OSCILLOSCOPE: The device which convert any electrical signal to visual.(waveform) The graph, usually called the trace, is drawn by a beam of electrons.

Similar presentations


Presentation on theme: "OSCILLOSCOPE: The device which convert any electrical signal to visual.(waveform) The graph, usually called the trace, is drawn by a beam of electrons."— Presentation transcript:

1

2 OSCILLOSCOPE: The device which convert any electrical signal to visual.(waveform) The graph, usually called the trace, is drawn by a beam of electrons striking the phosphor coating of the screen making it emit light, usually green or blue. This is similar to the way a television picture is produced. WHY DUAL TRACE? Unlike single trace oscilloscope,dual trace oscilloscope can display two traces on the screen, allowing you to easily compare the input and output. FOREXAMPLE: input and output of an amplifier.

3 Amplifier inputAmplifier output

4 Input and output comparison on DTO

5 Single trace oscilloscope Which this old box cant

6 Difference Between DTO and Dual Beam Oscilloscope  A dual-trace oscilloscope should not be confused with a dual- beam oscilloscope.  Dual-beam oscilloscopes produce two separate electron beams on a single scope, which can be individually or jointly controlled.  Dual-trace refers to a single beam in a CRT that is shared by two channels.

7 DTO WORKING MODES: VERT MODETRIGGER SIGNAL SOURCE CH 1 CH 2 ALTSwitches between CH1 and CH2 input signals after each sweep CHOPSwitches between Ch1 and CH2 input signal for each sweep ADDCombined signal of CH1 and CH2

8 Front panel

9 FRONT PANEL CAL TERMINAL INTENSITY CONTROL FOCUS POSITION CONTROL VOLTS/DIV INPUT JACK CH1 CH2 ALT CHOP ADD

10 Dual Trace Oscilloscope Works In Following Two Modes  alternate mode  chopped mode

11 Alternate mode In alternate mode an electronic switch alternates between signal A & B.

12  In this the electronic switch alternates between channels A and B, letting each through for one cycle of the horizontal sweep.  The display is blanked during the flyback and hold-off periods, as in a conventional oscilloscope.  The screen will show a stable display of both the waveform at channels A and B.  The alternate mode cannot be used for displaying very low frequency signals.

13 Chopped mode In the chopped mode the switch free runs at very high frequency.

14  In this mode the electronic switch free runs at a high frequency of the order of 100 kHz to 500 kHz.  The result is that small segments from channels A and B are connected alternately to the vertical amplifier, and displayed on the screen.  If the chopping rate is much faster than the horizontal sweep rate, the display will show a continuous line for each channel.  If the sweep rate approaches the chopping rate then the individual segments will be visible, and the alternate mode should now be used.

15

16 Typical Applications  Troubleshoot electro­optical and electrical systems.  Observe a triggered event separately or relative to the trigger itself.  Use with analog light meter output to visualize intensity of a source.  Check response of a silicon, InGaAs, or avalanche photod iode.  Analyze signals produced by function generators.  Evaluate system performance.

17 THANK u