TIMERS.

Slides:



Advertisements
Similar presentations
Producing an X-ray Exposure
Advertisements

Advanced Biomedical Imaging Lecture 3
Technique Guidance Systems
Chapter 11 Prime Factors.
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Rad Tech 110 Circuits.
Radiographic Quality Chapter 5.
Topic 3 Selection of kV High kV scattered radiation is energetic. Angle of rays is same as those of primary beam and is not deflected very much. Follows.
AUTOMATIC EXPOSURE CONTROL DMI 63 3/10/2012 online.
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Radiographic Grids II By Professor Stelmark.
The Grid Kyle Thornton DMI 50B.
MAMMO QC – covered in week 8
Analyzing the Image Density. Density Overall blackening of the image.
Prime Exposure Factors II By Professor Stelmark. Primary Factors The primary exposure technique factors the radiographer selects on the control panel.
Chapter 8 The X-ray Imaging System
The X-Ray Circuit.
Technical Artifacts and Errors
FILM ARTIFACTS.
Resident Physics Lectures
Basic quality control tests in medical radiography
Exposure Timers By Prof. Stelmark.
The X-ray Imaging System
X-ray Generators and Transformers
Conventional and Computed Tomography
Quality Control Rad T 110.
RADIOLOGIC IMAGING EQUIPMENT 1. THE X-RAY ROOM 2.
INSTRUMENTATION. ELECTRICAL REQUIREMENTS AND EFFICIENCY  Effective voltage  High frequency inverter technology (over 10,000Hz)  Constant potential.
Control of Scatter Radiation
Chapter 6: Digital Radiographic Imaging
Quality Control.
Dr Mohamed El Safwany, MD.
Components of Image Quality & Radiographic Artifacts Radiologic Technology A Spring 2010 Final Final.
December 2002 Endos DC has been designed from scratch. Every component of the system is completely new: –new generator –new tubehead –new arm –new timer.
Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low.
Factors affecting CT image RAD
RADIOLOGIC IMAGING EQUIPMENT 1. THE X-RAY ROOM 2.
Exposure Factors or Prime Factors
Preparing Variable kVp Technique Charts By Prof. Stelmark.
PRINCIPLES OF TECHNIQUE AND EXPOSURE
Radiographic Quality Visibility and Sharpness
Radiation Sources in medicine diagnostic Radiology
Beam And Scatter Radiation Restricting Devices DMI 50B Kyle Thornton.
Radiographic Equipment
Designing for Radiation Protection. Design Standards for Radiation Protection – Leakage Radiation.
Week 2: Radiographic Equipment
Chapter 6 X-ray Imaging System 3- main parts
IMAGE RECEPTORS. Follow the Sequence- Film System Tube Tube Patient Patient (bucky) or non-bucky (bucky) or non-bucky Cassette Cassette INTENSIFYING SCREEN.
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
DIGITAL IMAGING.
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 19.04: Optimization of protection in Mammography.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Lavin: Chapter 7 CTVT: pp
Direct Digital Radiography or Direct Capture Radiography
Tube Exposure Factors Math Technique Contrast and Density
Sergeo Guilbaud School of Radiologic Sciences
TECHNIQUE CHARTS SERGEO GUILBAUD CLINICAL COORDINATOR.
Automatic Exposure Control (AEC)
Radiographic Technique Sergeo Guilbaud Education Director Long Island College Hospital.
 Make and type of x-ray machine  Incoming line voltage  Kilovoltage  Milliamperage and exposure time  Focal film distance/source-imagine distance.
Week Seven.
8 November 1895, produced and detected electromagnetic radiation in a wavelength range today known as X- rays or Röntgen rays, electromagnetic radiationwavelengthX-
Exposure Factors or Prime Factors
Technique Guidance Systems By Prof. Stelmark. Anatomic Programming Anatomic programming, or anatomically programmed radiography (APR), refers to a radiographic.
Chapters 21 & 22.
Creating a Technique Chart
Pediatric Radiology By Dr. Wambani, J.S. Chief Radiologist
Exposure Factors Chapter 4
Resident Physics Lectures (Year 1)
Comparing Exposure Systems
Presentation transcript:

TIMERS

BASIC X-RAY SCHEMATIC

TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

TIMER A MECHANICAL OR ELECTRONIC DEVICE WHOSE ACTION IS TO MAKE OR BRAKE THE HIGH VOLTAGE ACROSS THE X-RAY TUBE

TYPES OF TIMING CIRCUITS TIMERS: MECHANICAL SYNCHRONOUS ELECTRONIC mAs BACKUP AUTOMATIC EXPOSURE CONTROL

MACHANICAL TIMER OPERATE BY CLOCKWORK. SPRING MECHANISMS IS WOUND UP. DURING EXPOSURE THE SPRING UNWINDS AND TERMINATES EXPOSURE.

SYNCHRONOUS TIMER IN SYNCH WITH FREQUENCY OF THE A/C CURRENT (60 Hz in the US). MINIMUM EXPOSURE TIME IS 1/60 s. TIMING INTERVALS INCREASE BY 1/60 s. MUST BE RESET AFTER EACH EXPOSURE

ELECTRONIC TIMER THE MOST SOPHISTICATED. THE MOST ACCURATE IT’S CIRCUITRY CONSISTS OF THYRATRON TUBE OR THYRISTOR. ALLOW WIDE RANGE OF TIME INTERVALS. ALLOW SERIAL EXPOSURES. NOWADAYS CONTROLLED BY MICROPROCESSOR

THYRATRON CIRCUIT

mAs TIMER MONITORS THE PRODUCT OF mA & EXPOSURE TIME AND TERMINATES EXPOSURE WHEN THE DESIRED mAs VALUE IS ATTAINED.

BACKUP SET TO TERMINATE EXPOSURE IN THE EVENT OF FAILURE IN THE AEC CIRCUIT. MAX. mAs ALLOWABLE – 600 mAs UNLESS TUBE POTENTIAL IS LESS THAN 50 kVp – 2,000 mAs IS THE LIMIT.

AUTOMATIC EXPOSURE CONTROL Automatic exposure control devices can assist the radiographer in producing consistent radiographic images from patient to patient, regardless of size or presence of pathology. The advantages of this consistency are numerous and include: decreased repeat rate; decreased patient exposure; and increased department efficiency. The most important benefit being decreased repeat rate

When an AEC device is used to terminate an exposure, the technologist sets the kVp and mA, but the time of the exposure is automatically determined by the machine. The AEC device differs from a manual timer because the AEC does not stop the exposure until the film has reached an appropriate density. Unlike manual timers, which simply stop the exposure after the preset time has elapsed.

A major benefit of the AEC device is its ability to consistently obtain accurately exposed radiographs, even in the presence of pathology. While manual timers terminate the exposure at the preset time, regardless of pathology or achievement of proper film density. The following example demonstrates the difference

AUTOMATIC EXPOSURE CONTROL (AEC) FLAT IONIZATION CHAMBER PHOTOTIME

IONIZATION CHAMBER AEC X-RAY TUBE X-RAY TABLE BUCKY

IONIZATION CHAMBERS ARE SITUATED BETWEEN THE GRID MECHANISM & THE CASSETTE

CHAMBERS

CHAMBER SELECTION ON THE CONSOLE DETECTOR SELECTION     Along with proper

AEC

IONIZATION CHAMBER AEC X-RAY TUBE X-RAY TABLE CASSETTE SCREEN

PM TUBE LIGHT ELECTRONS

PHOTOTIME MECHANISM IS SITUATED BEHIND THE CASSETTE CASSETTE NEEDS SPECIAL DESIGN IT HAS OPENING IN THE BACK

PHOTOTIME CASSETTE

PHOTOTIME SCREEN X

Even though the ionization chamber and the phototimer operate differently, they both have the same function: convert radiation into an electrical signal which will be used to automatically stop the exposure when the film has reached the proper density

FACTORS INFLUENCING AEC PATIENT POSITIONING DETECTOR SELECTION

POSITIONING!!!! AEC USES DETECTORS TO RECEIVE THE RADIATION THAT PASSED THROUGH THE PATIENT WE MUST POSITION THE ANAOMY OF INTEREST DIRECTLY ABOVE THE DETECTORS.

Without a technologist who is very knowledgeable about anatomy and positioning, automatic timers are worthless. In fact, they may actually decrease department efficiency because of the increased amount of repeat radiographs that will result if used improperly.

DETECTOR SELECTION

LAT. L-SPINE

C-SPINE

C-SPINE -AP

SHOULDER

KNEE

HIP

PELVIS

ABDOMEN

CHEST PA

LAT. CHEST

DENSITY SETTING D. SETTING % DENSITY CHANGE +4 +3 +2 +1 -1 -2 -3 -4 -1 -2 -3 -4 +100 +75 +50 +25 -25 -50 -75 -100

MINIMUM RESPONSE TIME SHORTEST EXPOSURE TIME POSSIBLE

EFFECTS OF mA, kVp, SID ON AEC mA AEC TIME kVp AEC TIME SID AEC TIME

PROBLEMS WITH AEC PATIENT PATHOLOGY PROSTHETIC DEVICE COLLIMATION STROBOSCOPIC EFFECT WRONG IMAGE RECEPTOR

CERTAIN PATHOLOGIES ALTER AEC READING

PROSTHESIS GIVES FALSE AEC READING

NOT ENOUGH COLLIMATION ALLOWS TOO MUCH SCATTER REACHING THE DETECTOR AND PREMATURE TERMINATION OF THE EXPOSURE TAKES PLACE

STROBOSCOPIC EFFECT TOO SHORT OF THE EXPOSURE TIME IS SELECTED BY THE AE, SHORTER THAN THE SPEED OF THE OSCILLATION OF THE GRID AND IT FREEZES THE GRID MOVEMENT.

WRONG SPEED OF IMAGE RECEPTOR!!!