Presentation is loading. Please wait.

Presentation is loading. Please wait.

Homogeneity of Small Chambers T4, T6, T7, T8 and Gas Gain Measurements of Frascati and Roma Chambers Emmanuel Angulo (CSU Fresno) Weekly Meeting ATLAS.

Similar presentations


Presentation on theme: "Homogeneity of Small Chambers T4, T6, T7, T8 and Gas Gain Measurements of Frascati and Roma Chambers Emmanuel Angulo (CSU Fresno) Weekly Meeting ATLAS."— Presentation transcript:

1 Homogeneity of Small Chambers T4, T6, T7, T8 and Gas Gain Measurements of Frascati and Roma Chambers Emmanuel Angulo (CSU Fresno) Weekly Meeting ATLAS Muon Micromegas 7/8/2012

2 Cu X-Ray Measurements 90 Positions (0.9cm x 1cm) HV of Strips @ 520 HV of Drift @ 300 X-Ray with 1mm collimator and cu foil cover @ 100000 event count, and @ 10mm distance from chamber Cu X-Ray12345678910 1 2 3 4 5 6 7 8 9

3 T4 (10mm to 5mm Drift) Unifortmity

4 T6 Uniformity Corner points 1cm x 1cm Middle point 6cm x 5cm 1 2 5 3 4 Connector 1 Connector 2

5 T6 Uniformity

6 T7 Uniformity Corner points 1cm x 1cm Middle point 4cm x 5cm 1 2 5 3 4 Connector 1 Connector 2

7 T7 Uniformity

8 T8 Uniformity Corner points 1cm x 1cm Middle point 4cm x 5cm 1 2 5 3 4 Connector 1 Connector 2 Noise up to Ch 143 (Bad signal with connector 2)

9 T8 Uniformity with 55 Fe 25 Positions (1.8cm x 1.8cm) 3mm Collimator Hole @ 50000 event count Strip HV @ 520 Drift HV @ 300

10 Roma and Frascati Chambers Gas Gain 55 Fe Measurements 1 3 5 2 4 5 Positions (1.8cm x 1.8cm) 3mm Collimator Hole @ 20000 event count Strip HV from 440- 560 Drift HV @ 300 HV Connectors X-Strips Y-StripsY-Strips

11 Frascati Gas Gain

12 Roma Gas Gain


Download ppt "Homogeneity of Small Chambers T4, T6, T7, T8 and Gas Gain Measurements of Frascati and Roma Chambers Emmanuel Angulo (CSU Fresno) Weekly Meeting ATLAS."

Similar presentations


Ads by Google