1 New pre study The setup results

2 Charge Preamplifier board and set-up
Side A New pre Old pre HV regulator B side

3 The setup

4 Raw energy distribuion

5 Results Old pre-amp signal/noise New pre-amp signal/noise
In our setup we gain on S/N a factor 2.6 without shaping. If we had a 12 mV peak Beam) from cosmic this would bring our ENE down to 1.5 MeV Single APD no shaping.

6 Single event study 1 MHz frequency cut Old pre amp New pre amp

7 Single event study In red pre-amp signal In green 100 kHz cut
In blue kHz cut In black 1Mhz cut

8 What happens if I apply the 200 kHz cut?
Brickwall filter sharp 200kHz cut After 200 kHz cut we get a signal to noise ratio better than 20 (1.5 MeV single apd)

9 What about butterworth filter ?
Sharp cut on frequency (200 kHz) N=2 Situation improves a little bit we are close to 1 MeV on a single APD!