Speech Audiometry. In addition to pure-tone audiometry, speech audiometry is a set of behavioral tests that provide information concerning sensitivity.

Slides:



Advertisements
Similar presentations
Strategies & Techniques Facilitating childrens listening and language skills in the classroom.
Advertisements

Audiograms What are they? Why have them? How do I read them?
Pure Tone Audiometry Training Course ——Marketing Dept.
Indications of non-organic hearing loss
Thresholds, Recognition Tests, MCL &UCL
Chapter 8 Auditory Training Perry C. Hanavan, Au.D.
Chapter Eleven Individuals with Hearing Impairments.
Room Acoustics: implications for speech reception and perception by hearing aid and cochlear implant users 2003 Arthur Boothroyd, Ph.D. Distinguished.
Pure Tone Audiometry Basic Principles of Sound The Audiometer
The nature of sound Types of losses Possible causes of hearing loss Educational implications Preparing students for hearing assessment.
Vocal Emotion Recognition with Cochlear Implants Xin Luo, Qian-Jie Fu, John J. Galvin III Presentation By Archie Archibong.
Pre-operative evaluation and post-operative rehabilitation for paediatric cochlear implantation Han Demin, M.D., Ph.D. Beijing Institute of Otolaryngology.
Profile of Phoneme Auditory Perception Ability in Children with Hearing Impairment and Phonological Disorders By Manal Mohamed El-Banna (MD) Unit of Phoniatrics,
CSD 2230 HUMAN COMMUNICATION DISORDERS Topic 2 Normal Communication Development and Communication Across the Lifespan.
Audiological Evaultions
Interrupted speech perception Su-Hyun Jin, Ph.D. University of Texas & Peggy B. Nelson, Ph.D. University of Minnesota.
TOPIC 4 BEHAVIORAL ASSESSMENT MEASURES. The Audiometer Types Clinical Screening.
1 New Technique for Improving Speech Intelligibility for the Hearing Impaired Miriam Furst-Yust School of Electrical Engineering Tel Aviv University.
CSD 5400 REHABILITATION PROCEDURES FOR THE HARD OF HEARING Auditory Training.
The use of FM systems with Cochlear Implants- How has research had an impact on practice? Sarah Flynn and Elizabeth Wood South of England Cochlear Implant.
PURE TONE AUDIOMETRY BALASUBRAMANIAN THIAGARAJAN DRTBALU'S OTOLARYNGOLOGY ONLINE.
CSD 3000 DEAFNESS IN SOCIETY Topic 6 The Deaf Adult.
Pseudohypacusis/ Nonorganic/ Functional Hearing Loss:
Measurement of Hearing and Audiogram Interpretation
Summer 2007 SPA 4302 Pure Tone Audiometry.
Modernising Children’s Hearing Aid Services Sound Field Testing MCHAS TEAM Wave 4 SFR 17/05/04.
Audiology Training Course ——Marketing Dept. Configuration of the ear ① Pinna ② Ear canal ③ Eardrum ④ Malleus ⑤ Incus ⑥ Eustachian tube ⑦ Stapes ⑧ Semicircular.
The Dream Team - The Playbook for Partnering Parents and Professionals in Early Intervention Monique Baker Jane Madell
A Child with a Hearing Impairment, Including Deafness ECEA Disability Category, Definition and Eligibility Criteria CDE Eligibility Training Slides March.
Fitting and Evaluation of FM Systems for Cochlear Implant Users Linda M. Thibodeau.
Speech and Language Impairments Erin Biermacher Kristen Yankoviak.
Comparisons of Word Recognition Performance in Normal-Hearing Children A Pilot Project by Tiffany Skinner and Stephanie Taylor Spring 1999.
Audiograms How to read them and what they are.. Terms to know: Audiogram An audiogram is a means of recording the results of a hearing test. It will include.
Cochlear Implants American Sign Language Children & Cochlear Implants Psychological Evaluation of Implant Candidates James H. Johnson, Ph.D., ABPP Department.
METHODOLOGY INTRODUCTION ACKNOWLEDGEMENTS LITERATURE Low frequency information via a hearing aid has been shown to increase speech intelligibility in noise.
Clinical Applications
Healthcare Communications Shannon Cofield, RDH. Essential Question How can communication affect patient care?
Dynamics of speech the diagnostic audiometer test environment patient’s and clinician’s role speech-threshold testing most comfortable loudness and uncomfortable.
Pure Tone Audiometry SPA 4302 Summer A, The Pure-Tone Audiometer Electronic device that generates tones for determining hearing thresholds Manufactured.
CSD 2230 HUMAN COMMUNICATION DISORDERS Topic 5 Hearing Disorders and Hearing Loss Assessment.
Speech Perception 4/4/00.
Speech Audiometry SPA 4302 Summer A, The Diagnostic Audiometer Equipped with Inputs for microphones, cassette tapes, or CDs Volume unit (VU) meters.
Pure Tone Audiometry most commonly used test for evaluating auditory sensitivity delivered primarily through air conduction and bone conduction displayed.
Autistic Spectrum Disorders Awareness Raising Information for health professionals.
 The infant will respond to stimuli.  They will become acquainted with their parent’s voices  They will become startled by loud noises, and be comforted.
Speech Audiometry SPA 4302 Summer The Diagnostic Audiometer Equipped with Inputs for microphones, cassette tapes, or CDs Volume unit (VU) meters.
Functional Listening Evaluations:
Chapter Eleven Individuals With Speech and Language Impairments.
Katherine Morrow, Sarah Williams, and Chang Liu Department of Communication Sciences and Disorders The University of Texas at Austin, Austin, TX
Speech Audiometry. Pure Tone Audiometer Schematic of Speech Audiometer.
AUDIOMETRY An Audiometer is a machine, which is used to determine the hearing loss in an individual.
Figures for Chapter 8 Candidacy Dillon (2001) Hearing Aids.
HEARING- 3. LEARNING OBJECTIVES LEARNING OBJECTIVES Discuss the principles used in performing tests of hearing Discuss the principles used in performing.
Dr Guita Movallali. How does Cued Speech help speech? Speech is much more complex than the ability to make speech sounds. It is necessary to know how.
Test procedure Several different methods followed. Several different methods followed. Most standard: Recommended by Carhart and Jerger (1959) and ASHA.
Speech Audiometry Lecture 8.
Equipment for Measuring Hearing and Calibration
Chapters 4 & 5 Behavioral Testing
LANGUAGE (Speech/Language Impaired)
Audiograms Degree, Type and Configuration
ABR measures Absolute latency Inter-peak latency Wave amplitudes
Chapters 4 & 5 Behavioral Testing
Chapters 4 & 5 Behavioral Testing
Masking for SRT Crosshearing suspected when SRTTE – IA ≥ Best BCNTE
Do I need an Implant Now? Or Wait? Things to Consider.
Masking for BC Initial masker level for BC = ACNTE + OE , where OE is the occlusion effect Use of insert earphones: Smaller occlusion effect, greater IA.
Spondees Word with two syllables pronounced with equal stress.
BEST SPEECH TESTING IN DELHI Contact us Source Article- testing/
Understanding the Audiogram
Tinnitus activities THERAPY
Presentation transcript:

Speech Audiometry

In addition to pure-tone audiometry, speech audiometry is a set of behavioral tests that provide information concerning sensitivity to speech materials and understanding of speech at supra-threshold levels. It can be used diagnostically to examine speech-processing abilities throughout the auditory system, and it can be used to crosscheck the validity of pure-tone thresholds.

Speech Audiometry Stimuli: Recorded or monitored live voice. Live is quicker and more flexible but recorded is more consistent and controlled. Live presentations also have the advantage of building quick rapport through verbal interplay. Patient's Role: Patient must be familiar with words being used and have ability to respond via verbal or written reply or identification of an object/ picture (especially for children or aphasics).

Speech Audiometry Clinician's Role: Give good directions! Don't give visual cues (such as lip movement) while doing testing especially for live speech presentations. Test environment: Two-channel audiometer with facilities for input from microphones and CD players/external devices. Two-room setting.

Types of tests Speech Threshold Testing: 2 types SDT (speech detection threshold) SRT (speech recognition threshold)

Speech Detection Threshold (SDT) Also called speech awareness threshold (SAT). Objective - to obtain the lowest level at which speech can be detected at least half the time. This test does not have patients repeat words; it requires patients to merely indicate when speech stimuli are present.

SDT The SDT is especially useful for patients too young to understand or repeat words. Sentences are preferred over isolated words. The speech should be presented monotonously Cold running speech The SDT may also be used for patients who speak another language or who have impaired language function and/or speech disorder.

SDT Present continuous discourse and raise and lower level by bracketing (down 10, up 5) until you find the level at which they detect it. Patient may respond verbally or by pushing a button or clinicians observes for facial/ body expressions (for infants).

SDT For patients with normal hearing or somewhat flat hearing loss, this measure is usually dB better than the speech-recognition threshold (SRT). For patients with sloping hearing loss, this measurement can be misleading with regard to identifying the overall degree of loss. The SDT will be closely related to the threshold for the lower frequency, and it will not indicate greater loss in higher frequencies.

The Ling Six-sound Check Is a behavioral listening check to determine a cochlear implant’s or hearing aid’s effectiveness. The sounds ah, ee, oo, sh, s, and mm indicate a child’s ability to detect all aspects of speech as these six sounds encompass the frequency range of all phonemes. This check can be used to determine what sounds the patient is able to detect, discriminate, and identify. Detection - Recognizing the presence or absence of sound Discrimination - Discerning if two or more sounds are the same or different Identification - Reproducing a sound or pointing to a picture of the sound heard

The Ling Six-sound Check Can be presented as live voice (calibrating the distance between the speaker and patient) or via an audiometer. Ling Sounds

Speech Recognition Threshold (SRT) Also referred to as the speech reception threshold. The objective of this measure is to obtain the lowest level at which speech can be identified at least half the time. Preferred over SDT. Typically use spondaic words. Spondee. Two syllables, both pronounced with equal stress and effort. E.g. baseball, hot dog, toothbrush, cowboy.

SRT Method Use live voice or prerecorded stimuli. May or may not use a carrier phrase "say the word". Familiarize patient with word list to ensure that all the words used are within the vocabulary of the patient. After you familiarize them, use the tracking procedure to determine the threshold. Although not commonly used, SRTs can be tested with BC. Children who do not like to ‘play’ with tones.

SRT Method SRT are used to validate pure-tone thresholds because of high correlation between the SRT and the PTA. In clinical practice, the SRT and PTA should be within +/- 6 dB. Other clinical uses of the SRT include establishing the sound level to present supra- threshold measures (such as word recognition test) and determining appropriate gain during hearing aid selection.

Relationship between Pure-tone thresholds and Speech Recognition performance - Articulation Index (AI) The Articulation Index or AI gives a measure of the intelligibility of hearing speech in a given noise environment. An adoption of the AI concept is to quantify communication difficulties based on pure-tone thresholds. The basic interpretation of the AI value is the higher the value then the easier it is to hear the spoken word.

Count the dot method - Mueller and Killion (1990) Each dot represents 1% of the information contributing to speech clarity. The number of dots that are audible predict how well one understands quiet speech from a six foot distance. The dots are unevenly distributed, with many more of them filling in the gray zone between 1000 and 3000 Hz than in the 250 to 500 Hz area.

Count the dot method - Mueller and Killion (1990) Count the dots that are below audiometric threshold. The more dots that are below threshold, the better one is able to hear normal conversational speech. The fewer dots that below threshold, the more difficulty hearing conversational speech.

Articulation Index At least three clinical uses of the Articulation Index: To predict from the audiogram the amount of disability for normal-level conversational speech, Predicting the benefit that will be obtained from a hearing aid, and Comparing the benefit from one aid to another.

Supra-threshold measures Most Comfortable Loudness level The intensity level of speech that is most comfortably loud is called the most comfortable loudness level (MCL) test. Typically dB above SRT. This sensation level is reduced for many patients who have sensorineural hearing loss (SNHL). Use continuous speech. Gradually increase levels, starting at SRT. At each level, ask "too soft", "too loud", "comfortable". Bracket around. Monaural/binaural headphones or loudspeakers

Uncomfortable Loudness Level (UCL) This level provides the maximum level at which word-recognition tests can be administered. UCL can also indicate maximum tolerable amplification. Another reason to establish UCL is to determine the dynamic speech range. Dynamic range represents the limits of useful hearing in each ear and is computed by subtracting SRT from UCL.

UCL For many patients with SNHL, this range can be extremely limited because of recruitment (abnormal loudness growth – commonly seen due to a cochlear pathology). For normals, usually dB SL (ref SRT). Testing similar to MCL but to indicate when speech level becomes uncomfortable. Patient is instructed to signal when level is intolerable.

Word Recognition Testing The primary purpose of supra-threshold word-recognition testing is to estimate ability to understand and repeat single-syllable words presented at conversational or another suprathreshold level. Another use is to verify speech-recognition improvements achieved by persons with hearing aids. Comparing performance with and without hearing aids

Stimuli Phonetically balanced words A word list containing all of the phonetic elements of connected English in their normal proportion to each other. A number of different standardized word lists are available. PB-50, CID W-22 lists, NU-6 lists, PBK words (K-kindergarten).

Word Recognition Scores Words can be presented via tape, CD, or monitored live voice. Patients are asked to repeat words to the audiologist. Present PB words at a fixed level either MCL or dB above SRT. Typically present lists of 25 or 50 words. Get a percent correct score.

What does the score mean? Evaluation of Word Recognition Scores: % = good 75-90%= slight difficulty 60-75%= moderate difficulty 50-60%=poor. Difficulty in conversational speech. Less than 50% = very poor. Probably unable to follow running speech.

What can we do with kids? Prefer a closed-set testing than a open-set testing (ex, PB-K word lists). Some closed-set tests include Word Intelligibility by Picture Identification (WIPI): Child is presented with 6 pictures on a card and asked to point to the one that is said. Northwestern University Children's Perception of Speech (NUCHIPS): Similar to WIPI except 4 pictures appear on each card. Pediatric Speech Intelligibility test: Pictures

Problems with word recognition testing Problems with Word Recognition: - Poor test-retest reliability for SNHL. -Difficult to relate to everyday life. Suggested to test word recognition with competition (noise). Need to specify the Signal-to-Noise ratio (SNR or S/N) – difference in speech level and noise level. Different noise stimuli such as modulated noise, speech noise, multitalker babble, etc.

Review Speech recognition threshold Speech detection threshold Word recognition UCL MCL Special tests for kids