Instrumented Walker Skyler Bullington Tommy Frankenberger Larson Stacy

Similar presentations


Presentation on theme: "Instrumented Walker Skyler Bullington Tommy Frankenberger Larson Stacy"— Presentation transcript:

1 Instrumented Walker Skyler Bullington Tommy Frankenberger Larson Stacy
Jairus Potts Rockwell Shields Chris Barnett

2 Purpose applied weight regardless of hand placement.
The purpose of this project is to develop, build, and test an instrumented walker that is capable of reporting applied weight regardless of hand placement.

3 Background The Mayo clinic had a need for an instrumented walker
Improve rehabilitation methods for patients Monitor reliance and walker use from patients Primarily used with Exoskeletons Research was conducted on how previous data was analyzed, procedures, and hardware used.

4 Design Criteria Improve existing designs Lightweight Removable
Able to communicate with Labview Accurately determine the forces being placed on the walker Uniform wiring

5 Load Cell Selection Single axis load cell
Determine forces being applied by the patient in the z-direction Load Cell selected that could measure weight up to 200 pounds with plus or minus 1lb accuracy.

6 Design Two custom attachments made of aluminum were designed that would allow the load cell to be mounted and screwed onto the bottom of each leg of the walker. Design is removable and capable of transferring forces without failure.

7 Design Continued A box made of plastic to house the circuit
board and battery. Mounted on the front crossbeam of the walker.

8 Stress Analysis

9 Load Cell Calibration Weight applied in 5 lb increments ranging from 0 to 60 Voltage read out from calibration board Entered into Excel Plotted and Curve Fitted Resulting functions for load

10 Load Cell Calibration (continued)

11 Electronics Criteria For Design
Read 4 Load Cell Outputs on 4 Separate Channels High Level of Accuracy Low Voltage Battery Conservation Functionality Small and Lightweight

12 Electronics Selection
Microcontroller - Arduino Mega 2560 Chosen for more Digital Inputs/Outputs Processing latency is less than Uno R3 Battery - Rechargeable 1.2V 3000 mAh HR6 Inexpensive Batteries NiMH Safe for Hospital Use Alkaline AA substitute

13 Electronics Selection
Analog To Digital Conversion - ADS 1234 Designed For Bridge Sensors 4 Channel Input Ultra Low Noise with Surface Mount Devices 24 Bit Resolution - 18 Effective Number Of Bits (ENOB) Wireless Module - XBee 1mW Series 1 Low Power Consumption 3.3V , 50mA Max 300 Feet Range

14 Printed Circuit Board Design
Add On Card for Arduino Mega ADC Requires SMD Components Manufacturer Recommended Parts Used Terminals For Load Cell Connections External MHz Crystal Oscillator Placement for Wireless Module

15 Arduino Code

16 Labview Programming Fig. Input Testing Setup
-Data read in from USB using VISA Serial -Input string split into four separate values, converted to decimal -Input voltages converted to individual forces using calibration formulae -Information displayed and plotted in GUI Fig. Input Testing Setup

17 LabView & Arduino Integration
Fig. Labview GUI Fig. Block Diagram

18 Results Full design prototype was built
Arduino was able to communicate with LabView Fully functional LabView code with counterfeit signal Fully functional Arduino code Fully developed PCB board

19 Gantt Chart

20 Questions ???