Autoclave Validation Presented by Paul Yeatman BSc. Microbiologist.

Slides:



Advertisements
Similar presentations
1 The Short Story on Long Cycles Why the length of your steam sterilization cycle may effect your sterility assurance.
Advertisements

Micro Risk Why do we clean so often? Does the cleaning make the product safer? Process Improvement & Product Services September
Sterilization - validation, qualification requirements
M.RAGHUVARAN M.PHARM 1 st SEM (PHARMACEUTICS) UNIVERSITY COLLEGE OF PHARMACEUTICAL SCIENCES KAKATIYA UNIVERSITY.
“Getting it right, every step of the way”
1 Utility Systems Process Systems – ‘Direct Impact Systems’ Contact the Product Contact Materials that will Ultimately Become Part of the Product Could.
KSVAS LTD1 WORLD CONGRESS OF STERILIZATION COMBINED MEETING WFHSS / AIOS Small Steam Sterilizers BS EN 1360:2004.
Onsite Biomedical Waste Management Environmental Health & Safety Environmental Protection Program
Drug Delivery & Tissue Engineering Laboratory
ETHYLENE OXIDE STERILIZATION VALIDATION
3.03 Understand support services
MEDICAL DEVICE STERILIZATION
Control of Microbial Growth Chapter Approaches to Control Physical methods Heat Irradiation Filtration Mechanical (e.g., washing) Chemical methods.
DENT 1260 IC Unit 5 Unit 5 Sterilization 1. 3 methods of HEAT Sterilization Autoclave- steam under pressure Chemclave- chemical heated under pressure.
Microbiology and Serology
World Health Organization
Autoclave Use and Safety
Micro Practical 3 PRINCIPLE AND WORKING OF AUTOCLAVE Dr. Shahzad Ali Assistant Professor Department of Wildlife and Ecology UVAS, Ravi Campus, Pattoki.
CSULB Department of Biological Sciences Common Use Equipment Training Autoclaves.
Domina PLUS B WORKING DIAGRAMS TROUBLESHOOTING INTERNAL VIEWS WIRING DIAGRAMS INSTALLATION 03 PREVIOUS MODEL.
Microbiology: Principles and Explorations Sixth Edition Chapter 12: Sterilization and Disinfection Copyright © 2005 by John Wiley & Sons, Inc. Jacquelyn.
Basic Principles of GMP
Disinfection and Sterilization.
STERILIZATION AND DISINFECTION
Autoclave. What is an Autoclave? Autoclave is a pressurized device designed to heat aqueous solutions above their boiling point at normal atmospheric.
Chapter 7: STERILIZATION
STERILIZATION AND DISINFECTION
Definition Sterilization: The freeing of an article from all living organisms including bacteria and their spores. Disinfection: Removal of some types.
Module 1, Part 3: Process validation Slide 1 of 22 © WHO – EDM – 12/2001 Validation Part 3: Process validation Supplementary Training Modules on Good Manufacturing.
Physical and Chemical Control of Microorganisms. Control of Microorganisms by Physical and Chemical Agents.
Sterilization of Medical Devices: Bodhisatwa Das.
WFHSS 2015 Lille/France Klaus Roth SMP GmbH Prüfen Validieren Forschen Hechingerstrasse Tuebingen Germany Validation of hydrogene peroxide Sterilizers.
Lab #10.
Sterilization and Disinfections. Sterilization Freeing of an environment from all living microorganisms includes bacteria and their spores, fungi, parasites.
CSULB Department of Biological Sciences Common Use Equipment Training
Chemical Sterilization
CONTROL OF MICROORGANISMS 1. TOPICS  Sterilization & Disinfection.  Antimicrobial definitions.  Factors influence the effectiveness of antimicrobial.
STERILITY the complete destruction of all living organisms and their spores or their complete removal from the formulation all parenterals must be sterile,
Biological indicators role in CSSD equipment qualification and monitoring Paulo Roberto Laranjeira 1.
The Improvement of ALARA for the Steam Generator Maintenance Radiation Safety Department The 3 rd Plant, Ulchin Nuclear Power Site.
Handling and Disposal of Infectious Wastes
Working safely with Biological materials Aseptic technique, sterilization and tissue culture techniques.
EQUIPMENT VALIDATION - Autoclave Dr. L. Prabakaran Prof. & Head R. R. College of Pharmacy Bangalore -90.
Onsite Medical Waste Management and Bloodborne Pathogens
Extrusion and canning.
Laboratory equipment Lecture (5).
Thermal and Non-Thermal Preservation
Sterilization &Disinfection
3.03 Understand support services
FDA Guide To Aseptic Processing
Practical Medical Bacteriology
DRAFT ONLY Extrusion and canning Extension.
November 16, 2005 University of Kentucky Department of Biosafety
VALIDATION OF MOIST HEAT STEILIZATION
3.03 Understand support services
Sterilization process
3.03 Understand support services
VALIDATION OF MOIST HEAT STEILIZATION
Sterilization and disinfection
VALIDATION OF MOIST HEAT STEILIZATION
Disinfection and Sterilization
3.03 Understand support services
The Improvement of ALARA for the Steam Generator Maintenance
 Define: The stirring heating plate or heated stirring heating plate has been developed to heat and mix fluids contained in laboratory receptacles such.
Sterile Technique.
3.03 Understand support services
Laboratory equipment Lecture (9).
3.03 Understand support services
Manufacture of Sterile Medicinal Products: ‘Annex 1’ - DRAFT
SID & GP MINPROMTORG OF RUSSIA Corporate Communication Center
Presentation transcript:

Autoclave Validation Presented by Paul Yeatman BSc. Microbiologist

Introduction What’s in it for you? If we did not have autoclaves (or steam in place), life would be different for the pharmaceutical industry. Sterilisation would involve fire (or heat), chemicals, filtration or radiation. Microbiological media would have a shorter shelf life and there would be a greater risk of contamination. Autoclave means “auto lock”. Interesting info: The man credited with initiating the project which lead to the invention of the autoclave was Charles Chamberland. He worked with Louis Pasteur and also invented a vaccine for chicken cholera by accident.

Presentation Aims/Outcomes What is an autoclave Why we use autoclaves Identify the two main types used by us Detail how to validate an autoclave (including loading patterns) What to look for in an audit

What is an Autoclave? How do they work –In general –In order to sterilize Types common to us Why use an autoclave What may affect sterilization Steam penetration

Hows Does An Autoclave Work? Steam enters the chamber jacket, passes through an operating valve and enters the rear of the chamber behind a baffle plate. It flows forward and down through the chamber and the load, exiting at the front bottom. A pressure regulator maintains jacket and chamber pressure at a minimum of 15 psi, the pressure required for steam to reach 121ºC (250ºF). Overpressure protection is provided by a safety valve. (from S.S. Block, Disinfection, Sterilization and Preservation, 2nd ed., Philadelphia, lea & Febiger, 1977)

Hows Does An Autoclave Sterilize? Steam held at elevated temperature and pressure for time is used to transfer moist heat. Steam has much greater heat transfer than boiling water. – 80 calories to boil water, 540 calories to produce steam. Heat acts to denature proteins, effectively killing all cells present.

Common Types Production autoclave. –Usually large –Loads one side, unloads the other –Used to sterilize production equipment –May be used to sterilize filled product (can have one opening) –If faulty, potential impact on sterile core or batch disposition Microbiology Laboratory Autoclave –May be large or small –Usually loads and unloads from same side - Sterilized items do not unload directly into production environment –Used to sterilize equipment as well as media. Also used to decontaminate materials before disposal

Why Use Autoclaves? Easiest way to sterilise large volumes of heat tolerant materials. –More effective than dry heat –Not as messy as chemicals –No need for radiation shielding Once validated, simple indicators used to tell autoclaved and non autoclaved material apart – the temp/time/pressure trace is used to confirm sterilization occurred.

What may affect Sterilization Sterilization is dependant on: –initial bioload –microbe sensitivity to heat –time A/C held at sterilizing temperature –Ability of steam to penetrate items being sterilized Steam Penetration: As steam is used to transfer heat, tightly wrapped items, or long tubing may not be properly penetrated. Would represent worse case for validation

What is Validation? Validation shows that a process (or item of equipment) does what it is claimed to do. Why validate? –Need to show that sterile equipment is being taken into sterile core - do not want to contaminate production (which could lead to adulterated product) –For micro testing (both bacteriological and Viable Environmental Monitoring), using sterile equipment helps reduce false positives. –Need to show that decontamination cycles are effective

How To Validate The usual – URS. IQ, OQ, PQ. –This discussion will focus on PQ. IQ: does equipment meet the URS requirements? Is everything that was on the box, in the box? Is the unit installed properly. Are support programs in place for ongoing operation of A/C? OQ: does the A/C operate properly? Does the unit hold temp and pressure correctly? PQ: validation of autoclave cycles and loading patterns – need to show sterilization.

PQ of Autoclave PQ: validation of autoclave cycles and loading patterns. What SAL do you need? –Need to show a reduction of microbes (WWQC ). What is your starting bioload? –Spore strips have >10 6 CFU. What is the microbe’s D value? –For Geobacillus stearothermophilus, this is around Can use physical, chemical or biological indicators. As microbiologist, tend to use BI’s (spore solution or impregnated strip). How does this all fit together?

D-Value, Z-Value and Fo What is D value? –refers to decimal reduction time - The time required at a certain temperature to kill 90% (eg reduce population by log 1) of the organisms being studied. Thus after an organism is reduced by 1 D, only 10% of the original organisms remain. Dependant on microbe and initial numbers. Eg D value of 1.5 = 1CFU left after 10.5 mins. What is Z value? –Refers to the temperature change required to produce a 1 log reduction in D value. What is F 0 ? –The number of minutes to kill a specified number of microbes with a Z value of 10 o C at a temp of o C. –Often confused with the time the chamber is held at elevated temperature and pressure and in practice is the same thing. Overkill –Use many more microbes than would find on items typically autoclaved. Negates the need to test sample for bioload before running the cycle. –Use a sterilisation time exceeding what is necessary to kill a large number of microbes. Negates the need to determine D value of microbe.

Loading Patterns Why important? –Sterilization relies on steam penetration. This may not occur in all cases –Very important to show what you put in an autoclave comes out sterile –When to use spore strips and when to use solutions How to validate? –3x successful runs each loading pattern –Place BI with each item in worse case spot. Place thermocouple next to BI, but not touching item. How often re qualify? –Directive says annually for terminal sterilisation. –For Micro A/C’s depends on use. If used as part of sterility testing…is used for EM sampling…is used for decon cycles… Loading patterns should be adhered to. –Worse case validated – can use less but not more equipment

Audit Considerations What to look for in an audit When auditing, look at: –Was validation conducted? –I would look mainly at PQ only if the unit is >15 years of age. If newer, the whole validation package. Also want to see a preventative maintenance program, SOPS, leak rate test data –Are all expiry dates current? –Cycle time – is it sufficient for tested D values? –Was validation equipment within calibration (pre and post use for thermocouples) –Temperature traces for validation and most recent cycle – Compare. Discrepancies? –Are vacuum cycles used appropriately? –Are the coolest and warmest positions clearly stated in the validation report? –Examine largest (bulkiest) loading pattern. Was the validation acceptable. Is anything thing not listed on the loading pattern present in the autoclave? Is there enough room for steam to circulate through the chamber? –Were there any deviations from the protocols. Are conclusions valid and justified? –*Can the site show that the product that has been terminally sterilized has retained efficacy? –*Could also check that the steam system is validated and that dye bathing (or some other integrity checking) occurs post sterilisation *Not exactly on A/C audit scope.

Conclusion You now know: –What is an autoclave and how it works –Why we use autoclaves –What the main autoclaves are used by microbiologists in the pharmaceutical industry –The basics behind autoclave validation –What to look for in an audit

Questions