Biophotonics lecture 11. January 2012
Today: -Correct sampling in microscopy -Deconvolution techniques
Correct Sampling
What is SAMPLING? Intensity [a.u.] X [µm] 1
Aliasing … suppose it is a sine-wave Intensity [a.u.] There are many sine-waves, SAMPLED with the same measurements. Which is the correct one?
Intensity [a.u.] X [µm] When sampling at the frequency of the signal, a zero-frequency is recorded!
Intensity [a.u.] X [µm]
Intensity [a.u.] X [µm] Problem: too high frequencies will be aliased, they will seemingly become lower frequencies
But … high frequencies are not transmitted well. Object: Microscope Image: Intensity Spatial Coordinate Intensity Spatial Coordinate OTF
Aliasing in Fourier-space Fourier-transform of Image Intensity Aliased Frequencies ½ Sampling Frequency Cut-off frequency =½ Nyquist Rate Sampling Frequency Nyquist Rate
Pixel sensitivity Intensity [a.u.] X [µm] 1 Convolution of pixel form factor with sample Multiplication in Fourier-space Reduced sensitivity at high spatial frequency
Optical Transfer Function |k x,y | [1/m] contrast Cut-off limit 0 1 rectangle form-factor OTF sampled
Consequences of high sampling Confocal: high Zoom more bleaching? No! if laser is dimmed or scan-speed adjusted bad signal to noise ratio? Yes, but photon positions are only measured more accurately binning still possible high SNR. Readout noise is a problem at high spatial sampling (CCD)
Optimal Sampling?
Regular sampling Reciprocal -Sampling Grid Real-space sampling: Multiplied in real space with band-limited information
Regular sampling Reciprocal -Sampling Grid Real-space sampling:
Widefield Sampling In-Plane sampling distance Axial sampling distance
Confocal Sampling In-Plane sampling distance (very small pinhole) else use widefield equation Axial sampling distance
Confocal OTFs WF 1 AU 0.3 AU in-plane, in-focus OTF 1.4 NA Objective WF Limit
Hexagonal sampling Advantage: ~17% + less ‚almost empty‘ information collected + less readout-noise approximation in confocal Reciprocal d-Sampling Grid Real-space sampling: Multiplied in real space with band-limited information
63× 1.4 NA Oil Objective (n=1.516), excitation at 488 nm, emission at 520 nm l eff = nm, a = deg widefield in-plane: d xy < 92.8 nm maximal CCD pixelsize: 63×92.8 = 5.85 µm confocal in-plane:d xy < 54.9 nm widefield axial: d z < nm confocal axial: d z < nm Fluorescence Sampling Example
OTF is not zero but very small (e.g. confocal in-plane frequency) Object possesses no higher frequencies You are only interested in certain frequencies (e.g. in counting cells, serious under-sampling is acceptable) Reasons for undersampling
If you need high resolution or need to detect small samples sample your image correctly along all dimensions Sampling Summary
Maximum Likelihood Deconvolution
Image:
The prior (requires prior knowledge; can imply contraints, e.g. positivity) Constant normalisation factor
Constant, therefore obsolete
MATLAB demonstration
Information & Photon noise Virtual Microscopy Only Noise? FT NO! 10 Photons / Pixel
Band Extrapolation? Object Mean Error Energy Mean Energy Relative Energy Regain
With Photon Noise
Is this always possible? White Noise Object
Is this always possible? Unfortunately NOT !