High Performance MALDI-TOF Mass Spectrometry Marvin Vestal and Kevin Hayden Virgin Instruments Corp. Sudbury, MA.

Slides:



Advertisements
Similar presentations
High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging
Advertisements

Kingdon Trap Hande Akbas Kealan Naughton Alexander Fuchs-Fuchs.
electrostatic ion beam trap
Experience with Bunch Shape Monitors at SNS A. Aleksandrov Spallation Neutron Source, Oak Ridge, USA.
Method of beam extraction from a synchrotron by the instrumentality of multilayer Cu-Fe shield Bondarenko Alexey.
Oligonucleotide analysis by Linear MALDI-TOF mass spectrometry Stephen J. Hattan , Kenneth C. Parker, Marvin L. Vestal; SimulTOF Systems, 60 Union Ave,
Introduction to MALDI-TOF MS
Current-Mode Multi-Channel Integrating ADC Electrical Engineering and Computer Science Advisor: Dr. Benjamin J. Blalock Neena Nambiar 16 st April 2009.
MALDI-TOF Mass Spectrometry and Introduction to Proteomics Dr. Steve Hartson Oklahoma State University Dept. Biochemistry and Molecular Biology Recombinant.
Analyzing Biological and Organic Polymers by MALDI-TOF Jonathan A. Karty, Ph.D.
Lecture Three: "MALDI-TOF MS, Up Close, and Personal Review of principles Delayed Extraction Reflector / reflectron Post-source Decay (PSD)
Development of TOF-MS from intellectual curiosity to practical technique Marvin Vestal SimulTOF Systems The Jim Waters Symposium at PittCon March 2, 2014.
Mass Spectrometry. How is MS Done? Basic MS Instrumentation.
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform.
PULSE REPETITION RATE OPTMIZATION IN SLR STATIONS TO PROVIDE MINIMUM SYSTEMATIC ERROR OF RANGING M.A. Sadovnikov M.A. Sadovnikov Institute for Precision.
Applied Sensor Technology. Outline Introduction Examples of Sensors Basic readout electronics Semiconductor detectors.
A Miniature Ion Mobility Spectrometer for Explosives Detection Andrew Goodin, William F. Siems, Christina L. Crawford, Prabha Dwivedi, and Herbert H. Hill,
Interference and Diffraction
Beam line characterization with the TOFs1 Demonstrating the emittance-momentum matrix Mark Rayner, CM26 California, 24 March Initial.
PROTEIN IDENTIFICATION BY MASS SPECTROMETRY. OBJECTIVES To become familiar with matrix assisted laser desorption ionization-time of flight mass spectrometry.
Spatial profiles of copper atom density in a Cu/Ne hollow cathode discharge P. Hartmann*, T.M. Adamowicz**, E. Stoffels and W.W. Stoffels, Department of.
Mass Spectrometry 101 An Introductory Lecture On Mass Spectrometry Fundamentals Presented to the Sandler Mass Spectrometry Users’ Group University of California.
MALDI-TOF-TOF with High Resolution Precursor Selection and Multiplexed MS-MS Poster Number ThP 618 Kevin Hayden, Stephen C. Gabeler, Mark Dahl and Marvin.
Janusz Podliński, Artur Berendt, Jerzy Mizeraczyk Centre for Plasma and Laser Engineering The Szewalski Institute of Fluid-Flow Machinery Polish Academy.
Proteomics Informatics – Overview of Mass spectrometry (Week 2)
Comparison of chicken light and dark meat using LC MALDI-TOF mass spectrometry as a model system for biomarker discovery WP 651 Jie Du; Stephen J. Hattan.
The ProteomeWorks System is the global alliance of Micromass and Bio-Rad Laboratories dedicated to furthering proteomics research.
Smoke Particle Velocity Measurements in Point to Plane Corona Discharges Noureddine ZOUZOU* Eric MOREAU Gérard TOUCHARD Laboratoire d’Etudes Aérodynamiques.
1 §5 Multiplexed and distributed sensors 1.Basic sensor array topologies 2.Time division multiplexing (TDM) 3.Wavelength division multiplexing (WDM) 4.Optical.
Marvin Vestal and Kevin Hayden Virgin Instruments Corp. Sudbury, MA
SACE Stage 2 Physics Motion in 2 Dimensions.
Tissue Imaging with New MALDI-TOF(TOF) Instrumentation Presented at MSIA 2013 Vanderbilt University April 17,2013 Marvin Vestal, CSO.
Tissue Imaging by 5 kHz High-Performance MALDI-TOF Poster Number TP191 Christina Vestal 1, Kenneth Parker 1, Kevin Hayden 1, George Mills 1, Marvin Vestal.
Mass Spectrometry I Basic Data Processing. Mass spectrometry A mass spectrometer measures molecular masses. The mass unit is called dalton, which is 1/12.
INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable.
-Aluminum housing with o-ring seal allows for washing (salt removal), plate cleaning and regeneration enabling for multiple use -Sample elution for MS.
Mass Analyzers IV Quadrupole Ion Traps Chem 5181 – Fall 2007 J. Kimmel.
Marvin Vestal and Kevin Hayden SimulTOF Systems
Overall efficiency and speed of analysis – key factors are ions per sample molecule desorbed and ions produced per laser pulse Precision and accuracy of.
MS Calibration for Protein Profiles We need calibration for –Accurate mass value Mass error: (Measured Mass – Theoretical Mass) X 10 6 ppm Theoretical.
A new "Molecular Scanner" design for interfacing gel electrophoresis with MALDI-TOF ThP Stephen J. Hattan; Kenneth C. Parker; Marvin L. Vestal SimulTof.
MS Intro. MS requires gas-phase ions, why? MS uses magnetic and electric fields to control the path of a compound based on mass to charge ratio (m/z)
For all CHEM5161 students: The first day of class for CHEM5161 (Analytical Spectroscopy) will be on TUE Sept 4 (following Labor Day). There will be no.
INF380 - Proteomics-51 INF380 – Proteomics Chapter 5 – Fundamentals of Mass Spectrometry Mass spectrometry (MS) is used for measuring the mass-to-charge.
Supplement material to Ref. JASMS Ms. No Detection of an amine impurity and quality assessment related to this impurity of the UV MALDI matrix.
Chemistry 367L/392N Macromolecular Chemistry Lecture 8.
Mass Spectrometry Instrumentation: Ionization methods and analyzers.
The Future of Time-of-flight Mass Spectrometry
Mass Analyzers : Time-of-flight Mass Spectrometry CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Joel Kimmel Fall 2007.
1 CH908 Structural Analysis by Mass Spectrometry revision lecture. Prof. Peter O’Connor.
RF source, volume and caesiated extraction simulations (e-dump)
Scintillator tile – SiPM development at ITEP Michael Danilov, ITEP CALICE Meeting, Casablanca, 23 Sep 2010.
Salamanca, March 16th 2010 Participants: Laboratori de Proteomica-HUVH Servicio de Proteómica-CNB-CSIC Participants: Laboratori de Proteomica-HUVH Servicio.
A new mobility analyzer for routine measurement of atmospheric aerosol in the diameter range of 0.4−7.5 nm Institute of Physics, University.
Quantitative MALDI-TOF for Clinical Applications.
RD Instruments Home of the ADCP Measuring Water in Motion and Motion in Water AVOIDING BIASED DATA DURING REAL TIME HADCP DATA COLLECTION.
2014 생화학 실험 (1) 6주차 실험조교 : 류 지 연 Yonsei Proteome Research Center 산학협동관 421호
What is Mass Spectrometry? Mass spectrometry could be considered as an analytical technique that involves the study in the gas phase of ionized molecules.
Yonsei Proteome Research Center Peptide Mass Finger-Printing Part II. MALDI-TOF 2013 생화학 실험 (1) 6 주차 자료 임종선 조교 내선 6625.
1 st : Mass Spectrometry on Heavy Molecules with Cryogenic Detectors - status Project Review 2004 PROTEOM 2 nd : Measurement of the Quenching Factors in.
Peptide Mass Finger-Printing Part II. MALDI-TOF
Acknowledgements Slides and animations were made by Dr. Jon Karty Mass Spectrometry Facility Indiana University, Bloomington.
Mass Spectrometry 101 Hackert - CH 370 / 387D
Time of Flight Analyzers
Presented by: Nigel Chou and Vivian Hecht December 9, 2011
Starter Complete the quiz and hand in..
An Introduction to the ProteinChip® Reader
Analyzing Biological and Organic Polymers by MALDI-TOF
Mass Spectrometry THE MAIN USE OF MS IN ORG CHEM IS:
Slope measurements from test-beam irradiations
Presentation transcript:

High Performance MALDI-TOF Mass Spectrometry Marvin Vestal and Kevin Hayden Virgin Instruments Corp. Sudbury, MA

Outline Goals: Design and build MALD-TOF Analyzers Optimized for Selected Applications Approach: –Modular construction using common elements –State-of-the-art technology, e.g. 5 khz laser –Focus on the needs of the application –Keep it small, simple, and reliable without sacrificing performance Analyzer geometries –Linear for broad mass range and sensitivity at high mass (to 500 kDa). –Reflector for high resolution, mass accuracy, and sensitivity at low mass (0.1-5 kDa) –Reversed geometry reflector for both resolution and sensitivity in intermediate mass range (1-50 kDa) Optimization Strategy Results

Time-of-flight analyzer geometries Ion SourceDrift TubeDetector Reflector Linear Reflecting Reversed Geometry Reflecting Reflectors

Reflector MS for MALDI mass fingerprinting peptides, metabolites, and other small molecules High sensitivity and dynamic range in mass range kDa Highest possible resolving power and mass accuracy (1 ppm RMS) High throughput Low cost, stable, reliable, and small No operator expertise required

V V1V1 0 d0d0 d1d1 DsDs DvDv Sample Plate Extraction Electrode Space focus Velocity focus y=V/(V-V 1 ) D s = 2d 0 y 3/2 D v =D s +(2d 0 y) 2 /(v n  t) Where v n =(2zV/m) 1/2 and  t is time lag For single field source, V 1 =0, y=1 Simultaneous 1 st and 2 nd order focus D v =6d 0, v n  t=d 0 Wiley and McLaren RSI (1955) 26, Vestal and Juhasz JASMS (1998) 9,

V d0d0 Sample plate Extraction electrode 0 V1V1 V2V2 d3d3 d40d40 d4d4 Drift tube velocity focus DmDm 2-stage reflector D m1 = 4d 4 w 3/2 +4d 3 [w/(w-1)][1-w 1/2 ] 3D m2 = 4d 4 w 5/2 +4d 3 [w/(w-1)][1-w 3/2 ] d 4 = d 4 0 (V-V 1 )/(V 2 -V 1 ) w=V/(V-V 1 ) Simultaneous 1 st and 2 nd order focus 4d 3 /D m =(w-3)/w 4d 4 /Dm = w -3/2 + (4d 3 /D m )/(w + w 1/2 ) D

Contributions to relative peak width,  m/m, with 1 st and 2 nd order velocity focusing Initial position,  x: R s1 = 2[(D v -D s )/2d 0 y](  x/D e ) Initial velocity,  v 0 : R v1 =(4d 0 y/D e )(  v 0 /v n )[(1-(m/m * ) 1/2 ] R v2 =2[2d 0 y/(D v -D s )] 2 (  v 0 /v n ) 2 =0 R v3 =2[2d 0 y/(D v -D s )] 3 (  v 0 /v n ) 3 R v = R v1 + R v3 Time error,  t: R t = 2  t/t = 2  tv n /D e Trajectory error,  L: R L = 2  L/D e Voltage error,  V: R V =  V/V Resolving power: R -1 = [R s1 2 + R v 2 + R t 2 + R L 2 + R V 2 ] -1/2

Instrument Parameters for Reflecting Analyzer (single-field source and two-field mirror) D e =3200, d 0 =3, D=2264, d 3 =114.2, d 4 0 =127.7 all in mm V=8.56, V 1 =6.31, V 2 =9.23 all in kV  t=1.5 nsec (0.5 nsec bins, 1 nsec single ion pulse width) Focus mass m * =1.6 kDa D v =6d 0 = 18 mm, v n (for m * )= mm/nsec Time lag  t=d 0 /v n = 93 nsec K=2d 0 /(D v -D s )=0.5 Initial Conditions for MALDI (typical)  v 0 =400 m/sec,  x=0.01 mm Trajectory error and voltage error assumed to be small

Calculated resolving power for reflecting analyzer.  v 0 =400m/s,  x=0.01 mm for I and III and 800 m/s and 0.02 mm for II.  t=1.5 nsec for I and II and 0.75 nsec for III, I III II

Conditions for minimizing relative peak width,  m/m Initial position,  x: R s1 = 2K -1 (  x/D e ) Initial velocity,  v 0 : R v1 =(4d 0 y/D e )(  v 0 /v n )[(1-(m/m * ) 1/2 ] R v3 =2K 3 (  v 0 /v n ) 3 R v = R v1 + R v3 Time error,  t: R t = 2  t/t = 2  tv n /D e Peak Width:R= [R s1 2 + R v 2 + R t 2 + R L 2 + R V 2 ] 1/2 Optimum focus: dR/dK =0, K=2d 0 /D v -D s Optimum velocity: dR/dv n =0, v n =C(V/m*) 1/2 (can be optimized at m* or any other mass)

nominal optimum R R s1 R v3 K=2d 0 /(D v -D s ) Minimum peak width as function of focusing parameter K. Optimum value corresponds to dR/dK=0

 t=1.5  t=0.75 R v3+ RtRt Minimum peak width as function of V/m*(indicated by arrows for  t= 1.5 and 0.75 nsec). For m*= 1600 Da optimum voltages are 2.4 and 3.4 kV, respectively.

Optimized resolving power for reflecting analyzer,  v 0 =400m/s,  x=0.01 mm.  t=1.5 nsec for I and II and 0.75 nsec for III and IV. Red lines nominal, black at optimum for focus mass. I III IV II

11,600 13,000BSA digest Resolution is 40% of calculated

Experimental resolving power for reflecting analyzer on BSA digest

Contributions to relative peak width,  m/m Initial position,  x: R s1 = [(D v -D s )/D e ](  x/d 0 y) Initial velocity,  v 0 : R v1 =(4d 0 y/D e )(  v 0 /v n )[(1-(m/m * ) 1/2 ] R v2 =2[2d 0 y/(D v -D s )] 2 (  v 0 /v n ) 2 R v3 =2[2d 0 y/(D v -D s )] 3 (  v 0 /v n ) 3 Time error,  t: R t = 2  t/t = 2  tv n /D e Trajectory error,  L: R L = 2  L/D e Voltage error,  V: R V =  V/V Resolving power: R -1 = [R s1 2 + R v 2 + R t 2 + R L 2 + R V 2 ] -1/2 Not small!!!

Calculated resolving power for reflecting analyzer with  V/V =70 ppm I II III

Comparison of calculated resolving power with experimental results With  V/V=70 ppm calculated

Comparison of calculated resolving power with experimental results With  V/V=70 ppm. Measured 5  channel plate ETP MagnetTOF TM Calculated  t=0.75 nsec Calculated  t=0.75 nsec

Contributions to relative peak width,  m/m, with 1 st and 2 nd order velocity focusing (size matters) Initial position,  x: R s1 = 2[(D v -D s )/2d 0 y](  x/D e ) Initial velocity,  v 0 : R v1 =(4d 0 y/D e )(  v 0 /v n )[(1-(m/m * ) 1/2 ] R v2 =2[2d 0 y/(D v -D s )] 2 (  v 0 /v n ) 2 =0 R v3 =2[2d 0 y/(D v -D s )] 3 (  v 0 /v n ) 3 Time error,  t: R t = 2  t/t = 2  tv n /D e Trajectory error,  L: R L = 2  L/D e Voltage error,  V: R V =  V/V Resolving power: R -1 = [R s1 2 + R v 2 + R t 2 + R L 2 + R V 2 ] -1/2

D e =3.2 m D e =6.4 m R v3 K=2d 0 /(D v -D s ) Minimum peak width as function of focusing parameter K for effective Lengths of 3.2 and 6.4 m. Optimum values correspond to dR/dK=0, And is proportional to D e 0.84 for these initial conditions.

Calculated maximum resolving power as function of effective length neglecting voltage error.  t=0.75 nsec  t=1.5 nsec Current instrument

 t=1.5 nsec  t=0.75 nsec (m) Calculated maximum resolving power as function of effective length neglecting voltage error.

Conclusions Resolving power of prototype reflecting analyzer is limited by voltage error or trajectory error If the total contribution of these two mass independent errors can be reduced to <10 ppm, then maximum resolving power of 50,000 is feasible with current prototype. Maximum resolving power is proportional to D e ,000,000 resolving power is theoretically feasible with an effective length of 100 m (physical length about 40 m) But!!! Errors independent of length must be less than 0.5 ppm (5 millivolts at 10 kilovolts)

The Virgins