1 From: Temperature Monitoring and Perioperative Thermoregulation
Anesthes. 2008;109(2): doi: /ALN.0b013e31817f6d76 Figure Legend: Fig. 1. The differences between the tympanic membrane thermocouple (Mon-a-therm, St. Louis, MO) and aural canal temperature measured by a Quickthermo infrared thermometer (Mie, Japan). The mean difference between core temperature and the infrared monitor was 1.1°C. Three other infrared monitors were evaluated in this study, but none proved sufficiently accurate for clinical use. From Imamura  et al.  15; used with permission from Wiley-Blackwell Publishing Ltd.  Date of download: 11/5/2017 Copyright © 2017 American Society of Anesthesiologists. All rights reserved.