full A-scan amplitude data, regardless of gate position Can be viewed live or after a scan has been captured A-scan storage must be ON in order to view

9 Required Parts 9

10 Parts Required for B-Scan
EPOCH 1000 B-Scan Software Option EP1000-BSCAN (U ) EPOCH 1000 Encoder Cable EP1000-ENC-CBL-10 (U ) Also available in 25, 50, or 100 foot lengths Encoder Buggy BSCAN-ENC (U ) Transducer Typically a D790-SM (U ) Transducer Cable: Choose length to match encoder cable BNC: BCMD B (U ) LEMO: L1CMD B (U )

11 B-Scan Kits Most customers choose to purchase B-Scan kits that contain the Instrument Software Option, B-Scan Encoder Buggy, and B-Scan Encoder Cable EP1000-BSCAN-KIT-10: 10 foot encoder cable (U ) EP1000-BSCAN-KIT-25: 25 foot encoder cable (U ) EP1000-BSCAN-KIT-50: 50 foot encoder cable (U ) EP1000-BSCAN-KIT-100: 100 foot encoder cable (U ) The transducer and transducer cable must be purchased separately.

12 B-Scan Encoder & Accessories
D790-SM Transducer Cable Encoder Cable Input Probe Holder Water Input Optional Extension Pole Bracket Optional Magnetic Wheels

13 Option Activation 13

14 Option Activation The EPOCH 1000 must be updated to Operating Software version or higher If the B-scan option has been purchased, it can be turned on using the B scan menu

15 Option Activation (cont.)
If the B-Scan Option is NOT activated in your instrument, contact your local Olympus NDT representative to purchase this option Once this software option is purchased it can be remotely activated on the EPOCH 1000i via a software code provided

16 Menus and Interface To turn on the B-scan option, select B Scan > BScan = On When B-scan is turned on, new sets of group menus appear

17 Menus and Interface (cont.)
While B-scan is ON, menu 5/6 will now contain additional parameters and controls for use in B-scan inspections Encoder Used to perform an encoder calibration Cursors Position controls for X and Y sizing cursors Tools Statistics and other post-analysis functions Datalogger and reporting functions will now be located in menu 6/6

18 B-scan Menu The controls in the B scan menu are as follows:
Screen: Selects views of A-scan/B-scan or B-scan only Mode: Starts or Stops B-scan acquisition Source: Selects type of B-scan image single value or cross-sectional Correlated View: Selects alignment of A-scan and image Source: Builds B-scan image from selected Gate

19 Additional Features A Minimum Depth Alarm may be used to alert the operator to areas where the part thickness has dropped below a user defined threshold. The Floating Gate option is included with the purchase of the B-scan software and may be used to set the gate height Gate can be set at a level from -1dB to -14dB from the height of the gated echo. Avoid echo drop out when scanning on heavily corroded parts. Maximize measurement consistency by measuring at the same point on the edge of the echo envelope. Floating gate can be set in the Gate Setup > Setup menu

20 Creating an Encoded B-scan
20

21 Creating an Encoded B-scan
The EPOCH 1000 must be properly setup and calibrated before the B-Scan option is turned ON. Pulser / Receiver Settings Calibrate for Zero Offset and Velocity Gate 1 should be setup to cover the minimum and maximum expected thickness values for the inspection. The screenshots below show a complete setup ready for B-Scan to be turned on.

22 Equipment Setup For encoded B-scans, a positional encoder is used to track distance traveled along the scan axis. The standard encoder for use with B-scan is the encoder buggy Insert your transducer into the B-scan buggy and tighten the thumb screws to secure Attach the transducer cable to the transducer and to the EPOCH 1000 Attach the encoder cable to the encoder buggy and EPOCH 1000

23 Encoded B-scan Setup Now you are ready to set up your B-scan using the following steps: Turn ON the option in the B Scan > BScan menu. Once the B-scan option is turned on, additional B-scan parameters will display in the menu.

24 Encoded B-scan Setup (cont.)
Select B Scan > Setup to enter the setup screen Choose an Encoder Mode (Uni or Bi-directional) Select a Direction (will determine the display view of the A-scan and B-scan image) Set your Scan Resolution (spacing between B-scan points) This value will determine the maximum scanning speed.

25 Encoded B-scan Setup (cont.)
The Encoder Resolution for the B-scan encoder buggy is 200 pulses/in. (7.87 pulses/mm) If you are using a different encoder, you should perform an encoder calibration to determine the encoder resolution. The value from the encoder calibration will automatically populate from the encoder calibration For instructions on how to calibrate the encoder, please use the following link: Encoder Calibration

26 Encoded B-scan Setup (cont.)
Set the desired Start and End Co-ordinates for the scan Select whether to Save Ascan images You must be saving A-scan images in order to create a cross-sectional B-scan If a cross-sectional B-scan for Gate 2 is being used, select the Amplitude Decimation factor Gate 1 amplitude compression is automatically set to maximum The current Scan File Size based on the selected parameters is displayed. This parameter cannot exceed the Max File Size of approximately 70Mb Press the [ESCAPE] key to return to the live screen.

27 Collecting B-scan Data
To acquire data for your B-scan Set the B Scan > Mode to Start Move your B-scan buggy and transducer along your test piece at the appropriate scan rate so that you do not receive an LOS measurement

28 Thickness limits defined by Gate1 Start and Width
A/B-Scan View Min Depth Alarm Areas of Scan below Minimum Depth Thickness limits defined by Gate1 Start and Width Encoded Scan axis

29 Creating a Timed B-scan
29

30 Creating a Timed B-scan
The EPOCH 1000 must be properly setup and calibrated before the B-Scan option is turned ON. Pulser / Receiver Settings Calibrate for Zero Offset and Velocity Gate 1 should be setup to cover the minimum and maximum expected thickness values for the inspection. The screenshots below show a complete setup ready for B-Scan to be turned on.

31 Timed B-scan Setup Now you are ready to set up your B-scan using the following steps: Turn ON the option in the B Scan > BScan menu. Once the B-scan option is turned on, additional B-scan parameters will display in the menu.

32 Timed B-scan Setup (cont.)
Select B Scan > Setup to enter the setup screen Set the Encoder Mode to Timed Select a Direction (will determine the display view of the A-scan and B-scan image) Select whether to Save Ascan images You must be saving A-scan images in order to create a cross-sectional B-scan

33 Timed B-scan Setup (cont.)
If a cross-sectional B-scan for Gate 2 is being used, select the Amplitude Decimation factor Gate 1 amplitude compression is automatically set to maximum The current Scan File Size based on the selected parameters is displayed. This parameter cannot exceed the Max File Size of approximately 70Mb. Press the [ESCAPE] key to return to the live screen.

34 Collecting B-scan Data
To acquire data for your B-scan Set the B Scan > Mode to Start Move your transducer along your test piece at the appropriate scan rate so that you do not receive an LOS measurement NOTE: If the scan is moving too quickly, lower your PRF to reduce the scan speed!

35 Thickness limits defined by Gate1 Start and Width
A/B-Scan View Min Depth Alarm Areas of Scan below Minimum Depth Thickness limits defined by Gate1 Start and Width Timed Scan axis

36 Cross-Sectional B-scan
36

37 What is a Cross-Sectional B-scan?
A-scans digitized and mapped to color scale based on amplitude values Points are plotted so that the colors are positioned at the proper depth A-scans from successive transducer positions are "stacked" to form a true few cross-sectional image 1 2 Transducer movement 3 Part depth

38 What is a Cross-Sectional B-scan? (cont.)
The depth axis on a cross-sectional B-scan uses the entire screen range, rather than just the gated region as on a single value B-scan Depth (screen range) Scan axis (encoded or timed)

39 Single Value vs. Cross Sectional
Single Value B-scan “Traditional” EPOCH B-scan view Same as EPOCH III, EPOCH 4, EPOCH XT, 37DLPlus, 38DLPlus, etc. Draws a profile view of the test piece based on the gated single thickness value per index Most commonly used in corrosion scanning applications

40 Single Value vs. Cross Sectional (cont.)
Cross Sectional B-scan Much more similar to phased array images (zero degree linear scan) Draws the B-scan image based on full A-scan amplitude data, regardless of gate position More commonly used in zero degree laminar applications

41 Reviewing Acquired B-scan Data
41

42 Completing a B-scan Once you have acquired all the data from a particular section or have finished your entire scan, you can choose to Stop or Freeze the B-scan. Stopping a scan will complete the scan. New B-scans will be started from the start coordinate Freezing a scan will bring you to Review mode. When the scan is unfrozen you can continue the scan from the point where the scan was frozen. vs. B Scan > Mode [FREEZE]

43 Saving Files B-Scan files may be stored in any file type in the EPOCH The most common are: Calibration (CAL) Files: Used to store B-Scan setups Incremental (INC) Files: Used for most inspection data storage When the [SAVE] key is pressed the instrument saves the entire B-Scan along with stored A-Scans and the instrument setup. Saved B-scan files can be recalled as a setup, if desired, through the Files > Open menu in the datalogger After recalling a file, press the [FREEZE] key to enter Review Mode

44 Start/Stop Function The Start/Stop function is used to end your B-scan
For convenience, the [CHECK] key can also be pressed when Start is highlighted to Stop a B-scan. Once a B-scan is stopped, you will need to save and recall the scan in order to review the data. When the B-scan is started again it will create a new scan from the start point rather than resume from the last captured point.

45 Freeze Function When you have finished your scan section, or to pause and review your acquired data, press the [FREEZE] key to enter Review Mode. An “F” icon will display to the left of the scan to indicate that you are reviewing a frozen scan. The Start/Stop Mode controls will be disabled when review mode is active.

46 Freeze Function (cont.)
Entering Review Mode allows use of the following functions/features, which are explained further in this section: Image cursors (D.T.) Sizing cursors (X, Y) Changing gate source data (Gate 1 or 2, single value or cross-sectional)

47 Cursors Once a B-scan is acquired, you can perform basic on-board analysis in Review Mode using available cursors Sizing cursors (X, Y) and DT cursor are available in the Cursor menu

48 Distance Traveled (D.T) Cursor
The distance traveled (D.T.) cursor selects a specific distance or timed point along the scan axis. For encoded B-scans, the D.T. cursor will select a specific position along the scan axis in inches or mm. For timed B-scans, this cursor will correspond to a specific screen pixel along the scan axis. To adjust the D.T. cursor select Cursors > D.T. The cursor will display as white lines on the B-scan image.

49 Distance Traveled (D.T) Cursor (cont.)
If selected in B-scan setup, a compressed A-scan will be stored for each point on the B-scan. The D.T. cursor is used to choose a particular B-scan point to view the A-scan image. To view stored A-scans along with the B-scan, set the B Scan > Screen view to A/B Scan

50 Sizing Cursors Sizing cursors are available for basic flaw sizing on the B-scan image Based on your screen view, the X and Y cursors will be correspond to part depth (thickness) and scan axis (distance or time) The X and Y cursors will form a shaded box around the selected area for easier viewing. To turn the sizing cursors ON, select Cursors > Cursors. The X and Y cursors can then be adjusted to take measurements in both directions along the scan. If you are selected in Auto mode, the relevant cursor measurements will be displayed in the measurement boxes above the scan.

51 Compressed View Since encoded B-scans allow for data collection for lengths longer than the screen width, you can compress a B-scan to view the entire scan on screen all at once. To turn compressed view ON, select B Scan > Compressed

52 Correlated View Aligns the thickness axis of the B-scan with the soundpath (depth) axis of the A-scan Best used with cross-sectional B-scan which is plotted over the full range of the A-scan The alignment of the A-scan and B-scan will match along the part thickness and depth axis

53 B-scan Analysis – Minimum Depth
In Single Value B-scan, you can display critical thinning areas in red by adjusting the Min Depth Alarm position To adjust the minimum depth select Gate 1 > Min Depth

54 B-scan Analysis – Statistics
NEW Tools menu allows the user to display Statistics on the current acquired B-scan To view the information for the current B-scan, select Tools > Statistics from Review mode

55 Encoder Calibration 55

56 Encoder Calibration The Encoder Resolution for the B-scan buggy is 200 pulses/in. (7.87 pulses/mm) If you are using a different encoder, you should perform an encoder calibration to determine the encoder resolution. Required Equipment: Encoder and encoder cable Transducer and cable Ruler, ideally at least 6 in. (150 mm)

57 Calibration Setup Once the encoder has been properly attached to your probe: Enter the Encoder > Calibration menu Press [P3] to Reset the encoder pulses and D.T. values Set the Start and End points to be used for the calibration Using a greater calibration distance will increase encoder resolution accuracy as error is averaged over a larger distance.

58 Encoder Calibration Position the front of the probe at the zero mark (start point) on your ruler Press [P1] Cal Start to begin the calibration Move the probe until the front is at the end point on your ruler Press [P1] End Cal to complete the calibration

59 Encoder Calibration (cont.)
After the calibration has been complete, the newly calculated Encoder Resolution will display. The resolution for the encoder buggy should be 200 pulses/in. This value will automatically populate under Encoder Resolution in the B Scan > Setup menu Press the [ESC] key to return to the live screen

60