8 / 6 / 2000Pixel 2000 /JPM1 X-Ray Medical Imaging and Pixel detectors PIXEL 2000 Genova, June 5-8 th 2000 J.P.Moy,, Moirans, France TRI ELLX.

Slides:



Advertisements
Similar presentations
DIGITAL RADIOGRAPHY.
Advertisements

Radiologic Technology
X-Ray Astronomy Lab X-rays Why look for X-rays? –High temperatures –Atomic lines –Non-thermal processes X-ray detectors X-ray telescopes The Lab.
Digital Radiography.
RAD 354 Chapt. 26 Digital Imaging Many types/names for the digital imaging to come – Types CR: Barium fluorohalide PSP SPR (scan projection RAD): Nal scintillator/photodiode.
Digital Radiographic Imaging 101
Alexander Klyachko, IU Cyclotron, PTCOG51, Seoul, South Korea, May 17-19, 2012 A.V. Klyachko 1, D.F. Nichiporov 1, L. Coutinho 2, C.-W. Cheng 2, 3, M.
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods
BMME 560 & BME 590I Medical Imaging: X-ray, CT, and Nuclear Methods X-ray Imaging Part 3.
BMME 560 & BME 590I Medical Imaging: X-ray, CT, and Nuclear Methods
tomos = slice, graphein = to write
Digital Technology 14.2 Data capture; Digital imaging using charge-coupled devices (CCDs)
8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS
Direct Digital Radiography or Direct Capture Radiography Bushong Ch. 27.
Chapter 1 Introduction to Digital Radiography and PACS
TERPSSC 2001Robert M. Gagne OVERVIEW OF DIGITAL IMAGING AND RADIATION PROTECTION ISSUES Robert M. Gagne MICAB/DECS/OST [From: Handbook.
CCD Detectors in High- Resolution Biology Jian Guan
Most of the images recorded during conventional radiography are obtained with film/screen combination image receptors. Which in lessens the patient dose.
Overview of Scientific Imaging using CCD Arrays Jaal Ghandhi Mechanical Engineering Univ. of Wisconsin-Madison.
ADC PRE-PATIENT COLLIMATION POST-PATIENT COLLIMATION.
Digital Image Characteristic
Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.
Digital Imaging Systems –I/O. Workflow of digital imaging Two Competing imaging format for motion pictures Film vs Digital Video( TV) Presentation of.
Photon detection Visible or near-visible wavelengths
X-Ray Medical Imaging Physics – IB Objectives
Chapter 6: Digital Radiographic Imaging
Medical Image Analysis Medical Imaging Modalities: X-Ray Imaging Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan,
Computers in Imaging Q & A Robert Metzger, Ph.D..
DIGITAL PROJECTION RADIOGRAPHY COMPUTED RADIOGRAPHY
Image Quality Radiographic Resolution.
Digital Radiology. 2 Aim : To become familiar with the digital imaging techniques in projection radiography and fluoroscopy.Aim : To become familiar with.
DIGITAL RADIOGRAPHY DA 105.
1 Components of Optical Instruments Lecture Silicon Diode Transducers A semiconductor material like silicon can be doped by an element of group.
Rad Tech 265 More Digital Imaging. Digital Fluoroscopy Radiation dose –Patient dose for DF is significantly less than conventional fluoro At 7.5 pulses/second.
Digital Imaging. acronyms 1.PSP 2.CRT 3.ADC 4.IP 5.CR 6.DR 7.PACS 8.SNR 9.CNR 10. CCD 11. FOV 12. LUT 13. DICOM 14. RIS 15. HIS 16. TFT 17. DQE 18. DAP.
Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to.
Digital Radiography and PACS By Professor Stelmark.
© Jimoid.com 2005 Imaging Basics A medical image is a 2D or 3D distribution of signals which represent properties of an object. The purpose of medical.
Elements of DR Imaging Systems
1 INTRODUCTION TO THE PHYSICS OF DIAGNOSTIC IMAGING Outline of Course Brief History Common Terminology Imaging Modalities.
High resolution X-ray analysis of a proximal human femur with synchrotron radiation and an innovative linear detector M.Bettuzzi, R. Brancaccio, F.Casali,
Digital imaging By : Alanoud Al Saleh. History: It started in 1960 by the National Aeronautics and Space Administration (NASA). The technology of digital.
Medical applications of particle physics General characteristics of detectors (5 th Chapter) ASLI YILDIRIM.
IMAGE RECEPTORS. Follow the Sequence- Film System Tube Tube Patient Patient (bucky) or non-bucky (bucky) or non-bucky Cassette Cassette INTENSIFYING SCREEN.
Fluoroscopy. Real-time imaging Most general-purpose fluoroscopy systems use TV technology, operating at 30 frames/sec May be recorded (barium swallow.
Neutron Imaging for the Hydrogen Economy Neutrons see material differently than x-rays The fine details of the water in this Asiatic Lily are clear to.
DIGITAL IMAGING.
Fluroscopy and II’s. Fluroscopy Taking real time x-ray images Requires very sensitive detector to limit the radiation needed Image Intensifier (II) is.
Digital Radiology Dr M A Oghabian Medical Physics Group Tehran University of Medical Sciences
Image Receptor Unsharpness By Professor Stelmark.
Nuclear Medicine Instrumentation 242 NMT 1 Dr. Abdo Mansour Assistant Professor of radiology
DIGITAL RADIOGRAPHY.
Direct Digital Radiography or Direct Capture Radiography
Acronyms. A SEQUENCE OF COMPUTER OPERATIONS FOR ACCOMPLISHING A SPECIFIC TASK Algorithm.
Analog X-ray Imaging Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS By: Maisa Alhassoun
Medical Physics.
Copyright © 2012, 2006, 2000, 1996 by Saunders, an imprint of Elsevier Inc. 1 Digital Imaging Instead of Radiography.
Introduction to Medical Imaging Week 2: X-ray and CT
History of digital radiology 1980 Kinos L.T. ET all developed a portable radiographic X-ray camera made of tantalum and aluminum. Digital radiography.
(Chapter 7) CHEM–E5225 Electron microscopy
Imaging Characteristics
Digital Imaging.
Rad Tech 265.
Overview of CR and DR Gyeongsang National University Hospital
כיצד נרכשת התמונה בסרט הרנטגני?
COMPUTED RADIOGRAPHY -Dr.SHEFALI MESHRAM
Chapter 6 Cassetteless Equipment and Image Acquisition
Digital Imaging.
Computed Tomography (C.T)
Presentation transcript:

8 / 6 / 2000Pixel 2000 /JPM1 X-Ray Medical Imaging and Pixel detectors PIXEL 2000 Genova, June 5-8 th 2000 J.P.Moy,, Moirans, France TRI ELLX

8 / 6 / 2000Pixel 2000 /JPM2 OUTLINE - X-ray medical imaging. The requirements, some particular features - Present detectors. - The new X-ray Flat detectors scintillator and photoconductor approach - How can pixel detectors help medical imaging? The detecting material, the readout circuit CONCLUSIONS

8 / 6 / 2000Pixel 2000 /JPM3 X-ray Imaging in Medicine : Radiography, Fluoroscopy, Computed Tomography (1) The oldest medical imaging technique : projection radiography discovered by Röntgen in 1895 : About systems in the world: best for bones, but also widely used for soft tissues, often with contrats agents, such as barium sulfate for gastro-intestinal imaging. Mammography is a particular case, as it concerns soft tissues and aims at the detection of very fine calcifications.

8 / 6 / 2000Pixel 2000 /JPM4 X-ray Imaging in Medicine : Radiography, Fluoroscopy, Computed Tomography (2) Fluoroscopy : Originally visual observation of the fluorescent screen. Now with electronic image converters : XRII Angiography is a particularly important application of fluoroscopy : imaging blood vessels after injection of an iodine compound in an artery to detect stenosis or other pathologies. Computed tomography is a 3D imaging technique based on the reconstruction of the object from many linear projections. At present, it does not rely on imaging detectors

8 / 6 / 2000Pixel 2000 /JPM5 X-ray Imaging in Medicine : competition with Ultra-Sound, Magnetic Resonance Imaging ? Imaging techniques without ionizing radiation will certainly develop at the expense of X-rays : - US is easy to use and cheaper than other modalities. - MRI provides invaluable information on soft tissues, and is becoming fast enough to adress cardiac imaging, but will remain expensive. X-rays will definitely remain for many years the most practical and cost effective imaging technique for bones, joints, and mammography.

8 / 6 / 2000Pixel 2000 /JPM6 Physical limitations Absorbed photon flux = 10 6 photons /mm² 0,01 0, Object size (µm) contrast (%) 10 6 absorbed photons/mm² for S/N = 30 S/N = 10 S/N = absorbed photons /mm² S/N = 3 After M.Arques, JRI 97 Poisson statistics imply a trade off between size-dose-contrast. For instance, a 100 µm detail with 10 % contrast will be detected with a 10:1 Signal to Noise ratio only if the photon flux exceeds 10 6 photons /mm² (with an ideal detector)

8 / 6 / 2000Pixel 2000 /JPM7 X-ray image sampling The image from a digital detector is spatially sampled, and therefore must comply with the laws of sampling : Neither signal nor noise spectra should exceed (Nyquist) Failure to comply with this law results in aliasing. A spatial response of the converter layer smaller than the pixel is deceptive : the noise spectrum extends well beyond the Nyquist limit, so that it piles up in the [0- ] range. When the spatial response stops at the Nyquist limit, signal and quantum noise are filtered by the same MTF, and the input S/N is preserved as long as the other noises remain small.

8 / 6 / 2000Pixel 2000 /JPM8 Simulated images : photoconductor and scintillator based detectors Photoconductor, PSF = Pixel 500 µm CsI, measured PSF

8 / 6 / 2000Pixel 2000 /JPM9 DETECTIVE QUANTUM EFFICIENCY : A measure of how well X-rays are used MTF Readout noise Quantum Noise Dose X-ray Energy X-ray absorption DQE

8 / 6 / 2000Pixel 2000 /JPM10 Compared requirements for RADIOGRAPHY and FLUOROSCOPY

8 / 6 / 2000Pixel 2000 /JPM11 The present detectors in Radiography (1) Film At present, the most widely used detection scheme is the screen-film. A light sensitive silver halide film is sandwiched between two radioluminescent screens, usually made of Gd 2 O 2 S:Tb powder in a binding agent. The sensitivity vs resolution trade-off results from : - the thickness of the absorbing screen, - the light absorption or reflection of the backing layer, - the size of the grains in the screen.

8 / 6 / 2000Pixel 2000 /JPM12 The present detectors in Radiography (2) Screen - Optics - CCD Based on existing elements. Possible extension to dynamic imaging. The basic obstacle is to get more than 1 el. /X-ray in the CCD ("Quantum sink" situation ) - According to the laws of optics the collection of light decreases as 1/demagnification². Coupling a 20 cm screen to a 2 cm CCD results in a very poor light collection - Fiber optics are the best way to couple a screen to a CCD (but the most expensive...) Some optical gain is necessary : X-ray Image Intensifiers

8 / 6 / 2000Pixel 2000 /JPM13 The present detectors in Radiography (3) Storage Phosphors Electrons created by the absorption of X-rays are stored as a latent image in a screen. It is then read by laser scanning Provides a digital image with a very broad dynamic range. Handled like film : Thin, identical formats and read time, disposable if damaged. Image quality and resolution comparable to that of sreen-films. Single reading station for several units. Not suitable for fluoroscopy

8 / 6 / 2000Pixel 2000 /JPM14 The present detectors in fluoroscopy X-ray Image Intensifiers are widely used. They offer an unequaled range of performance : X-ray detection efficiency close to the theoretical limits, Excellent S/N, even for very low X-ray flux Large size, up to Ø 400 mm Dynamic imaging capability, Zooming Mature technology, affordable However, they are bulky, especially for large diameters, and suffer from strong geometrical and magnetic distortion.

8 / 6 / 2000Pixel 2000 /JPM15 Operation of an X-ray image intensifier G1 G2 G3Anode Metal vacuum bottle CsI input screen Aluminum input window Output window P20 output screen Gain : input screen = 200 el. / X-photon Gain : output screen = 1000 vis. photons / el. Total gain = vis. photons / X-photon vis. photons / X-photon Lens Camera X-ray photocathode

8 / 6 / 2000Pixel 2000 /JPM16 X ray Image Intensifiers from TTE

8 / 6 / 2000Pixel 2000 /JPM17 X-ray Flat Detectors, the emerging technology Two approaches : - The scintillator/visible image sensor - The photoconductor/charge sensor Both have led to commercial systems. So far, only amorphous silicon can be obtained in the required sizes. Image sensors as well as charge detection arrays can be built with a technology derived from that of LCD active matrices An assembly of standard single crystal Si circuits is also possible, but such tiling results in challenging technical obstacles.

8 / 6 / 2000Pixel 2000 /JPM18 Readout Architecture bias Line drivers PD PC bias Line drivers Photoconductorscintillator / Photodiode Charge amplifiers Multiplex, coding

8 / 6 / 2000Pixel 2000 /JPM19 The Photoconductor based pixel Data column TFT a-Si gate Se -HV bias e-e- h

8 / 6 / 2000Pixel 2000 /JPM20 Cross-section of a scintillator-photodiode-TFT pixel Data column a-Si TFT Photodiode gate CsI:Tl Bias column

8 / 6 / 2000Pixel 2000 /JPM21 Photodiode quantum efficiency and CsI:Tl fluorescence spectrum 0% 20% 40% 60% 80% 100% wavelength (nm) Photodiode quantum efficiency CsI:Tl emission (nb photons)

8 / 6 / 2000Pixel 2000 /JPM22 Energy absorption of different materials (standard DN spectra, escape taken into account)

8 / 6 / 2000Pixel 2000 /JPM23 Compared Performance of scintillator based detectors N.A.= not available. 1 Varian 99, 2 Philips 98, 3 dpiX 98, 4 Canon 98, 5 Trixell 98, 6 General Electric 2000 RQA5 is a standard for X-ray quality : 70 kV DC on the X-ray tube, 23 mm of Al filtration to simulate the patient.

8 / 6 / 2000Pixel 2000 /JPM24 Compared performance of photoconductor based detectors N.A.= not available. 1 Toshiba 99, 2 Sterling 98, 3 University of Toronto 98. RQA5 is a standard for X-ray quality : 70 kV DC on the X-ray tube, 23 mm of Al filtration to simulate the patient.

8 / 6 / 2000Pixel 2000 /JPM25 Commercial devices Clinical tests have been performed for many years, and several manufacturers are now starting the production : Scintillator screen / a-Si array : TRIXELL, GEMS, CANON Selenium : KodaK-HOLOGIC (formerly STERLING) According to the manufacturers, various applications are (or will soon be) covered : General and chest radiography, mammography, cardiac angiography

8 / 6 / 2000Pixel 2000 /JPM26 Pixium 4600 and radiographic table

8 / 6 / 2000Pixel 2000 /JPM27 Thorax image with a pixium TRI ELLX

8 / 6 / 2000Pixel 2000 /JPM28 The benefits of the new X-ray Flat detectors Improved conditions : Immediate readout. The patient no longer waits in painful positions for the development of the film, and a new shot. The clinician has easier access to the patient during intervention. Reduced running cost (increased throughput of radiology rooms, no film, no chemicals, no waste processing, cheaper storage of data). Less dose: depending on the device, the required dose is 100 to 40% of the film dose (for a given S/N in the image). All the advantages of a digital image : processing, transfer, archiving, access to Computer Aided Diagnosis,....

8 / 6 / 2000Pixel 2000 /JPM29 The weak points of the new X-ray Flat detectors High investment costs (detector + image display & process.), because the detector relies on specific techniques (a-Si photodiodes, converter material,...) and requires the assembly of many expensive components. Difficult image corrections : offset and gain correction accuracy limited by small non-linearities and drifts. At very low dose (fluoroscopy), obtaining a S/N comparable to that of XRII requires extreme care (costly!)

8 / 6 / 2000Pixel 2000 /JPM30 What next? Development time extremely long : The work on the present generation started in the mid eighties => It is time to prepare the next generation ! Which improvements are worth a new development? A spectacular improvement in resolution or dose is unlikely. Reduce manufacturing costs without compromising on performance : make it simpler! Increase the S/N in fluoroscopy Open new modalities dual energy, tomosynthesis,...

8 / 6 / 2000Pixel 2000 /JPM31 The detecting material Obviously the cornerstone of future devices. Should combine : –Strong X-ray absorption from 20 to 150 keV, –large area deposition technique, –Chemical, thermal compatibility with Si, –high resistivity, –high µ  E, preferably with a low E, –low e-h creation energy (50 eV in Se, 5 eV desirable), –environmentally acceptable (HgI 2...?) At present, there is no consensus on a potential workhorse.

8 / 6 / 2000Pixel 2000 /JPM32 Can pixel detectors meet the requirements of medical imaging (1)? Single crystal silicon technology will soon reach the point where elaborate functions can be implemented in a µm pixel, with a realistic yield over a large area. A suitable converter material is still to be found : CdTe, PbI 2, HgI 2, PbO,.... Discrimination and counting in a pixel would open new possibilities : suppression of offset correction, dual energy, better linearity, no longer escape noise... However, it should be borne in mind that the counting rate will be huge : in the worst (but common) case where the patient does not cover the whole detector, ~10 7 photons/s hit each pixel.

8 / 6 / 2000Pixel 2000 /JPM33 Can pixel detectors meet the requirements of medical imaging (2)? Tiling will still be required for general radiography. Integration of driving and readout circuits will help to reduce manufacturing costs, but redundancy will be mandatory in order to reach reasonable yields. The better linearity should alleviate the task of matching the different tiles to avoid the checkerboard effects. Realistic assembly techniques are still to be found COST will most likely be the driving force, more than performance, as the pressure on health budgets will undoubtedly increase.

8 / 6 / 2000Pixel 2000 /JPM34 Conclusions X-ray imaging may benefit from the development of pixel detectors : - Simpler devices thanks to higher integration. - Rely on the standard Si technology. - Better S/N at very low doses. Besides the work on Si circuits, the need for a good X-ray converter is a prerequisite.