3

4

5 Gomer, Marc

6 Gomer, Marc Gomer, Marc

7 Gomer, Marc

8 Cardiac pacing Temporary transcutaneous pacing for any unstable bradyarrythmia Anterolateral or AP pads Need cardiac leads if demand mode used Don’t forget sedation and analgesia if patient has awareness!

9 Check DEFIBRILLATOR has pacing mode!
Activate pacing mode Select demand or fixed mode Set rate 70bpm + output 1mA/kg (70mA) Press “Start pacing” Check your defib has a pacing function Must have cardiac leads for “sensing” in demand mode.

10 Check for electrical capture. by 10mA until electrical capture
Confirm mechanical capture: Add 10-15mA safety margin Signs of perfusion

11 CHANGE TO OTHER POSITION
No capture by 130mA? CHANGE TO OTHER POSITION