Electrosurgical Unit.

Slides:



Advertisements
Similar presentations
Effect of Electricity on Living Tissue
Advertisements

Cutting and Dissecting
Session IV B, Slide #1 Contraceptive Implants Session IV B: One-Rod Implant Removal.
Safe Laparoscopy Reducing Complications Jonathan Frappell FRCS.FRCOG.
Two-mode Safety and Battery System Service
Surgical Diathermy Machine
 Electrosurgery:  use of electricity to cause thermal tissue destruction, most commonly in the form of tissue Dehydration coagulation vaporization.
Kenny Low Dec 2009 Urology registrar Thursday teaching
EE 4BD4 Lecture 25 Electrosurgery Biomedical Device Technology: Principles and Design, Charles C. Thomas Publisher 2007 Anthony K. Chan 1.
Shortwave and Microwave Diathermy part(2). Induction Electrodes (Drum Electrode) One Or More Monopolar Coils Rigidly Fixed In A Housing Unit May Use More.
Electrocautery.
Preparation for Surgery Sterile Field Maintenance
© 2011 The McGraw-Hill Companies, Inc. All rights reserved Learning Outcomes 41.1Define the medical assistant’s role in minor surgical procedures.
Surgical Instruments and Aseptic Technique
Electrical Surgical Unit
Suture Materials ABSORBABLE: lose their tensile strength within 60 days. NON- ABSORBABLE:
Surgical Instruments and Terminology 4-H Veterinary Science Extension Veterinary Medicine Texas AgriLife Extension Service College of Veterinary Medicine.
 Requires a working knowledge of the sequential steps for a specific surgical procedure based upon four concepts:  Approach  Procedure  Possible.
What is a Watt? Everything you ever wanted to know about Electrosurgery. Fahad bamehriz, MD.
What is electrosurgery?
Diathermy.
Assisting with minor surgery and suture removal. Minor Surgery includes Removal of warts, cysts, tumors, growths, foreign objects Performing biopsies.
Hemostasis & Emergency Situations
Veterinary Clinical Procedures Surgeries. The Surgery Team Surgery Team consists of a surgeon, an anesthetist (or a few sterile assistants-scrub nurse),
Without reference, identify principles about Electrosurgical Units with at least 70 percent accuracy.
Session VI, Slide #1 Contraceptive Implants Session VI: Infection Prevention.
Session V B, Slide #1 Contraceptive Implants Session V B: Two-Rod Implant Removal.
Chapter 28 Surgical Instruments and Aseptic Technique.
BASICS OF ELECTROSURGERY
Electrosurgery & Gyn Surgery: Get the Point Across
Intraoperative Case Management, Anticipation, Routines, & Counts ST230 Concorde Career College.
Asepsis Copyright © 2012 by Mosby, an imprint of Elsevier Inc.
Computer controlled I NSTANT R ESPONSE ™ system provides safer and more precise electrosurgery. TM.
Equipment Source- Alexander’s Care of the Patient in Surgery Patty Pavlikowski RN CST MA.
Principles of Surgery EFE Veterinary Science. For best surgical outcomes Minimize contaminationMinimize blood loss Minimize surgical timeStable anesthetic.
Electro surg i cal unit  PURPOSE ESU are used for surgical cutting and for controlling bleeding by causing coagulation “ hemostasis ” at the surgical.
Asepsis Copyright © 2012 by Mosby, an imprint of Elsevier Inc.
Electro surgery in Gynaecology
Chapter Three - AED. The Heart’s Electrical System - The heart’s electrical system sends out signals that tell the heart to pump blood. The signal travels.
This presentation may contain some forward looking statements, particularly regarding operational prospects in 2009 and beyond, which involve a number.
Institute of Surgical Research „A” Module - Surgical Technics
RDT Hospital, Dept of Gen Surgery
Minor Surgery In The Medical Office
TERMINAL OBJECTIVE OPERATE SELECTED OR EQUIPMENT IAW MANUFACTURER’S USER MANUAL.
Technological Sciences for the Operating Room
Dissection Lab Safety. What will we dissect? First---a pickle (to learn technique)
Electrosurgery Units ELECTROSURGICAL UNITS (ESUs)
SAFE USE OF DIATHERMY RACHEAL AFOLABI, RN, RM, H. Dip. PERI-OPERATIVE NURSING, H.Dip. HEALTH SYSTEM MANAGEMENT.
Common Medical Procedures for Treating Varicose Veins.
TOPIC: PATIENT SAFETY BASED ON SUCTION AND ELECTROSURGICAL DIATHERMY MACHINES. PRESENTER: ALCARD FIDELIS BIOMEDICAL TECHNOLOGIST. ENGINEERING DEPARTMENT.
Electro surgical unit.
Basics of Medical Equipment Lecture 6
Electrosurgical/Surgical Diathermy Units
VUSE Medical and Electromechanical Design Lab
A. In monopolar electrosurgery, current from an electrosurgical generator passes from an active electrode (the “Bovie” tip) through the patient to a return.
Electrosurgery in Periodontics
Soft Tissue Injuries.
Cutting and Dissecting
Soft Tissue Injuries.
Cutting and Dissecting
LIFEPAK® 1000 Defibrillator Orientation Guide
Diathermy and its safe use
Soft Tissue Injuries.
Contraceptive Implants Session V B: Two-Rod Implant Removal
Surgical Instruments Grasping and Clamping Retracting Cutting and Dissecting Probing and Dilating.
Fundamental Use of Surgical Energy
Contraceptive Implants Session V B: Two-Rod Implant Removal
Ebola Facts October 16, 2014.
Ebola Facts October 16, 2014.
Presentation transcript:

Electrosurgical Unit

Overview History Components Placement of active and inactive electrode. Monopolar/Bipolar Functions Safety considerations

Electrosurgical Unit Used to apply electrical current through the patient’s tissue to cut and/or coagulate the tissue. Monopolar mode may cut or coagulate. Bipolar mode is used for coagulate only.

History By the late 1800’s D’arsonval and Tesla applied high frequency energy to the human body to generate heat. 1908 electromagnetic waves are used to cut tissue called surgical diathermy. Results were highly unsatisfactory.

History Cushing and Bovie collaborated to produce a device that used radio frequency electric current for coagulation. Worked much smoother than the spark gap generator, allowing the cutting of tissues.

History In the 1970’s solid state generators were developed. Modern electrosurgical units are equipped with safety features to reduce the chance of burns. Although many changes have been made to the original design, most electrosurgical units are still referred to as a bovie.

Components Generator Provides source of electrical current to the active electrode. Control panel allows for power level and blend adjustments.

Components Active electrode Referred to as the electrosurgical pencil. May be hand or foot controlled.

Components Dispersive electrode Also known as the grounding pad or inactive electrode. Placed on the patient to return the electrical current to the generator.

Placement Active electrode Packaged sterile, with a protective holster. Once the patient has been draped it is secured with a non penetrating clamp. End is passed off to circulator that is plugged into the generator.

Placement Inactive electrode (grounding pad) Available in adult and pediatric sizes. Disposable, although metal plates and gel are occasionally used. Placed on a large fleshy area, close to the incision site. Must be clean and dry. Relatively free of scars or hair. Not to be placed over a joint, metal prosthesis, or bony prominence.

Placement Inactive electrode (grounding pad) Errors in placement may result in a patient burn.

Monopolar Electrosurgical pencil. Various interchangeable tips. Blade Needle Ball Hand or foot controlled. Dispersive electrode is necessary in monopolar mode.

Bipolar Uses a reusable Forcep tip with a disposable cord. Dispersive electrode is not necessary in bipolar mode.

Current Flow Monopolar Generator Active electrode Patient’s body Inactive electrode

Current Flow Bipolar Generator Active electrode One prong of the Forcep. Patient’s tissue Inactive electrode Other prong of the Forcep.

Principle of Electrosurgery High frequency current is produced and applied to tissues to cut or coagulate. Advantages include: Reduced blood loss Reduced surgical time Using the cutting feature allows you to dissect tissue without stopping to control bleeding.

Coagulation and Cutting Functions Coagulation function allows the surgeon to individually coagulate blood vessels by applying the tip directly to the tissue or by grasping the tissue with a forcep and applying the tip to the forcep.

Coagulation and Cutting Functions Cutting function allows the surgeon to make scalpel like incisions and simultaneously control hemostasis. Control panel offers “spray” or “blend” configurations to control the amount of coagulation.

Coagulation and Cutting Functions The active electrode blade must be free of charred tissue to work efficiently. Special abrasive tip cleaners are placed on the sterile drape near the operative site.

Safety Considerations Flammable prep solutions must be allowed to completely dry, and not be allowed to pool under the patient. Flammable anesthetics should not be used during the procedure. EKG electrodes with metal tips should be placed carefully to avoid burning the patient.

Safety Considerations Patients with pacemakers or internal defibrillators may malfunction. All jewelry must be removed from the patient before surgery to prevent burns.

Safety Considerations Burns to the surgical team may result by a hole or weak spot in the surgeons surgical gloves. Example: If surgeon applies a hemostat onto a bleeding vessel and then applies the bovie pencil to the instrument, with a hole or weak spot in their glove.

Review History Components Placement of active and inactive electrode. Monopolar/Bipolar Functions Safety considerations