Presentation is loading. Please wait.

Presentation is loading. Please wait.

Workshop on Providing the traceability of measurements

Similar presentations


Presentation on theme: "Workshop on Providing the traceability of measurements"— Presentation transcript:

1 Workshop on Providing the traceability of measurements
Application practice of provisions GUM concerning calibration of measuring instruments Workshop on Providing the traceability of measurements in the national metrology institutes and accredited calibration and testing laboratories Kyiv, Ukraine – January 2011 Workshop on Providing the traceability of measurements

2 Uncertainty and Calibration
ISO/IEC 17025 A calibration laboratory, or a testing laboratory performing its own calibrations, shall have and apply a procedure to estimate the uncertainty of measurement for all calibrations and types of calibration. Workshop on Providing the traceability of measurements

3 Uncertainty and Calibration
When estimating the uncertainty of measurement, all uncertainty components which are of importance in the given situation shall be taken into account using appropriate methods of analysis. Workshop on Providing the traceability of measurements

4 Uncertainty and Calibration
Definitions Uncertainty is a parameter, associated with the result of a measurement, that characterizes the dispersion of the values that could reasonably be attributed to the measurand. Result of a measurement must always be a measurand and its uncertainty (for a stated interval of confidence). Can be expressed as absolute or relative. Workshop on Providing the traceability of measurements

5 Uncertainty and Calibration
Uncertainty of type A is obtained from the statistics of the measurement results. Uncertainty of type B comes from all other sources. Combined uncertainty combines type A and type B uncertainty. Expanded uncertainty transforms the combined uncertainty for the requested interval of confidence. Workshop on Providing the traceability of measurements

6 Uncertainty and Calibration
Type A is obtained from the statistics of the measurement results. Several (often n = 3) independent observations are needed. Observable scattre in the values obtained. Standard deviation of the mean (68% confidence interval): Xi are the measured values, X is the arithmetic mean of these values. Workshop on Providing the traceability of measurements

7 Uncertainty and Calibration
Type B is obtained from the remaining (other then statistical) sources, need not have normal distribution. Sources uncertainty of the used standard, uncertainty due to the resolution of the calibrated device, uncertainty due to the conditions (temperature) of calibration, Workshop on Providing the traceability of measurements

8 Uncertainty and Calibration
Inputs can be obtained from: calibration certificates of the used measuring instruments, manufacturer‘s specifications, reference data, previous measurement data and experience, They always must be renormalized to the normal distribution. The sensitivity coefficients c must be considered. For m non-correlated input uncertainties, we obtain: Workshop on Providing the traceability of measurements

9 Uncertainty and Calibration
Combined uncertainty Combines both types into one value of 68% confidence: Expanded uncertainty Transforms the result to the needed confidence interval by multiplying with a coverage factor: Mostly to 95 % for k = 2, sometimes to 99.7 % for k = 3. Workshop on Providing the traceability of measurements

10 Uncertainty and Calibration
Coverage factor Usually for 95% confidence interval it can be taken from table. veff 1 2 3 4 5 6 7 8 10 20 50 k 13,97 4,53 3,31 2,87 2,65 2,52 2,43 2,37 2,28 2,13 2,05 2,00 Workshop on Providing the traceability of measurements

11 Uncertainty and Calibration
Welch-Satterthwait equation ui (i = 1, 2, …, N) - uncertainty of determination of error from input xi (considered non correlated), vi = n – 1 - degrees of freedom for standard uncertainty of type A, for standard uncertainty determined by method type B assume vi → ∞. Calculated value of veff is rounded to the nearest smaller value given in table. Workshop on Providing the traceability of measurements

12 Uncertainty and Calibration
Reporting the results Uncertainty should have one or two significant digits. If it begins with the number 1 or 2, then two significant digits. Measurand must be rounded to the same position as the uncertainty The coverage factor must be reported. Workshop on Providing the traceability of measurements

13 Measurement results and uncertainties calibration of a NAWI
Examples Measurement results and uncertainties calibration of a NAWI Workshop on Providing the traceability of measurements

14 Workshop on Providing the traceability of measurements
Examples Errors of indication – subject of the calibration Discrete values Characteristic of the weighing range In addition, or as an alternative to the discrete values a characteristic, or calibration curve may be determined for the weighing range, which allows to estimate the error of indication for any indication I within the weighing range Workshop on Providing the traceability of measurements

15 Workshop on Providing the traceability of measurements
Examples Discrete values For each test load, the error of indication is calculated as Ej = Ij – mref,j Workshop on Providing the traceability of measurements

16 Examples Characteristic of the weighing range A function E = f(I)
may be generated by an appropriate approximation based on the “least squares” approach: vj² = (f(Ij) – Ej)² = minimum with vj = residual f = approximation function Workshop on Providing the traceability of measurements

17 Examples Uncertainty of the measurement
Standard uncertainty for discrete values The basic formula is E = I – mref with the variances u²(E) = u²(I) + u²(mref) Workshop on Providing the traceability of measurements

18 Examples I = IL + IdigL+ Irep + Iecc - I0 - Idig0
Uncertainty of the measurement Standard uncertainty of the indication I = IL + IdigL+ Irep + Iecc - I0 - Idig0 Workshop on Providing the traceability of measurements

19 Examples Uncertainty of the measurement
Standard uncertainty of the reference mass mref = mN + mc + mB +mD + mconv + m... Workshop on Providing the traceability of measurements

20 Workshop on Providing the traceability of measurements
Examples Uncertainty of the measurement Standard uncertainty of the error u²(E) = d0²/12 + dI²/12 + s²(I) + u²(Iecc) + u²(mc) + u²(mB) + u²(mD) + u²(mconv) Workshop on Providing the traceability of measurements

21 Workshop on Providing the traceability of measurements
Examples Uncertainty of the measurement Expanded uncertainty at calibration U(E) = ku(E) Workshop on Providing the traceability of measurements

22 Workshop on Providing the traceability of measurements
Examples Uncertainty of NAWI in use Standard uncertainty of a weighing result The user of an instrument should be aware of the fact that in normal usage of an instrument that has been calibrated, the situation is different from that at calibration in some if not all of these aspects: the indications obtained for weighed bodies are not the ones at calibration the weighing process may be different from the procedure at calibration: a. certainly only one reading for each load, not several readings to obtain a mean value b. reading to the scale interval d of the instrument, not with higher resolution c. loading up and down, not only upwards – or vice versa d. load kept on load receptor for a longer time, not unloading after each loading step – or vice versa e. eccentric application of the load, f. use of tare balancing device Workshop on Providing the traceability of measurements

23 Workshop on Providing the traceability of measurements
Examples Uncertainty of NAWI in use the environment (temperature, barometric pressure etc.) may be different on instruments which are not readjusted regularly e.g. by use of a built-in device, the adjustment may have changed, due to ageing or to wear and tear. Unlike the items 1 to 3, this effect is usually depending on the time that has elapsed since the calibration, it should therefore be considered in relation to a certain period of time, e.g. for one year or the normal interval between calibrations. It is therefore difficult to conclude from the results of a calibration as given in the calibration certificate: Workshop on Providing the traceability of measurements

24 Workshop on Providing the traceability of measurements
Examples Uncertainty of NAWI in use Uncertainty of the error of a reading Uncertainty from environmental influences Uncertainty from the operation of the instrument Workshop on Providing the traceability of measurements

25 Workshop on Providing the traceability of measurements
Examples Uncertainty of NAWI in use Standard uncertainty of a weighing result Workshop on Providing the traceability of measurements

26 Workshop on Providing the traceability of measurements
Examples Uncertainty of NAWI in use Repeatability (standard deviation of single reading) Workshop on Providing the traceability of measurements

27 Workshop on Providing the traceability of measurements
Examples Uncertainty of NAWI in use Digital rounding Workshop on Providing the traceability of measurements

28 Workshop on Providing the traceability of measurements
Examples Uncertainty of NAWI in use Temperature effect Workshop on Providing the traceability of measurements

29 Workshop on Providing the traceability of measurements
Examples Uncertainty of NAWI in use Temperature effect Workshop on Providing the traceability of measurements

30 Workshop on Providing the traceability of measurements
Examples Uncertainty of NAWI in use Temperature effect (coeffitient C ) Workshop on Providing the traceability of measurements

31 Workshop on Providing the traceability of measurements
Examples Uncertainty of NAWI in use Expanded uncertainty of a weighing result Errors accounted for by correction Errors included in uncertainty Workshop on Providing the traceability of measurements

32 Calibration of a digital manometer
Examples Calibration of a digital manometer Workshop on Providing the traceability of measurements

33 Workshop on Providing the traceability of measurements
Examples Calibrated digital pressure meter Measuring range: (0 to 35) kPa Resolution: 0,01 kPa Thermal error: 0,02 % FS/°C Reference temperature: 20 °C Temperature of enviroment: (20 ± 1) °C Used standard manometer Measuring range: (1,5 to 100) kPa Accuracy: 0,01 % MV Difference of reference levels of calibrated and measuring standart pressure meter is insignificant Workshop on Providing the traceability of measurements

34 Workshop on Providing the traceability of measurements
Examples Measured values Workshop on Providing the traceability of measurements

35 Workshop on Providing the traceability of measurements
Examples Components of uncertainty Repeatability Uncertainty of measuring standart pressure meter Resolution of calibrated pressure meter Thermal error of calibration pressure meter Workshop on Providing the traceability of measurements

36 Evalution of uncertainty of measurement
Standard uncertainty of type A: Uncertainty of the standard: Uncertainty of resolution of calibrated device:

37 Evaluting of uncertainty of measurement
Uncertainty of thermal error of the calibrated device: Standard uncertainty of type B: Combined standard uncertainty:

38 Expanded uncertainty of measurement
Effective degrees of freedom: Coverage factor: Expanded uncertainty of measurement:

39 Calibration of digital multimeter
Examples Calibration of digital multimeter Workshop on Providing the traceability of measurements

40 Calibration of digital multimeter
Definition of measurement Direct measurement Multimeter 6 ½ digit DC Voltage 10 V Workshop on Providing the traceability of measurements

41 Calibration of digital multimeter
Uncertainty type A Reading Meas value [ V ] 1 2 3 4 5 6 7 8 9 10 u A Reading n – number of reading r – average xi – single reading Average Workshop on Providing the traceability of measurements

42 Calibration of digital multimeter
Uncertainty type B Error of indication EX of calibrating multimeter is: EX = UDMM – UE + URES – USPEC where: UE is the output voltage of calibrator (standard) UDMM is average value indicated by calibrated DMM URESL is correction due to finite resolution of DMM USPEC is corection of output voltage of calibrator from his specification (included temp. depence, load depence, drift since last cal, linearity) Workshop on Providing the traceability of measurements

43 Calibration of digital multimeter
Output voltage of calibrator UE Output voltage of calibrator is 10, V Uncertainty from Calibration certificate of the calibrator for 10 V is ±72 µV for k = 2 and normal distribution

44 Calibration of digital multimeter
Resolution of DMM URES The least significant digit is 10 µV (10, V) Uncertainty of resolution is ±(5 µV / √3) = ±2,9 µV (rectangular distribution) Workshop on Providing the traceability of measurements

45 Calibration of digital multimeter
Specificatin of calibrator USPEC Range ppm from reading + ppm from range 200 mV 20 + 0, µV 2 V 12 + 4 20 V 7 + 0,5 200 V 20 + 4 1000 V 25 + 4

46 Calibration of digital multimeter
Specification of calibrator USPEC 10 V * 7E V * 0,5E-6 = 70E-6 V + 10E-6 V = 80E-6 V = 80 µV If not specified, we assume the rectangular distribution: 80 µV / √3 = 46,2 µV Workshop on Providing the traceability of measurements

47 Calibration of digital multimeter
The combine uncertainty Workshop on Providing the traceability of measurements

48 Calibration of digital multimeter
Expanded measurement uncertainty U = k * u(Ex) = 2 * 59,7 µV = 119,4 µV We assume normal distribution, than k = 2 for coverage probability of approximately 95%. The error of multimeter is: (0, ± 0,000 12) V The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k = 2, which for a normal distribution corresponds to a coverage probability of approximately 95%. Workshop on Providing the traceability of measurements

49 Expression of the uncertainty of Measurement in calibration
MICROMETR CALLIPER Expression of the uncertainty of Measurement in calibration Workshop on Providing the traceability of measurements

50 MICROMETER CALLIPER calibration method
The error of the micrometer calliper is determined by set of gauge blocks. There are recommended calibration points in the technical standard ISO 3611. The values are as follows: (2,5; 5,1; 7,7; 10,3; 12,9; 15; 20,2; 22,8; 25) mm Workshop on Providing the traceability of measurements

51 MICROMETER CALLIPER calibration method
Workshop on Providing the traceability of measurements

52 MICROMETER CALLIPER calibration method
Preparation (marking, cleaning, visual inspection, rework of slight damage) Preliminary tests and activities (smooth run, function of the locking device, surface, acclimatisation) Calibration (checking of the lower limit of the measuring range, determination of the errors) Workshop on Providing the traceability of measurements

53 MICROMETER CALLIPER calibration method
Workshop on Providing the traceability of measurements

54 MICROMETER CALLIPER input data
Range : (025) mm Reading : analog Division : 0.01 mm Standard : Set of gauge blocks Expanded uncertainty of the standard U = 0, k = 2 Temperature conditions : (20 ± 2) °C Workshop on Providing the traceability of measurements

55 MICROMETER CALLIPER input data
Measured values [mm]: 25,000; 25,001; 25,000; 25,002; 25,001 25,001; 25,002; 25,000; 25,000; 25,003 The result of the measurement is (corrected) arithmetical mean of these values Workshop on Providing the traceability of measurements

56 MICROMETER CALLIPER contributors to resulting uncertainty
To resulting uncertainty consider following influences: dispersion of the measured values uncertainty of the standard resolution of the calliper thermal expansion Workshop on Providing the traceability of measurements

57 MICROMETER CALLIPER basic equation - mathematical model
LK = A + E + M + L (a · Dt + Da · Dt20) LK ... corrected result of the maeasurement A arithmetic mean of the measured values E ... standard correction M ... reading correction L ... nominal length Workshop on Providing the traceability of measurements

58 MICROMETER CALLIPER basic equation - mathematical model
LK = A + E + M + L (a · Dt + Da · Dt20) a = (a1 + a2) / mean value of the expansion coefficient Dt = (t1 - t2) ... difference in material temperatures Da = (a1 - a2) ... difference in expansion coefficients Dt20 = 20 – (t1 + t2) / 2 deviation of the mean temperature from 20 °C Workshop on Providing the traceability of measurements

59 MICROMETER CALLIPER basic equation - mathematical model
Notes 1: For making out the mathematical model not only mathematical knowledge is sufficient Experience with particular measurement task is necessary Not enough experience leads to insufficient mathematical model Workshop on Providing the traceability of measurements

60 MICROMETER CALLIPER basic equation - mathematical model
Notes 2: Equation for resulting uncertainty is derived from the mathematical model. Only mathematical knowledge is needed for this derivation Workshop on Providing the traceability of measurements

61 MICROMETER CALLIPER resulting uncertainty equation
u =  (uA2 + uE2 + uM2 + + (L · a· uDt)2 + (L · uDt 20 · uD)2 ) uA ... type A uncertainty uE ... uncertainty of the standard uM... uncertainty of the reading L ... nominal length Workshop on Providing the traceability of measurements

62 MICROMETER CALLIPER resulting uncertainty equation
u =  (uA2 + uE2 + uM2 + + (L· a· uDt)2 + (L · uDt 20 · uD)2 ) a = (a1 + a2) / mean value of the expansion coefficient uDt uncertainty of the difference in material temperatures uDt uncertainty of the deviation of the mean temperature from 20 °C uD ... uncertainty of the difference in expansion coefficients Workshop on Providing the traceability of measurements

63 MICROMETER CALLIPER type A uncertainty
uA = s/√n s … experimental standard deviation n … number of measurement uA … standard uncertainty associated with the arithmetical mean Workshop on Providing the traceability of measurements

64 MICROMETER CALLIPER type A uncertainty
The measured values are as follows: (25,000; 25,001;25,000;25,002;25,001;25,001; 25,002; 25,000; 25,000; 25,003): Using scientific calculator or Excel sheet, the value of the experimental standard deviation is: s = 0,00105 mm = 1,05 µm Workshop on Providing the traceability of measurements

65 MICROMETER CALLIPER type A uncertainty
uA = s/n = 1,05/10 = 0,332 µm The value of the type A uncertainty is 0,332 µm By increasing of the number of measurement n, the type A uncertainty associated with the arithmetical mean can be decreased Workshop on Providing the traceability of measurements

66 MICROMETER CALLIPER uncertainty of the standard
Reported expanded uncertainty for the length 25 mm and coverage factor from the certificate: U = 0,62 µm coverage factor k = 2 The value of the standard uncertainty of the Standard for the length 25 mm is: uE = U/k = 0,62/2 = 0,31 µm Workshop on Providing the traceability of measurements

67 MICROMETER CALLIPER uncertainty of the reading
The division of the scale Workshop on Providing the traceability of measurements

68 MICROMETER CALLIPER uncertainty of the reading
Experienced operator is able to estimate result of the measurement with resolution 0,001 mm Assume rectangular distribution with bounds ± 1µm uM = 1/3 = 0,577 µm The value of the standard uncertainty of the reading is: 0,577 µm Workshop on Providing the traceability of measurements

69 MICROMETER CALLIPER thermal expansion
there are three facts: There is the difference in material temperatures of the calliper and gauge block There is the difference in expansion coefficients of the calliper and gauge block There is deviation of the mean temperature from the reference temperature (20 °C) Workshop on Providing the traceability of measurements

70 MICROMETER CALLIPER thermal expansion
There is the difference in material temperatures of the calliper and gauge block... Estimate the maximum difference in material temperatures of the calliper and gauge block to ± 2 °C Standard uncertainty of the difference: uDt = 2/3= 1,15 °C Workshop on Providing the traceability of measurements

71 MICROMETER CALLIPER thermal expansion
There is the difference in expansion coefficients of the calliper and gauge block Estimate the maximum difference in expansion coefficients of the calliper and gauge block to ± 2 µm/m°C Standard uncertainty of the difference: uDa = 2/3 = 1,15 µm/m°C Workshop on Providing the traceability of measurements

72 MICROMETER CALLIPER thermal expansion
There is deviation of the mean temperature from the reference temperature (20 °C) The temperature in the laboratory is in kept in limits (20 ± 2) °C Standard uncertainty of the deviation from reference temperature: uDt 20 = 2/3 = 1,15 °C Workshop on Providing the traceability of measurements

73 MICROMETER CALLIPER thermal expansion
Contribution to uncertainty by thermal expansion: uLt =  ((L · a · uDt )2 + (L · uDt 20 · uD)2) = = ((0,025 · 11,5 · 1,15)2 + (0,025· 1,15 · 1,15)2) = 0,332 µm Workshop on Providing the traceability of measurements

74 MICROMETER CALLIPER resulting standard uncertainty
u = √(uA2 + uE2 + uM2 + uLt2 ) = = √(0, , , ,3322)= = 0,806 µm Workshop on Providing the traceability of measurements

75 MICROMETER CALLIPER resulting expanded uncertainty
U = k · u = 2 · 0,806 = 1,61 µm   1,6 µm  2 µm k … coverage factor k = 2 ( 5.1 and 5.2 EA 4/02) u … resulting standard uncertainty Workshop on Providing the traceability of measurements

76 MICROMETER CALLIPER evaluation in excel
Workshop on Providing the traceability of measurements

77 System of prepackages control
Thank you for your attention Ivan Kříž ČMI Workshop on Providing the traceability of measurements


Download ppt "Workshop on Providing the traceability of measurements"

Similar presentations


Ads by Google