Fluorescence Spectroscopy

Slides:



Advertisements
Similar presentations
PCSIR Labs. Karachi Pakistan
Advertisements

Big Question: We can see rafts in Model Membranes (GUVs or Supported Lipid Bilayers, LM), but how to study in cells? Do rafts really exist in cells? Are.
HORIBA Jobin Yvon Fluorescence Division Presents: Time-Resolved Fluorescence Spectroscopy Edison, NJ March 15, 2005.
Fluorescence Fluorescent corals.
Lecture 12 Molecular Photophysics
UV / visible Spectroscopy
Molecular Fluorescence Spectroscopy
1 Molecular Luminescence Spectroscopy Lecture 31.
The iron content of runoff from a banana ranch is a necessary analytical parameter to analyze. A 25.0mL sample of the runoff was acidified with HNO3 and.
Chapter 19 Spectrophotometry : Instruments & Application.
Lecture 3 Kinetics of electronically excited states
Luminescence (Miklós Nyitrai; 27 th of February, 2007)
Methods: Fluorescence Biochemistry 4000 Dr. Ute Kothe.
Molecular Luminescence
Luminol Chemoluminescence
Molecular Luminescence Spectrometry
Molecular Luminescence Spectrometry Chap 15. Three Related Optical Methods Fluorescence Phosphorescence Chemiluminescence } From excitation through absorption.
Luminescence is rare in nonaromatic hydrocarbons. Possible if highly conjugated due to  –  * transitions. Seyhan Ege, Organic Chemistry, D.C. Heath and.
Fluorescence. Topics Sensitivity (nM-pM) –Contrast UV-vis measurements Derivatization Laser-Induced Fluorescence.
Light and Fluorescence
Luminescence Spectrometry
Fluorescence. Topics Definition Instrumentation Sensitivity (nM-pM) –Contrast UV-vis measurements Derivatization Laser-induced Fluorescence.
Atomic Absorption Spectroscopy Elemental analysis Dissolved metals Source –Hallow Cathode Lamp –Cathode of specific element Sample –Acetylene-air flame,
Chapter 15 Molecular Luminescence Spectrometry Molecular Fluorescence  Optical emission from molecules that have been excited to higher energy levels.
INTRO TO SPECTROSCOPIC METHODS (Chapter 6 continued ) Quantum-Mechanical Properties Of Light Photoelectric Effect Photoelectric Effect Energy States of.
Atomic Absorption Spectroscopy
4-1 Chap. 7 (Optical Instruments), Chap. 8 (Optical Atomic Spectroscopy) General design of optical instruments Sources of radiation Selection of wavelength.
Chem. 133 – 3/12 Lecture. Announcements I HW 2.1 problem due today Quiz 3 also today Lab – Term Project Proposal due next Thursday Sign Up (see details.
Illumination and Filters Foundations of Microscopy Series Amanda Combs Advanced Instrumentation and Physics.
Wickstrom PR 613 Fluorescence. Source Dispersing Sample Detector Computer Lens Dispersing Instrumentation.
Dr. Hisham E Abdellatef 2. Spectrofluorimetry Dr. Hisham E Abdellatef
Fluorescence, Phosphorescence, & Chemiluminescence
FLUORESCENCE Joanna Piechowska. 1. Colors of the visible spectrum 2. Absorption of UV-VIS light –Types of electronic transitions in polyatomic molecules.
Molecular Luminescence Spectroscopy Chapter 15 Fluorescence, Phosphorescence and Chemiluminescence.
IPC Friedrich-Schiller-Universität Jena 1 6. Fluorescence Spectroscopy.
Optical Characterization methods Rayleigh scattering Raman scattering transmission photoluminescence excitation photons At a glance  Transmission: “untouched”
CHAPTER 15: MOLECULAR LUMINESCENCE. Chapter LUMINESCENCE TECHNIQUES Emission of light is used to determine certain properties,e e.g.structure and.
Fluorescence spectroscopy Komal Choudhary Lecturer School of Biotechnology DAVV Indore.
Chapter 15 Molecular Luminescence Spectrometry. A very sensitive and selective instrumental technique with some of the lowest LOD's for molecules that.
19 장 Fundamentals of Spectrophotometry Fundamentals of Spectrophotometry –19-1Properties of Light –19-2Absorption of Light –19-3 The Spectrophotometer.
Chapter 15 Molecular Luminescence Spectrometry
Fluorescence Absorption of light occurs within ~ seconds, leaving a molecule in an excited state What happens next? –If no photon is re-emitted,
Fluorometric Analysis
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 19 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Molecular Luminescence
Relationship between A(OD) and %T Transmittance, T = P / P 0 % Transmittance, %T = 100 T Absorbance, A = log 10 P 0 / P A = log 10 1 / T A = log
Chapter 15 Molecular Luminescence Spectrometry Three types of Luminescence methods are: (i) molecular fluorescence (ii) phosphorescence (iii) chemiluminescence.
The iron content of runoff from a banana ranch is a necessary analytical parameter to analyze. A 25.0mL sample of the runoff was acidified with HNO3 and.
Fluorescence Spectroscopy
1.1Excited electronic states Each electron has unique set of quantum numbers (Pauli Exclusion Principle) n principle (1s, 2s, 3s,…) langular momentum (l.
Fluorescence, Phosphorescence, & Chemiluminescence
1.Stable radiation source 2.Wavelength selector 3.Transparent sample holder: cells/curvettes made of suitable material (Table 7- 2) 4.Radiation detector.
MOLECULAR FLUORESCENCE SPECTROSCOPY. Fluorescence is a form of photoluminescence; and this later is a type of luminescence that occurs when certain molecules.
Molecular Luminescence Spectroscopy
Fluorescence spectroscopy, Einstein’s coefficients Consider a molecule with two energy levels S a and S b The rate of transition.
1.1 What’s electromagnetic radiation
UV SPECTROSCOPY Absorption spectra.
1 Molecular Luminescence Spectroscopy Lecture 29.
PHCM561t WS Lecture 4 LUMINESCENCE Quinine Dr. Rasha Hanafi
Molecular Fluorescence Spectroscopy
1 Instrumental Analysis Tutorial 2. 2 Objectives By the end of this session the student should be able to: 1.Describe the grating principle of work. 2.Describe.
Chem. 133 – 3/9 Lecture.
Chem. 133 – 3/14 Lecture.
Ultrafast Spectroscopy
Transient Absorption (Courtesy of Kenneth Hanson, Florida University): The technique applied to molecular dynamics Source hn Sample Detector.
Lecture 4 LUMINESCENCE PHCM561t WS Quinine © Dr Fazal
For B.Pharm IIIrd yr students
Overview of spectroscopy transitions
Molecular Luminescence Spectrometry
Fluorescence Spectroscopy
Presentation transcript:

Fluorescence Spectroscopy CHM 5175: Part 2.5 Fluorescence Spectroscopy Source hn Sample Detector Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00

Fluorescence Spectroscopy First observed from quinine by Sir J. F. W. Herschel in 1845 Filter Church Window 400nm SP filter Yellow glass of wine 400 nm LP filter hn Quinine Solution (tonic water) Observe Blue emission Herschel concluded that “a species in the solution exert its peculiar power on the incident light and disperses the blue light.”

Fluorescence Spectroscopy Measuring the light given off by an electronically excited state. Ground State (S0) Singlet Excited State (S1) hn Fluorescence hn Excitation Emission Intersystem Crossing hn Phosphorescence Emission Triplet Excited State (T1)

Fluorescence Spectroscopy Singlet Excited State (S1) Fluorescence Spin allowed Fast (ns) Organic molecules hn Emission Triplet Excited State (T1) Phosphorescence Spin “forbidden” slow (ms to s) Transition metal complexes hn Emission

Jablonski Diagram Excitation Internal Conversion Fluorescence Non-radiative decay Intersystem Crossing Phosphorescence S1 T2 Energy T1 S0

Fluorescence 2 1 3 1) Excitation -Very fast (< 10-15 s) -No structure change 2) Internal Conversion -Fast (10-12 s) -Structure change 3) Fluorescence -”Slow” (10-9 s) - No structure change 2 1 S1 Energy 3 S0 Geometry

Fluorescence Internal Conversion (sprinter) “always” wins! Sprinter (7 m/s) Snail (0.005 m/s) n3 S2 n2 n1 IC n3 S1 n2 n1 Internal Conversion (sprinter) “always” wins! Absorption Fluorescence Kasha’s Rule: Emission predominantly occurs from the lowest excited state (S0 OR T1) S0 Internal Conversion (1012 s-1) S2 Fluorescence (109 s-1)

Fluorescence Kasha’s Rule: Kasha Laboratory Building AKA Institute of Molecular Biophysics 1920-2013 Kasha’s Rule: Emission predominantly occurs from the lowest excited state (S0 OR T1)

Fluorescence Kasha’s Rule: Emission predominantly occurs from the lowest excited state (S0 OR T1) Blue Higher E Red Lower E Internal Conversion S0 S1 Eabsorption > Eemission Emission is red-shifted (bathochromic) relative to absorption Absorption is blue-shifted (hypsochromic) relative to emission

Mirror Image Rule Vibrational levels in the excited states and ground states are similar An absorption spectrum reflects the vibrational levels of the electronically excited state An emission spectrum reflects the vibrational levels of the electronic ground state Fluorescence emission spectrum is mirror image of absorption spectrum S0 S1 v=0 v=1 v=2 v=3 v=4 v=5 v’=0 v’=1 v’=2 v’=3 v’=4 v’=5

Mirror Image Rule n4 n3 S1 n2 n1 n4 n3 S0 n2 n1

Mirror Image Rule fluorescein ethidium bromide Anthracene

Stokes Shift Stokes Shift: Internal Conversion Difference in energy/wavelength between absorption max and emission max. Internal Conversion S0 S1 Sensitivity to local environment: Solvent polarity Temperature Hydrogen bonding

Solvent Dependence Stokes Shift: Solvatochromism Difference in energy/wavelength between absorption max and emission max. 4-dimethylamino-4'-nitrostilbene (DNS) Solvatochromism

Solvatochromism

Singlet Excited State (S1) Triplet Excited State (T1) Jablonski Diagram S2 Excitation Internal Conversion Fluorescence Non-radiative decay Intersystem Crossing Phosphorescence S1 T2 Energy T1 S0 hn Intersystem Crossing Emission Singlet Excited State (S1) Triplet Excited State (T1) Ground State (S0)

Phosphorescence 2 3 2 1 4 2 1) Excitation -Very fast (10-15 s) -No structure change 2) Internal Conversion -Fast (10-12 s) -Structure change 3) Intersystem Crossing -No Structure change 4) Phosphorescence -”Slow” (10-6 s) - No structure change T2 2 3 S1 2 1 T1 E 4 2 S0 Geometry

Emission Fluorescence Phosphorescence Rates: Lifetime: Dl: O2 sensitive: Fast (10-9s-1) nanoseconds <100 nm no Slow (10-6 – 0.1 s-1) >microseonds >100 nm Yes

Fluorescence vs Phosphorescence Internal Conversion (10-12 s) S2 Intersystem Crossing w/ Heavy atom (< 10-12 s) w/o Heavy atom (> 10-9 s) S1 E T1 Excitation (10-15 s) Fluorescence (10-9 s) Phosphorescence (10-6 s) S0

Emissive Molecules Phosphorescent Fluorescent Perylene OEP PtOEP Ir(ppy)3 BODIPY Fluorescein Rose Bengal [Ru(bpy)3]2+ Coumarin Anthracene Anthracene + ICH3 C60

Fluorometer Source Excitation Detector Sample Emission hn hn Variables Excitation Wavelength Excitation Intensity Emission Wavelength Filters

Fluorometer 3 1 2 4 2 Components 1) Light source 2) Monochrometer 3) Sample 4) Detector 5) Filters 6) Slits 7) Polarizers 4 2

Fluorometer: Simple Diagram Xenon Lamp Grating Mirrors Excitation Monochromator Emission Monochromator PMT Two light sources = Two monochromators! 1 for excitation 1 for emission Sample Grating

Fluorometer: Medium Diagram Grating Mirror Mirror Lens Sample

Fluorometer: Hard Mode Grating Mirrors Mirror Grating

Fluorometer: Hard Mode 2 450 W Xe 300 nm blaze 1200 g/mm exit slit iris NIR: 9170-75=950-1700 nm 1000 nm blaze 600 g/mm grating shutter polarizer slit r V UV-VIS: R928 = 250-850nm 500 nm blaze 1200 g/mm grating V V

(Materials Characterization) Horiba JY Fluoromax-4 Horiba JY Fluoromax-4 MAC Lab (Materials Characterization) Dr. Bert van de Burgt CSL 116

Measuring Emission Spectra Procedure 1) White light source on 2) Shift excitation grating to desired wavelength (excitation wavelength) 3) Light enters sample chamber 4) Light Hits the Sample 5) Emission from the sample enters emission monochromator 6) Set emission grating 7) Detect emitted light at PMT 8) Raster emission grating PMT Xenon Lamp Excitation Monochromator Emission Sample Ex Grating Em Grating 1 2 3 7 4 5 6 8

Measuring Emission Spectra Absorption Spectrum Procedure 1) White light source on 2) Shift excitation grating to desired wavelength (excitation wavelength) 3) Light enters sample chamber 4) Light Hits the Sample 5) Emission from the sample enters emission monochromator 6) Set emission grating 7) Detect emitted light at PMT 8) Raster emission grating Emission Spectrum Excitation at 450 nm Emission from 550 – 900 nm

Excitation Spectrum S3 S3 S2 S1 S1 S2 IC n3 S1 S1 n2 S2 n1 Absorption Fluorescence Fluorescence emission spectrum is the same regardless of the excitation wavelength! S0

Excitation Spectrum But intensity changes! S3 S2 S1 S0 Absorption Absorbance n3 S3 n2 n1 n3 S2 n2 n1 IC n3 S1 n2 n1 Fluorescence emission spectrum is the same regardless of the excitation wavelength! Absorption Fluorescence But intensity changes! S0

Excitation Spectrum Monitor emission (Fixed l) Absorbance Scan Through Excitation l

Measuring Excitation Spectra Procedure 1) Shift emission grating to desired wavelength (monitor emission max) 2) Shift excitation grating to stating wavelength 3) Light source on 4) Light Hits the Sample 5) Emission from the sample enters emission monochromator 6) Detect emitted light at PMT 7) Raster excitation grating PMT Xenon Lamp Excitation Monochromator Emission Sample Ex Grating Em Grating 3 2 7 6 4 5 1

Excitation Spectrum If emitting from a single species: Absorption Spectrum If emitting from a single species: Excitation spectrum should match absorption spectrum!

Fluorometer 3 1 2 4 2 Components 1) Light source 2) Monochrometer 3) Sample 4) Detector 5) Filters 6) Slits 7) Polarizers 4 2

Samples Solutions Thin Films Powders Crystals

Solution Fluorescence Top View Source hn Sample Detector Excitation Emission Excitation Beam Emission non-emitting molecules filter effect “self”-absorption

For Fluorescent Samples: Filter Effect Anthracene For Fluorescent Samples: Absorbance < 1.0

Real emission spectrum + Solid Samples Emission Spectrum Thin Films/Solids Ex: 380 nm Source Sample Detector Real emission spectrum + Second Order

Real emission spectrum + Solid Samples Emission Spectrum Thin Films/Solids Ex: 380 nm 2d λ = 2d(sin θi + sin θr) Detector at 760 nm sees 380 nm light! Source Sample Detector Real emission spectrum + Second Order

Filters

Filters Band Pass Filter

Fluorometer 3 1 2 4 2 Components 1) Light source 2) Monochrometer 3) Sample 4) Detector 5) Filters 6) Slits 7) Polarizers 4 2

Fluorometer: Slits Entrance Slit Mirrors Exit Slit

Fluorometer: Slits

Slit widths But…resolution decreases! Entrance Slit Wider Slits: More light hitting sample More emission More light hitting the detector More signal Greater signal-to-noise But…resolution decreases! Exit Slit Entrance Slit Source hn Sample

Slit widths Small Slit Large Slit bandpass (nm) = Entrance Slit Source hn Sample Small Slit Large Slit bandpass (nm) = slit width (mm) x dispersion (nm mm-1) for a 4.25 nm mm-1 grating

Excitation Slit widths Single Component: Wider slit: Larger bandwidth Intensity increase No emission spectra change Absorbance

Excitation Slit widths Multi Component : Wider slit: Larger bandwidth Intensity increase Emission ratio changes (1:2) -small slit less of dye 2 -large slits more of dye 2

Ex: For 8 nm bandwidth set emission acquisition to 4 nm per step. Emission Slit widths Wider slit: Larger bandwidth More light hitting the detector More signal Lower Resolution Exit Slit Sample Detector hn Grating doubled slits = intensity2 570 nm emission Small Slit (0.5 mm) Large Slit (2.0 mm) summing 569-571 nm (2.125 nm bandwidth) summing 566-574 nm (8.5 nm bandwidth) Nyquist Rule: scanning increment should be greater than 1/2 slit widths Ex: For 8 nm bandwidth set emission acquisition to 4 nm per step.

Always report your slit widths (in nm)! Emission Slit widths Emission Intensity Emission Intensity Always report your slit widths (in nm)!

Fluorometer 3 1 2 4 2 Components 1) Light source 2) Monochrometer 3) Sample 4) Detector 5) Filters 6) Slits 7) Polarizers 4 2

Fluorometer: Polarizer Mirrors Polarizer Polarizer

Fluorescence Anisotropy Absorption is polarized Fluorescence is also polarized

Absorption Probablity

Fluorescence Anisotropy Detector End View Unpolarized Light

Fluorescence Anisotropy Detector End View Unpolarized Light

Fluorescence Anisotropy Detector End View End View Unpolarized Light Unpolarized Light

Fluorescence Anisotropy Polarizer Detector End View Polarized Light

Fluorescence Anisotropy Polarizer Detector End View Polarized Light

Fluorescence Anisotropy Polarizer Detector End View End View I|| I^ Slightly Polarized Light Polarized Light

Fluorescence Anisotropy Sample I|| I^ Detector Polarized Excitation r = anisotropy factor I|| and I^ are the intensities of the observed parallel and perpendicular components

Fluorescence Anisotropy r = anisotropy factor I|| and I^ are the intensities of the observed parallel and perpendicular components

Monitor Binding

Reaction Kinetics

Other Sampling Accessories Cryostat Spatial Imaging Integrating Sphere Microplate Reader

Potential Complications With Sample Solvent Impurities -run a blank Raman Bands Concentration to high - A > 1 - Self-absorption Scatter (2nd order or spikes) With the Instrument Stray light Slit Widths Signal/Noise

Fluorescence Spectroscopy End Any Questions?