CMBEC 36 / APIBQ 42 Joint Conference 1 Technological assessment of laparoscopic monopolar electrosurgery instruments at CSSS du Lac-des-Deux-Montagnes.

Slides:



Advertisements
Similar presentations
High voltage testing of laparoscopic accessories
Advertisements

Part III: Safe Work Practices
1 The Short Story on Long Cycles Why the length of your steam sterilization cycle may effect your sterility assurance.
CESO Education Day November 30, 2004 Shawn Kenny Manager, Central Processing Department University Health Network – Toronto General Hospital.
PREVENTION OF CATHETER ASSOCIATED URINARY TRACT INFECTIONS (CAUTIs)
National ReUse Conference 2009 Atlanta, GA Presented by Chris Brand President/CEO Friends of Disabled Adults and Children, Too!
VALIDATION – ARE WE DOING ENOUGH David McNamara Manager – CSSD Holy Spirit Northside Private Hospital SRACA (NSW) Conference 10 th November 2011.
SAFE-ITSM STANDARDS.
Bookwalter® Retractor System Assembly Instructions
1 The aim…. ‘to enable assessors to objectively assess a laboratory’s compliance with the new standards’
Controlling CDI — Environmental services cleaning and disinfecting in the operating and procedure rooms Contents courtesy of Allina Healthcare.
Surgical Diathermy Machine
EFFECTIVE DELEGATION AND SUPERVISION
The Perioperative Nursing Role January 12th, 2009.
Audits in Infection Prevention and Control
CS2 – Cross Country Check-up Around the Table What Issues Are Filling Your Days? Québec Gnahoua Zoabli, P.Eng., M.Eng., Ph.D. Chief of biomedical engineering.
FMEA Applied to the Phenomenon of Retained Objects After Surgery Project Managers Dr. Joan Burtner and Dr. Laura Moody Mercer University School of Engineering.
MEDICAL DEVICE STERILIZATION
LOCKOUT/TAGOUT.
Medical Device Quality Assurance د فضل الاكوع المدير العام أستاذ مساعد بقسم الهندسة الطبية دكتوراه في الأجهزة الطبية – جامعة القاهرة
Controlling Hazardous Energy
Mechanical and fluidic integration of scintillating microfluidic channels into detector system 1 Davy Brouzet 10 th September 2014.
This presentation may contain some forward looking statements, particularly regarding operational prospects in 2008 and beyond, which involve a number.
What is the autoclave thing again. Sanitization: fancy word for Cleaning Removing of foreign material Dirt, body fluids, lubricants Using detergents,
Cleaning, Packaging & Sterilization of Dental Instruments
Frankfurt (Germany), 6-9 June 2011  Speaker Jean LAVALLÉE  Authors Janislaw TARNOWSKI, Jacques CÔTÉ, André GAUDREAU, Pierre GINGRAS, Mircea IORDANESCU.
CESO SEMINARS 2002 Insulation and Continuity Testing for Laparoscopic Instruments Presented by Stephen Moule CARSEN GROUP INC.
Integrated Fluorescent Probe and Radiofrequency Ablator Rachel Riti and Alex Walsh Advisers: Bart Masters and Anita Mahadevan-Jansen Department of Biomedical.
Flexible Robotics Presented by: Autum Artz. Objectives: Understand Flexible Robotics and the growth of tele-surgical devices. Describe and evaluate hardware.
2015 AORN Fire Safety Tool Kit
Emergency Nurses’ Knowledge and Attitudes Regarding Pain Keri Dillon, BSN, RN, CEN; Virginia Morse, PhD, RN; Sharon Ward, MS, RN, CEN Introduction Purpose.
The Joint Commission’s 2011 National Patient Safety Goals.
10/07/2009 Matthias Schuetzeberg Slide 1 Calibration Management In Food & Beverage.
Hazard Communication Presentation Presented by: Western Assurance Corp Desert Utility & Paving, LLC and Century Club Construction, LLC Design © 2005, 2012.
Theme 2: Hospital design to reduce adverse events CEO, Jane Kraglund Odense University Hospital.
OVERVIEW Cervical canal dilation - problem and its traditional solution Gynedil - new solution for dilation About product Advantages of the new method.
Checklist Modification 101 Make Me Your Own. The Basics One size doesn’t fit all. Every hospital should modify the checklist. Checklist modification creates.
TurkUrolap Nurse Laparoscopy Training Concepts TurkUrolap Nurse Laparoscopy Training Concepts Dr. Cenk Gurbuz Assocıate professor of urology ıstanbul,
Cost Benefit Analysis Using SteriTite ® Containers & MediTray ® Inserts Copyright 2003, Revised 2013.
Sr.Panchavarnam.  The theatre manager is responsible for the effective day to day running of the theatre, and maintaining high standards.  Responsibilities.
This presentation may contain some forward looking statements, particularly regarding operational prospects in 2009 and beyond, which involve a number.
Chemical Sterilization
Electromagnetic Interference (EMI) in the Hospital Lake Regional Health System.
TF_Arusha_Jun10/1 Framework for Evaluation of Male Circumcision Devices Tim Farley Department of Reproductive Health and Research World Health Organization.
Wet Packs: Strategies for Prevention and Resolution
SAFE USE OF DIATHERMY RACHEAL AFOLABI, RN, RM, H. Dip. PERI-OPERATIVE NURSING, H.Dip. HEALTH SYSTEM MANAGEMENT.
NANCY CHOBIN, RN, AAS, ACSP, CSPM, CFER CHOBIN & ASSOCIATES CONSULTING HOT TOPICS IN STERILIZATION AND DISINFECTION.
SteriTite Instructions for use
EFFECTIVE DELEGATION AND SUPERVISION
“NOTHING LEFT BEHIND” THE SURGICAL COUNTING in ISRAEL`S O.R A NATIONAL RESERCH EDNA LAVI, RN, ORN, BA RAMBAM HEALTH CARE CAMPUS HAIFA, ISRAEL.
Methods The initial audit was carried out retrospectively, looking at the acute paediatric presentations from January 2014 to May 2014 inclusive. Patient.
Management of Surgical Smoke Tool Kit Part III: An Overview of the Health Care Regulations, Standards, and Guidelines Related to Surgical Smoke.
IS THERE A SPECIFIC PROFESSIONAL ACT FOR CLINICAL ENGINEERS? Gnahoua Zoabli, P.Eng, M.Eng., Ph.D. 10:30 am on Friday May 27 th Performance and Quality.
2017 SPRING EDUCATION CONFERENCE
The Joint Commission’s 2011 National Patient Safety Goals
TOPIC: THE VALUE OF BIOMEDICAL ENGINEERS AND TECHNICIANS IN HOSPITAL PRESANTED BY: KWIZERA ABRAHAM STUDENT IN: BMET /FINALIST STU INSTITUTION: IPRC KIGALI.
Chad Keller, MSEd, RN, BSN, CRMST
Electrosurgical/Surgical Diathermy Units
Chapter 15 Safe Patient Handling.
Prepared by: Mohammed J. Mghari
Clinical engineering Lecture (8).
Clinical engineering Lecture (7).
Project Cascade – A simple technique to improve dissemination of learning points from Serious incidents and Never events Gowrishankar S1, Meadows S2, Ameerally.
Improving Simulations in the Post Anesthesia Care Unit
Clinical Alarm Systems - NPSG Goal # 6 -
Laparoscopy To examine peritoneal cavity and its viscera
Sterilization September 2018.
Control of Hazardous Energy
Continuous Blood Gas Monitor
Fundamental Use of Surgical Energy
Presentation transcript:

CMBEC 36 / APIBQ 42 Joint Conference 1 Technological assessment of laparoscopic monopolar electrosurgery instruments at CSSS du Lac-des-Deux-Montagnes Gnahoua Zoabli 1, Geneviève Foley 2, Pierre Cadieux 1, Éric Perreault 1, Sylvain Pomerleau 1 et Frédéric Ouellet 1 1 CSSS du Lac-des-Deux-Montagnes, St-Eustache, Québec; 2 École polytechnique, Montréal, Québec Service du génie biomédical

CMBEC 36 / APIBQ 42 Joint Conference 2 INTRODUCTION Thermal injuries observed during two recent interventions at the CSSS LDDM’s surgical unit Long term shift to Single Use Devices could be costly to CSSS LDDM This study aims to identify the cause of the incidents reported at the operating rooms at CSSS LDDM and To issue recommendations regarding the purchase of new reusable electrosurgical laparoscopic instruments Results and recommendations acceptable by medical doctors Sharing of this experience with peers

CMBEC 36 / APIBQ 42 Joint Conference 3 MATERIALS AND METHODS Scope of the assessment –Duration: 2 months during which single use devices are purchased –Laparoscope assessment was limited to 8 electrosurgical hooks and 2 spatulas Factors analyzed to determine the specific problem causing the reported thermal injuries –Instrument handling and wear –Electrode insulation testing –Isolative sheathing thickness measurement –Current CSSS LDDM sterilization parameters vs. surgical hooks/spatulas’ manufacturer instructions

CMBEC 36 / APIBQ 42 Joint Conference 4 Instrument handling and wear Two members of the biomedical engineering department witnessed a laparoscopic cholecystectomy to observe the normal usage of the instruments. Observations focused on contact between internal organs and instruments during cauterization (local zoom) Normal wear of instruments during an intervention. Friction of the laparoscope during its insertion into the trocar is also observed for resulting in potential mechanical damage. The number of procedures done with each instrument is unknown (no traceability system for surgical instruments) –This information could have helped to determine if normal insulation wear led to the reported incidents.

CMBEC 36 / APIBQ 42 Joint Conference 5 Electrode insulation testing Insulation layer was tested with an insulation tester (Atlas Technology Inc., ATI-013) –A beep indicates a leakage if any Only four (4) models of instruments tested for isolative sheathing thickness measurement –3 surgical hooks –1 spatula

CMBEC 36 / APIBQ 42 Joint Conference 6 Isolative sheathing thickness measurement Among the ten reusable instruments, four models of laparoscopes (3 hooks and 1 spatula) A section of the insulation material was removed on a limited mid-to-distal portion of the diameter of the laparoscope –representing the area outside the field of view and outside the trocar –possibly in contact with surrounding organs Insulation thickness was defined as the difference, measured by a caliper, between the unaltered instrument and the section missing a portion of insulation material

CMBEC 36 / APIBQ 42 Joint Conference 7 Current sterilization parameters vs. manufacturer instructions Steam SterilizationEthylene Oxide Sterilization Enzymatic Solution Pasteurization  Equipment : Amsco Evolution  Available Cycles  Vacuum  Gravity  Liquid  SFPP  Rapid SFPP  121 à 135°C  Cycle Validated à 132°C, 5 min  Equipment : 3M Steri- Vac / 5XL aerator  Hot Cycle  55°C  1 h exposition  10 h aeration  Cold Cycle  37°C  3 h exposition  20 h aeration  35 à 80% relative humidity  OE 100% Cartridge  Neutral Ph Detergent  Equipment : Olympic PMC 85-3  Thermal Disinfection  74°C  30 min exposition 132 to 135°C during 3 to 4 min

CMBEC 36 / APIBQ 42 Joint Conference 8 RESULTS Visual inspections of laparoscopes have revealed that the material used for joints in the instrument tends to become degraded first. Electrode insulation testing –A few designs were considered to test the insulation of laparoscopes before every sterilization cycle. –Even though only the distal third of the instruments would need to be verified, none of the designs were simple and efficient enough to ensure it could be successfully added to the nursing staff’s workload. –The major issue in optimizing insulation testing is the need to insulate the tip of the electrode in order to test exclusively the insulative sheathing. Insulative sheathing thickness –Insulation layer thickness for the four tested instruments were of 0.27 mm, 0.39 mm, 0.49 mm and 0.53 mm. The model involved in the incidents has the thinnest insulation thickness (0.27 mm). Sterilization parameters vs. Manufacturer recommendations –The instructions provided by the manufacturer of the instrument involved in the incidents are to steam sterilize it from 132 to 135°C during 3 to 4 minutes. The cycle used by the CSR of the hospital is 132°C during 5 minutes 1 minute more. –The sterilization settings currently under validation to replace the current cycle are 134°C during 4 minutes.

CMBEC 36 / APIBQ 42 Joint Conference 9 DISCUSSION The availability of single use laparoscopes – during the assessment of the reusable laparoscopes – has re-established surgeons’ confidence in the instruments and reduced the risk of a third incident occurring. Considering the necessity of safely cauterizing even when the insulative sheathing is in contact with patient organs, this layer must imperatively be of great quality especially in the distal third of the instrument ( a very small portion of the instrument is visible during cauterization) The insulation layer must be strong enough to withstand normal usage due to intra- and peri-operative handling. Problems reported with the instrument with the thinnest insulative layer (0.27mm). –is almost half the thickness of the thickest layer (0.53 mm) –sheathing thickness is critical in assuring patient safety. –A thicker ( ≥ 0.4 mm ) and more flexible insulative layer would reduce the probability of electrical breakdown when the instrument heats up from the flow of high intensity current. –Risks associated with wear are also reduced since the same chink in a thicker layer is, overall, a smaller flaw in the material –Thicker insulative layer may result in a longer useful life for the instrument.

CMBEC 36 / APIBQ 42 Joint Conference 10 DISCUSSION (Cont’d) The absence of visible joints should also be favoured since these tended to degrade faster than the other materials in the instrument. The sterilization procedure at the CSSS LDDM (132°C, 5 min) is a minute longer than the manufacturer’s instructions. –Even though this difference is small, it could compromise material integrity over a large number of cycles. The sterilization cycle of 134°C during 4 minutes currently under analysis could allow to follow the manufacturer’ instructions. –Ultimately, it is essential to consider the hospital’s sterilization cycles when purchasing reusable devices. The establishment of a surgical instrument traceability system would allow quantification of laparoscope usage. –A threshold, slightly more restrictive than manufacturer recommendations, could be established to dictate when an instrument should be retired. For example, an annual replacement of instruments could be considered –In the absence of a traceability system, we recommend replacing the monopolar instruments after two years on an amortization basis and not on their useful life.

CMBEC 36 / APIBQ 42 Joint Conference 11 RECOMMENDATIONS & CONCLUSION Patient safety can be compromise by: –Quality of the electrode insulation sheathing material, –handling of instruments during surgery and –non-compliant sterilization cycle parameters can. The following guidelines should be followed when purchasing monopolar surgical instrument in laparoscopy: –minimal insulation thickness of 0.4 mm, –compatibility with the institution’s sterilization cycles, and –visual inspection. A periodic insulation testing in the normal validation cycle of the laparoscopic instruments

CMBEC 36 / APIBQ 42 Joint Conference 12 ACKNOWLEDGEMENTS We would like to thank Mrs. Laurette Aubé, assistant head nurse in the central sterilization room, for sharing helpful knowledge regarding sterilization issues.