Reduction of effective and organ dose to the eye lens in cerebral MDCT scans using iterative image reconstruction Zizka J, Jandura J, Kvasnicka T, Klzo.

Slides:



Advertisements
Similar presentations
Radiation safety in CT.
Advertisements

Jessica Kishimoto The University of Western Ontario Medical Biophysics Undergrad April 7, 2010 The Effect of Dose on Image Quality When Using an Image.
CT Scanning: Dosimetry and Artefacts
Radiation safety in CT.
CT Organ Dose - Part 1 Estimating organ dose for patients Michael F. McNitt-Gray, Ph.D., DABR David Geffen School of Medicine at UCLA.
Improved Conspicuity of Abdominal Lesions with Single-Source Dual-Energy MDCT Hadassah Hebrew University Medical Center Jerusalem, Israel Ruth Eliahou.
State of Michigan Computed Tomography Regulations
Radiation Dose in CT Scanning: A Team Approach
Managing Patient Dose in Computed Tomography (CT)
Key CT Parameters - What Are They Called and What Do They Mean?
Lower Dosage CT-guided Lung Biopsy Protocol Maintains Quality, Minimizes Exposure Jeremy Collins, MD Pegah E, Lewandowski RJ, Yaghmai V, Nemcek jr AA,
Seeram Chapter 13: Single Slice Spiral - Helical CT
RAD309 Patient Dose.
Radiation Protection and Diagnostic Imaging Physicist
1 Physics Testing for Performance Based Protocol version 3.0 (Now based on ACRIN 6678) QIBA Group 1C.
5 th September 2005 Paul Collins Computed Tomography Dosimetry Assessment of Effective Dose in Computed Tomography using an Anthropomorphic Phantom Paul.
Novel single-source high-pitch protocol for CT angiography of the aorta: comparison to high-pitch dual-source protocol in the context of TAVI planning.
Business: Cell/Text: Presented by: Thomas J. LaRocca, M.S., D.A.B.R. Medical Physicist
Kalpana Kanal, Ph.D., DABR Associate Professor, Diagnostic Physics
Dose Survey in Computed Tomography DS /CM Kampala IAEA/RCA Kampala.
International Atomic Energy Agency L 4 PROTECTION ISSUES IN CLINICAL METHODOLOGY.
Radiation Dosimetry of the Patient
Optimizing Patient Radiation Dose
Radiation Protection of Patients Unit
Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.
Strategies for Reducing Radiation Dose in CT. Source: IMV Medical Information division 2004 CT Census.
Pratik Mehta, B.A. 1 ; Kate MacGregor, M.P.H. 2 ; Robert Gould, D.Sc. 2 ; Diana Miglioretti, Ph.D. 3 ; J. Anthony Seibert, Ph.D. 4 ; Rebecca Smith-Bindman,
CT physics and instrumentation
Catherine Ashmore Lorna Sweetman North Western Medical Physics Elaine Holt.
BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina.
THE DICOM 2013 INTERNATIONAL CONFERENCE & SEMINAR March 14-16Bangalore, India On handling low-dose CT images in the absence of reliable DICOM header information.
Low Dose Head Computed Tomography Protocol for Evaluation of Non-traumatic Emergencies in Children Rishi Mhapsekar, MD Marilyn J. Siegel, MD Robert McKinstry.
Image Gallery: Lesion detection on low dose head CT Sarabjeet Singh, MD Mannudeep K. Kalra, MD *Eugene J. Mark, MD *James Stone, MD James H. Thrall, MD.
JUSTIFICATION OF COMPUTERIZED TOMOGRAPHY EXAMINATIONS AND RADIATION RISKS IN EVERYDAY RADIOLOGICAL PRACTICE Darka R. Hadnađev, Olivera Nikolić, Sanja Stojanović.
IAEA International Atomic Energy Agency Optimization of Protection in Computed Tomography (CT)-What can radiographers do? IAEA Regional Training Course.
QIBA CT Volumetrics - Cross-Platform Study (Group 1C) March 18, 2009 Interclinic Comparison of CT Volumetry Quantitative Imaging Biomarker Alliance.
How we set a DRL An example using CT David Sutton / Colin Martin Kampala IAEA/RCA Kampala.
Copyright © 2015 Lippincott Williams & Wilkins. Unauthorized commercial reproduction of this slide is prohibited Supplemental PowerPoint Slides Spine CT.
MEASUREMENTS OF RADIATION DOSES IN MULTISLICES COMPUTED TOMOGRAPHY EXAMINATIONS A. ELMAHDI*, A. SULIEMAN *Presenting author 1 Sudan Atomic Energy Commission,
Part No...., Module No....Lesson No
Quality Assurance.
1 Physics Testing for Performance Based Protocol version 2.1 QIBA Group 1C.
Part No...., Module No....Lesson No
What are the dose quantities in CT ? IAEA/RCA Kampala Kampala.
Impact of Contrast Media Concentration and kVp settings on Image Quality in CT Angiography of the Intracranial Vessels Birgitta Ramgren MD Roger Siemund.
Jordan Russell, Medical Student, FRCP (C), Brent Burbridge, MD, FRCP (C), Jennifer Tynan, MD.
RADIATION DOSE ISSUES.
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
Blinded by the Light The irradiation of lenses during routine CT head examinations across a NHS Trust; a quality improvement project Dr Jack Roberts, Dr.
Background Trauma Patients undergo an initial, “on admission” CT scan which includes: Non contrast brain Arterial phase full body scan Portal venous phase.
Introduction of a low dose protocol for thoracic ablation under CT guidance Thea Buchan, Jason Godbold, Dr Steve Bandula Interventional Oncology Service,
CT Multi-Slice CT.
Head computed tomography doses in four selected hospitals in Nigeria
Patient exposure trends and problems in implementing ALARA
Role of Adaptive Statistical Iterative Reconstruction (ASIR) in lowering radiation dose for pediatric head CT Electronic Poster-eP 140 Azadeh Tabari, MD.
European Journal of Radiology
How we set a DRL An example using CT
Sunday Case of the Day Physics
Neuro Best Contrast Filter for Head CT
Image acquisition in a modern CT scanner
Sample General Electric CT Dose Report from a CT of the chest, abdomen, and pelvis. The dose report contains information about the patient (deleted), type.
Optimisation of Chest Computed Tomography Using a Phantom: Impact of mAs and Reconstruction Techniques on Image Quality C.S. Reis1; T. Faqir2; V. Harsaker3;
CT dose optimization software synergy – A clinical perspective
Computed Tomography (CT) Dose Optimization
Physics Testing for ACRIN 6678 Protocol version 3.0 QIBA Group 1C
Exhibit Number: C19 Evaluation and Reduction of Head Computed Tomography Dose Because of full frame graphic use title slide sparingly because too many.
Wednesday Case of the Day
David J. Brenner, Maria A. Georgsson  Gastroenterology 
Yogesh Thakur, PhD, MCCPM, Thorarin A
What Is the Intrinsic Dose Reduction?
Presentation transcript:

Reduction of effective and organ dose to the eye lens in cerebral MDCT scans using iterative image reconstruction Zizka J, Jandura J, Kvasnicka T, Klzo L, Grepl J Dept. of Radiology Charles University Teaching Hospital Hradec Kralove Czech Republic ECR 2012

Introduction medical radiation exposure steadily risesmedical radiation exposure steadily rises CT accounts for > 2/3 cumulative exposureCT accounts for > 2/3 cumulative exposure it reached the level of natural radiation background in the USAit reached the level of natural radiation background in the USA ALARAALARA dose reductiondose reduction –iterative reconstruction Source:

Iterative reconstruction (IR) significantly lower image noise significantly lower image noise radiation dose reduction (25 – 60 %) radiation dose reduction (25 – 60 %) FBP IR FBP IR full dose 50 % dose

Purpose to compare effective radiation dose and dose to the eye lens in MDCT brain examinations utilizing either standard filtered back projection (FBP) or iterative reconstruction in image space (IRIS)

Material and Methods 400 routine adult brain CT exams:400 routine adult brain CT exams: –200 with FBP –200 with IRIS single source MDCT scannersingle source MDCT scanner –collimation: 64 x 2 x 0.6 mm –tube voltage: 120 kV (FBP & IRIS) –reference mAs: 300 (FBP) and 200 (IRIS) –rotational period 1 s –pitch 0.55 –automatic dose modulation switched on

Material and Methods recorded wererecorded were –CT Dose Index (CTDIvol) –Dose Length Product (DLP) –effective dose calculated by ImPACT software –organ dose to the lens was derived from the actual tube current-time product value applied to the slices with lens included –image noise (SD) in a 1.5 cm 2 ROI in the centrum semiovale qualitative visual assessment: two experienced radiologists blinded to the type of image reconstruction using a visual analogue scalequalitative visual assessment: two experienced radiologists blinded to the type of image reconstruction using a visual analogue scale 1 = very low noise, optimal diagnostic quality 2 = low noise, good diagnostic quality 3 = increased noise, diagnostic quality 4 = high level noise, limited diagnostic quality 5 = unacceptable noise, non-diagnostic scan

Results: radiation dose FBP (n = 200) IRIS (n = 200) Reduced against FBR by CTDIvol [mGy] % DLP [mGy.cm] % Effective dose [mSv]1.47 ± ± % Organ dose to the eye lens [mGy] 40.0 ± ± %

CTDI = 60 mGy European Commission Quality Criteria for MDCT % - 58 %

Results: image quality FBPIRISp-value image noise [median] < 0.01 image quality [mean] (kappa = 0.76)

FBP IRIS FBP IRIS dose reduced by 34 % dose reduced by 34 % Image quality

Discussion ICRP ref

eye lens radiosensitivity is higher than previously thoughteye lens radiosensitivity is higher than previously thought the lens is frequently exposed to the primary beam in spiral MDCT scansthe lens is frequently exposed to the primary beam in spiral MDCT scans the organ dose to the lens leading to cataract formation (0.5 Gy) might be acquired in as low as 7 (!) non- optimized CT head scans (CTDI ≈ 60 mGy)the organ dose to the lens leading to cataract formation (0.5 Gy) might be acquired in as low as 7 (!) non- optimized CT head scans (CTDI ≈ 60 mGy) with optimized iterative reconstruction acquisition the number of CT head scans leading to lens opacities formation rises approx. 3-fold (20 MDCT head scans)with optimized iterative reconstruction acquisition the number of CT head scans leading to lens opacities formation rises approx. 3-fold (20 MDCT head scans) Discussion

even in the settings of optimized FBP based CT protocols, iterative reconstruction is capable of further substantial radiation dose savings which, in our population, further reduced both the effective and eye lens dose by more than 33 % compared to optimized FBP protocols and by 58 % compared to reference standardeven in the settings of optimized FBP based CT protocols, iterative reconstruction is capable of further substantial radiation dose savings which, in our population, further reduced both the effective and eye lens dose by more than 33 % compared to optimized FBP protocols and by 58 % compared to reference standard Conclusion