Recent Advances in Small Animal SPECT instrumentations and techniques F. Cusanno 1, M. Ballerini 1, E. Cisbani 1, S. Colilli 1, R. Fratoni 1, F. Garibaldi.

Slides:



Advertisements
Similar presentations
Instruments for Radiation Detection and Measurement Lab # 4.
Advertisements

MOIRA (MOlecular Imaging with RAdionuclides) F. Garibaldi, gr. Coll. Sanita What: Design (and implement) a SPECT detector with submillimeter spatial resolution.
A Novel High Resolution, High Sensitivity SPECT Detector for Molecular Imaging Of Cardiovascular Diseases F. Cusanno (1), R. Accorsi (6), E. Cisbani (1,3),
Preliminary pinhole SPECT reconstruction results from the NaI detector Images are displayed as MIP (maximum-intensity-re-projections) animations Please.
“From man to mouse” … Human PET microPET human mouse rat rat infant
Imaging Molecolare con radionuclidi: un potente mezzo di indagine di processi biologici in vivo F. Garibaldi - Fisica e tecnologie Nucleari per la Salute.
Sezione di Napoli Univ. “Federico II” Experimental study of beam hardening artefacts in photon counting breast computed tomography M.G. Bisogni a, A. Del.
 Nuclear Medicine Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT Kathleen Vunckx Johan Nuyts Nuclear Medicine,
F. Garibaldi - ISS-NIH, Rome, 4-6 June 06
16 November 2004Biomedical Imaging BMEN Biomedical Imaging of the Future Alvin T. Yeh Department of Biomedical Engineering Texas A&M University.
OVERVIEW NEDA Introduction to the Simulations – Geometry The Simulations Conclusions 3.7% This work summarizes the introduction to the simulations of.
Special Imaging Techniques Chapter 6 Bushong. Dynamic Computed Tomography (DCT) Dynamic scanning implies 15 or more scans in rapid sequence within one.
PHYSICS IN NUCLEAR MEDICINE: QUANTITAITVE SPECT AND CLINICAL APPLICATIONS Kathy Willowson Department of Nuclear Medicine, Royal North Shore Hospital University.
Single Photon Computerized Tomography SPECT neuroimaging Seyed Kazem Malakouti, MD Faculty of Iran University of Medical Sciences Seyed Kazem Malakouti,
Planar scintigraphy produces two-dimensional images of three dimensional objects. It is handicapped by the superposition of active and nonactive layers.
8/18/2015G.A. Fornaro Characterization of diffractive optical elements for improving the performance of an endoscopic TOF- PET detector head Student: G.
Very High Resolution Small Animal PET Don J. Burdette Department of Physics.
Novel high resolution detectors for Positron Emission Tomography (PET)
F. Garibaldi – INFN Roma and ISS
FP 7 Collaborative Project Novel MR-compatible PET detectors for simultaneous PET/MRI imaging FP7-HEALTH-2009-single-stage -Scope -Issues -Actions -Deadline(s)
Prostate probe with SPECT technique
New Perspectives in Molecular Imaging of Cardiovascular Diseases F. Garibaldi - INFN – Roma1 - molecular imaging: the role of radionuclides techniques.
Design and simulation of micro-SPECT: A small animal imaging system Freek Beekman and Brendan Vastenhouw Section tomographic reconstruction and instrumentation.
High Resolution and High Efficiency Open SPECT Detector for Molecular Imaging Studies of Cardiovascular Diseases on Mice MIOCARDIAL PERFUSION MEASUREMENTS.
November 8, 2005JLAB Awake Animal Project New Instrumentation for JLAB Awake Animal Project Smaller Pixel Size requires higher granularity in PMT (2.4.
Characterization of LaBr3:Ce scintillator optimized for spatial resolution in low-energy gamma detection F. Cusanno, E. Cisbani, S. Colilli, R. Fratoni,
FP7 FMTXCT Project UMCE-HGUGM second year activity report Partner FIHGM Laboratorio de Imagen Médica. Medicina Experimental Hospital Universitario Gregorio.
M. Alnafea1*, K. Wells1, N.M. Spyrou1 & M. Guy2
Biomedical Mechatronics Lab M ONTE C ARLO M ODELING AND A NALYSIS OF S TRUCTURED C S I S CINTILLATOR - COUPLED P IXEL D ETECTORS Chang Hwy Lim, Ho Kyung.
Md.Kausher ahmed Electrical department. Biomedical engineering Code:6875.
BMI I FS05 – Class 4 “Nuclear Imaging: Math” Slide 1 Biomedical Imaging I Class 5 – Radionuclide Imaging (PET, SPECT), Part 3: Attenuation and Scatter.
Professor Brian F Hutton Institute of Nuclear Medicine University College London Emission Tomography Principles and Reconstruction.
Concept of the scanning table in Strasbourg François DIDIERJEAN Tatjana FAUL, Fabrice STEHLIN Strasbourg AGATA week July 2008 Uppsala, Sweden.
Active Pixel Sensors in Nuclear Medicine Imaging RJ Ott, N Evans, P Evans, J Osmond, A Clark, R Turchetta Physics Department Institute of Cancer Research.
O AK R IDGE N ATIONAL LABORATORY U.S. DEPARTMENT OF ENERGY Image Reconstruction of Restraint-Free Small Animals with Parallel and Multipinhole Collimation:
New techniques for early detection of diseases: imaging of small biological structures with amplifying PET/SPECT probes using SiPm photodetectors An important.
Increase in Photon Collection from a YAP:Ce Matrix Coupled to Wave Lenght Shifting Fibres N. Belcari a, A. Del Guerra a, A. Vaiano a, C. Damiani b, G.
Nuclear Medicine Principles & Technology_I
R. Pani Department of Experimental Medicine and Pathology University of Rome La Sapienza-Italy. Flat Panel PMT: advances in position sensitive photodetection.
A Single Photon Emission Computer Tomograph for breast cancer imaging S. Vecchio a, N. Belcari a, P. Bennati b, M. Camarda a, R. Campanini c, M. N. Cinti.
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Prostate probe with SPECT technique NSS – MIC November 5 - KnoxvilleF. Garibaldi- INFN – Roma1 – gr. Coll. ISS  the medical problem  the proposal.
A. G. Argentieri 1), E. Cisbani 2), S. Colilli 2), F. Cusanno 2), R. De Leo 1), R. Fratoni 2), F. Garibaldi 2), F. Giuliani 2), M. Gricia 2), M. Lucentini.
1 Tri-Modality Imaging of Small Animals Mark B. Williams, PhD University of Virginia DOE Awake Animal Meeting, May 20, 2004.
Scintillators suitable for PET applications must be characterized by a high efficiency for gamma-ray detection, determined by a high density and atomic.
Active Pixel Sensors in Medical and Biologi The application of Large Area Active Pixel Sensor (LAS) to high resolution Nuclear Medicine imaging Bob Ott.
1 Nuclear Medicine SPECT and PET. 2 a good book! SR Cherry, JA Sorenson, ME Phelps Physics in Nuclear Medicine Saunders, 2012.
Acquisition time6 min1 min 12 s Collimator height25 mm (Anger)12 mm (HiSens) Detector1 layer, 1 pixel / hole3 layers, 1 pixel / hole3 layers, 4 pixels.
F. Garibaldi 1,, F. Cusanno 1), S. Majewski 3), N. Clinthorne, P. Musico 4),…………………………. TOPEM: a PET TOF probe, compatible with MRI and MRS for diagnosis.
Nuclear Medicine Physics and Equipment 243 RAD 1 Dr. Abdo Mansour Assistant Professor of radiology
Nuclear Medicine Instrumentation 242 NMT 1 Dr. Abdo Mansour Assistant Professor of radiology
Biomarkers from Dynamic Images – Approaches and Challenges
Introduction to Medical Imaging SPECT, Introduction to Medical Imaging SPECT, PET and Lesion Detection Guy Gilboa Course
Journal of Nuclear Cardiology | Official Journal of the American Society of Nuclear Cardiology The high matrix acquisition technique for imaging of atherosclerotic.
Segmentation of 3D microPET Images of the Rat Brain by Hybrid GMM and KDE Tai-Been Chen Department of Medical Imaging and Radiological Science,
Prostate probe with SPECT technique NSS – MIC November 5 - KnoxvilleF. Garibaldi- INFN – Roma1 – gr. Coll. ISS  the medical problem  the proposal.
Abstract The shape of the current pulse produced by a SiPM in response to an incident photon is sensibly affected by the characteristics of the FE used.
IMAGE QUALITY. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Molecular Breast Imaging – state of art of design, analysis and optimization processes in industrial context F. Cirillo 1, F. Garibaldi 2,3, S. Gioia 1,
Figure 4 PET imaging in experimental pancreatic cancer
Imaging molecolare ad alta risoluzione spaziale ed alta efficienza
Quantitative Nuclear Medicine Imaging
Role of Apoptosis in Vulnerable Plaques
Development of a High Precision Axial 3-D PET for Brain Imaging
HIGH RESOLUTION, HIGH MOLECULAR IMAGING CODED APERTURE THE DETECTOR
Imaging the event-prone coronary artery plaque
Circ Cardiovasc Imaging
Application of Nuclear Physics
Preliminary pinhole SPECT reconstruction results from the NaI detector
A multichannel compact readout system for
Presentation transcript:

Recent Advances in Small Animal SPECT instrumentations and techniques F. Cusanno 1, M. Ballerini 1, E. Cisbani 1, S. Colilli 1, R. Fratoni 1, F. Garibaldi 1, F. Giuliani 1, M. Gricia 1, M. Lucentini 1, M. L. Magliozzi 1,2, S. Majewski 3, G. P. K. Mok 4, F. Santavenere 1, S. Torrioli 1,2, B. M. W. Tsui 4, P. Veneroni 1, Y. Wang 4 Frontiers in Imaging Science: High Performance Nuclear Medicine Imagers for Vascular Disease Imaging, Rome, 13 – 14 November Istituto Superiore di Sanità, Rome, ITALY 2. Università La Sapienza, Rome, ITALY 3.Jefferson Lab, Newport News VA, USA 4.Johns Hopkins University, Baltimora MD, USA

CT Tissue Density, Z A µm  -galactocidase 0.1 µmole H / µmole 31 P MRIA H Concentration MF BOLD, DCE 0.1 mm UltrasoundStructure A F Doppler Optical (Bioluminescence, fluorescence) A Topography M ~10 3 cells  quantitative µm to mm PET/SPECTRadiotracer M ~1-2 mm < mole = quantitative F Molecular Imaging : the Role of the SPECT Unique!

Monte Carlo to Design Advanced Detectors Collimation plays a fundamental role (B. Tsui, R. Accorsi) GEANT4-based simulations: For the intrinsic gamma-detector different designs are possible Quality of the position reconstructions depends on amplitude and spread of light at the photo-detection plane and reconstruction algorithm Electronics should be adequate to really take advantage of chosen design and photodetector (A. Walenta)

 Smaller pixel in principle may provide better position reconstruction, provided sufficient light sampling  Pixel dimension, light guide and photodetector(s) coupling impacts on performances, they are not indipendent Monte Carlo can predict the limit of the intrinsic spatial resolution for a given configuration Light Sampling

Array Image and Map Reconstruction (1) The Present

Possible Future Array Image and Map Reconstruction (2)

Application: Atherosclerotic Plaques in Mice 1.ApoE -/- Mice Model for studying patogenesis of the atherosclerosis If feeded with lipidic food, it spontanously develops atherosclerotic lesions very similar to human plaques 2.Correlation between the presence of apoptotic macrofages and level of vulnerability of the plaque 3. 99m Tc-Annexin V: radiotracer with very high uptake in apoptotic cells Foto e Autoradiografia di Aorta di topo dopo 37 settimane di dieta ricca di lipidi Apolipoprotein E Knockout Mice Detection using advanced small-animal SPECT system Plump A, Ann Med 29, (1997) Kolodgie FD et al, Am J Pathol 157, (2000) Mari C et al, J Nucl Med 43, (2002) P702

Why Studying Atherosclerosis in Mice First cause of death in the western countries Plaques develop slowly, silently and asimptomatically since youth, as accumulation of lipids, colesterhol, connective tissue on the vascular walls Rupture of vulnerable plaques manifests suddendly as cardiac attack Discrimination of vulnerable/stable plaques is fundamental for effective diagnosis Conventional technique (angiography) are not specific and invasive Conventional techniques (angiography):  are not specific  are invasive Goal: design a detection technique  specific for vulnerable plaque  not invasive

Parameters for Simulation Blankenberg FG et al, PNAS 95, (1998) Collimator: Pinhole, 0.3˚mm aperture, magnification factor 3 Scintillator: CsI(Tl) pitch 0.5/0.8˚mm, 3˚mm thickness CsI(Na) pitch 0.8˚mm, 3˚mm thickness LaBr 3 (Ce) continuous, 3 − 5 mm thickness PS-PMT (“offline”): Flat Panel H9500, anode size 3.0˚mm Burle Planacon 85021, anode side 1.5 mm Assumed radiotracer distribution: Liver 12%, Kidneys 56 %, Spleen 2%, 5% ejected  25 % is distributed in the tissue H 3.0 mCi di Tc 99m -Annexin-V, H Phantom size 8×3×2 cm 3  Surrounding Tissue Uptake ~ 600 Bq/mm 3  Plaque Uptake ~ 10÷20*600 Bq/mm 3 H Plaque size ~ 0.5 × 1 × 4 mm 3

Results of Simulation T/B ratio “Shallow” Plaque Mid-deep Plaque Evaluation of S/N ratio using analytic algorithm

Preliminary JHU (1) Spatial Resolution  0.6 mm Detector setup: tungsten PinHole 0.3 mm aperture, magnification factor 3 Scintillator: CsI(Tl) 1.0 mm pitch Photodetector: 2x2 H9500 array Electronics: IDEAS F.E.C and data acquisition

ApoE -/- mouse, 15-week old Detector setup: tungsten PinHole 0.3 mm aperture, magnification factor 3 Scintillator: CsI(Tl) 1.0 mm pitch Photodetector: 2x2 H9500 array SPECT Reconstruction of 64 one-min views using MIP (maximum- intensity-re-projections) alghoritm Preliminary JHU (2)  Setup of SPECT with alive mouse

Stem Cell Studies

Conclusions Molecular imaging with radionuclides is a very powerful technique allowing studies of “in-vivo” biological process; Single-Photon modality generally allows longer studies than PET and it could use different tracers Spatial resolution does not have theoretical limit, better detector technology is continuosly improving performances Advanced SPECT systems are required (for ex.) in the study of atherosclerotic plaques with mice Others interesting studies (for ex. stem cells) could benefit the use of advanced small-animal SPECT devices THANK YOU FOR YOUR ATTENTION!