Presentation is loading. Please wait.

Presentation is loading. Please wait.

Trekview® Desktop Software Training

Similar presentations


Presentation on theme: "Trekview® Desktop Software Training"— Presentation transcript:

1 Trekview® Desktop Software Training

2 TrekView Tag Basics

3 TrekView Desktop Reader (DTR)
Once the desktop reader is connected to a PC and TrekView software is open, hold the tag within 2 inches of the snowflake on the DTR. Be sure that the desktop reader is not near or on any metal objects (file cabinets, etc.) as this will cause interference. The Desktop Reader plugs into your PC with a USB “B” to USB cord. It must be plugged into a PC running TrekView software to read and download tag data.

4 TrekView Handheld Reader (HHR)
The HHR must be in the “L” shape shown below, and the grey bulls-eye must be within 5 inches of a tag in order to read the tag data. The HHR features a touchscreen user interface The HHR plugs into your PC with a mini USB to USB cord. The HHR features an SD card slot to store tag data.

5 TrekView Report Software
Report Options (Language, Time Zone, °C/ °F, Report Details) Save Location (View / Open Files Saved) Close Program Regenerate Report Note: TrekView user and report software can be downloaded from

6 TrekView User Software
Shortcut Toolbar – allows you to read / configure tags and export reports HHR Icon – available when handheld reader connected Tag Tree – displays tags ID when tag connected / scanned New Work Session Open Save Export PDF Export CSV Tag Mode Configure Tag Tag Name Current Tag Temp Download Log Snyc HHR Connect Icons – show if tag / reader has been connected

7 Opening Screen – Tag Connected
Select Configure Tag Icon

8 Tag Setup Use default setting, a custom setting previously created, or click add button to create a new custom setting Click the configure button to transfer a setup to a tag

9 Build a New Configuration
Ascending thresholds (red) Descending thresholds (blue) Threshold conditions Temperature Time Single or Cumulative LED Status (solid, blink, none) Add / remove thresholds Logging interval Sets max logging time Max tracking time ALWAYS 1095 days (3 years) Soaking countdown Time delay before start

10 Examination of Tag Data
Overview of Tag Data LRT (last recorded temp) MKT (mean kinetic temperature) Min / Max Temperature Status – Check or Ok

11 Temperature Curve

12 Rollover Icon Use mouse to roll over the curve and get specific time / temperature information

13 Histogram Tab Histogram of Data – rollover bars to see time in each temperature bin

14 Pie Chart Tab

15 Excursion Tab

16 Info Tab Check battery life and use count

17 History Tab Displays a history of tag scans

18 Tag Alerts Alerts user if tag has been damaged or compromised

19 Ordering Information Part Number Description TV43800 INT USB
TrekView Wireless Temperature Recorder with Internal USB TV43900 EXT USB TrekView Wireless Temperature Recorder with External USB TV43800 SP INT USB TrekView Wireless Temperature Recorder Internal USB & Stainless Probe TV43800 RP INT USB TrekView Wireless Temperature Recorder Internal USB & Ribbon Probe TV43900 SP EXT USB TrekView Wireless Temperature Recorder External USB & Stainless Probe TV43900 RP EXT USB TrekView Wireless Temperature Recorder External USB & Ribbon Probe TV43700 TrekView Handheld Reader TV43705 TrekView Desktop Reader

20 For additional information email info@shockwatch. com or visit www
For additional information or visit