Nitric Oxide Set-up for Patients on Ventilator
2
Injector Module Place the injector module
directly to ventilator on the inspiratory limb of the ventilator circuit, where the gas flow exits the ventilator. This is the same for any ventilator. Inspiratory Injector Module
3
Sample line non- heated circuit set up
Is placed in line on inspiratory limb, 6-12” after the injector module and 6-12” from patient. Will need to cut circuit and use blue adaptor to add T-piece for sample line. S A M P L E I N
4
Sample Line for heated circuit
Sample line placed on dry side of heater 6-12” after injector module . Will need blue adaptor to place in-line and T-piece where sample line attach. Sample Line
5
Nitric Machine Big white power button on back of Nitric. Touch Screen
To adjust NO level touch and turn knob. Constant sample being measured for accuracy. Photo from dartmed.dartmouth.edu Sample measurement NO Level Delivery line
6
Connecting the Nitric tank
Remove the shrink wrap and red lock (push down on lock to remove.) Remove cap on tank and connect hose to tank (one hose for each tank.)
7
Turn tank on and purge line
Turn on tank: Turn tank on (open completely). When tank is off, dashes and the # Zero should be flashing. The dashes disappear when tank is on.
8
Delivery of Nitric Resuscitation Bag
Can bag with same amount of NO you can deliver in-line. Must be hooked up to Oxygen source. Attach Ambu bag to front of nitric where flow meter is located. Run flow meter at 15LPM. Picture from dartmed.dartmouth.edu NO dial O2 line Connect ambu
9
Nitric via Transport Ventilator
Sample Line Injector Module
10
Common alarms and what to check
Low nitric oxide Tank closed Sample line blocked Recalibrate Make sure tank is on Make sure sample line is not blocked
11
The big picture
12
When you are all done with Nitric
Turn off the gauge when Nitric is discontinued. You will see Dashes ( ) on the gauge when gas is completely off. No dashes and the tank is not off. Turn paper work in to Mike, also call mike and Cyndi when you turned it off.