NIST validation studies on the 3500 Genetic Analyzer

Similar presentations


Presentation on theme: "NIST validation studies on the 3500 Genetic Analyzer"— Presentation transcript:

1 NIST validation studies on the 3500 Genetic Analyzer
Erica L.R. Butts, Margaret C. Kline, David L. Duewer, Carolyn R. (Becky) Hill, John M. Butler, Peter M. Vallone  Forensic Science International: Genetics Supplement Series  Volume 3, Issue 1, Pages e184-e185 (December 2011) DOI: /j.fsigss Copyright © Terms and Conditions

2 Fig. 1 STR typing results for the D7S820 locus with the Identifiler multiplex PCR kit showing dropout of the 8 allele when using (a) a single analytical threshold of 140 RFU across all dyes. (b) The use of dye-specific analytical thresholds allows for the 8 allele to be called in a peak height ratio of 47%. Forensic Science International: Genetics Supplement Series 2011 3, e184-e185DOI: ( /j.fsigss ) Copyright © Terms and Conditions


Download ppt "NIST validation studies on the 3500 Genetic Analyzer"