From: Continuous Measurement of Glucose:Facts and Challenges
Anesthes. 2012;116(1): doi: /ALN.0b013e318236abf6 Figure Legend: Fig. 1. A Bland-Altman plot using repeated measures to present the agreement between the continuous glucose-monitoring system and reference, with the data points of individual patients connected by a line. The limits of agreement are −21 mg/dl and 25 mg/dl. CGMS = continuous glucose-monitoring system. (Reproduced with permission from Brunner R, Kitzberger R, Miehsler W, Herkner H, Madl C, Holzinger U: Accuracy and reliability of a subcutaneous continuous glucose-monitoring system in critically ill patients. Crit Care Med 2011; 39:659–64.) Date of download: 10/17/2017 Copyright © 2017 American Society of Anesthesiologists. All rights reserved.