X-rays Ouch!.

Slides:



Advertisements
Similar presentations
Advanced Biomedical Imaging Lecture 3
Advertisements

Wilhelm Conrad Röntgen
Line focus principle Heal effect Ratings Tube failure
CHAPTER 6 THE X-RAY BEAM SPECIFIC OBJECTIVES NOTED IN THIS POWER POINT BEGIN ON PAGE
The Generation of X-ray:
The X-Ray Tube Bushong Ch 7.
Quality Control in Diagnostic Radiology
X-ray tube and detection of X-rays Lecture 5. Reminder: The rough schematics of an X-ray tube filament cathod target anode photon flux e-e- electron kinetic.
Dental X-ray Machine 118 Radiology.
MAMMO QC – covered in week 8
Physics of Radiation Therapy
ACVR Artifacts Artifacts of Diagnostic Radiology
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Resident Physics Lectures
X-radiation. X-rays are part of the electromagnetic spectrum. X-radiation (composed of X-rays) is a form of electromagnetic radiation. X- rays have a.
Radiation Physics II.
Medical Imaging X-Rays I.
PRODUCTION OF X-RAY & X-RAY SPECTRUM
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods X-ray Instrumentation Part 1.
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods
Types of Radiation Interactions All or Nothing Many Small There is a finite probability per unit length that the radiation is absorbed. If not, there is.
Resident Physics Lectures
8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS
X-Ray Production & Emission
The X-ray Imaging System
X-Ray Production & Emission
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
ADC PRE-PATIENT COLLIMATION POST-PATIENT COLLIMATION.
Radiation therapy is based on the exposure of malign tumor cells to significant but well localized doses of radiation to destroy the tumor cells. The.
Stopping Power The linear stopping power S for charged particles in a given absorber is simply defined as the differential energy loss for that particle.
Resident Physics Lectures Christensen, Chapter 6Filters George David Associate Professor Department of Radiology Medical College of Georgia.
Topic 29: Remote Sensing 29.1 Production and use of X-rays 29.2 Production and uses of ultrasound 29.3 Use of magnetic resonance as an imaging technique.
INSTRUMENTATION. ELECTRICAL REQUIREMENTS AND EFFICIENCY  Effective voltage  High frequency inverter technology (over 10,000Hz)  Constant potential.
Name each component of the x-ray circuit and
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
Ch. 2 – Anatomy of the X-ray Machine
HABIS X-RAY PRODUCTION AND EXPOSURE FACTORS X-RAY PRODUCTION AND EXPOSURE FACTORS PREPARED BY PREPARED BY Dr fahad albadr radiology chairman radiology.
Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low.
Equipment Operation and Maintenance. Grid-controlled tube Designed to turn on and off rapidly, providing short precise exposures Use in fluoro and in.
Chapter 2 The X-ray Beam.
Physics of Radiography X-ray production and Spectra.
Medical Image Analysis Interaction of Electromagnetic Radiation with Matter in Medical Imaging Figures come from the textbook: Medical Image Analysis,
Radiographic Equipment
X-rays discovered in 1895 by Wilhem Roentgen Roentgenology – Branch of medicine dealing with x-ray or gamma rays in diagnosis and treatment. Radiology.
Attenuation As x-rays pays through matter, the exit beam will contain less photons than the entrance beam. This reduction in the quantity of photons is.
Fluoroscopy. Real-time imaging Most general-purpose fluoroscopy systems use TV technology, operating at 30 frames/sec May be recorded (barium swallow.
Medical Equipment Technology Department 1 Introduction to Biomed. Imaging Systems Lecture No.5-6 Dr. Yousif Mohamed Y. Abdallah.
Fluroscopy and II’s. Fluroscopy Taking real time x-ray images Requires very sensitive detector to limit the radiation needed Image Intensifier (II) is.
Interaction of x-ray photons (and gamma ray photons) with matter.
Resident Physics Lectures Christensen, Chapter 2C Production of X-Rays George David Associate Professor Department of Radiology Medical College of Georgia.
DIGITAL RADIOGRAPHY.
By Dr. Deepa Gautam 1st yr Resident ,Radiotherapy
Introduction By Dr. Nimer Khraim DVMS,BVMS,MVSc.
X-ray tube.
An early x-ray by Wilhem Rontgen
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
3 July July July Conventional X-rays Generator Basic components of an X-ray machine:  Electron source.  Vacuum where electrons were.
Patient is placed between X-ray tube and silver halide film.
Diagnostic Radiology II X-ray Tubes. Anode angle Anode angle defined as the angle of the target surface with respect to the central ray in the x-ray field.
01 X-ray Tube.
Computed Tomography Data Acquisition
X RAYS.
X-RAY PRODUCTION AND EXPOSURE FACTORS
X-ray Production Sharif Qatarneh Medical Physics Division
Part No...., Module No....Lesson No
X-Radiation.
Resident Physics Lectures
Equipment Operation and Maintenance
Generation of X-rays Q&A
Presentation transcript:

X-rays Ouch!

X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage 0.1-20 kV soft x-rays 20-120 kV diagnostic x-rays 120-300 kV orthovoltage x-rays 300 kV – 1 MV intermediate energy x-rays > 1MV megavoltage x-rays

Bremmstrahlung Bremsstrahlung x-rays occur when electrons are (de)accelerated in the Coulomb field of a nucleus

Bremsstrahlung

Bremsstrahlung The power radiated from an accelerating charge is given by Larmor’s equation In the case of an electron in the Coulomb field of a nucleus

Bremsstrahlung The probability of bremsstrahlung goes as Z2, hence high Z targets are more effective than low Z The energy of the x-rays varies from zero to the maximum kinetic energy of the electron (x-ray tube kVp) The energy spectrum from a thick target goes as 1/E but inherent (1mm Al eq) plus additional (few mm Al) filtration removes the lower energy x-rays Here I am referring to diagnostic x-rays

Bremsstrahlung The unfiltered energy spectrum is approximately given by Kramer’s law which was an early application of quantum mechanics

Bremsstrahlung

Characteristic x-rays After excitation, ions with a vacancy in their inner shell can de-excite Radiatively through x-ray fluorescence Non-radiatively through the emission of Auger electrons

Characteristic X-rays Thus an x-ray spectrum will also show characteristic x-rays arising from L to K and M to K transitions after ionization of a K electron Usually transitions to higher shells absorbed by the filtration or are not x-rays

Characteristic X-rays The probability of K shell fluorescence increases with Z

Characteristic X-rays

Characteristic X-rays Sometimes the characteristic x-rays are emphasized using the same material for target and filter Characteristic x-rays from molybdenum are effective in maximizing contrast in mammography

Characteristic X-rays Mo target, filter, and result

Directionality For MeV electrons, bremsstrahlung x-rays are preferentially emitted in the electron’s direction For keV electrons, bremsstrahlung x-rays are emitted at larger angles Characteristic x-rays are emitted isotropically since there is no angular correlation between the incident electron that causes the ionization and the fluorescent photon

X-ray Tube A simplified x-ray tube (Coolidge type) shows the idea behind most x-ray tubes today

X-ray Tube In addition to bremsstrahlung and characteristic x-ray production, electrons also loose energy through collisions Collision losses dominate in this energy region For 100 keV electrons in W Thus >99% of the electron energy goes into heating the target rather than x-rays Removing heat from the anode in a vacuum is an issue

X-ray Tube Efficiency of x-ray production depends on the tube voltage and the target material W (Z=74) in this example kVp (V) Heat (%) X-rays (%) 50 99.7 0.3 200 99 1 6000 65 35

X-ray Tube X-ray tubes

X-ray Tube More detail

X-ray Tube Housing for shielding (Pb) and cooling (oil)

X-ray Tube More detail

X-ray Tube The main parts of the x-ray tube are Cathode/filament Typical electron current is 0.1-1.0 A for short exposures (< 100 ms) Anode/target Glass/metal envelope Accelerating voltage Typical voltage is 20-150 kVp

Cathode Cathode consists of Typically there are two filaments Low R tungsten wire for thermionic emission Tungsten has a high melting point (3370C) and minimum deposit on the glass tube Tube current is controlled by varying the filament current which is a few amps A focusing cup Uses electric field lines to focus the electrons Typically there are two filaments Long one: higher current, lower resolution Large focal spot Short one: lower current, higher resolution Small focal spot

Cathode Dual focus filament is common

Anode Usually made of tungsten in copper because of high Z and high melting point Molybdenum and rhodium used for soft tissue imaging Large rotating surface for heat distribution and radiative heat loss Rotation of 3k-10k revolutions/minute Resides in a vacuum (~10-6 torr) Thermally decoupled from motor to avoid overheating of the shaft Target is at an tilted angle with respect to axis Bremsstrahlung is emitted at ~ right angles for low energy electrons Determines focal spot size

Anode

Anode

Anode The heating of the anode limits the voltage, current, and exposure time An exposure rating chart gives these limits

Anode Power = V x I (watts) Energy = Power x time = V x I x s (joules) HU (Heating Unit) ~ J Damaged anodes

Anode The angle determines the projected focal spot The smaller the angle the better the resolution Typically 7-20 degrees  Angle Incident electron beam width Apparent focal spot size Actual focal spot size Film Increased apparent focal spot size ‘ 

X-rays The energy of the photons depends on the electron energy (kVp) and the target atomic number Z The number of photons depends on the the electron energy (kVp), Z, and the beam current (mA) A typical number / area is ~ 1013 / m2 About 1% will hit the film ~ 1011 / m2 Absorption and detection efficiency will further reduce this number

Automatic Exposure Control AEC detectors can ionization chambers or solid-state detectors X Ray tube Collimator Beam Soft tissue Bone Air Patient Table Grid Cassette AEC detectors

Automatic Exposure Control Most modern x-rays machines are equipped with automatic exposure control also called a phototime The AEC sets the technical parameters of the machine (kV, mA, time, …) in order to avoid repeated exposures AEC is used to keep the radiographic quality (film density) equal on all patients AEC detectors can be ionization chambers or solid state detectors

Grid To reduce the number of secondary scattered photons making it to the film, a grid between the patient and film is used

Grid Details Grid bars are usually lead whereas the grid openings are usually made of aluminum or carbon Grid thickness is typically 3 mm Grid ratio is H/W and 10/1 is typical Grid frequency of 60 lines / cm is typical B/W/H on the figure might be 0.045, 0.120, 1.20 in mm The Bucky factor is the entrance exposure w/wo the grid while achieving the same film density – 4 is average

Accelerating Voltage The potential difference between cathode and anode must be generated by 60 Hz 220V AC power High voltages are produced using a transformer

Accelerating Voltage Electrons are accelerated when the filament is at a negative potential with respect to the target Diode circuits can be used to provide rectification (AC to DC voltage) Three phase power (6 pulse or 12 pulse) can be used to reduce ripple Constant potential operation can be achieved by using constant potential (voltage regulations) or high frequency x-ray generators

Half-wave Rectifier Not very efficient

Full-wave Bridge Rectifier This circuit allows the entire input waveform to be used

Accelerating Voltage Single phase single pulse kV ripple (%) 100% 13% 4% Line voltage Single phase single pulse Single phase 2-pulse Three phase 6-pulse Three phase 12-pulse 0.02 s 0.01 s kV ripple (%)

Images Analog radiography Film based – still widely used Fluorescent screens are used to convert x-rays into visible light that is then recorded on film Screens are more efficient at stopping x-rays than the film (CaWO4 or Gd2O2S:Tb or other rare earth)

Analog Radiography The film itself has excellent spatial resolution but Film detects 0.65% of incident x-ray energy Gd2O2S detects 29.5% of incident x-ray energy Thus using phosphor screens greatly reduces the radiation dose to the patient And also reduces load on the x-ray tube

Analog Radiography There are two efficiency considerations Absorption efficiency or QDE Fraction of incident x-rays that interact with the screen Depends on kVp and screen thickness Gd2O2S has a QDE of ~ 60% for 80 kVp beam, 20 cm patient, 120 mg/cm2 screen thickness

Analog Radiography Conversion efficiency Fraction of absorbed x-ray energy that is emitted as light 5% for CaWO4 15% for Gd2O2S 50,000 eV x 0.15 = 7500 eV 7500 eV / 2.7 eV = 2800 photons produced per absorbed x-ray 50-90% reduction in photon diffusion to film

Analog Radiography Film is an emulsion containing silver-halide grains (AgBr and AgI) coated on mylar Body Film X-Ray source Dark Light

Analog Radiography

Film Badge A film badge consists of a photographic film with various filters The film is a gelatin emulsion containing silver-halide grains (95% AgBr and 5% AgI) on a supporting material Grain diameter is ~ 1mm

Film Badge The film is exposed by light by An electron is released from Br- and moves about the 1m diameter crystal The electron may be captured by a trap such as a crystal imperfection or AgS speck The trapped electron attracts mobile Ag+ ions where it is subsequently neutralized Additional Ag atoms are formed by repeated trapping and neutralization These Ag atoms are called a latent image center The developing process effectively amplifies this process turning the grains with latent image centers into a visible silver deposit

Film Badge

Film Badge Silver atoms at latent image centers

Images Digital radiography Detector based

Digitial Radiography CCD systems CCD systems use a scintillator like gadolinium disulphide to convert x-rays to visible light Light is collected by optics to demagnify the 35x45cm2 film to 2-4 cm2 CCD We’ll talk about CCD’s much later in the course but essentially visible light is converted into charge that is amplified and readout A negative is the thickness of the detector system because of the optical system

Digital Radiography Indirect or direct conversion thin-film transistor (TFT) arrays Also called FPD (flat panel detectors) We’ll cover these later in the course as well – probably through a student talk The idea is that charge proportional to the x-rays received is stored on a capacitor The charges are conducted out by transistors one row at a time and subsequently amplified, multiplexed, and digitized The readout is very fast

Digital Radiography Indirect or direct conversion thin-film transistor (TFT) arrays Indirect conversion uses a scintillator layer (like CsI:Tl) to convert x-rays to visible light and amorphous silicon photodiodes to convert visible light into charge Direct conversion uses an x-ray photoconductor layer (usually amorphous selenium) to convert x-rays to charge An applied electric field directs the charges to the charge collection electrodes

Digital Radiography

Digital Radiography

Digital Radiography Readout

Images Digital radiography “The battle over image quality, however, may be incomprehensible to anyone without a background in high-energy physics.”

X-rays For bone tissue, the linear attenuation coefficient is much greater than that for soft body tissue