Filter Set Selection. It is best to follow this presentation with a Fluorophore excitation/emission maxima chart to hand.

Slides:



Advertisements
Similar presentations
Fluorescence and Confocal Microscopy
Advertisements

Bio-Rad Labs. Chromatic Aberration Sources and Measurement.
Avoiding bleed-through artifacts on the confocal microscope Fluorescent dyes are molecules that, when exposed to light of a specific range of wavelengths,
Absorbance spectroscopy
1 Molecular Luminescence Spectroscopy Lecture 31.
The Modern Atom Figure: 05-00CO Caption:
VISUALIZATION OF HYPERSPECTRAL IMAGES ROBERTO BONCE & MINDY SCHOCKLING iMagine REU Montclair State University.
Dave Piston May 18, 2014 Optical Sectioning 2: Confocal Designs and “Pseudo-confocals” Outline 1.PMT detectors 2.Laser Scanning Design Features 3.Pseudo-confocals.
Some structures Dansyl chloride 1,5-I-AEDANS Fluorescein isothiocyante ANS Ethidium bromide 5-[2-[(2-iodoacetyl)amino]ethylamino] naphthalene-1-sulfonic.
Introduction to LIDAR Mapping Technology
HFT 405, nm 405nm 488nm 405nm Only 405 and 488 are reflected All other wavelengths are transmitted 488nm 405nm This is the first beam splitter that.
USE AND CARE OF THE MICROSCOPE LECTURE 1. MICROSCOPY u Light Microscopy: any microscope that uses visible light to observe specimens u Compound Light.
STORM and PALM How to get started. June 2014 Christa Walther.
Applications of SPCE to Pharmaceutical Research Kathleen Hamilton, Tom Laue, and James Harper Presentation at the 2007 BITC Meeting University of New Hampshire.
Study of Protein Association by Fluorescence-based Methods Kristin Michalski UWM RET Intern In association with Professor Vali Raicu.
Diffraction.
Basics of Flow Cytometry Holden Maecker. Outline Definitions, what can be measured by flow cytometry Fluidics: Sheath and sample streams, flow cells,
Scanning and Image Processing -by Steve Clough. GSI Lumonics cDNA microarrays use two dyes with well separated emission spectra such as Cy3 and Cy5 to.
Practical aspects of fluorescence microscopy
Illumination and Filters Foundations of Microscopy Series Amanda Combs Advanced Instrumentation and Physics.
Are You FRETting? Find Out for Sure With FLIM Frequency Domain FLIM for Your Scope Intelligent Imaging Innovations.
Two-photon absorption standards in the nm excitation range: establishing a correction curve for accurate cross section calibration Nikolay S.
Preparation of Buffers - 1 Calculate the volume of sulfuric acid (H 2 SO 4 ) necessary to prepare 600 milliliter 0.5M H 2 SO 4 from concentrated H 2 SO.
Camera Excitation Filter Wheel Emission Filter Wheel Automated Interchangeable Dichroic Mirror Splitter Microscopy Body Lens Xenon Arc Lamp Sutter Controller.
Miriam Israelowitz 1 and Dr. David L. Wilson 2 1 Department of Physics, Case Western Reserve University, Cleveland OH, 2 Deparment of Biomedical Engineering,
- 1 - LSM 510 META AIM Carl Zeiss - PMLSM 510 META EF-Guided Tour April 2002 Guided tour through:  Configuration  Defining spectral range  Scan control.
Optical Characterization methods Rayleigh scattering Raman scattering transmission photoluminescence excitation photons At a glance  Transmission: “untouched”
Bios 532 Fluorescence. Vocabulary Fluorophore - molecule that emits fluorescence. Excitation - absorption of a photon. Emission - release of a photon.
Fluorometric Analysis
Relationship between A(OD) and %T Transmittance, T = P / P 0 % Transmittance, %T = 100 T Absorbance, A = log 10 P 0 / P A = log 10 1 / T A = log
Microscopes The invention of the microscope in the 17 th century led to the discovery of the cell. Robert Hooke described cells using this light microscope.
- 1 - LSM 510 META AIM Carl Zeiss - PMLSM 510 META EF-Guided Tour April 2002 Guided tour through:  Configuration  Defining spectral range  Scan control.
Lab 5 – Emission Spectra.
An instrument for magnifying very small objects
Molecular Luminescence Spectroscopy
Electron Microscope. How do they work Instead of using light they fire a beam of electrons (which have a wavelength less than 1nm compared to light which.
Status of PREMOS data SOLID Workshop 14th October 2013 G. Cessateur for the PREMOS team PMOD/WRC, Switzerland.
Constructing a Filter Cube
Weight Encoding Methods in DNA Based Perceptron 임희웅.
23W!~ Anti-vWF Immunofluorescence in Mouse Brain Minjung Choi, Katie Stockdale, James O. McNamara Department of Internal Medicine, The University of Iowa,
Date of download: 6/26/2016 Copyright © 2016 SPIE. All rights reserved. Horizontal noncontact FMT imaging system. (a) The FMT setup is illustrated, where.
Fluorescence and Fluorochromes Peter O’Toole Tel:
Biology 227: Methods in Modern Microscopy Andres Collazo, Director Biological Imaging Facility Yonil Jung, Graduate Student, TA Week 10 Multispectral Imaging.
Wave theory predicts diffraction of light (the spreading of light into a region behind an obstruction), but this is not easily observed unless the obstruction.
Starter: Microscopes Which image is from the light microsope? How do you know?
From: Coherent Anti-Stokes Raman Scattering (CARS) Microscopy: A Novel Technique for Imaging the Retina Invest. Ophthalmol. Vis. Sci ;54(5):
Absorbance spectroscopy
Anti-vWF Immunofluorescence in Mouse Brain
OPTICAL MONITORING OF PHOTOSENSITIZER DIFFUSION INTO TISSUE
Fluorescence Lifetimes
Which of the following graphs is the graph of the derivative of the function {image} . 1. {applet}
Indicate all x- and y-intercepts on the graph of the function y = x Choose the correct answer from the following: x-intercept (4,0), y-intercept.
Molecular Dynamics in Living Cells Observed by Fluorescence Correlation Spectroscopy with One- and Two-Photon Excitation  Petra Schwille, Ulrich Haupts,
Quantum Dots for Molecular Pathology
Vital imaging: Two photons are better than one
Volume 95, Issue 7, Pages (October 2008)
Volume 99, Issue 8, Pages (October 2010)
Diffraction and the Wave Model
Supraresolution Imaging in Brain Slices using Stimulated-Emission Depletion Two- Photon Laser Scanning Microscopy  Jun B. Ding, Kevin T. Takasaki, Bernardo.
Francesca Pennacchietti, Travis J. Gould, Samuel T. Hess 
Flow Cell Injector Tip Fluorescence signals Focused laser beam Sheath
Volume 90, Issue 10, Pages (May 2006)
Aquiles Carattino, Veer I.P. Keizer, Marcel J.M. Schaaf, Michel Orrit 
Volume 101, Issue 10, Pages (November 2011)
Harvard Medical School Neurobiology Imaging Facility
Characterization of the Photoconversion on Reaction of the Fluorescent Protein Kaede on the Single-Molecule Level  P.S. Dittrich, S.P. Schäfer, P. Schwille 
Polarized Fluorescence Resonance Energy Transfer Microscopy
Comparison of fluorescence spectra of cells expressing the FL-IFN-γR2/EBFP and FL-IFN-γR2/GFP chains in the presence and absence of the IFN-γR1 chain and.
A, Arrangement of LEDs at the excitation port of the microscope fluorescence illuminator. a, Arrangement of LEDs at the excitation port of the microscope.
Flow Cell Injector Tip Fluorescence signals Focused laser beam Sheath
Presentation transcript:

Filter Set Selection

It is best to follow this presentation with a Fluorophore excitation/emission maxima chart to hand.

Preliminary The MRC series has two dichroics They are numbered 1 and 2

Preliminary Dichroic number 1 is to the rear of the scan head It filters the excitation beam They are identified with names such as K1 or I1 etc. Also VHS/BHS

Preliminary Dichroic number 2 is to the front of the scan head It filters the emissions from the sample It is identified with a name such as K2 or I2 etc.

The MRC1000/1024 The MRC1000/1024 has extra filters They have excitation filters and emission filters as well as dichroics The dichroics are changed manually The extra filters are computer controlled

The Radiance The Radiance has similar filters to the MRC1000/1024 series They are all computer controlled. No manual filter changes are necessary Instead of changing dichroics, dichroic filter wheels are rotated by computer control.

To select the Correct Filter Set First identify the fluorescent dye used in your sample

Spectra The excitation and emission wavelengths of fluorophores are not single wavelengths They have spectra

An example. Imaging CY3

From the spectra, the useful excitation and emission ranges are seen to be: Excitation = nm maximum at 550 nm Emission = nm maximum at 565 nm

An example. Imaging CY3 To successfully image CY3, first look to the left hand side of the chart. Locate CY3 on the chart.

Excitation Filter Selection This shows the excitation maximum

Excitation Filter Selection Identify the laser line which is closest to and (if possible) has a shorter wavelength than the dye excitation maximum.

Excitation Filter Selection Project the chosen laser line horizontally across the chart to the RIGHT hand side

Excitation Filter Selection Note the dichroic number 1 it intercepts. This determines which laser excitation filter and dichroic number 1 are best

Emission Filter Selection From the excitation maximum, follow the solid white line on the chart from left to right This will identify your emission maximum

Emission Filter Selection Continue the line horizontally to the right of the chart Note the emission filter and the dichroic number 2 it intercepts

Emission Filter Selection Ensure the emission filter is NOT intercepted by the chosen laser line

Emission Filter Selection If it is, you must choose an emission filter which passes longer wavelengths Or a laser line which has a shorter wavelength Or both

Conclusion Once these conditions are met, you have successfully chosen the best filter set for your sample