Geneva, 12/06/2014 1 Results of Magnetic Measurements on MQXC 02 L. Fiscarelli on behalf of TE/MSC/MM section 12.06.2014 1
2
Geneva, 12/06/2014 Overview Introduction: – Measurement system – Measurement plan DC characterization – Stair step cycle (load-line) Dynamic characterization – Ramp-rate study Simulated machine cycle – Repeatability and Decay / Snapback Comparison with ROXIE model (data provided by Per Hagen) Cold-warm correlation Measurement resolution Conclusions
3
Geneva, 12/06/2014 The two setups 5 coils Support tube R=45 mm – Flexible software Framework for Magnetic Measurements (FFMM) – PXI crate – Fast Digital Integrator (FDI) – Motor + encoder + slip-ring units – Large diameter rotating shaft
4
Geneva, 12/06/2014 Measurement plan A short magnetic measurement campaign was requested (limited availability of cryogenic test station) Minimal set of measurement cycles was planned (“standard program”) – Stair-step (“load-line”) – Ramp-rate study (10, 20, 40, 80 A/s) – Simulated machine cycle 3x (plateau of 1000 s at 820 A and then PELP ramp) Problems with Power Converter (trips on high current plateaus) wasted some time – Simulated machine cycle repeated only 2 times
5
Geneva, 12/06/2014 Results: stair-step cycle Stair-step cycle to measure the DC transfer function and DC field quality quench pre-cycle (ramp-up at 11 A/s up to 12.8 kA, plateau of 300 s, ramp-down at -11 A/s to 100 A) from 100 A to 12.8 kA and back with a total of 35 steps plateaus (140 s) linear ramps (11 A/s)
6
Geneva, 12/06/2014 Results: stair-step cycle, TF Transfer function Strong saturation -180 units on straight section -160 units on integral
7
Geneva, 12/06/2014 Results: stair-step cycle, allowed multipoles
8
Geneva, 12/06/2014 Results: stair-step cycle, not allowed multipoles
9
Geneva, 12/06/2014 Results: ramp-rate study Ramp-rate study cycle to measure the effect of eddy currents and cable coupling currents pre-cycle (ramp-up at 11 A/s up to 12.8 kA, plateau of 300 s, ramp-down at -11 A/s to 100 A) current cycles from 820 A to 12.8 kA and back different ramp rates (10, 20, 40, 80 A/s)
10
Geneva, 12/06/2014 Results: ramp-rate study As expected very small effects of ramp-rate on TF and multipoles
11
Geneva, 12/06/2014 Results: simulated machine cycle Simulated machine cycles to measure multipoles in machine simulated conditions and the contribution of the persistent current decay and snapback quench pre-cycle (ramp-up at 11 A/s up to 12.8 kA, plt of 300 s, ramp-down at -11 A/s to 100 A) plateau at injection current of 1000 s ramp to the nominal field with slow parabolic start (no exponential) repeated two times
12
Geneva, 12/06/2014 Results: simulated machine cycle, repeatability
13
Geneva, 12/06/2014 Results: simulated machine cycle, decay and snapback
14
Geneva, 12/06/2014 Comparison with ROXIE model Transfer Function – -25 units measured on geometric – Exact match for saturation b 6 – Good agreement b 10 – Constant offset * data provided by Per Hagen
15
Geneva, 12/06/2014 Cold-warm correlation of multipoles Multipoles at 1.9 K are smaller than at 300 K Correlation factor of 0.91 Elastic deformation ???
16
Geneva, 12/06/2014 Measurement precision High precision measurement results better than 0.01 units at 40 mm on multipoles ±5 units on TF
17
Geneva, 12/06/2014 Conclusions Complete MM test campaign at cryogenic temperature on MQXC 02 Saturation Low ramp-rate dependency Decay and snapback as expected High precision results both on TF and multipoles Disagreement model/measurements on TF solved
18
Geneva, 12/06/2014 Appendix
19
Geneva, 12/06/2014 Appendix 1: Correction of thermal contraction of shaft at 1.9 K [1] http://cryogenics.nist.gov/MPropsMAY/G-10%20CR%20Fiberglass%20Epoxy/G10CRFiberglassEpoxy_rev.htm Material Properties: G-10 CR (Fiberglass Epoxy) [1] We range from 300 K to 1.9 K thermal coefficient is ‘a’ A coil is a 3D object (width, length, radius) a normal * a warp * a warp → -1.21 %
20
Geneva, 12/06/2014 Appendix 2: Parameterization of Transfer Function [2] L. Bottura, Standard Analysis Procedures for Field Quality Measurement of the LHC Magnets - Part II: Transfer Functions and Parameterization