BLM Test Results 22 October SIC, Faraday cup and 2 cables mounted on Girder 5.

Similar presentations


Presentation on theme: "BLM Test Results 22 October SIC, Faraday cup and 2 cables mounted on Girder 5."— Presentation transcript:

1 BLM Test Results 22 October SIC, Faraday cup and 2 cables mounted on Girder 5

2 Note: The SIC and the Faraday cup give almost the same signal. But since they have different collective areas (Faraday cup 8X sic), and that the Faraday cup is only sensitive to the charge particles, we must conclude that the amount of Xrays contributing to the signal of the SIC is about 8 times as much as the signal on the SIC from charged particles. The difference in the cable areas is: …

3 Beam loss scan in Z on Girder 6, data 1

4 Beam loss scan in Z Girder 6 data 2