S P E A R™ decompression needle system
Lateral Thoracic Decompression The following program was created specifically for the SPEAR™ decompression needle system orientation and training. For questions, concerns or additional information please contact: North American Rescue®, LLC Tel: Mail: 35 Tedwall Court Greer, SC Fax: S P E A R™ decompression needle system Simplified Pneumothorax Emergency Air Release Rev D
2
The following materials were developed for the purpose of
SPEAR™ decompression needle system orientation and training Warranty: The SPEAR™ decompression needle system is a medical device, the use of which requires specific education and training. North American Rescue, LLC. warrants the SPEAR™ as merchantable expressly for the indication detailed. North American Rescue disclaims all other implied warranties relating to this product, to include use beyond this product’s identified purpose, and utilization by untrained personnel or legally unauthorized parties. Caution: Federal Law restricts the SPEAR™ to sale by, or on the order of, a licensed physician. As with most modern medical devices, the SPEAR™ (Simplified Pneumothorax Emergency Air Release) decompression needle system meets rigid safety standards, governmental mandates, and the demanding expectation of clinicians worldwide. The SPEAR™ decompression needle system was developed with attention to the latest presented and published scientific evidence, as well as guidance from today’s leading experts in the fields of Emergency Medicine, Trauma Surgery, Pulmonology, Critical Care, Combat Care, and Pre-Hospital Emergency Medicine. The SPEAR™ Warranty and Advisory should be read and understood as a component of this program.
3
Indication: For relief of tension pneumothorax in the adult patient.
The following materials were developed for the purpose of SPEAR™ decompression needle system orientation and training Key Term: Tension Pneumothorax. A known life-threatening medical emergency which, if left untreated, may result in death. Indication: For relief of tension pneumothorax in the adult patient. Contraindications: Not intended for treatment of simple pneumothorax or hemothorax. The efficacy of this device has not been established in pediatric patients. Warning: Failure to utilize this device properly may result in injury to cardiac, pulmonary, or vascular structures. The Key Term, Indication and Contraindications are further reinforced within this program. WARNING: As with all invasive medical devices, usage is predicated on proper orientation and training, careful consideration of the indications, patient condition, review of potential contraindications, and a specific RISK versus BENEFIT analysis. Failure to utilize this device according to the directions for use, didactic materials provided, and NAR recommended hands-on training MAY CAUSE FURTHER INJURY.
4
Identify components and function
Objectives At the conclusion of didactic and hands-on training, you should be able to: Identify components and function List indications, contraindications, and expected therapeutic benefits of thoracic decompression Identify (right and left) lateral landmarks for thoracic decompression Identify (right and left) anterior landmarks for thoracic decompression List indications of successful thoracic decompression Define potential complications of improperly performed thoracic decompression procedure Discuss current scientific evidence as it relates to thoracic decompression covered in anterior insertion program At the conclusion of this specifically developed SPEAR™ didactic, and NAR recommended hands-on training program, the clinical provider should be able to: Identify the SPEAR™ components and function. List the indications, contraindications, and expected therapeutic benefits of thoracic decompression. Identify the (right and left) lateral landmarks for thoracic decompression. Identify the (right and left) anterior landmarks for thoracic decompression. Objective is GRAY on this slide because it will be covered in anterior insertion program. Suggestion – review and become competent with both insertion locations. List the indications (signs and symptoms) of successful thoracic decompression. Define the potential complications of an improperly performed thoracic decompression procedure. Discuss current scientific evidence as it relates to thoracic decompression. At the completion of clinical training consider offering both a written and skills competency evaluation.
5
SEVEN (7) key system features
Objective 1: components and function decompression needle system Simplified Pneumothorax Emergency Air Release Spin-Lock SPEAR™ (Simplified Pneumothorax Emergency Air Release) decompression needle system and training materials meet and exceed 11 April 18 TCCC Guidelines Change 17-02”. The SPEAR™ is a robustly packaged complete assembly created to address documented failures and current management challenges in the field and critical care settings. Length: 3.75 inches. KEY Point: Specific research (of both civilian and military patients) notes that the 3.75 inch length will improve thoracic decompression and chest injury management success rates (*based on an average 2.1 inches of skin to rib depth AND an additional average of 0.5 inches of rib to pleura depth). Size: 10ga (10 French) catheter with distal fenestrations (holes) and soft distal tip. The fenestrations provide a 314% increase in air outflow when compared to a 14ga catheter, and a 74% increase in air outflow over a non-fenestrated 10ga. The catheter’s soft tip has been shown to lessen the likelihood of inadvertent penetration, or injury to the visceral pleura. IMPORTANT POINT: SPEAR™ catheter walls (external and internal lumen combined) have been shown to remain patent (open) during routine usage. Catheter length has been shown to remain in position (on both larger and smaller adult patients) during treatment and transport (movement). Needle tip has been modified (to a bi-beveled cutting blade with an adjusted angle of entry) to improve penetration through varying types of tissue. This modification offers clinicians enhanced entry control (which prevents “explosive entry” and penetration failure, thus improving patient safety). The catheter has been designed with NVG-friendly centimeter markings along its length for precise depth control (a key factor to improve patient safety) during placement. Needle and catheter are conjoined by a novel spin-lock preventing, premature separation during SPEAR™ placement (a documented failure point in ALL other needle decompression systems). A one-way valve is contained within this system and attached (via slip-tip) to the proximal end of the needle hub. This one-way valve is designed for clinician directed attachment to the catheter during patient treatment. The one-way valve may also be connected to a low volume suction system as needed. SEVEN (7) key system features
6
Objective 1: components and function (removal from protective case)
2 in protective case (sterile) sterile seal broken (twist off) 3 The SPEAR™ decompression needle system is packaged in a proven, robust hard-shell case, with an assurance seal indicating sterility and lot number. As with all medical devices, labeling provides additional information regarding device length and size, as well as contact information for North American Rescue®, LLC. Instructions for use accompany EACH needle set and should be reviewed during training efforts and emergent usage when needed. To open: Ensure integrity of safety seal “Twist off” red cap, breaking safety seal Remove red case cap, exposing SPEAR™ case cap removed (exposing proximal needle set)
7
removed from case (pictured on sterile field)
Objective 1: components and function (removal from protective case) 4 Continued from previous slide: The SPEAR™ decompression needle system is packaged in a proven robust hard-shell case with an assurance seal indicating sterility and Lot number. As with all medical devices, labeling provides additional information regarding device length and size as well as contact information for North American Rescue®, LLC. Instructions for use accompany EACH needle set and should be reviewed during training efforts and emergent usage when needed. To open: Ensure integrity of safety seal “Twist off” red cap, breaking safety seal Remove red case cap, exposing SPEAR™ Remove SPEAR® and inspect device prior to usage removed from case (pictured on sterile field)
8
Objective 1: components and function
catheter disconnects from needle by releasing spin-lock (with ¼ turn) The SPEAR™ decompression needle system has a purposefully designed Slip-Lock between the catheter and the needle hub. The novel Spin-Lock is designed to prevent premature separation during insertion (a documented insertion failure point). Once proper insertion depth has been achieved (≤ 3 cm or ≤ to 1.2 inches), and SPEAR™ has been oriented toward the mid-clavicle, separation of needle and catheter can be easily accomplished with a ¼ turn disconnect (“twist off”) and release. Catheter may then be threaded (guided) into position along parietal pleural wall for maximal therapeutic benefit. catheter threads into position (while holding needle stationary)
9
Objective 1: components and function
remove entire needle (after catheter has been threaded into position) Note that SPEAR™ needle should be removed ONLY when the catheter has been threaded into position (an established failure point for needle decompression). Note that 10ga catheter has improved flexibility to promote continued relief of thoracic pressure during critical patient management. note catheter flexibility
10
Objective 1: components and function
disconnect one-way valve from needle hub (safely secure needle in bio-hazard sharps container) A one-way valve is attached to the SPEAR™ needle hub. The one-way valve is attached with a “slip-tip connection” and can simply be “pinched off” with one hand (in preparation for placement on the catheter’s proximal end).
11
Objective 1: components and function
attach one-way valve to catheter hub (auditory signal may be heard) Attach one-way valve to the catheter’s proximal hub as indicated. Attachment can be achieved by standard medical connection. Secure catheter per protocol or standard. One-way valve may be attached to a low-pressure suction system as indicated.
12
tension pneumothorax:
Objective 2: indication tension pneumothorax: a life-threatening emergency, which if left untreated, may result in death A tension pneumothorax is known to be a life-threatening emergency, which if left untreated, may result in death. NOTE: Illustration depicts left lateral view of patient WITH left-sided tension pneumothorax. left lateral illustration of left sided tension pneumothorax
13
indications: consider treatment of tension pneumothorax
Objective 2: indication indications: consider treatment of tension pneumothorax when one or more of the following are present: severe or progressive respiratory distress severe or progressive tachypnea absent or markedly decreased breath sounds oxygen saturation less than 90% shock traumatic cardiac arrest without obvious fatal wounds (consider immediate placement of bilateral SPEAR™ decompression needles and consider ALL other possible causes) Indications: Treatment of a tension pneumothorax should be considered when one or more of the following are present: Severe or progressive respiratory distress. Severe or progressive tachypnea. Absent or markedly decreased breath sounds. oxygen saturation less than 90%. Shock. Traumatic cardiac arrest without obvious fatal wounds. (consider immediate placement of bilateral SPEAR™ decompression needles and consider ALL other possible causes).
14
Objective 2: contraindications
contraindications: The S P E A R™ is not intended for treatment of simple pneumothorax or hemothorax efficacy of the S P E A R™ has not been established in pediatric patients Contraindications: The S P E A R™ is not intended for treatment of simple pneumothorax or hemothorax. Efficacy of the S P E A R™ has not been established in pediatric patients.
15
thoracic decompression should improve one or more of the following:
Objective 2: therapeutic benefits thoracic decompression should improve one or more of the following: respiratory distress relief of restrictive pressure between the parietal and visceral pleura (secondary to injury or significant medical complication) oxygen saturation (≥ 90% may be dependent on use of supplemental oxygen) return of radial pulse or vital signs thoracic decompression should improve one or more of the following: respiratory distress. Relief of restrictive pressure between the parietal and visceral pleura (secondary to injury or significant medical complication). oxygen saturation (≥ 90% may be dependent on use of supplemental oxygen). Return of radial pulse or vital signs.
16
Objective 3: lateral insertion site
numbered Intercostal spaces 5 4 3 2 1 5 4 3 2 1 6 5 4 3 2 1 numbered ribs 6 5 4 3 2 1 Lateral land-marking for thoracic decompression is critical for patient safety. Note that thoracic decompression may be relieved at either the 4th or 5th intercostal space (consult your protocol or standards for specific verification). IMPORTANT POINT: The images presented on this slide depict intercostal spaces and ribs that are numbered for clinician clarification and discussion. Providers MUST ensure proper land-marking prior to SPEAR™ insertion. NOTE: Insertion site is superior to the rib that is initially targeted. THIS IS A CRITICAL STEP to ensure proper placement location, avoidance of neurovascular bundle beneath each rib, and represents a known, safe starting position, from which the provider can control precise placement and insertion depth. 4th intercostal space - anterior axillary line 5th intercostal space - anterior axillary line site located superior to 5th rib site located superior to 6th rib
17
identify 4th or 5th intercostal space
Objective 3: lateral insertion insertion site identification anterior-axillary mid-axillary 4th ICS posterior-axillary safe zone mark & cleans site area over 4th & 5th intercostal spaces anterior axillary line The insertion sites commonly accepted for lateral decompression include the 4th or 5th intercostal space on the anterior axillary line (traditionally, insertion has also been viewed as acceptable between the anterior and midaxillary lines. Consult protocol or standards for verification). Once site has been correctly identified, it should be thoroughly cleansed according to protocol and applicable standards of care. identify 4th or 5th intercostal space 5th ICS anterior axillary line (or between anterior and mid axillary lines)
18
site verification and framing site reverification insertion site is
4th ICS Once site has been properly cleansed, preparations for SPEAR™ insertion can be completed. Note site “framing” and “reverification” in the image provided. Framing tightens the skin to improve penetration. Site reverification (of the targeted rib) ensures that the SPEAR™ needle tip will FIRST be placed against the rib. insertion site is superior to targeted rib 5th ICS Objective 3: lateral insertion
19
needle set stabilization improves precision placement and
4th ICS needle set stabilization improves precision placement and depth control During initial penetration through the skin, it is imperative that the tissue be stretched and that the provider is targeting the rib BELOW the intended insertion site. NOTE: In the provided image, the clinician is stabilizing the needle set and will provide DEPTH CONTROL with their index finger. 5th ICS Objective 3: lateral insertion
20
Objective 3: lateral insertion
insertion site framing needle set must remain perpendicular to chest wall during insertion insertion site is superior to targeted rib Once the SPEAR™ has penetrated the skin and adipose tissue, it should be firmly placed against the rib (this is a critical identification step to ensure patient safety). When the SPEAR™ needle tip has been placed against the targeted rib, the provider should MARK and HOLD ≤ 3 centimeters above the skin. Depth marking is critical to the next phase of insertion (which is entry of the needle set within the pleural space). hand stabilization needle set stabilization with finger tip depth control Objective 3: lateral insertion
21
needle set stabilization with finger tip depth control
Illustration on this slide is for additional clarity of previous point. Once the SPEAR™ has penetrated the skin and adipose tissue, it should be firmly placed against the rib (this is a critical identification step to ensure patient safety). Once the SPEAR™ needle tip has been placed against the targeted rib, the provider should MARK and HOLD ≤ 3 centimeters above the skin. Depth marking is critical to the next phase of insertion (which is entry of the needle set within the pleural space). insertion site is superior to targeted rib Objective 3: lateral insertion
22
needle set stabilization with finger tip depth control
needle set must remain perpendicular to chest wall during insertion KEY POINT: stabilization and control target rib and precisely penetrate chest wall to prevent further injury With ≤ 3 centimeters identified, the provider must carefully move the SPEAR™ needle tip superiorly over the targeted rib and guide the assembly into the thoracic cavity. Images on this slide depict two methods. Note that each image in this slide demonstrates the SPEAR™ being advanced a specific distance with “finger tip control.” This step is safely accomplished by pinning ≤ 3 centimeters above the tissue WHILE the needle tip is against the rib. Note that the provider ASSURES proper insertion depth is achieved (and NOT exceeded). Critical Point: Ensure that SPEAR™ placement is accomplished with the needle set perpendicular to the thoracic wall. It is well established that some experienced providers can appreciate (feel) needle set penetration of the parietal pleura (thus indicating the appropriate initial needle set penetration depth). Regardless of depth control method - DO NOT “BURY” or fully extend ANY decompression needle into the thoracic cavity. *Air under pressure may release from the thoracic cavity at this point. insert SPEAR™ superiorly over targeted rib insert ≤ 3 cm into pleura cavity Objective 3: lateral insertion DO NOT “BURY” OR “HUB” NEEDLE SET
23
insert superiorly over targeted rib insert ≤ 3 cm into pleura cavity
Objective 3: lateral insertion KEY POINT: stabilization and control target rib and precisely penetrate chest wall to prevent further injury The illustration on this slide is for additional clarity of previous point. With ≤ 3 centimeters identified, the provider must carefully move the SPEAR™ needle tip superiorly over the targeted rib and guide the assembly into the thoracic cavity. Illustrations on this slide depict needle set entry over the targeted rib. Note that this image shows the SPEAR™ advanced the correct distance with “finger tip control.” This step was safely accomplished by pinning ≤ 3 centimeter above the tissue WHILE the needle tip was against the rib. Note that the provider ASSURED proper insertion depth was achieved (and NOT exceeded). Critical Point: Ensure that SPEAR™ placement is accomplished with the needle set perpendicular to the thoracic wall. It is well understood and established that some experienced providers can appreciate needle set penetration of the parietal pleura (thus indicating appropriately achieved depth). Regardless of depth control method - DO NOT “BURY” or fully extend ANY decompression needle completely into the thoracic cavity. *Air under pressure may release from the thoracic cavity at this point. insert superiorly over targeted rib insert ≤ 3 cm into pleura cavity
24
Objective 3: lateral insertion
direct catheter toward middle of clavicle Prepare to advance catheter while retaining needle in stationary position release needle from catheter release spin-lock ¼ turn Once the SPEAR™ has been placed into the thoracic cavity, DIRECT the assembly toward the middle of the clavicle. Release the needle from the catheter by rotating the spin-lock ¼ turn. (This is a standard medical connection or “twist off” maneuver). Critical Point: Ensure that SPEAR™ placement is accomplished with the needle set perpendicular to the thoracic wall. *Air under pressure may release from the thoracic cavity at this point. adjust needle set angle cephalad Objective 3: lateral insertion
25
in stationary position
thread catheter toward middle of clavicle advance catheter while retaining needle in stationary position Once catheter has been released from the needle, advance the catheter while keeping the needle stationary as a guide. Critical Point: Ensure that SPEAR™ placement is accomplished with the needle set perpendicular to the thoracic wall. *Air under pressure may release from the thoracic cavity at this point. advance catheter needle set must remain stationary as a catheter guide needle set must remain at 90° angle to thoracic cavity Objective 3: lateral insertion
26
completely thread catheter into pleural cavity thread catheter
Objective 3: lateral insertion completely thread catheter into pleural cavity thread catheter toward middle of clavicle The illustrations on this slide are for additional clarity of previous point. Once catheter has been released from the needle, advance the catheter while keeping the needle stationary as a catheter guide. *Air under pressure may release from the thoracic cavity at this point. catheter released from spin-lock advance catheter while maintaining needle in stationary position
27
Objective 3: lateral insertion
Advance the catheter completely into the pleural cavity. Critical Point: Ensure that SPEAR™ placement is accomplished with the needle set perpendicular to the thoracic wall. *Air under pressure may release from the thoracic cavity at this point. advance catheter completely into pleural cavity maintain needle in stationary position
28
remove one-way valve from proximal end of needle hub
secure needle in bio-hazard sharps container Remove one-way valve from proximal end of needle hub. The one-way valve is attached with a slip-tip and can simply be pinched off. Secure the needle in a bio-hazard sharps container. *Air under pressure may release from the thoracic cavity at this point. Objective 3: lateral insertion
29
attach one-way valve to proximal end of catheter (standard lock)
Objective 3: lateral insertion catheter and one-way valve assemble may be attached to low pressure suction as needed secure catheter according to local protocol and standards Attach the one-way valve to the catheter hub (standard medical connection). Secure the catheter according to local protocol and standards. The catheter and one-way valve assembly may be attached to low pressure suction as needed. Monitor the patient. attach one-way valve to proximal end of catheter (standard lock)
30
catheter and one-way valve assemble may
be attached to low pressure suction as needed The illustration on this slide is for additional clarity of previous point. Attach the one-way valve to the catheter hub (standard medical connection). Secure the catheter according to local protocol and standards. The catheter and one-way valve assembly may be attached to low pressure suction as needed. Monitor the patient.
31
improvement of respiratory distress
Objective 5: indications of successful thoracic decompression successful thoracic decompression may have occurred if one or more of the following is observed: improvement of respiratory distress relief of air from catheter or one-way valve (valve may produce auditory signal) improvement of oxygen saturation (≥ 90% may be dependent on use of supplemental oxygen) return of radial pulse or vital signs Successful thoracic decompression may have occurred if one or more of the following is observed: Improvement of respiratory distress. Relief of air from catheter or one-way valve (valve may produce auditory signal). Improvement of oxygen saturation (≥ 90% may be dependent on use of supplemental oxygen). Return of radial pulse or vital signs.
32
IS THIS A CIRCULATORY PROBLEM?
Objective 6: complications following thoracic decompression procedure, continually assess patient for complications: Hemodynamic instability Respiratory distress Unilateral chest expansion Decreased oxygen saturation Bleeding Catheter occlusion Hematoma if two needle decompression attempts fail to relieve condition consider other causes and potential treatments IS THIS A CIRCULATORY PROBLEM? A patient should be monitored following any medical procedure. Thoracic decompression (by definition) is a life-saving procedure that is NOT without risk. With continuous training, as well as safeguards and careful considerations, patient and provider risk can be mitigated. Following thoracic decompression, continually assess the patient for complications. These should include (but not be limited to): Hemodynamic instability Respiratory distress Unilateral chest expansion Decreased oxygen saturation Bleeding Catheter occlusion Hematoma If two needle decompression attempts fail to relieve the suspected problem consider other causes and potential treatments. IS THIS A CIRCULATORY PROBLEM?
33
potential adverse complications: of improper thoracic decompression
Objective 6: complications potential adverse complications: of improper thoracic decompression Death secondary to cardiac penetration Lung injury Vascular injury Nerve damage Pain Numbness Paralysis of intercostal muscle Infection As with all invasive medical devices, potential adverse complications are a risk that MUST be considered. Most often, complications are the result of improperly performed interventions. Thoracic decompression is NOT WITHOUT RISK. The FIRST step to avoiding injury is a complete understanding of the indications and contraindications for SPEAR™ usage. Each of the listed complications SHOULD be discussed WITH specific mitigation strategies integrated into the training program. Complications are bolded in BLACK Mitigation strategies presented in GRAY Potential complications include (but are not limited to): Death secondary to cardiac penetration (clinician should control thoracic penetration and restrict depth of needle set insertion) Lung injury (insertion of needle set should be ≤ 3 cm past the parietal pleura) Vascular injury (1. insertion site restrictions to the specific locations identified in this training program. 2. maintain proper needle set angle during insertion. 3. avoidance of needle set over penetration. Nerve damage (ensure proper insertion site location “superior to targeted rib”) Pain (if time permits, consider local anesthetic per protocol or standard) Numbness (avoid insertion on the inferior aspect of the rib) Paralysis of intercostal muscle (avoid insertion on the inferior aspect of the rib) Infection (use aseptic technique)
34
Scientific Evidence The S P E A R™ decompression needle system, and this presentation, were developed utilizing the latest published evidence, independent research, and the support of dedicated Military and Civilian medical professionals in Emergency Medicine, Trauma Surgery, Pulmonology, Radiology, and Pathology. Clinical providers, regardless of their position, must dedicate themselves to the unrelenting truth that critical care is an evolution on behalf of those in need. Butler F, Holcomb J, Shackelford S, et al. Management of the Suspected Tension Pneumothorax in Tactical Combat Care, TCCC Guidelines Change J Spe Op Med. 2018; 18: *The aforementioned publication references ninety-six additional papers worthy of careful review. The S P E A R™ decompression needle system, and this presentation, were developed utilizing the latest published evidence, independent research, and the support of dedicated Military and Civilian medical professionals in Emergency Medicine, Trauma Surgery, Pulmonology, Radiology, and Pathology. Clinical providers, regardless of their position, must dedicate themselves to the unrelenting truth that critical care is an evolution on behalf of those in need. Butler F, Holcomb J, Shackelford S, et al. Management of the Suspected Tension Pneumothorax in Tactical Combat Care, TCCC Guidelines Change J Spe Op Med. 2018; 18: *The aforementioned publication references ninety-six additional papers worthy of careful review.
35
Objective 7: scientific evidence
Nine Key Facets of Tactical Combat Care Guidelines Change 17-02 Continuation of aggressive approach to suspecting and treating tension pneumothorax Emphasis of bilateral decompression in traumatic arrest Addition of 10 Gauge catheter (length indicated in current guidelines differs from the S P E A R™) Designates either Lateral or Anterior sites as acceptable for thoracic decompression Addition of procedural elements (critical procedural differences are included within this material) Defines successful thoracic decompression Recommends ONLY two needle decompressions be attempted before moving on to circulation Addition of materials that recommend consideration of tension pneumothorax in presentations of shock Addition of finger thoracostomy (if presentation warrants - following two unsuccessful needle decompression attempts - and provider is trained) There are nine Key Facets to thoracic decompression in the Tactical Combat Care Guidelines Change that should be considered: Slide numbers correspond to Lateral Insertion Program Continuation of aggressive approach to suspecting and treating tension pneumothorax. There is discussion and some evidence that unwarranted needle decompressions have presented complications. While this is a valid concern, failure to recognize and treat tension pneumothorax remains a lethal reality. SPEAR™ training materials list documented indications for thoracic decompression. Emphasis of bilateral decompression in traumatic arrest. See slide #13. Failure to completely address ALL possible causes of traumatic arrest continues to present as a lethal concern. See slides #13 & 32. Adds 10 Gauge. Length indicated in guidelines differs from the S P E A R™. Key points are found within the notes on slides #5 which offer details and specific clinical rationale for SPEAR™ length. Designates either Lateral or Anterior sites as acceptable for thoracic decompression. SPEAR™ training materials [slides, video, and NAR recommended hands-on skills training] comprehensively review [right and left] lateral and anterior thoracic decompression sites. Addition of procedural elements. SPEAR™ insertion differs significantly from that which is included within current TCCC guidelines and some institutions. See SPEAR™ procedural insertion techniques, for lateral and anterior sites, to obtain complete details and specific clinical rationale. Defines successful thoracic decompression. SPEAR™ training materials identify key components of successful thoracic decompression. See slides # 15 & 31. Recommends ONLY two needle decompressions be attempted before moving on to circulation. See slides #13 & 32. Addition of materials that recommend consideration of tension pneumothorax in presentations of shock. See slide #13. Addition of finger thoracostomy when presentation warrants. Provider has attempted two needle decompressions without success, and is authorized to perform procedure. See slides #32. Butler F, Holcomb J, Shackelford S, et al. Management of the Suspected Tension Pneumothorax in Tactical Combat Care, TCCC Guidelines Change J Spe Op Med. 2018; 18:
36
S P E A R™ decompression needle system
For additional information about the Tel: Mail: 35 Tedwall Court Greer, SC Fax: Thank you for participating in the SPEAR™ decompression needle system orientation and training. If you have any questions, concerns or would like additional information please contact: North American Rescue®, LLC Tel: Mail: 35 Tedwall Court Greer, SC Fax: S P E A R™ decompression needle system Simplified Pneumothorax Emergency Air Release