Multidisciplinary Engineering Senior Design Project 6218 Miniature Membrane Press Critical Design Review 5/16/2006 Project Sponsor: Bausch & Lomb Team.

Slides:



Advertisements
Similar presentations
Interfacing to the Analog World
Advertisements

Interfacing to the Analog World
© NMISA 2010 Implementing a Shock Calibration System Using a Vibration Exciter and Pendulum Presented by Ian Veldman.
Data Acquisition Risanuri Hidayat.
A/D Conversion and Interfacing Physics 270. Voltmeters.
Automated Electric Guitar Tuner Abstract In the music industry, there has been a recent trend of increased research with regard to self-tuning guitars.
Transmission Machine Components
R·I·T Kate Gleason College of Engineering Design of a Tensile Load Frame for a Scanning Electron Microscope Senior Design Project Project Manager.
Alexander Moulton Marie Hammer Xingwang Gao Andrew Robertson Team Lead Mechanical Engineer Electrical Engineer.
Members: Morgan Muren Matt Nusbaum.  To develop an apparatus that can test the spin down characteristics of a cup anemometer and record and display the.
Multidisciplinary Engineering Senior Design Project 6508 Controls Lab Interface Improvement Preliminary Design Review 11/11/05 Team Members: Michael Abbott,
Multidisciplinary Engineering Senior Design Automated Plasma Generator Test System Preliminary Design Review 11/11/05 Project Sponsor: MKS, ENI Incorporated.
Lecture 31 Electrical Instrumentation. Lecture 32 Electrical Instrumentation Electrical instrumentation is the process of acquiring data about one or.
Lecture161 Instrumentation Prof. Phillips March 14, 2003.
R·I·T Kate Gleason College of Engineering Design of a Tensile Load Frame for a Scanning Electron Microscope Senior Design Project Project Manager.
Controls Lab Interface Improvement Project #06508Faculty Advisors: Dr. A. Mathew and Dr. D. Phillips Project Objectives This work focused on the improvement.
Design of a Control Workstation for Controller Algorithm Testing Aaron Mahaffey Dave Tastsides Dr. Dempsey.
Multidisciplinary Engineering Senior Design Project 6218 Soft Tissue “Tensile” Tester Preliminary Design Review 2/24/2006 Project Sponsor: Bausch & Lomb.
Multidisciplinary Engineering Senior Design Project 6218 Soft Tissue “Tensile” Tester Preliminary Design Review 2/24/2006 Project Sponsor: Bausch & Lomb.
GE Energy Silicon Wafer Measurement System Team 10 Olin Biddy Scott Johnson Chetwyn Jones Rob McCoy Tim Weber.
Wireless Data Acquisition for SAE Car Project by: J.P. Haberkorn & Jon Trainor Advised by: Mr. Steven Gutschlag.
UBI >> Contents Chapter 9 Data Acquisition Introduction MSP430 Teaching Materials Texas Instruments Incorporated University of Beira Interior (PT) Pedro.
Automated Feeding Solutions for Dog Owners Preliminary Design Review October 19, 2007 Project Automated Dog Feeder Project Advisor Dr. Hongwei Wu The Canine.
Rochester Institute of Technology Kate Gleason College of Engineering Presents …
30 day money back guarantee if you are not satisfied Coupled with a no-hassle, 12 month warranty against manufacturer’s defects Zaber’s Money Back Guarantee.
Stability Control System for a Propeller Powered by a Brushless DC Motor (BLDC) Codey M. Lozier Christian A. Thompson Advisor: Dr. Mohammad Saadeh.
Oscilloscope Tutorial
3D Micromanufacturing Lab. School of Mechatronics Gwangju Institute of Science and Technology 3D Micromanufacturing Lab. School of Mechatronics Gwangju.
DESIGN OF MODERN INTEGRAGTED LABORATORY-STATION (ILS) FOR ENGINEERING EXPERIMENTAL COURSES M. Kostic, Srinivasa Rao Vaddiraju, Srinivas.
Digital to Analog Converters (DAC)
Digital to Analog Converters
A Mobile Wireless Electrocardiogram System for Health Care Facilities John Farner Jason Fritts Julian Jaeger Joe Richard Georgia Institute of Technology.
3-Point Bending Device to Measure Transmural Strains for Multilayer Soft Tissue Composite Jennifer Olson Sarah Rivest Brian Schmidtberg Sponsor: Dr. Wei.
Chapter 5 Engineering Tools for Electrical and Computer Engineers.
Team: James Mazza, Andre Berwin, Peter Dunning, Nathan Fulcher, Andrew Watson, Travis Bardsley, Anthony Parker April 5, 2013.
Wearable Sensors Final Presentation Problem Background Inferred Transmission (short range) Bulky Design, hard to wear 20 Samples per second.
ACOE2551 Microprocessors Data Converters Analog to Digital Converters (ADC) –Convert an analog quantity (voltage, current) into a digital code Digital.
Accessible Manufacturing Equipment Phase 2 Team 2 Nicholas Neumann Ralph Prewett Jonathan Brouker Felix Adisaputra Li Tian December 10 th, 2010 ECE 480.
Wireless Intelligent Sensor Modules for Home Monitoring and Control Presented by: BUI, Phuong Nhung, 裴芳绒 António M. Silva1, Alexandre Correia1, António.
Power Control System for a Concrete Durability Test Cabinet – Phase 2 Jacob Jameson Madhav Kothapalli Thomas Persinger Andrew Versluys.
EPICS Developments at the Australian Synchrotron DSP EPICS driver for the General Standards 16AIO analog card EPICS driver for the Galil range of motor.
Readout ASIC for SiPM detector of the CTA new generation camera (ALPS) N.Fouque, R. Hermel, F. Mehrez, Sylvie Rosier-Lees LAPP (Laboratoire d’Annecy le.
Disturbance Rejection: Final Presentation Group 2: Nick Fronzo Phil Gaudet Sean Senical Justin Turnier.
Anti-Fraud System Design: Authenticating Magnetic Fingerprint Signal Amanda Spencer AND Raja Timihiri Dr. Robert E. Morley, Jr., Edward J. Richter, Washington.
HARDWARE INTERFACE FOR A 3-DOF SURGICAL ROBOT ARM Ahmet Atasoy 1, Mehmed Ozkan 2, Duygun Erol Barkana 3 1 Institute of Biomedical Engineering, Bogazici.
P16221 – FSAE Shock Dynamometer System Level Design Review September 29, 2015.
Engineering H192 - Computer Programming Gateway Engineering Education Coalition Lab 1P. 1Winter Quarter Data Acquisition System Fundamentals Lab 1.
The palm was created using a modular cavity design. It was designed using ProEngineer and printed using Rapid Prototype. The fingers were made using Polymorph.
Automated Maze System Development Group 9 Tanvir Haque Sidd Murthy Samar Shah Advisors: Dr. Herbert Y. Meltzer, Psychiatry Dr. Paul King, Biomedical Engineering.
Path The purpose of this project is to design a universal data recording device to monitor the health of a Dresser-Rand compressor during operation. In.
Automated Maze System Development Group 9 Tanvir Haque Sidd Murthy Samar Shah Advisors: Dr. Herbert Y. Meltzer, Psychiatry Dr. Paul King, Biomedical Engineering.
P – , TOW TANK FOR TETHERED HYDROFOIL TEAM MEMBERS: Hope Alm (ME)Tim Buckner (ME) William Lentlie (ME) Andres Santizo Matheu (ISE) Shauna.
Digital to Analog Converter (DAC)
By Dr. Hong Zhang For Freshman Engineering Clinic.
Motor Assisted Wheelchair Design Review 10/01/15.
INSTRUMENT CONTROLLED MICROSCOPY Mason Jellings, Justin Gearing, Sarah Reichert, John Byce Client: Dr. Joshua Medow Advisor: Professor Thomas Yen Introduction.
P10203 LV1 MOTOR CONTROLLER FINAL REVIEW MAY 14, 2010 Electrical: Kory Williams, Adam Gillon, Oladipo Tokunboh Mechanical: Louis Shogry, Andrew Krall.
BIOMECHANICS & BIOMEDICAL ENGINEERING LOW COST FORCE SENSOR
Introduction to Discrete-Time Control Systems fall
Instrumented Walker Skyler Bullington Tommy Frankenberger Larson Stacy
MECH 373 Instrumentation and Measurement
Electronic Instrumentation Lectrurer Touseef Yaqoob
1 2 3 INDIAN INSTITUTE OF TECHNOLOGY ROORKEE PROJECT REPORT 2016
Multidisciplinary Engineering Senior Design Project P06441 See Through Fog Imaging Preliminary Design Review 05/19/06 Project Sponsor: Dr. Rao Team Members:
Test Rig Development Group
Eye Movement Tracking Device Senior Design Project: P09004
P09045: Membrane Characteristic Test Stand
Anti-Rollback & Brake Assist
P19045 Essential Tremor Test Rig II Final Review April 7th, 2019
Presentation transcript:

Multidisciplinary Engineering Senior Design Project 6218 Miniature Membrane Press Critical Design Review 5/16/2006 Project Sponsor: Bausch & Lomb Team Members: Ryan Schkoda – Team Lead Christopher Kudla – Lead engineer Kristina Schober – Mechanical Engineer Robert Mc Coy – Electrical Engineer Team Mentor: Dr. Elizabeth DeBartolo Kate Gleason College of Engineering Rochester Institute of Technology

Agenda Introduction Recap of Project Planning Fabrication and Build LabView™ Development Data Analysis Routine Testing & Results Error Recommendations Demonstration

Introduction Purpose is to gather material properties Employing a nontraditional method of property determination Showing much promise

Design Layout

Design (cont.)

Component Selection Motorized Translating Stage: ±0.1 μm resolution Built in linear encoder LabVIEW compatible CMM Style Probe Tip Tightly toleranced diameter 10 gram Load Cell ±0.03 gram resolution Safety pins help prevent damage LabVIEW compatible

Electrical Components Connect Load Cell to User Interface –Signal Conditioner Accuracy: ±0.05% of FS Noise and Ripple: Less than 5 mV P-P at gain=1000 –Data Acquisition 8 Single-Ended, 12-Bit Analog Inputs LabVIEW Drivers and Examples

Electrical Components Connect Linear Stage to User Interface –Single-Axis Low-Power Motion Controller/Driver Compatible for plug-and-play operation 1000x programmable micro-step resolution for ultra smooth low speed stepper positioning RS232 communications link for easy user interfacing LabVIEW Compatible with Drivers

Control and Display LabVIEW interface USB and Serial port links for data input DLL files LabVIEW drivers Load Cell Signal Conditioner/ Amplifier USB Data Acquisition / A/D Converter Motorized Linear Stage/Encoder Motion Controller/Driver LabVIEW ®

Alignment VI

Test VI

Data Operations Purpose –Necessary to extract material properties –To account for not sensing origin

Representative Data Set Smoothness & Continuity Noise contaminated zero values Characteristic curve Both Small and large deflection data

Replication Replication of area of interest Systematically scatter along x- axis

Replication, Curve Fitting & Placement Theory dictates P  3 Cubic is fit to each curve Best fit means best placement

“Small” Deflections Question: What is “small?” Answer: Data points corresponding to the part of the data curve that have yet to deviate from the small deflection equations’ predictions

Investigation of Plastic Wrap

Investigation of Artificial Rhexis

Implementation Methods Two methods of Data Analysis –MATLAB® More Robust Fast Easily Customizable –Microsoft Excel® Easier to adjust Microsoft Excel® is more readily available

MATLAB® Implementation

Excel® Implementation

Poor Data Obvious anomaly in the data curve Severe impact on modulus readings Suspected to be from wrinkles or poor sample loading

Concentricity Testing (BLP0001) Objective: –To evaluate the membrane press alignment. It is critical that the probe tip and stage be centered. Procedure –Set alignment prior to each test and record position per SOP#: TD-PPD-PRO-001. –Use 10 different samples. Results: –The data provided a range for the center point location. –Favorable data results of Young’s Modulus.

Concentricity Testing Results Radius: Mean= Stdev=0.0024

Concentricity Testing Results

Offset Testing (BLP0002) Objective: –To evaluate the membrane press with an offset alignment. –A worst case scenario. Procedure: Determine the offset value, running 10 tests at the offset position. –Determining the offset value: Used centered data collected from the Concentricity Tests. Found the 95 th and 50 th percentiles of the centered x and y positions.

Offset Testing Results Radius: Mean= Stdev=0.0031

Offset Testing Results Young’s Modulus Values:

Realignment Testing (BLP0003) Objective: –To reduce the number of times the stage must be realigned. Procedure: –Run 10 tests, 5 tests, and 2 tests. Acceptance Criteria: Results: After each test the device must be realigned

Realignment Testing Results

Gage R&R (BLP0004) Objective: Measure the Repeatability and Reproducibility of the device. Procedure: –3 operators –10 different samples a piece Acceptance Criteria: Less than or equal to 25% Results: –Gage R&R= 383%

Testing for Misalignment Induced Error 95 th percentile

Error Analysis

Results PVDC (plastic wrap) –352.7 MPa ±50.2 MPa Artificial Rhexis –MPa ±14.3%

Recommendations Eddy current or capacitance sensors Negative pre-strain testing Saline testing grip redesign Continued silicone testing Fitting non-linear constants Shorter

Questions & Discussion

Objectives & Specifications Compact and easily transportable design. Incorporate ergonomics. Overall weight < 25 lbs. Carrying case. Compliant with FDA, GMP, and OR regulations. Easy to operate user interface. Sterilization using autoclave. Load cell and encoder capable of continuous data acquisition. Repeatable mounting of specimens. Repeatable method of generating a sample out of the original rhexis. Powered by 110V outlet. Ability to test a sample that is submerged in a saline solution. Cost to be on the order of $5,000.