Beam Restricting Devices

Slides:



Advertisements
Similar presentations
S. Guilbaud Education Director School of Radiologic Technology
Advertisements

Advanced Biomedical Imaging Lecture 3
Control of Scatter Radiation
Chapter 11 Prime Factors.
Kilovolt Peak (kVp) kVp determine the quality of the beam.
Formation Characteristics
X-Ray Beam Restrictors
X-Ray Interaction with Matter & Human Biology
Primary Exposure Factors IV
Radiographic Quality Chapter 5.
Topic 3 Selection of kV High kV scattered radiation is energetic. Angle of rays is same as those of primary beam and is not deflected very much. Follows.
RADIOLOGY. NEXT GENERATION SCIENCE / COMMON CORE STANDARDS ADDRESSED! CCSS.ELA-Literacy.RST Determine the central ideas or conclusions of a text;
Chapter 6 Control of Secondary and Scattered Radiation.
Control of Scatter Radiation Beam Restricting Devices
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Radiographic Grids II By Professor Stelmark.
Chapter 6, Fauber Supplement-Chapter 14, Bushong
The Grid Kyle Thornton DMI 50B.
Analyzing the Image Density. Density Overall blackening of the image.
X-ray Beam Composition and Collimation
DESIGNING FOR RADIATION PROTECTION. TUBE HOUSING  REDUCES LEAKAGE TO LESS THAN 100 mR PER HOUR AT A DISTANCE OF ONE METER FROM HOUSING  One meter is.
Resident Physics Lectures
Lecture 4 Beam restricting Devices, their Use and Maintenance
Factors affecting the X-Ray output
Radiographic Grids I By Professor Stelmark.
Resident Physics Lectures Christensen, Chapter 6Filters George David Associate Professor Department of Radiology Medical College of Georgia.
Radiation Exposures Unit 2
Control of Scatter Radiation
Quality Control.
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
HABIS X-RAY PRODUCTION AND EXPOSURE FACTORS X-RAY PRODUCTION AND EXPOSURE FACTORS PREPARED BY PREPARED BY Dr fahad albadr radiology chairman radiology.
Dr Mohamed El Safwany, MD.
Resident Physics Lectures
Image Quality Radiographic Resolution.
Exposure Factors or Prime Factors
Photon Tissue Interactions
PRINCIPLES OF TECHNIQUE AND EXPOSURE
Radiographic Quality Visibility and Sharpness
Beam And Scatter Radiation Restricting Devices DMI 50B Kyle Thornton.
Radiographic Equipment
Attenuation As x-rays pays through matter, the exit beam will contain less photons than the entrance beam. This reduction in the quantity of photons is.
Radiographic Grids I By Professor Stelmark.
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Lavin: Chapter 7 CTVT: pp
Tube Exposure Factors Math Technique Contrast and Density
Sergeo Guilbaud School of Radiologic Sciences
TECHNIQUE CHARTS SERGEO GUILBAUD CLINICAL COORDINATOR.
RADIOGRAPHIC GRIDS.
RADIOGRAPHIC QUALITY S. Guilbaud, Education Director School of Radiologic Sciences.
Automatic Exposure Control (AEC)
Radiographic Technique Sergeo Guilbaud Education Director Long Island College Hospital.
Optimizing The Image RVT: Chapter 7
Exposure Factors or Prime Factors
Control of Scatter Radiation Beam Restricting Devices By Prof. Stelmark.
Chapters 21 & 22.
Optimisation of Patient Protection for Radiography
Protection of Patients
Radiographic Contrast
Part No...., Module No....Lesson No
X-Radiation.
Factors Effecting the Production
Exposure Factors Chapter 4
Resident Physics Lectures
Resident Physics Lectures (Year 1)
Resident Physics Lectures
Resident Physics Lectures (Year 1)
Chapter 5, Fauber Supplement-Chapter 11, Bushong
GRIDS PRESENTER-Dr.Vishwanath patil PG resident
Presentation transcript:

Beam Restricting Devices Sergeo Guilbaud School of Radiologic Technology

Abstract Beam limiting devices are used to control the amount of scatter radiation that is created as a result of Compton’s interaction with the body’s tissues. Scatter hinders the visualization of detail by adding additional density to the radiographic image. Scatter is also responsible for increasing the patient’s radiation dosage, decreasing image contrast and impairing the visibility of detail. By controlling the amount of scatter radiation that is produced, there is improvement in detail, enhancement of contrast and a reduction in radiation dosage to the patient. Beam limiting devices are types of radiographic tools used to reduce the production of scatter radiation.

Objectives At the conclusion of this lecture, each participant should be able to: Identify the principle factors that are responsible for the creation of scatter radiation. Identify the three types of Beam limiting devices. Identify the advantages and disadvantages of each beam limiting device. Identify the rationale for using Positive Beam Limitation (PBL). Describe the ancillary devices that may be used to further reduce the amount of scatter radiation reaching the image receptor system.

Principle factors creating scatter Kilovoltage (kV) Irradiated Tissues

Principle factors creating scatter Kilovoltage (kV) Affects beam penetrability As kVp increases, the percentage of x-ray photons that undergo Compton’s interaction increases. B/C Compton interactions are responsible for creating scatter, as the kVp increases, the percentage of primary photons that will scatter also increases.

Principle factors creating scatter As kVp increases, the percentage of photons that undergo photoelectric absorption decreases. Thus, the patient receives significantly less radiation.

Principle factors creating scatter Scattered photons from Compton’s interaction are of no value to the radiographic image b/c they do not assist in demonstrating the anatomic structures of interest. (add additional density to film) The majority of the overall radiographic density on a film is created by scattered photons. Thus, a decrease in Compton’s scatter will improve the image quality by removing the scatter that reaches the film.

Principle factors creating scatter Irradiated Tissue Tissue volume and atomic number affect the amount of scatter produced during a radiographic exposure. Tissue volume irradiated is controlled by the radiographer. When the amount of tissue volume increases, so does the amount of scatter radiation. The larger the field size, the more photons to interact with tissue, this results in greater amount of scatter production.

Principle factors creating scatter Remember: More scatter is created in bone than in soft tissue. When the beam is restricted, less scatter radiation is produced thus, less scatter reaches the film. The technical factors may also have to be manipulated. There may need to be an increase in the technical factors to compensate for the reduction in the overall density on the film.

Beam Restrictors Three types: 1. Aperture diaphragms. 2. Cones/cylinders. 3. Collimators.

Aperture Diaphragm A flat sheet of metal with a hole cut in the middle attached to the tube head. The simplest of beam limiting devices both in design and ease of use. This was the original type of beam limiting device. Use is now limited to dedicated chest units, where field size and SID do not vary.

Aperture Diaphragm

Aperture Diaphragm Disadvantages 1. Increase in penumbra. Advantages: This is a result of the diaphragm’s close proximity to the target. An increase in off-focus (stem radiation) radiation reaching the film. This can result in images, similar to shadows beyond the field. Advantages: 1. Simple design 2. Low cost 3. Ease of use

Aperture Diaphragm Formula for determining diaphragm field size. Projected image size = SID X diameter of diaphragm dx from focal spot to diaphragm e.g. A diaphragm is placed 5” from the FS. The diameter of the opening in the aperture is a 2” circle. What would be the projected image size if the SID is 40”? 40 x 2 = 80 -------- --- = 16 inches 5 5

Cones/Cylinders Circular aperture diaphragms with metal extensions. Cone – has a flaring extension w/ upper diaphragm smaller than bottom (flared end). Cylinder – does not flare but, may be equipped w/ an extension sleeve which can expand or collapse to cavy the beam restriction.

Cones/Cylinders

Cones/Cylinders Advantages: Inexpensive. Simple to use. Reduce penumbra and off-focus radiation. Provide better beam restriction at a greater distance from the focal spot.

Cones/Cylinders Disadvantages: Fixed field sizes unless equipped with extension sleeve. Flared cones are no better at reducing penumbra than aperture diaphragms.

Cones/Cylinders Formula for determining Cone field size. Projected image size = SID x lower diameter of cone Dx from focal spot to bottom of cone e.g. The diameter of the lower rim of a cone is 3”. The bottom of the cone is 16” from the focal spot. What will be the size of the projected image at 40” SID? 40 x 3 120 Projected image = -------- = ----- = 7.5 “ circle 16 16

Collimators

Collimators Consists of two sets of lead shutters at right angles to one another which move in opposing pairs. Each set moves symmetrically from the center of the field. They can be adjusted to correspond to an infinite number of squares or rectangular field sizes. Bottom shutters serve to reduce penumbra along the periphery of the beam. Upper shutters help reduce amount of off-focus radiation.

Collimators Construction: Two sets of shutters moving in opposite pairs. Light field provided by light source. Mirror mounted at 45 degrees within path of beam to reflect light source. Provides additional beam filtration.

Collimators Advantages: Permit infinite number of field sizes. Provides light source as an aid to proper tube, part and film alignment. Only one device providing all said functions.

Collimators Disadvantages: Expensive to manufacture. Light requires replacement with heavy usage. Whenever light source is changed, proper alignment must be checked (light-field congruency test. Collimator accuracy testing must be performed at regular intervals.

Positive Beam Limitation (PBL) Devices Automatic Collimators Automatically reduce the collimated area to the size of the film placed in the Bucky tray. When operating properly, they leave a small unexposed area on all four borders of the film. State regulations allows the collimation to fall within plus or minus 2% of the SID. Can be overridden by the radiographer. Assure that area exposed does not exceed image receptor size.

Ancillary devices used to reduce the scatter reaching the film Normally designed to fill a special need. Lead blockers Lead Masks

Ancillary Devices Lead Blockers A sheet of lead-impregnated rubber that can be cut to a size or shape. Especially useful in preventing scatter radiation from reaching the film especially when imaging the Thoracic & Lumbar spines in the projection and the shoulder when done on the table.

Lead Masks A sheet of lead-impregnated rubber that can be cut to a size or shape to correspond to a particular field size then, secured to the end of a collimator. Most common: lead mask used during cerebral angiography.

References Bushberg et al, The Essentials of Physics and Medical Imaging, Williams & Wilkins Publisher. Bushong, S., Radiologic Science for Technologists, Physics, Biology and Protection, 8th Edition, C.V. Mosby Company. Carlton et al, Principles of Radiographic Imaging, An Art and Science, Delmar Publishing. Curry et al, Christensen’s Introduction to the Physics of Diagnostic Radiology. 4th Edition, Philadelphia, PA, Lea and Febiger, 1990. Selman, J., The Fundamentals of X-Ray and Radium Physics, 8th Edition, Charles C. Thomas Publisher.