Presentation is loading. Please wait.

Presentation is loading. Please wait.

OVERVIEW ANALYTICAL SAMPLES AND METHODS

Similar presentations


Presentation on theme: "OVERVIEW ANALYTICAL SAMPLES AND METHODS"— Presentation transcript:

1 ANALYTICAL PROPERTIES PART III ERT 207 ANALYTICAL CHEMISTRY SEMESTER 1, ACADEMIC SESSION 2017/18

2 OVERVIEW ANALYTICAL SAMPLES AND METHODS
STANDARDIZATION AND CALIBRATION EXTERNAL STANDARD CALIBRATION INTERPRETATION OF LEAST-SQUARE RESULTS TRANSFORMED VARIABLES QUALITY ASSURANCE OF ANALYTICAL RESULTS

3 ANALYTICAL SAMPLES AND METHODS
Macro analysis: Is used for samples whose masses are > 0.1 g. Semimicro analysis: Is performed on samples in the range of 0.01 to 0.1g. Micro analysis: Is used for samples whose mass is 10-4 – 10-2g. Ultramicro analysis: Is used for samples whose mass is < 10-4g.

4 ANALYTICAL SAMPLES AND METHODS
Major constituents: Those present in 1 – 100% by mass. Minor constituents: Species present in % by mass. Trace constituents: Those present in amount between 100 ppm (0.01%) and 1 ppb. Ultratrace constituents: Those present in amounts < 1 ppb.

5 ANALYTICAL SAMPLES AND METHODS
Fig 1: Classification of analytes by sample size Fig 2: Classification of constituent types by analyte level

6 ANALYTICAL SAMPLES AND METHODS
Note: Samples are analysed, but species or concentrations are determined. Determination of glucose in blood serum analysis or the analysis of blood serum for glucose concentration determination

7 ANALYTICAL SAMPLES AND METHODS
The composition of the gross sample and the laboratory sample must closely resemble the average composition of the total mass of material to be analyzed. The items chosen for analysis are often called as sampling units or sampling increments. In sampling, a sample population is reduced in size to an amount of homogeneous material that can be conveniently handled in the laboratory and whose composition is representative of the population.

8 ANALYTICAL SAMPLES AND METHODS
Statistically, the goals of the sampling process are: To obtain a mean analyte concentration that is an unbiased estimate of the population mean. It can be realized if all members of the population have an equal probability of being included in the sample. To obtain variance in the measured analyte concentration that is an unbiased estimate of the population variance so that valid confidence limits can be found for the mean, and various hypothesis tests can be applied. It can be realized if every possible sample is equally likely to be drawn.

9 ANALYTICAL SAMPLES AND METHODS
The number of particles required in a gross sample ranges from a few particles to particles. Based on the Bernoulli equation: The standard deviation of the number of A particles drawn: The relative standard deviation of drawing A type particles: Probability of randomly drawing an A type particles

10 ANALYTICAL SAMPLES AND METHODS
The number of particles needed to achieve a given relative standard deviation: If 80% of the particles are type A (p=0.8) and the desired relative standard deviation is 1% (σr=0.01), the number of particle making up the gross sample should be:

11 ANALYTICAL SAMPLES AND METHODS
In reality: The type A particles contain a higher percentage of analyte, PA and the type B particles a lesser amount, PB. The average density d of the particles differs from the densities dA and dB of these components. Overall average percent of active ingredient (%)

12 ANALYTICAL SAMPLES AND METHODS
Rearrange the equation: The relative standard deviation: If we make the assumption that the sample mass m is proportional to the number of particles and the other quantities in the equation are constant, the product of m and σ, should be a constant. σr x 100% = % relative standard deviation

13 ANALYTICAL SAMPLES AND METHODS
When σr = 0.01, σr x 100% = 1%, Ks = m. Ks is the minimum sample mass required to reduce the sampling uncertainty to 1%.

14 ANALYTICAL SAMPLES AND METHODS
EXAMPLE 1 A column packing material for chromatography consists of a mixture of two types of particles. Assume that the average particle in the batch being sampled is approximately spherical with a radius of about 0.5 mm. Roughly 20% of the particles appear to be pink in color and are know to have about 30% by mass of a polymeric stationary phase attached (analyte).

15 ANALYTICAL SAMPLES AND METHODS
The pink particles have a density of 0.48 g/cm3. The remaining particles have a density of about g/cm3 and contain little or no polymeric stationary phase. What mass of the material should the gross sample contain if the sampling uncertainty is to be kept below 0.5% relative?

16 ANALYTICAL SAMPLES AND METHODS
Number of laboratory samples: How many samples should be taken for the analysis? The number of samples N,

17 ANALYTICAL SAMPLES AND METHODS
EXAMPLE 2. The determination of copper in a seawater sample gave a mean value of μg/L and a standard deviation ss of 1.74 μ g/L. How many samples must be analysed to obtain a relative standard deviation of 1.7% in the results at the 95% confidence level?

18 STANDARDIZATION AND CALIBRATION
It determines the relationship between the analytical response and the analyte concentration. The relationship is usually determined by the use of chemical standards. Interference could be reduced from other constituents in the sample matrix, called concomitants by using standards added to the analyte solution or by matrix matching or modifications.

19 STANDARDIZATION AND CALIBRATION
The absolute method (e.g. gravimetric method) do not rely on calibration with chemical standards. Comparison with standards: 2 types of comparison methods: Direct comparison techniques Titration procedures Null camparison or isomation methods: Comparison a property of the analyte with standards such that the property being tested matches or nearly matches that of the standard.

20 STANDARDIZATION AND CALIBRATION
With some modern instruments, a variation of this procedure is used to determine if an analytes concentration exceeds or is less than some threshold level. Comparator can be used to indicate that the threshold has been exceeded. Titration is a type of chemical comparison. The amount of the standardized reagent needed to achieve chemical equivalence can be related to the amount of analyte present by stoichiometry.

21 EXTERNAL STANDARD CALIBRATION
A series of standard solutions is prepared separately from the sample. The standards are used to establish the instrument calibration function. It is obtained from analysis of the instrument response as a function of the known analyte concentration. Ideally, 3 or more standard solutions are used in the calibration process, although in some routine determinations, 2 point calibration can be reliable.

22 EXTERNAL STANDARD CALIBRATION
The calibration function can be obtained graphically or in mathematical form. A plot of instrument response versus known analyte concentrations is used to produce a calibration curve, sometimes called a working curve. It is often desirable that the calibration curve be linear in at least the range of the analyte concentrations.

23 EXTERNAL STANDARD CALIBRATION
Fig. 1: Calibration curve of absorbance versus analyte concentration for a series of standards.

24 EXTERNAL STANDARD CALIBRATION
The linear relationship is then used to predict the concentration of an unknown analyte solution. (1) The Least-squares Method The investigator must try to draw the ‘best’ straight line among the data points. Regression analysis provides the means for objectively obtaining such a line and also for specifying the uncertainties associated with its subsequent use.

25 EXTERNAL STANDARD CALIBRATION
Assumptions of the least- square method: There is actually a linear relationship between the measured response y and the standard analyte concentration x. The mathematical relationship that describes this assumption is called the regression model (y = mx + c). Fig. 2: The slope-intercept form of a straight line.

26 INTERPRETATION OF LEAST-SQUARE RESULTS
The closer the data points are to the line predicted by a least-square analysis, the smaller are the residuals. The sum of the squares of the residuals, SSresid, measures the variation in the observed values of the dependent variables (y values) that are not explained by the presumed linear relationship between x and y:

27 INTERPRETATION OF LEAST-SQUARE RESULTS
A total sum of the squares, SStot is a measure of the total variation in the observed values of y since the deviations are measured from the mean value of y. The coefficient of determination (R2) measures the fraction of the observed variation in y that is explained by the linear relationship:

28 INTERPRETATION OF LEAST-SQUARE RESULTS
The closer R2 is to unity, the better the linear model explains the y variations. The difference between SStot and SSresid is the sum of the squares due to regression, SSreg. In contrast to SSresid, SSreg is a measure of the explained variation. &

29 INTERPRETATION OF LEAST-SQUARE RESULTS
EXAMPLE 3 Find the coefficient of determination for the chromatographic data below. The least-square line: SStot = Mole percent isooctane, xi Peak area, yi 0.352 1.09 0.803 1.78 1.08 2.60 1.38 3.03 1.75 4.01

30 TRANSFORMED VARIABLES
Transformations to linearize function: Linear least squares gives best estimates of the transformed variables, but these may not be optimal when transformed back to obtain estimates of the original parameters. Nonlinear regression methods may give better estimates.

31 QUALITY ASSURANCE OF ANALYTICAL RESULTS
Control chart: A sequential plot of some characteristic that is a criterion of quality (quality assurance). It shows the statistical limits of variation that are permissible for the characteristic being measured. Upper control limit: Lower control limit: μ = population mean σ = population standard deviation

32 QUALITY ASSURANCE OF ANALYTICAL RESULTS
Fig 3: A control chart for a modern analytical balance. For example, from independent experiments, estimates of the population mean and standard deviation were found to be μ = g and σ = g, respectively.

33 QUALITY ASSURANCE OF ANALYTICAL RESULTS
The mean of 5 measurement, = UCL = g & LCL = g. (see Fig 3). As long as the mean mass remains between the LCL and the UCL, the process is said to be in statistical control.

34 QUALITY ASSURANCE OF ANALYTICAL RESULTS
Fig. 4 shows the results of 89 production runs of a cream containing a nominal 10% benzoyl peroxide measured on consecutive days. Each sample is represented by the mean percent benzoyl peroxide determined from the results of five titrations of different analytical samples of the cream. The chart shows that, until day 83, the manufacturing process was in statistical control with normal random fluctuations in the amount of benzoyl peroxide.

35 QUALITY ASSURANCE OF ANALYTICAL RESULTS
Fig 4: A control chart for monitoring the concentration of benzoyl peroxide in a commercial acne preparation On day 83, the system went out of control with a dramatic systematic increase above the UCL.

36 QUALITY ASSURANCE OF ANALYTICAL RESULTS
This increase caused considerable concern at the manufacturing facility until its source was discovered and corrected. These examples show how control charts are effective for presenting quality control data in real world problems.

37 EXAMPLE 3 Finding the sum of the squares of the residuals:
SStot = , = xi yi ẏi yi-ẏi (yi-ẏi)2 0.352 1.09 0.803 1.78 1.08 2.60 1.38 3.03 1.75 4.01 5.365 12.51


Download ppt "OVERVIEW ANALYTICAL SAMPLES AND METHODS"

Similar presentations


Ads by Google