Method Comparison A method comparison is done when: A lab is considering performing an assay they have not performed previously or Performing an assay.

Slides:



Advertisements
Similar presentations
Properties of Least Squares Regression Coefficients
Advertisements

Errors in Chemical Analyses: Assessing the Quality of Results
Regression Analysis Once a linear relationship is defined, the independent variable can be used to forecast the dependent variable. Y ^ = bo + bX bo is.
Probabilistic & Statistical Techniques Eng. Tamer Eshtawi First Semester Eng. Tamer Eshtawi First Semester
World Health Organization
Laboratory Quality Control
Summary 1 l The Analytical Problem l Data Handling.
Quality Control Procedures put into place to monitor the performance of a laboratory test with regard to accuracy and precision.
Limitations of Analytical Methods l The function of the analyst is to obtain a result as near to the true value as possible by the correct application.
Week 2: Spectrophotometry Spectrophotometry Beer’s law Standard curve Dilution problems Quality control Control samples Levey-Jenning chart Shift and trend.
Quality Assurance.
InterQC Delivering Simplicity in the Quality Control Management for your Laboratory Developed by Vitro.
Quality Assurance in the clinical laboratory
QUALITY CONTROL OF PHYSICO-Chemical METHODS Introduction :Validation توثيق المصدوقية.
CALIBRATION METHODS.
Chemometrics Method comparison
Unit #7 - Basic Quality Control for the Clinical Laboratory
Quality Assurance.
Quality Assessment 2 Quality Control.
Method Selection and Evaluation Method Selection and Evaluation D. Kefaya EL- Sayed Mohamed Prof. Of Clinical Pathology (Clinical Chemistry), Mansoura.
Inference for regression - Simple linear regression
Correlation and regression 1: Correlation Coefficient
ABC of Quality Control A problem based approach RT ERASMUS NHLS / FACULTY OF HEALTH SCIENCES, UNIVERSITY OF STELLENBOSCH 20th July, 2007, Bela Bela.
PGCC CHM 1030 Sinex/Gage Three archers each with three arrows to shoot. Can you hit the bull's-eye? Both accurate and precise Precise but not accurate.
The following minimum specified ranges should be considered: Drug substance or a finished (drug) product 80 to 120 % of the test concentration Content.
Analytical Goals. Valid data are essential in making medical decisions, the most important concepts used in judging analytical performance :- 1)Analytical.
Statistics Chapter 9. Statistics Statistics, the collection, tabulation, analysis, interpretation, and presentation of numerical data, provide a viable.
Unit 1 Accuracy & Precision.  Data (Singular: datum or “a data point”): The information collected in an experiment. Can be numbers (quantitative) or.
Biostatistics: Measures of Central Tendency and Variance in Medical Laboratory Settings Module 5 1.
Version 2012 Updated on Copyright © All rights reserved Dong-Sun Lee, Prof., Ph.D. Chemistry, Seoul Women’s University Chapter 5 Errors in Chemical.
LECTURER PROF.Dr. DEMIR BAYKA AUTOMOTIVE ENGINEERING LABORATORY I.
Correlation and Linear Regression. Evaluating Relations Between Interval Level Variables Up to now you have learned to evaluate differences between the.
Chapter 5 Errors In Chemical Analyses Mean, arithmetic mean, and average (x) are synonyms for the quantity obtained by dividing the sum of replicate measurements.
Automated CBC Parameters
Automated CBC Parameters
Quality Control Lecture 5
Make observations to state the problem *a statement that defines the topic of the experiments and identifies the relationship between the two variables.
Choice of Methods and Instruments
Laboratory QA/QC An Overview.
Quality Assurance How do you know your results are correct? How confident are you?
CALIBRATION METHODS. For many analytical techniques, we need to evaluate the response of the unknown sample against the responses of a set of standards.
Why do we need to do it? What are the basic tools?
1 Exercise 7: Accuracy and precision. 2 Origin of the error : Accuracy and precision Systematic (not random) –bias –impossible to be corrected  accuracy.
Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Lecturer of Biochemistry 2015 Introduction to Quality Control.
28. Multiple regression The Practice of Statistics in the Life Sciences Second Edition.
Measurement and Error PGCC CHM 103 Sinex In small groups... Each group must invent a new unit of measurement Use the new unit to determine the area of.
RESEARCH & DATA ANALYSIS
QC/QA.
Quality Control: Analysis Of Data Pawan Angra MS Division of Laboratory Systems Public Health Practice Program Office Centers for Disease Control and.
Quality Control Internal QC External QC. -Monitors a test's method precision and analytical bias. -Preparation of quality control samples and their interpretation.
LECTURE 13 QUALITY ASSURANCE METHOD VALIDATION
 Remember Chemistry panel Quality Control:-  In a medical laboratory, it is a statistical process used to monitor and evaluate the analytical process.
 Routine viral diagnostics: indirect and direct detection of viruses. ◦ Indirect detection: serological tests; ◦ Direct detection:  Viral antigens;
Diagnostic clinical chemistry
Quality Assessment.
Clinical practice involves measuring quantities for a variety of purposes, such as: aiding diagnosis, predicting future patient outcomes, serving as endpoints.
Quality Assurance in the clinical laboratory
CHOICE OF METHODS AND INSTRUMENTS
Quality Assurance in the clinical laboratory
Practical clinical chemistry
به نام خدا تضمين کيفيت در آزمايشگاه
Analytical Method Validation
Chapter 5 Quality Assurance and Calibration Methods
Choice of Methods and Instruments
Method Selection and Evaluation
ANALYTICAL METHOD VALIDATION
Introduction To Medical Technology
Quality Control Lecture 3
▪Internal quality control:
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
Presentation transcript:

Method Comparison A method comparison is done when: A lab is considering performing an assay they have not performed previously or Performing an assay using a different instrument, method or reagent system

Examples include New method not offered be that lab previously Lab is currently performing an assay using a manual method and would like to use an automated method Lab is performing an assay on an older instrument but would like to move that assay to a newer instrument New producer of reagents or in some cases even a new lot of reagents

Method Comparison includes the following Precision Linearity Correlation Accuracy

Precision Study Precision refers to the reproducibility (repeatability) of the method Two Types of Precision are examined Intra run precision Inter run precision (Total Imprecision)

Intra run Precision Usually two levels of controls are used Not the same controls used to check QC One should be in the normal range and the other at or near “Clinical Decision” level Each is run about 20 times all in one run The mean, Std Deviation and CV are calculated

Inter run Precision or Total Imprecision The same controls used for the Intra run Precision study are used They should be run over a course of days or longer, a couple of times on each shift the assay will be normally be performed For example an assay that is done on the day and evening shift, the control would be run twice on each of those shifts for 5 days for a total of twenty results

Statistics for Precision Studies Both methods should be about the same Mean The average of the results Standard Deviation A measure of the amount of scatter about the mean Coefficient of Variation S.D./Mean x 100 = ___% C.V. allows for a better comparison of results expressed in different units or at different levels

Linearity To perform a linearity study the lab must have a substance at various levels Either a substance can be bought, or made, or sometimes a patient sample with a very high value is used Each level is run at least in duplicate and then graphed to look for linearity The highest value and the lowest value are the range that the lab can report Specimens above the highest must be diluted. Specimens below the lowest value are reported as < that value

Correlation A correlation study is done to determine if the result you get using one method will be the same as you get using the other method. To perform a correlation study about 60 specimens are run with both methods Samples are collected that have values that represent the entire assay range The slope, y intercept and r value are determined and the line is drawn on a graph

Correlation Statistics The black line shows “perfect” correlation It has a slope of 1 and a Y intercept of 0 The result using the Test Method will be the same as the result using the Comparative Method

Slope Correlation Statistics Slope The slope of the line should be 1 The black line has a slope of 1 The red line has a slope less than 1 maybe “0.92” showing proportional systematic error

Systematic Error Proportional Systematic Error: Varies with the magnitude of the value, often linearly, and corresponds to a non-unity slope of the correlation plot. Generally reflects a failing of the methodology. Sometimes, may be simply fixed by recalibration and, other times, may be more insidious.

Y intercept Correlation Statistics Y intercept The y intercept of the line should be 0 The blue line has a y intercept > 0 maybe “10” showing constant systematic error

Systematic Error Constant Systematic Error: A constant difference between the measured and true values and corresponds to the “vertical offset” of a correlation plot. Often caused by an interfering substance. Can be practically eliminated by a proper “blank” as long as the interference is constant.

Correlation Graph

Accuracy Accuracy is how close to the real value To perform an accuracy study a substance with a known value is run and the results you get are compared to the expected value

Steps Required to Validate an Assay (in no particular order) Establish Accuracy Establish Precision Establish the Reportable Range (linear range) Determine the Analytical Sensitivity and the Lower Limit of Detection Investigate the Analytical Specificity (search for interferences) Establish the Reference Range (define medically relevant decision points).

Correlation Coefficient -1 to +1

Standard Error Put simply, the standard error of the sample is an estimate of how far the sample mean is likely to be from the population mean, whereas the standard deviation of the sample is the degree to which individuals within the sample differ from the sample mean. If the population standard deviation is finite, the standard error of the sample will tend to zero with increasing sample size, because the estimate of the population mean will improve, while the standard deviation of the sample will tend to the population standard deviation as the sample size increases.