Photoelectron Spectroscopy Lecture 7 – instrumental details –Photon sources –Experimental resolution and sensitivity –Electron kinetic energy and resolution.

Slides:



Advertisements
Similar presentations
X-ray Photoelectron Spectroscopy
Advertisements

Focused ion beam (FIB) 1.Overview. 2.Ion source and optics. 3.Ion-solid interaction, damage. 4.Scanning ion beam imaging. ECE 730: Fabrication in the nanoscale:
X-Ray Astronomy Lab X-rays Why look for X-rays? –High temperatures –Atomic lines –Non-thermal processes X-ray detectors X-ray telescopes The Lab.
Photoelectron Spectroscopy Straightforward evidence for the validity of orbital diagram Konsler 2013.
Lecture 16. LIPS. Introduction a type of atomic emission spectroscopy which uses a highly energetic laser pulse as the excitation source. The laser is.
LCLS Atomic Physics with Intense X-rays at LCLS Philip H. Bucksbaum, University of Michigan, Ann Arbor, MI Roger Falcone, University of California, Berkeley,
First of all, do you know any methods to check chemical composition? Or how you know what is what? First of all, do you know any methods to check chemical.
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
Lecture 18. Chemical: XPS.
Mass Spectroscopy Mass Spectrometry ä Most useful tool for molecular structure determination if you can get it into gas phase ä Molecular weight of.
How to explore a system? Photons Electrons Atoms Electrons Photons Atoms.
Emre Ertuğrul Emin Şahin Seçkin Gökçe KMU 396 Material Science and Technology.
Generation of short pulses
Sub-THz Component of Large Solar Flares Emily Ulanski December 9, 2008 Plasma Physics and Magnetohydrodynamics.
Excitation processes during strong- field ionization and dissociatation of molecules Grad students: Li Fang, Brad Moser Funding : NSF-AMO November 29,
Instrumental Chemistry Chapter 11 Atomic Mass Spectrometry.
2 AB AB + + e AB* AB +* + e n h or n 1 h 1 + n 2 h 2 + : -absorption 1h  n h  -ionization Energy.
Workshop SLAC 7/27/04 M. Zolotorev Fluctuation Properties of Electromagnetic Field Max Zolotorev CBP AFRD LBNL.
4-1 Chap. 7 (Optical Instruments), Chap. 8 (Optical Atomic Spectroscopy) General design of optical instruments Sources of radiation Selection of wavelength.
Seminar Author: Bojan Hiti Mentor: doc. dr. Matjaž Kavčič Determination of trace impurities on Si wafers with x-ray fluorescence.
What Are Some Types of Spectroscopy ?
Atomic Emission Spectroscopy
Atomic Emission Spectrometry
 PART Requirements for Spectroscopic Techniques for Polymers 1. High resolution 2. High sensitivity (>1%) 3. High selectivity between molecular.
1 Waves, Light & Quanta Tim Freegarde Web Gallery of Art; National Gallery, London.
Photoemission Fundamentals of Data Acquisition and Analysis J. A. Kelber, June Texts: PHI handbook, Briggs and Seah Outline: I.Photoemission process.
1 Lecture: Solid State Chemistry (Festkörperchemie) Part 2 (Further spectroscopical methods, ) H.J. Deiseroth, SS 2004.
2D Position Sensitive Detector for Plasma diagnosis
References Hans Kuzmany : Solid State Spectroscopy (Springer) Chap 5 S.M. Sze: Physics of semiconductor devices (Wiley) Chap 13 PHOTODETECTORS Detection.
 How does the wavelength of a light beam and the size of a slit it is going through control the amount of diffraction? DO WORK STOP.
TRIGGERING EXCIMER LASERS BY PHOTOIONIZATION FROM A CORONA DISCHARGE* Zhongmin Xiong and Mark J. Kushner University of Michigan Ann Arbor, MI USA.
Scanning Electron Microscope (SEM) – Shoots a high energy beam of electrons (waves of electrons) at a target. Electron gun Focusing coil Objective lens.
Profile Measurement of HSX Plasma Using Thomson Scattering K. Zhai, F.S.B. Anderson, J. Canik, K. Likin, K. J. Willis, D.T. Anderson, HSX Plasma Laboratory,
IPC Friedrich-Schiller-Universität Jena 1 6. Fluorescence Spectroscopy.
Tools for Nuclear & Particle Physics Experimental Background.
Kinetic Investigation of Collision Induced Excitation Transfer in Kr*(4p 5 5p 1 ) + Kr and Kr*(4p 5 5p 1 ) + He Mixtures Md. Humayun Kabir and Michael.
Yiting Zhangb, Mark Denninga, Randall S. Urdahla and Mark J. Kushnerb
Negative Ions in IEC Devices David R. Boris 2009 US-Japan IEC Workshop 12 th October, 2009 This work performed at The University of Wisconsin Fusion Technology.
1 My Chapter 28 Lecture. 2 Chapter 28: Quantum Physics Wave-Particle Duality Matter Waves The Electron Microscope The Heisenberg Uncertainty Principle.
Energy-Dispersive X-ray Microanalysis in the TEM Anthony J. Garratt-Reed Neil Rowlands.
Plasma diagnostics using spectroscopic techniques
LRT2004 Sudbury, December 2004Igor G. Irastorza, CEA Saclay NOSTOS: a spherical TPC to detect low energy neutrinos Igor G. Irastorza CEA/Saclay NOSTOS.
References Hans Kuzmany : Solid State Spectroscopy (Springer) Chap 5 S.M. Sze Physics of semiconductor devices (Wiley) Chap 13 PHOTODETECTORS.
In this experiment, C 60 was photoionized with single photons with specified energies between the energy range of 37 to 160eV. The photons used came from.
Experimental set-up for on the bench tests Abstract Modeling of processes in the MCP PMT Timing and Cross-Talk Properties of BURLE/Photonis Multi-Channel.
Ionization Energy Measurements and Spectroscopy of HfO and HfO+
Molecular Deceleration Georgios Vasilakis. Outline  Why cold molecules are important  Cooling techniques  Molecular deceleration  Principle  Theory.
Max Zolotorev AFRD LBNL 1 Max Zolotorev CBP AFRD LBNL Radiation & Acceleration Tutorial.
1 Data Acquisition What choices need to be made?.
SEM- Schematic Overview. Electron Detection Tungsten Filament Electron Source.
Molecular Triplet States: Excitation, Detection, and Dynamics Wilton L. Virgo Kyle L. Bittinger Robert W. Field Collisional Excitation Transfer in the.
Electron Energy Loss Spectroscopy (EELS) Suggested Reading: Used electron spectrometer Used TEM Review of theory for elastic + inelastic scattering Examples.
Secondary Ion Mass Spectrometry A look at SIMS and Surface Analysis.
X-ray absorption spectroscopy (XAS)
PHYSICS 225, 2 ND YEAR LAB NUCLEAR RADIATION DETECTORS G.F. West Thurs, Jan. 19.
Lecture 3-Building a Detector (cont’d) George K. Parks Space Sciences Laboratory UC Berkeley, Berkeley, CA.
Results using molecular targets Linear-circular comparison of the intense field ionization of simple molecular targets (N 2, CO 2 ): evidence of nonsequential.
Results using atomic targets Suppression of Nonsequential ionization from an atomic ion target (comparison of double ionization of Ar and Ar + ). Determination.
REVISION PHOTOELECTRIC EFFECT. the process whereby electrons are ejected from a metal surface when light of suitable frequency is incident on that surface..
Acquisition of major research instrumentation for advanced photoelectron spectroscopy with spin, angle and spatial resolution Acquisition of major research.
ESS | Non-Invasive Beam Profile Measurements| | C. Böhme Non-Invasive Beam Profile Measurement Overview of evaluated methods.
1© Manhattan Press (H.K.) Ltd Continuous spectra Spectra Sun’s spectrum and Fraunhofer lines.
X-ray photoelectron spectroscopy (XPS)
Acknowledgements Slides and animations were made by Dr. Jon Karty Mass Spectrometry Facility Indiana University, Bloomington.
Mass Spectronomer.
Atomic Absorption Spectroscopy
Photoelectron Spectroscopy
High Harmonic Analysis Using a COLTRIMS Technique
AMO Early Science Capability
Surface analysis techniques part I
Presentation transcript:

Photoelectron Spectroscopy Lecture 7 – instrumental details –Photon sources –Experimental resolution and sensitivity –Electron kinetic energy and resolution –Electron kinetic energy analyzers

Laboratory Photon Sources Gas discharge VUV sources: ~ eV resolution (40 cm-1) –He I: 21.2 eV (most common for UPS) –He II: 40.8 eV –Ne I: 16.7 eV 1s 2s 2p 3s 3p HV He I  h = 21.2 eV He I  h = 23.1eV

Related (sort of): Metastable Atoms Rare gas in high voltage can also form a metastable state –He* 2 3 S: 19.8 eV, lifetime ~ 10 sec –M + He*  M + He + e- –Transition probability depends on spatial overlap –Penning Ionization Electron Spectroscopy (PIES) or Metastable Atom Electron Spectroscopy (MAES) 1s 2s 2p HV E = 19.8 eV

Laboratory Photon Sources X-ray guns, ~ 1 eV resolution –Most used are: Mg K  ( eV); Al K  ( eV) –other sources from 100 – 8000 eV available

Laboratory Photon Sources Laser sources, ~ 8 eV max, very high resolution and intensity –pulsed source; not continuous flux of photons –photoelectron spectroscopy of negative ions Two or more photon ionization –Using powerful laser source, even these very low probability events can be observed. –Complete separate field of study is multi-photon ionization (MPI) spectroscopy. –Advantage: extremely high resolution. –We will discuss these in last lecture if we have time.

Synchrotron Radiation Source range of resolutions with various monochromators continuous range of photon energies additional cross section, resonance, polarization information The Advanced Photon Source, Argonne National Lab

Why does the photon source chosen matter? We know that we need to select a photon source with sufficient energy to cause ionizations of interest to occur. Choice of photon source “sets” the kinetic energy of the photoelectrons of interest. Now we need to consider how to measure the kinetic energy of these electrons.

Electron Kinetic Energy Analyzers A few important concepts: –Throughput: What % of photoelectrons produced are detected –Resolution: How close in kinetic energy can two electrons be, and still be separated by the analyzer Resolving Power: E/  E higher kinetic energy, lower resolution –electrons with higher kinetic energy are faster than electrons with lower kinetic energy

Deflection (Electrostatic) Analyzers Electrons can be separated, focused by kinetic energy using an electric field Most common is the hemispherical analyzer Resolving power E/  E >1,000

Throughput of Deflection Analyzers Analyzer Entrance steradian: solid angle subtended by a circular surface A sphere subtends 4  steradians

More about kinetic energy and deflection analyzers: Resolving power: E/  E –This means resolution is dependent upon kinetic energy –Scanning through kinetic energy range to collect spectrum: different working resolutions for different portions of the spectrum Measured photoelectron count rate (intensity) –Also dependent upon kinetic energy How do get around these difficulties? –Slow down electrons before they get to analyzer

Rather than scanning through electron kinetic energies with a deflection analyzer: Use an electron-optics lens to slow electrons to a “pass energy” Gain better resolution, but lose sensitivity Hemispherical Analyzer with Electron Optics

Time-of-Flight Analyzers Resolving power ~100 Need to have “packets” of electrons Hence useful with lasers: low photon energy (therefore low kinetic energy), pulsed source Magnetic Bottle: Magnetic field in ionization region allows a large solid angle of photoelectrons to be collected, increasing spectrometer sensitivity. In principle, 2  steradians of photoelectrons can be collected.