Catalytic Tests I: Flow systems and product analysis Rolf Jentoft, November 15, 2002 Modern Methods in Heterogeneous Catalysis Research.

Slides:



Advertisements
Similar presentations
OFFLINE COMPOSITION MEASURING SENSORS
Advertisements

D e t e c t o r s f o r H P L C.
Gas Chromatography Vaporization of sample Gas-solid
Gas Chromatography 427 PHC.
Lecture 8b Gas Chromatography.
Gas Chromatography & Gas-Liquid Chromatography
Mass Spectrometry.
Instrumental Analysis
GAS CHROMATOGRAPHY.
Lab Methods Day June 25, 2014 Gas Chromatography
The Foundations: Classical Split and Splitless Injection
6890/7890 GC Hardware and Technology Overview. 2 In this Section, We Will Discuss:  The major components of the gas chromatograph.  The typical chromatogram.
Flame Ionization Detector Most common detector Carbon atoms (C-C bonds) are burned in a hydrogen flame. A small portion of carbon atoms are ionized (about.
Flame photometry.
1 Gas Chromatography Lecture a. Thermal Conductivity Detector (TCD) This is a nondestructive detector which is used for the separation and collection.
1 Chapter 24 GC Gas Chromatography. 2 GC Mechanism of separation is primarily volatility. Difference in boiling point, vapor pressure etc. What controls.
Monitoring the desorption of analytes from nonpolar SPME fibers using high speed gas chromatography Authors: Kimberly Jasch, Tony Borgerding* Department.
Instant Notes Analytical Chemistry
Introduction to Gas Chromatography Written by Bette Kreuz Produced by Ruth Dusenbery University of Michigan-Dearborn 2000.
Gas chromatography is used in many research labs, industrial labs (quality control), forensic (arson and drug analysis, toxicology, etc.), environmental.
Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences.
Gas Liquid Chromatography
Gas Chromatography. Gas Chromatography Basics Gas Liquid Chromatography (GLC) Gas Solid Chromatography (GSC) Mobile phase does not interact with analyte.
CHAPTER 29 Supercritical Fluid Chromatography The mobile phase is a supercritical fluid (a fluid above its critical T and critical pressure) Supercritical.
Chemical Analysis. Analytical Techniques When chemical evidence is collected at a crime scene, it must be run through an instrument. These instruments.
Gas Chromatography Carrier Gas Flow Control Injector Port Column Column Oven Detector Recorder The GC system consists of gas supplies for the mobile phase.
Gas volume measurement Benedict Roth Spirometer Vitalograph Dry gas meter Wright respirometer (calibration inaccurate if used for continuous flow)
Chapter 27 Gas Chromatography 1. Principles
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 22 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
PAS-Intro1 Dated 24th Aug Copyright © Siemens AG 1999 All Rights Reserved Siemens Advanced Engineering Pte Ltd Process Analytical Services Division.
Year 12 Chemistry Unit 3 – AOS 1 Chemical Analysis.
An Introduction to Chromatography What is chromatography? The separation of a mixture by distribution of its components between a mobile and stationary.
Advanced Method for Renewable Ethanol by Direct Synthesis from Syngas for Renewable Fuel Applications Julia Fisher, Chemical Engineering, Arizona State.
Interfacing Gas Chromatography with Mass Spectroscopy and Infra Red Spectroscopy Is this a large expensive detector…. Or a separation prior to analysis….
Chapter 13 - Analytical Instrumentation 1 Chapter 14 Analytical Instrumentation.
Gas Chromatography Lecture 38.
Measurements in Fluid Mechanics 058:180:001 (ME:5180:0001) Time & Location: 2:30P - 3:20P MWF 218 MLH Office Hours: 4:00P – 5:00P MWF 223B-5 HL Instructor:
Analytical Separations
Basic Gas Chromatography. History Separation of dyes by Runge Separation of plant pigments by Tswett Theoretical gc (Martin & Synge)
LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY
HPLC.
Chapter 28 High Performance Liquid Chromatography.
Lecture 4a Gas Chromatography.
GAS CHROMATOGRAPHY In gas chromatography (GC), the sample is injected onto the head of a chromatographic column and immediately vaporized. The components.
Chromatography and Instrumentation. Chromatography Separate Analyze Identify Purify Quantify Components Mixture Chromatography is used by scientists to:
CHAPTER 9 Control System Instrumentation
Instrumentation. Chromatography Chromatography is the collective term for a set of laboratory techniques for the separation of mixtures. It involves.
Hiden HPR-20 QIC EGA for evolved gas analysis in TGA-MS.
Presented by – ANAMIKA BANERJEE (RESEARCH SCHOLAR)
Chapter 1: The Nature of Analytical Chemistry
Instrumental Analysis
Gas Chromatography. In Gas Chromatography (GC), a gaseous mobile phase transports a gaseous solute through a long, thin column containing solid or liquid.
Gas Chromatography Chap 27 Types: Gas-solid chromatography (GSC)
Reading assignment: section 26E(p781) Chapter 26 # 2, 3, 14, 15, 16 Chapter 27 # 7(a,d,f), 22, 23, 24, 25.
Fundamentals of Analytical Chemistry Eighth Edition.
Analytical Chemistry II ChEm 321
GAS CHROMATOGRAPH. GAS CHROMATOGRAPH Principle: Which type of compounds can analyse? Volatile Thermostable.
Gas Chromatography.
BASICS OF SUPERCRITICAL FLUID CHROMATOGRAPHY
VAPOUR PHASE CHROMATOGRAPHY
Chromatography Daheeya Alenazi.
Gas Chromatography.
CHAPTER 6 Gases and Vapors.
PROCESS AND PROCESS VARIABLE
GAS CHROMATOGRAPHY.
Electron Capture Detectors
Instrumentation.
CHROMATOGRAPHY.
High performance liquid chromatography (HPLC)
Suppose we make a new compound containing of C,H, and N
Presentation transcript:

Catalytic Tests I: Flow systems and product analysis Rolf Jentoft, November 15, 2002 Modern Methods in Heterogeneous Catalysis Research

Gas flow control Gas supply Product analysis Reactor Flow Control Flow Measurement Pressure Measurement Pressure Control Flow systems and product analysis

Important parameters for reactor feed flow Gas flow measurement and control Valves Mass Flow controllers Pressure measurement and control Saturators and vaporizers Product analysis Gas chromatography Mass spectroscopy Infra Red spectroscopy Outline

Parameters for reactor feed flow Flow rate: ml / minute or micromoles / minute PV = nRT use it! Volumetric flow calibration must specify conditions STP (0 °C, 760 mm Hg) Non-deal gas mixtures Correct for pressure: 1 ml in Berlin = 1.12 mlDenver Colorado

Pressure Do you need to control the pressure Pressure drop in reactor Is your flow meter independent of pressure Quartz reactor body: safety Condensation (products or reactants) Uncertainty of measurement During the design of a gas flow system consider the precision that you will need Calculate uncertainty before you build (Partial derivative analysis)

Laboratory reactors Flow: sccm/min 5 mg to 5 grams catalyst Pressures bar Minimize volume of tubing Minimize “dead volume”

Dry piston flow meters Not for in-line use Pressure and temperature correction necessary Solubility of gas in bubble liquid Inexpensive Often used for calibration standard Digital versions available Bubble flow meter Flow measurement

Rotameter (variable area flow meters) Can be installed in line Pressure and temperature correction necessary Gas specific calibration Flow measurement

Measures temperature increase for a set energy input Usually no temperature or pressure correction needed Thermal mass flow meter (Electronic mass flow meters)

Thermal mass flow meter (continued) Calibrated for specific gas and flow range Conversion factor, K, can be used for measurement of other gases Where flow at standard conditions density at standard conditions heat capacity at standard conditions Flow measurement

Thermal mass flow meter (continued) Flow measurement More recently, flow meters which have polynomial conversion factors are available Expensive Gas must be filtered

Flow Control Valve constant: C V, K V, f used to select valves Control Valves If Where specific gravity (Kg/dm 3 ) absolute temperature (K) absolute pressure (Kg/cm 2 )

Mass Flow Controller Combines measurement of mass flow with control valve

Minimize tube volume Check valves Reactor He O2O2 C3H6C3H6 Add check valves to stop back diffusion No “dead” volume (gasses are not flushed away)

Pressure measurement and control Bourden-tube gauge Pressure transducers Back pressure regulators Tend to pulse at low flows Specification should include minimum K V

Multiple control valves Reactor Pressure reduction Mass flow control Back pressure regulator Check valve Danger of interaction between control valves Oscillations due to interaction between pressure reducer and mass flow controller

Multiple control valves Reactor Pressure reduction Mass flow control Back pressure regulator Check valve De-couple regulators by increasing volume if possible Change control parameters of valves Add restrictions to change response of the system to valves

Saturators and vaporizers Reactants that are liquids at room temperature and atmospheric pressure If design is correct, gas is in equilibrium with liquid reagent and contains Saturators: carrier gas is bubbled through a volume (height) of the liquid to be evaporated Gas in Gas out Vapor content of outgoing gas function of temperature and pressure Need to control temperature and measure pressure

Analyze to determine saturation and stability Saturators and Vaporizers Carrier gas inert to liquid, (saturates liquid at very low concentration) Bubbles should be small (frit on bottom of inlet) Height of liquid must be controlled No foaming or formation of spray Tubing down stream of saturator must be at higher temperature than liquid

What is gas concentration in exit stream? saturation of methanol in N 2 at 21.2 °C and 1 bar Vapor pressure methanol = = bar Flow N 2 = 20 sccm/min bar 23.0 sccm/min Saturators and Vaporizers

Vaporizors: liquid flow is controlled and liquid is mixed with carrier gas and evaporated Liquid Pump Reactor Mass flow controller Heated Liquid flow are very low (5 microliters methanol / minute) The vaporization process tends to oscillate at low flows ? Configuration of vaporizer is not trivial Commercial products are available

Heating (tracing) tubes and valves Reactor Pressure reduction Mass flow control Product analysis With saturator or vaporizer, tubes must be heated Products tubes should also be heated Tubes need only be heated to stop condensation

Reactor Bypass Reactor Pressure reduction Mass flow control Check valve Pressure reduction Mass flow control Check valve Reactor Check for reactant stream purity / stability Particularly useful for systems with saturators

Product analysis Products are usually not unknowns Quantification is necessary Rapid analysis may be necessary Describe in more detail: Gas chromatography Mass spectroscopy Infrared spectroscopy

Product analysis: GC Best for quantification and stability Slower than MS or IR Not as flexible as MS GLC and GSC

Product analysis: GC

Injector system: Split / splitless Direct on-column Paralyzing Liquid injector usually included Useful for product identification and troubleshooting Not good for calibration

Gas sampling valves Sample loop adjusted to sample concentration and detection limit Disadvantage: can increase pressure upstream Product analysis: GC Injector system:

Product analysis: GC Columns: Capillary vs. Packed Separation of complex mixtures may require more than one column Vendors can be sent a product list and asked for solutions

Product analysis: GC Detectors:

Product analysis: GC Detectors: High sensitivity for hydrocarbons Destructive Does not detect water or many permanent gases Flame Ionization Detector

Product analysis: GC Detectors: Thermionic Emission Detector High sensitivity for nitrogen and phosphorus

Product analysis: GC Detectors: Thermal Conductivity Detector Changes in conductivity due to product in the carrier stream Non-destructive, often combined with FID

Product analysis: MS Fast scanning speeds Flexible and portable Less stable and more difficult to quantify Not suitable for complex mixtures Difficult or impossible to separate isomers Inlet Ionization Separation Detection

Product analysis: MS Inlet: Usually a capillary in the product stream takes continuously samples products Heating is necessary About 2 ml / min required Capillary leads to a pinhole or to a concentration device

Product analysis: MS Ionization: Electron or hard Ionization Fragments molecules Positive ions are directed to analyzer

Product analysis: MS Ionization: Electron or hard Ionization

Product analysis: MS Ionization: Soft ionization (Chemical ionization) Reagent gases are ionized (Hard ionization) Ionized reagents (CH 4 ) are mixed with product molecules Products MH + for amines or ethers Saturated hydrocarbons often give (M-1) +

Product analysis: MS Ionization:

Product analysis: MS Ionization:

Product analysis: MS Seperation: Quadrupol mass spectrometer

Product analysis: MS Ionization: Time of flight

Product analysis: MS Detectors: Faraday cup Stabile and reliable

Product analysis: MS Detectors: Multichannel electron multiplier Signal gain of 10 5

Product analysis: Analysis: After identification of products: Selected ions are monitored with time For simple mixtures a calibration matrix can be measured and concentrations can be calculated

Product analysis: IR spectroscopy A more specific IR analyzer Detector contains analyte, responds only to radiation absorbed by analyte

References: F.W. McLafferty and F. Turecek, Interpretation of Mass Spectra, Fourth Edition, University Science Books, 1993 H. H. Willard, L. L. Merritt, Jr., J. A. Dean, and F. A. Settle, Jr., Instrumental Methods of Analysis, Seventh Edition, Wadsworth, 1988