VAD Resource Team University of Washington Medical Center

Slides:



Advertisements
Similar presentations
Managing the LVAD patient in Cardiac Rehabilitation
Advertisements

Community Preparation for Caring for Mechanical Circulatory Device Patients University of Wisconsin Hospital And Clinics Ventricular Assist Device Program.
Left Ventricular Assist Devices “LVAD”
BACKGROUND Ventricular assist devices (VADs) are a proven therapy as bridge-to-cardiac transplantation in Class IIIB and Class IV heart failure patients.
Power Module – Setup Before using the Power Module (PM) to power the HeartMate II or XVE LVAS and optional Display Module or System Monitor, prepare the.
SECTION 11 LEVEL B SUIT OUT. SECTION 12 AIR MONITORING.
HeartMate® GoGear™ System Components Training
 Description of PressurePro System  Identify system components  Display and Operational Modes  Sensor alerts  Checking of tire pressures  Manual.
1 CAUTION: Federal Law (USA) restricts this device to sale by or on the order of a physician. Refer to the “Instructions For Use” for complete Indications.
1 CAUTION: Federal Law (USA) restricts this device to sale by or on the order of a physician. Refer to the “Instructions For Use” for complete Indications.
BACKGROUND Ventricular assist devices (VADs) are a proven therapy as bridge-to-cardiac transplantation in Class IIIB and Class IV heart failure patients.
Syringe Pumps. Topics  Diagrams  Principles of Operation  Applications  Safety  Operation  Preventive Maintenance  Common Failure Modes  Troubleshooting.
Clinical Assessment of Pulsatile and Non-Pulsaltile VADs Diana Joseph RN, BSN, CCTC VAD/Heart Transplant Coordinator OSF St. Francis Medical Center Peoria,
Startup Training – TPS3 15. Siemens TPS3 Startup Training 2 Topics Covered  Visual Inspections  Verify environmental requirements are met  TPS3 15.
1 LIFECARE PLV-100 by Bryce Younger. 2 Classification ► Electrically powered  microprocessor controlled ► Volume-control, time cycled ► Rotary drive.
Initiating Positive Pressure CMV A generic approach!
SUPER PASS II Personal Alert Safety System Charles Montgomery Fire Chief Course Developed by Lt. Ricky Phillips Sam Johnson Assistant Chief.
EMS Management of the SynCardia Temporary Total Artificial Heart (TAH-t) Patient Clinical-xxx Rev001.
3100B Theory of Operation and Controls. VIASYS Healthcare, Inc. 3100B Theory of Operation and Controls Approved for sale outside the US in 1998 for patients.
Clean Air Box Usage & Care CABs RAMs CO Monitors.
Suction Pumps.
RSPT 2414 Mechanical ventilation Review Unit 3 classifications By Elizabeth Kelley Buzbee AAS, RRT- NPS.
Powered Air-Purifying Respirator (PAPR) Copyright © 2015 by The S.E.A. Group.
Newport HT50 Ventilator Orientation & Training
1 Batteries & Battery Clips – Setup for Operation Follow these steps to prepare for battery-powered operation: 1.Place two battery clips, two charged batteries.
Community Preparation for Caring for Mechanical Circulatory Device Patients University of Wisconsin Hospital And Clinics Ventricular Assist Device Program.
AccuMax Multi-Point Injection Mechanics
Heart Failure, HF CHF develops when plasma volume increases and fluid accumulates in the lungs, abdominal organs (liver especially), and peripheral tissues.
Ventricular Assist Devices Brian Schwartz, CCP February 25, 2003.
A Case Study. The background You admitted MB after a LIMA-LAD, SVG x2, initial vital signs were stable – HR 75, BP 114/73, CVP of 5, Cardiac index was.
RT 408 Repairing manual. Chapter one The whole machine problems.
Introduction to Ventricular Assist Device (VAD)
Ventricular Assist Devices
B757 Review Questions.
Ventricular Assist Device: An Advanced Surgical Intervention for the Treatment of End Stage Heart Failure Laura Coyle, MSN, ACNP-BC VAD Coordinator Advocate.
Nokia Bluetooth Headset BH-108 for Contact Centers
Erik Simpanen September 27 th, 2011 BME 281. Heart Disease  A generic term for referring to numerous different forms of sicknesses that effect the heart.
Design of Ventricular Assist Device Basic Components of VAD 1. Blood pump 2. Controller 3. Power source Two Categories of VAD’s Pump 1. Pulsatile flow.
Gauge Operation and Software by Scott A. Ager. Computer Recommendations 750 MHz Pentium III 64 Meg SRAM 40 Gig Hard Drive 1024 x 768 graphics CD Writer.
Community Preparation for Caring for Mechanical Circulatory Device Patients University of Wisconsin Hospital And Clinics Ventricular Assist Device Program.
Top Ten things you need to know… About VAD’s Kim Byrum Chappell Mechanical Assist Coordinator.
3100B Theory of Operation and Controls. SensorMedics 3100B u Electrically powered, electronically controlled piston-diaphragm oscillator u Paw of 5 -
BURNS Incidence and Causes 8,000-10,00 burns per year in the U.S.A.
BMC Medical Co., Ltd. Nov Training Course. 1 Table of Contents 1. CATALOGUE 2. INSTALL& DISASSEMBLE 3. SET UP.
Copyright © 2000 Agilent Technologies, Inc.
Ventricular Assist Devices A Ventricular assist device, or VAD, is a mechanical circulatory device that is used to partially or completely replace the.
Unit 5 Learning Objective(s) L5.1a To understand the need for safe working practices when carrying out practical electronics L5.1b To understand relevant.
1 Respironics BiPAP-Vision. 2Classification Electrically powered – internal battery powers Vent Inoperative & audible alarms if AC power lost. Error code.
BMC Medical Co., Ltd. Nov Training Course. 1 Table of Contents 1. CATALOGUE 2. INSTALL& DISASSEMBLE 3. SET UP.
Paramedic Ventilator Management
Heart Failure Cardiac Insufficiency. What is Heart Failure? Heart failure is a progressive disorder in which damage to the heart causes weakening of the.
The Basics Electronics. To familiarise yourself with electronics To have an understanding of how and why we test electronic circuits. To complete a test.
STEP 1 Locate the transmission cooler line connection on the vehicle. Identify the correct adapters to be used by referring to the vehicle application.
1 Automated External Defibrillator. 2 Sudden Cardiac Arrest Sudden cardiac arrest is the most common cause of death in Australia. Sudden Cardiac Arrest.
The Incredible Cardiovascular System How the
Hemodynamic Monitoring
MSA Passport  Personal Alarm Combustible gases and vapors Oxygen enriched Oxygen deficient Specific Toxic gases Alarm levels are set at the factory.
10/11/2011 Ashton Pepple Kyler Sanders
SHORT & LONG TERM EFFECTS OF EXERCISE
Ventricular Assist Device
Hemodialysis Lecture 3.
University of Wisconsin Hospital And Clinics
SB Packing Automatic Downpuller Power Unit
Fuel Supply System Service
LIFEPAK® 1000 Defibrillator Orientation Guide
Heart Lung Machine Lecture (9).
Digital Chart Recorder Operation
Durable Mechanical Circulatory Support in Advanced Heart Failure
Presentation transcript:

VAD Resource Team University of Washington Medical Center HeartMate XVE LVAD VAD Resource Team University of Washington Medical Center

VAD Resource Team Pager/Phone Email Cardiac Surgery Shauna Andrus 994-6986 sandrus 5NE/5SE Sheryl Greco 994-4535 sagreco 5NE Kristina Green 598-4500 kegreen 5SE Nicola Kaye 598-6500 kiwi 5SE Janie Shively 598-7369 janie 5NE Eileen Suver esuver 5SE Teddy Villamarin pinoy

Heart Failure Heart failure is defined as the inability of the heart to effectively pump oxygenated blood to meet the metabolic demands of the body. This devastating condition affects 4.7 million people a year in the United States alone. When conventional medical therapies are unsuccessful, cardiac transplantation is an option for treatment and to prolong life. Unfortunately, only 2300 patients each year receive heart transplants, because the number of patients awaiting transplants far exceeds the number of organs available.

HeartMate Candidates Transplant candidate: dependent on vasopressor support OR incomplete response to continued vasopressor support Approved for ‘Destination Therapy’ BSA of 1.5 or greater

Pump Percutaneous Tube Vent Adapter Weight: 3.74 lbs Pump volume: 83 ml Rate: Up to 120 Flow: Up to 10 L/min Titatanium Motor with 2 bearings Vent port Two tissue valves: Inflow & Outflow

Inside surface of pump housing

Pump weighs 2.63 pounds Titanium case Inside surfaces textured to encourage natural formation of a lining that mimics intima Clot formation inside the pump not a concern because of the lining Coumadin not used ASA daily to prevent platelet aggregation which helps to prevent antibody formation Pump stroke volume is 83 ml Normal volume of blood pumped each cycle is about 90% of the 83 ml Typical stroke volume for a patient is in the low 70s Pump works via microprocessing chip that runs a motor encased inside the pump. Motor drives a pusher plate up, forcing blood out of the pump Percutaneous lead is tunneled under the skin from the pump on the left side to the exit site on the right This helps prevent infection at the exit site from traveling to the pocket where the actual pump is More comfortable for the patient Allows easier activity

Percutaneous Tube & Vent Adapter AVOID: WATER & POWDER AROUND FILTER Percutaneous lead Tunneled from pump to the surface of the skin Contains an electrical cable and an air tube Once the lead emerges from the body, it separates at the Y-connector into an electric connector and an air vent Our patients have had problems with broken Y-connectors The company has developed a new Y-connector that will be introduced sometime this year Air Vent Purpose is to allow free moving air in the system Without free air movement, a vacuum would be created in the pump, causing malfunction The motor causes the pusher plate to move up, pushing blood into the aorta Air moves in under the pusher plate to fill the space When the pusher plate returns to its starting pooint, it pushes the air out of the way If the air didn’t move in and out, a vacuum would be created in the empty space and the pusher plate would not be able to return to its starting place Because this is an open line directly into the pump, it is vital that no fluids enter this line!!! Could cause the pump to seize and stoop working Like any piece of metal equipment, a little bit of fluid sitting in a motor can cause corrosion resulting in a malfunction Either way, this creates a major problem and can result in needing to emergently replace the implanted pump at high risk to the patient and a cost of $65,000! NO FLUID SHOULD EVER BE NEAR THE PORT! A simple filter is attached to the end of the port Non-occlusive to allow the air to flow Should never have an occlusive cap of any sort Will not prevent fluids from entering the line Collects dust Change weekly Electrical connector attaches to a cable on the system controller and provides transmission of information between the controller and the pump

System Controller “Quick” Connector

System Controller Cables

System Controller ‘Brains’ of system (computer) Alarms: Audible & Visual Two cables: connect to power source ‘Quick connect’ to percutaneous lead

System Controller Change Mode Alarm Reset Has three functions Controls the motor of the pump Assures that the pump is working correctly Provides a warning when things are not working correctly Worn at the waist on the right side of the body Three cables Single cable attaches to the electric cable of the percutaneous lead, allowing information to be passed back and forth Two cables are used to attach to the power source Two buttons Mode button (lower left corner of the controller) Selects auto and fixed modes of operation 1 beep = Fixed (one syllable) Present rate that VAD pumps at regardless of activity Used when the controller is unable to get information from the pump (switches into fixed mode automatically) When you want to control the pump rate to keep it from pumping too fast or too slow When the batteries are low 2 beeps = Auto (two syllables) Normal mode of operation Pump will adjust the rate to meet the demands of the patient, just like the normal heart rate adjusts Systems Check: Done daily and charted in CIS Press and hold the Mode button for 3 seconds Yellow wrench, red heart, red battery and fuel gauge lights will come on and a steady alarm will sound After 3 seconds, release the mode button. Lights remain lit and the alarm will sound for another 3 seconds All warning lights will turn off except for the yellow wrench. Warning lights will turn back on in less than 5 seconds and will stay on for another 5 seconds. The alarms continue to sound. If lights and alarms turn off after 5 seconds it means the controller cell needs to be replaced Courtesy dictates shutting the patient’s door and letting those around you know what is going on. Alarm Reset Button Resets alarms 5 minutes for yellow wrench 2 minutes for red heart Check battery life Press and hold while looking at the number of green battery lights showing Amount of battery life is approximate, since the amount of power used is based on pump workload and the age and condition of the batteries At < 25% available battery power, you should change to another power source! Battery Symbols If < 15 minutes of power remains, yellow battery symbol lights with audible beep 1/sec If < 5 minutes of power remains, red battery symbol will appear and a steady tone will sound Controller initiates Power Saver Mode, telling the pump to beat at 50 bpm This uses less battery energy and allows time to immediately switch to another power source A patient with a sudden decrease in HR may be symptomatic! Change Mode Alarm Reset System Check Battery Power Check

Power Base Unit (PBU) & System Monitor

Power Base Unit PBU Internal battery Battery charger AC Fail Low Batt Alarm reset

Screen of System Monitor Pump Rate: Rate PUMP is beating Flow: Cardiac output Stroke volume: ml/beat FUNCTION BUTTONS Auto/Fixed Fixed Rate  Fixed Rate 

FIXED vs. AUTO MODE FIXED MODE is a set rate the pump will beat at regardless of other conditions. This rate can be adjusted by using the system monitor, and should always be set at a rate close to the patient’s baseline. AUTO MODE is responsive to the filling and emptying of the LVAD pump. The pump fills to a fixed volume at a variable rate depending on the patient’s activity and volume status. NO RELATIONSHIP TO RATE OR RHYTHM OF HEART

Display Module DISPLAY Provides information ONLY No button interface

Components:Tethered Quick Connect: percutaneous tube to system controller Power Base Unit Cable: connecting cable from PBU to system controller Twist together connection: system controller to power base unit cable VE LVAD Percutaneous lead Y-connector External air vent Electrical connection System Controller PBU cable Power base unit System monitor Power cord Rear panel of PBU Display Module Hand Pump Emergency Power Pack Stroke Volume Limiter Pneumatic console

Components: Battery Powered Requires 2 batteries 4-6 hours of power/set Worn in holster Quick Connect Twist together connect: system controller to battery

Hand Pump: Goes with patient EVERYWHERE Emergency Power Pack

Battery Clips & Batteries

Changing Power Sources

YELLOW WRENCH: ADVISORIES 1 beep per second May be silenced for 24 hours RED HEART: HAZARD ALARMS Steady tone alarm Preceded and accompanied by a Yellow Wrench alarm YELLOW WRENCH ALERTS 1 beep per second May be silenced RED HEART ALARMS Steady tone alarm Preceded and accompanied by a yellow wrench alarm May be reset for 2 minutes May be reset for 2 minutes

YELLOW WRENCH Power Cable Disconnect MOST COMMON Routine alarm that sounds when power source (battery or outlet) is changed or only one power source connected Check the connections to the power source

YELLOW WRENCH 2. Controller Malfunction Caused by a malfunction in the 2nd MOST COMMON Caused by a malfunction in the system controller *Inoperative Controller = ½ wrench* Change the system controller

Occurs when pump has higher than normal pressure to pump against YELLOW WRENCH 3. Power Limit Advisory RARE Occurs when pump has higher than normal pressure to pump against Kinked or occluded vent line  replace the extension High blood pressure to pump against  lower blood pressure Outflow obstruction  requires surgery Kink or obstruction at vent line  check for and remove kinks or obstructions, change the vent filter

Occurs when pump is changed to Fixed Mode and the preset rate is: YELLOW WRENCH 4. Rate Control Fault RARE Occurs when pump is changed to Fixed Mode and the preset rate is: Less than 50 or Greater than 120 Pump defaults to 72 beats per minute  correct by attaching the system monitor and changing the backup rate

YELLOW WRENCH: ADVISORIES 1 beep per second May be silenced for 24 hours RED HEART: HAZARD ALARMS Steady tone alarm Preceded and accompanied by a yellow wrench alarm YELLOW WRENCH: ADVISORIES 1 beep per second May be silenced for 24 hours RED HEART: HAZARD ALARM Steady tone alarm Preceded and accompanied by a yellow wrench alarm May be reset for 2 minutes May be reset for 2 minutes

THIS IS A CRITICAL ALARM RED HEART Low Beat Rate or No Op Low Beat Rate: Pump slows to < 30 bpm No Op: may be stopped THIS IS A CRITICAL ALARM AND CANNOT BE RESET

Check connections most likely culprit is the percutaneous tube connection to the system controller Check your patient  you may need to hand pump Check the vent port to insure it is not occluded Try changing the power source Try changing the system controller

If pumping does not resume immediately, remove all power sources and start hand pumping STAT page the perfusionist on call STAT page Surg C Attending or Fellow Call for pneumatic console (5NE) with Stroke Volume Limiter (SVL) Do NOT restart the pump if it has been off > 5 minutes except with order from Attending or Fellow

2. Low Stroke Volume: < 25 ml RED HEART 2. Low Stroke Volume: < 25 ml 3. Low Flow: < 1.5 L/min STAT page perfusion STAT page Surg C Attending or Fellow Consider changing to Fixed Mode to  flow Assess whether this is a patient or equipment problem

RED HEART CAUSES Severe RV failure  consider inotropes Extreme low blood volume  assess for bleeding or fluid loss, replace fluids VT, VF or Asystole  ACLS protocols EXCEPTION: NO chest compressions (may dislodge outflow conduit) LV inflow cannula obstruction  requires surgical intervention

disconnected from the percutaneous tube FLASHING YELLOW BATTERY RED HEART YELLOW WRENCH Steady tone CANNOT BE SILENCED XVE System Controller disconnected from the percutaneous tube Check system controller cable connection to percutaneous tube

BATTERY ALARMS Flashing Yellow Battery No tone Yellow Battery Once-per-second Red Battery Continuous Audio Tone

FLASHING YELLOW BATTERY No tone Battery Module The battery that controls the alarm lights and tones on the system controller is low This battery controls ONLY the alarm tones and lights, NOT the pump Change the battery, then do a controller self-test to clear the alarm

YELLOW BATTERY Once-per-second BEEP Low Voltage Advisory Less than 15 min of battery power remain Change to alternate power source

RED BATTERY: Steady tone Low Voltage Alarm Less than 5 minutes of battery power remains Pump defaults to POWER SAVER MODE: 50 bpm Patient may be symptomatic @ this rate if used to faster rate Immediately switch to an alternate power source or prepare to hand pump

EMERGENCIES VT/VF BLOOD FROM EXIT SITE RED HEART Monitor BP; Flows; Cardiovert/Defibrillate BLOOD FROM EXIT SITE STAT page CT surgeon &/or fellow RED HEART Trouble shoot; be prepared to hand pump

IF LVAD MOTOR STOPS < 5 minutes > 5 minutes Troubleshoot and get it working again Try hand pumping Replace system controller Try pneumatic console (with SVL) > 5 minutes DO NOT attempt to restart motor STAT page perfusionist STAT page Surg C Attending/Fellow

Pneumatic Console Can drive XVE HeartMate if pump failure Must use STROKE VOLUME LIMITER Pneumatic tubing with flutter diaphragm FIXED only available mode No electric connections to provide pump information Consider transfer to 5SE No alarms on IP console

Three months after implant

University of Washington Medical Center 1997-2006 Device Total TX Waiting/ Current IP HeartMate 19 17   VE HeartMate 22 20 XVE HeartMate 26 23 1 Thoratec PVAD 24 5 Thoratec IVAD 2 XVE HeartMate DT/HM II DT 4 HM II 7 TOTAL 102* 66 *4 Combo HeartMate + Thoratec