 The annular space. The clearance between the head of the float and the flow tube is known as the annular space.

Slides:



Advertisements
Similar presentations
Safety Requirements of the Anesthesia Workstation
Advertisements

Anesthesia Machine Presented by Gil Soto C.R.N.A.
INTERMEDIATE PRESSURE SYSTEM PRESENTED BY: Dr CHITTRA MODERATED BY: Dr JYOTI PATHANIA.
AMIR SALAH MODERN ANAESTHETIC MACHINE MODERN ANAESTHETIC MACHINE 1 of 4.
Presented by: Dr. manoj kumar Moderated by: Dr. Aparna Sharma
Use of the Inhalant Anesthetic Machines
Anesthesia Apparatus Checkout Recommendations
Inhaled Anesthetic Delivery Systems
Regulators and Flowmeters
© Blue Graphics Concept Sauer-Danfoss START-UP PROCEDURE PROPEL PUMP Instruction Sauer-Danfoss Belgium – T. Van Mossevelde.
Check Valves and Manufacturer-Recommended Leak Test
Section 2 Safety, Tools and Equipment, Shop Practices Unit 4 General Safety Practices.
pNeuton Transport Ventilator
Presented by : Dr Rashmi Moderator: Dr Gian Chauhan.
Anesthetic Equipment.
Renaissance in Anesthesia Machine Safety
Hydraulics.
Inhaled Anesthetic Delivery Systems
 Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety.
By Dr. Ahmed Mostafa Assist. Prof. of anesthesia and I.C.U.
By Dr. Ahmed Mostafa Assist. Prof. of anesthesia and I.C.U.
Flow Meters Flow: Volume of a fluid passing certain point over a given time Flow Rate. over a given time Flow Rate. Flow meters:sets used to measure flow.
Parts of the Anesthetic Machine
The Anesthesia Gas Machine
Anesthesia Machine VAPORIZERS. Vaporizers Convert liquid anesthetic into a volatile inhalation agent Based on laws of physics You must memorize the chemical.
A). When the carrier gas is quickly switched to 100% nitrous oxide (B). the halothane concentration decreases to 3% within 8 to 10 seconds A new steady-state.
Dr. S. Parthasarathy MD., DA., DNB, MD (Acu), Dip. Diab. DCA, Dip. Software statistics PhD(physiology) Mahatma gandhi medical college and research institute,
STEAM HEATING.
Inhaled anesthetic Delivery Systems
By Dr. Ahmed Mostafa Assist. Prof. of anesthesia and I.C.U
Lease Automatic Custody Transfer
Pressure reducing valve By Dr. Ahmed Mostafa Assist. Prof. of anesthesia & I.C.U.
CHAPTER 6 Moving Heat: Heating and Air Conditioning Principles
Chapter 10 Fluid Power Systems.
Introductory Lecture Series: The Anesthesia Machine
Waste Anesthetic Gases. The anesthetic gas and vapors that leak out into the surrounding room during medical and surgical procedures are considered waste.
Dr. subhash technical campus
Gas Metal Arc Welding (Metal Inert Gas) (MIG)
Pro-Science 4 th International Conference of Hydrogen Safety, September 12-14, 2011, SAN FRANCISCO, USA EXPERIMENTAL STUDY OF IGNITED UNSTEADY HYDROGEN.
Without reference, identify principles about Anesthesia Units with at least 70 percent accuracy.
Evaporative Emissions Control Systems (EVAP)
Anesthesia Machine Circuits
Experimental and numerical studies on the bonfire test of high- pressure hydrogen storage vessels Prof. Jinyang Zheng Institute of Process Equipment, Zhejiang.
Manual Density Business Today
Matter and Composition What is matter?  MATTER is anything which has mass and occupies space.  Matter is all things that we can see, feel, and smell.
Parts of the Anesthetic Machine A&A Pages
Flux Cored Arc Welding (FCAW)
1 Elsevier items and derived items © 2010 by Saunders, an imprint of Elsevier Inc. Chapter 20 Neonatal and Pediatric High-Frequency Ventilation.
Vaporizers – principles Dr. S. Parthasarathy MD., DA., DNB, MD (Acu), Dip. Diab. DCA, Dip. Software statistics- PhD ( physiology)
Vocabulary Set #1. Condensation the process of changing from a gas to a liquid.
GOVERNMENT ENGINEERING COLLEGE, BHARUCH (014) Chemical Engineering department SEM-iii Sub: fluid flow operation topic: orifice meter & rotAmeter Guid by:
Introductory Lecture Series: The Anesthesia Machine
The Anesthesia Machine: Introduction
MECH1300 Pneumatic Components Topics Pneumatic Cylinders Pneumatic Motors Other Pneumatic Actuators Pneumatic Directional Control Valves Pneumatic Flow.
GOVERNMENT ENGINEERING COLLEGE DAHOD
Anesthesia Equipment Session 3
Gas Laws Jed Wolpaw MD, M.Ed Sept 2, 2015.
Heat Exchangers and Condensers
Non-Rebreathing System
General Safety Practices
Vapor pressure, partial pressure, and concentration of inhaled anesthetics: The impact of atmospheric pressure on vaporizer output. Top. Variable bypass.
GOVERNMENT ENGINEERING COLLEGE, BHARUCH.
Parts of the Anesthetic Machine Parts 1 & 2: A&A pages
Pneumatic schematic of the Dräger Infinity V500 intensive care ventilator. A. Gas-mixture and gas-metering assembly. Gas from the supply lines enters the.
Vapor pressure, partial pressure, and concentration of inhaled anesthetics: The impact of atmospheric pressure on vaporizer output. Top. Variable bypass.
VALVES CONTREN Series #
Pressure Regulator, Pressure Gauge, Flowmeters
Sophie Young 21st February 2007
Levin P.D. , Levin D. , Avidan A.   British Journal of Anaesthesia 
Presentation transcript:

 The annular space. The clearance between the head of the float and the flow tube is known as the annular space.

 upper portion › orificial › Density is dominant  lower portion: › tubular › Viscosity is dominant

 The oxygen flow control knob is physically distinguishable: › distinctively fluted › projects beyond the other knobs › larger in diameter  All knobs: › color-coded › chemical formula or name marked on each › protected with a shield or barrier

 FLOW TUBES: › tapered glass tubes › double flow tubes  better visual discrimination at low flow rates 200ml to 1 L/min and 1 to 10 to 12 L/min  connected in series  total gas flow is that shown on the higher flow meter INDICATOR FLOATS AND FLOAT STOPS:  floats: plumb-bob, rotating skirted, ball  float stops: visible  SCALE: › marked directly on the flow tube or to the right of the tube › Rib guides are used in some flow tubes  SAFETY FEATURES: › individually hand-calibrated › The tube, float, and scale make an inseparable unit ( The entire set must be replaced if any component is damaged)

 An oxygen leak from the flow tube can produce a hypoxic mixture regardless of the arrangement of the flow tubes

 Sticking of the indicator float  damaged float can cause inaccurate readings  Backpressure from the breathing circuit  not aligned properly in the vertical  ambiguous scales

 The output represented in L/minute:  graphically  numerically  or both  Even when electrical power is totally interrupted, oxygen should continue to flow (flow control valves are non electronic)  Small conventional pneumatic “fresh gas” or “total flow” indicator

 prevent the creation and delivery of a hypoxic mixture  minimum oxygen concentration at the common gas outlet is between 23% and 25%  Limitations:  Wrong Supply Gas  Defective Pneumatics or Mechanics  Leaks Downstream  Inert Gas Administration  Dilution of Inspired Oxygen by Volatile Anesthetics

 Ohmeda Link-25 Proportion-Limiting Control system

 North American Dräger Oxygen Ratio Monitor Controller

 Allows direct communication between the oxygen high-pressure circuit and the low-pressure circuit  Actuation of the valve delivers 100% oxygen at 35 to 75 L/min to the breathing circuit  Provide a high-pressure oxygen source suitable for jet ventilation in ??  Hazards: › Barotrauma (if the anesthesia machine does not incorporate fresh gas decoupling or an appropriately adjusted inspiratory pressure limiter) › Patient awareness

 “old” copper kettle–like technology  “new” computerized control technology Physics:  Vapor Pressure (boiling point: desflurane 22.8°C, isoflurane 48.5°C, halothane 50.2°C, enflurane 56.5°C, and sevoflurane 58.5°C)  Latent Heat of Vaporization (number of calories required to change 1 g of liquid into vapor without a change in temperature)  Specific Heat ( the number of calories required to increase the temperature of 1 g of a substance by 1°C)  Thermal Conductivity (the speed at which heat flows through a substance)

 Flow Rate Output is less than the dial setting at very low and very high flow rates (<250 mL/min and 15 L/min)  Backpressure (pumping effect, is more pronounced at low flow rates, low dial settings, and low levels of liquid anesthetic)  Temperature (Automatic temperature-compensating mechanisms)  Carrier Gas Composition (sudden transient decrease in vaporizer output)

Simplified schematic of the Ohmeda Tec–type vaporizer

Simplified schematic of the North American Dräger Vapor 19.n and 20.n vaporizers