Filmless MLC Quality Assurance in Robotic Radiosurgery
Presented at the 2016 Radiosurgery Society Scientific Meeting Jacob A. Gersh, PhD DABR Filmless MLC Quality Assurance in Robotic Radiosurgery
2
Presented at the 2016 Radiosurgery Society Scientific Meeting
Disclosures Speaker is the owner and president of Spectrum Medical Physics, LLC, which has contracts or financial interest with: Standard Imaging Inc. Siemens Medical Integrated Medical Technologies Speaker received financial reimbursement for travel associated with the presentation presented herein (Standard Imaging Inc).
3
QA StereoCheckerTM Filmless CyberKnife® QA System
Presented at the 2016 Radiosurgery Society Scientific Meeting QA StereoCheckerTM Filmless CyberKnife® QA System High-Resolution Flat-Panel Imager 200 micron Resolution High Frame Rate Portable and Compact
4
QA StereoCheckerTM Aligned using implanted high-Z markers
Presented at the 2016 Radiosurgery Society Scientific Meeting QA StereoCheckerTM Aligned using implanted high-Z markers Automated alignment using treatment localization system Fast and consistent setup No manual manipulation of robot No attachment to collimator housing Designed to be used by physicists and therapists
5
The InCiseTM MLC Available for CyberKnife M6 26 leaf pairs
Presented at the 2016 Radiosurgery Society Scientific Meeting The InCiseTM MLC Available for CyberKnife M6 26 leaf pairs 0.385 mm wide at 800 mm Full over-travel Allows for treatment of larger lesions 100 mm x 115 mm Leaf Position Accuracy ≤0.95 mm
6
MLC QA – The Garden Fence Test
Presented at the 2016 Radiosurgery Society Scientific Meeting MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film-based tests Replaces monthly Garden Fence Tests Capable of detecting 0.2 mm leaf variations
7
MLC QA – The Garden Fence Test
Presented at the 2016 Radiosurgery Society Scientific Meeting MLC QA – The Garden Fence Test
8
MLC QA – The Garden Fence Test
Presented at the 2016 Radiosurgery Society Scientific Meeting MLC QA – The Garden Fence Test
9
MLC QA – The Garden Fence Test
Presented at the 2016 Radiosurgery Society Scientific Meeting MLC QA – The Garden Fence Test
10
MLC QA – The Garden Fence Test
Presented at the 2016 Radiosurgery Society Scientific Meeting MLC QA – The Garden Fence Test
11
MLC QA – The Garden Fence Test
Presented at the 2016 Radiosurgery Society Scientific Meeting MLC QA – The Garden Fence Test
12
MLC QA – The Garden Fence Test
Presented at the 2016 Radiosurgery Society Scientific Meeting MLC QA – The Garden Fence Test
13
MLC QA – The Garden Fence Test
Presented at the 2016 Radiosurgery Society Scientific Meeting MLC QA – The Garden Fence Test 0.1mm mm mm mm mm mm mm Induced Individual leaf errors 0.1 mm to 1.0 mm Increments of 0.1 mm “Detection” occurs if a cluster of pixels is simultaneously above a threshold with SNR above 10. Mean pixel difference values are used as “Warning” and “Error Thresholds” as well as for trend analysis.
14
Thank You J. David Myers, CMD Michael Starnes, CMD
Presented at the 2016 Radiosurgery Society Scientific Meeting J. David Myers, CMD Michael Starnes, CMD Ryan S. Dubose, MS DABR Kevin L. Shay, MS DABR Daniel B. Fried, MD PHD Melanie Rhyne, RT Abby Sammut, RT Lauren Miles, RT LinLee Khoune, RT Thank You
15
Presented at the 2016 Radiosurgery Society Scientific Meeting
Questions?