Medical Laboratory Instrumentation (Blood cell counter) 2011-2012 Third Year Dr Fadhl Alakwa www.Fadhl-alakwa.weebly.com UST-Yemen Biomedical Department.

Slides:



Advertisements
Similar presentations
OFFLINE COMPOSITION MEASURING SENSORS
Advertisements

Particle Size Sizing Technique 1: Coulter principle
Reticulocyte Count.
Geiger-Muller detector and Ionization chamber
Complete Blood Count ( CBC). Complete Blood Count ( CBC)
Page 1 © J.Paul Robinson - Purdue University Cytometry Laboratories BMS 631 – LECTURE 8 Flow Cytometry: Theory J.Paul Robinson Professor of Immunopharmacology.
PARTICLE SIZE ANALYSIS
Electrical Impedance and Colorimetric Measurements Joanna Ellis, MLS(ASCP)
Automated Hematology Cell Counters
CHAPTER 37 MLAB 1415 HEMATOLOGY JOANNA ELLIS, MLS(ASCP) Optical Light Scatter and Flow Cytometry.
Chapter 5 Part Three Blood Pressure and flow by Ibrhim AlMohimeed BMTS /3/2013.
بسم الله الرحمن الرحيم UMM AL-QURA UNIVERSITY Faculty of Medicine Dep. Of BIOCHEMISTRY.
FLOW CYTOMETRY Dr. MOHAMMED H SAIEMA LDAHR KAAU FACULTY OF APPLIED MEDICAL SCIENCES MEDICAL TECHNOLOGY DEPT. 2 ND YEAR MT INSTROMINTATION EXT
Automation in hematology Until Slide 23
Monroe L. Weber-Shirk S chool of Civil and Environmental Engineering Closed Conduit Measurement Techniques ä Pipeline systems ä pipe networks ä measurements.
-Automation blood count -Red and White blood count and differential count (Manual blood count) Prepared by: Mona alharbi Experiment:
Pharos university Faculty of Allied Medical SCIENCE Clinical Laboratory Instrumentation (MELI-201) Dr. Tarek El Sewedy.
Physiology Presentation Roll No.# 218, 224, 230, 236, 242, 248 Muhammad Mohsin Ali Dynamo.
-Automation blood count -Red and White blood count and differential count (Manual blood count) Experiment:
Objectives 1. Methods used to measure the different hematological values, and compare it with the normal values. 2. The calculation of RBC indices, their.
Mrs. Ibtisam H. Alaswad Mr. Mohammed A. Jaber
Honors Physics Final Review An image that appears to form within the reflective surface is a ________ image? ans: virtual The reference line perpendicular.
Cell viability studies Sepideh Khoshnevis. The Goal To distinguish live cells from dead and apoptotic cells in order to calculate the the percentage of.
Week 5: Electronic Cell Counters
METR125: Light, Color, and Atmospheric Optics et130/notes/chapter19/color.html.
Flow Cytometry at Boston University Medical Campus Introduction to some methods that we offer Yan Deng (X4-5225), Gerald Denis (X4-1371),
Dr Gihan Gawish Hydrodynamic focusing is a technique used to provide more accurate results from flow cytometers or Coulter counters for determining the.
Basic Principles in Flow Cytometry
 Flow cytometry is a technique for counting, examining, and sorting microscopic particles suspended in a stream of fluid.
CLS 541-Advanced Medical Haematology M. sc. Program Lab-2-
Sarmishtha Ghosh Physiology
Automation in haematology
Complete Blood Count.
Growth of bacteria in culture
Chapter 5 Hematology. Figure 5.1 (a) A double ended needle partially pierces the rubber stopper in an evacuated tube. (b) After the left needle pierces.
Flow Cytometry Principles & practice of “Fluorescence Spectroscopy in Biological Diagnosis & Research” Dr.Hekmatimoghaddam Assistant professor of pathology.
Interference Introduction to Optics Coherent source Two Slit Interference Thin film interference Interference from a Grating.
Measurements in Fluid Mechanics 058:180:001 (ME:5180:0001) Time & Location: 2:30P - 3:20P MWF 218 MLH Office Hours: 4:00P – 5:00P MWF 223B-5 HL Instructor:
Sources of noise in instrumental analysis
Microscope parts:.
Flow Cytometry Basic Training. What Is Flow Cytometry? Flow ~ cells in motion Cyto ~ cell Metry ~ measure Measuring properties of cells while in a fluid.
Typically refers to an increase in population rather than in size
Electric Circuits. Electric circuit: a complete path from the positive terminal to the negative terminal.
Nuclear Medicine Instrumentation 242 NMT 1 Dr. Abdo Mansour Assistant Professor of radiology
HAEMATOLOGY LINE A.C.KAROBIA
Complete Blood Count (CBC)
Hematology Unit 2 Chapter 8 Automated Analyzers Copyright © 2015 by Mosby, an imprint of Elsevier Inc. All rights reserved.
Light Chapter 12. Light and Matter Objects can absorb, reflect, or transmit light Depends on what material the object is made up of Typically, “shiny”
General Approach in Investigation of Haemostasis
Radiation detectors Ion chamber 2. Geiger Muller counter (GM).
Electrical Impedance and Colorimetric Measurements
Automatic Electronic Blood Cell Counting
Medical Laboratory Instrumentation
Flowcytometry.
Asst. Prof. Dr. Aseel B. Al- Zubaidi
Osmotic Fragility Test October 2017 Physiology Lab-3
White blood cell (WBC) count
Particle Size Analysis
Determination of Hematocrit (Hct) (Packed Cell Volume; PCV)
Measurement of cell growth by direct methods
(Fluid-Particles Systems)
White blood cell count.
Kap.11 Clinical laboratory
Kap.11 Clinical laboratory
Measurement of cell growth by direct methods
Medical Laboratory Instrumentation Third Year
Circuit Components.
Complete Blood Count (CBC)
Presentation transcript:

Medical Laboratory Instrumentation (Blood cell counter) Third Year Dr Fadhl Alakwa UST-Yemen Biomedical Department

Normal Range Encyclopedia of Medical Devices and Instrumentation - Vol. 2 page 117

Count and size A typical symptom of common anemia is an inadequate amount of RBCs. The increase or decrease in the numbers of the different types of WBCs may indicate infection and inflammation as discussed above. In addition to the cell numbers, other information regarding blood cells, such as cell size, is also important. For example, in patients with anemia caused by vitamin B12 deficiency, the average size of the RBCs is larger than normal and this disease state is called macrocytic anemia. On the contrary, if red blood cells are smaller than normal, as in the case of microcytic anemia, the condition may be indicative of iron deficiency.

Cell Counter Manual – Hemocytometer Instrumental – Wallace coulter (beer law) – Electric impedance or aperture impedance – Flow cytometer

Hemocytometer

Haemocytometer grid: red square = mm 2, nl green square = mm 2, nl yellow square = mm 2, 4.00 nl blue square = mm 2, 0.25 nl Haemocytometer at a depth of 0.1 mm

Electrooptic Measurements

Electronic Cell Counter: Wallace coulter >> Electric impedance

Electric impedance It is based on the measurement of changes in electrical resistance as cells pass through a small orifice that separates two electrodes. cells are suspended in an electrically conductive diluent, such as saline. Low frequency electrical current is applied between two electrodes; one of them being placed in the cell medium and the second within the aperture tube. The aperture tube has a small orifice or sensing aperture that is typically of size 50–200mm in diameter. During the measurement, cells are drawn through the aperture using a pressure gradient that is either generated by a mercury manometer or oil displacement pump. Cells are assumed to be non- conducting.

Electric impedance Electrical resistance between the two electrodes or impedance in the current occurs as the cells pass through the sensing aperture, causing voltage pulses that are measurable. The number of pulses is proportional to the number of cells counted. The size of the voltage pulse is directly proportional to the size (volume) of the cell. This principle allows discrimination between cells of different sizes. Counting of specific-sized cells is also possible using threshold circuits that cut-off voltage pulses above and below predetermined values. The quantity of suspension drawn through the aperture is precisely controlled to allow the system to count and size particles precisely. Finally, several thousand particles are individually counted within seconds in this device. Measurements are independent of particle shape, color, and density.

First, the aperture size is critical. The instrument is set to count only particles within the proper size range. The upper and lower levels of the size range are called size exclusion limits. Any cell or material larger or smaller than the size exclusion limits will not be counted.

Laser Light Scattering and Fluorescence Detection: Flow cytometer

An external sheath flow allows alignment of blood cells one-at-a-time in the path of a light beam, usually within a quartz flow cell. Incident light on cells within this flow stream are scattered or redirected in a manner that is dictated by the size of the cell and the intracellular distribution of refractive index. Lasers are generally preferred as the light source since it produces monochromatic light that has a small spot size.

Photomultiplier tubes (PMTs) are used to collect the weak signal of scattered light. The light scattered at angles from 5–108 (forward scatter) in general correlates with cell size. A cell with many complex intracellular organelles will also give a larger side-scatter signal than a cell with fewer intracellular organelles. In general, these scatter data together allow reliable identification of distinct populations, such as platelets, RBCs, monocytes, and neutrophils in a mixture.

AUTOMATED CELL COUNTERS

Chromatography Analysis of complex mixtures often requires separation and isolation of components, or classes of components. Examples in noninstrumental analysis include extraction, precipitation, and distillation. A procedure called chromatography automatically and simply applies the principles of these “fractional” separation procedures. Chromatography can separate very complex mixtures composed of many very similar components. Electrophoresis is also separation technique but the separation principle is different.

Chromatography The Russian botanist Mikhail Tswett invented the technique and coined the name chromatography.