Ultrasound Modes A Mode presents reflected ultrasound energy on a single line display. The strength of the reflected energy at nay particular depth is.

Slides:



Advertisements
Similar presentations
Chapter3 Pulse-Echo Ultrasound Instrumentation
Advertisements

Foundations of Medical Ultrasonic Imaging
إعداد : أ. بلسم فهد صوفي 1 Ultrasound in Medicine Ch.3 Ultrasound pictures of the body.
Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc.
إعداد : أ. بلسم فهد صوفي،،،المصدر:محاضرات د.حنان 1 Ultrasound in Medicine Ch.4 Ultrasound pictures of the body.
Resident Physics Lectures
Krystal Kerney Kyle Fontaine Ryan O’Flaherty
Hospital Physics Group
Ultrasound Physics Have no fear Presentation by Alexis Palley MD
Pegasus Lectures, Inc. Volume II Companion Presentation Frank Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition.
“ULTRASONIC SYSTEM OF DISTANT DETERMINATION COORDINATES”
Basic Physics of Ultrasound
PRINCIPLES OF MEASUREMENT AND INSTRUMENTATION EKT 112
ECE 501 Introduction to BME
Ultrasound Imaging Atam Dhawan.
Ultrasonography: Where to start?
Ultrasonic Testing This technique is used for the detection of internal surface (particularly distant surface) defects in sound conducting.
Sound Waves and Ultrasound
Welcome.
Sound In Medicine. Sound waves Sound is a mechanical waves travel in air as longitudinal waves in which vibrations in air causes increases (compressions)
Ultrasound – Physics & Advances
Medical Imaging By Prof. M.M. Mohamed.
DOPPLER SONOGRAPHY By Dr/ Dina Metwaly.
THE ULTRASOUND IMAGE: GENERATION AND DISPLAY
Oscilloscope Tutorial
Ultrasound Physics Sound is a mechanical, longitudinal wave that travels in a straight line Sound requires a medium through which to travel Ultrasound.
Ultrasonography.
Digital Image Characteristic
3/13/2009IB Physics HL 21 Ultrasound Medical Imaging Physics – IB Objectives I.2.7Describe the principles of the generation and the detection of ultrasound.
GENERAL ULTRASONIC SYSTEM  TRANSMITTER CIRCUIT  ULTRASONIC TRANSDUCERS  RECEIVER CIRCUIT  MEASURING CIRCUIT  ANALYZER CIRCUIT  SIGNAL PROCESSOR 
Display of Motion & Doppler Ultrasound
George David Associate Professor Ultrasound Physics 04: Scanner ‘97.
Resident Categorical Course
1 March 2013 PGI Chandigarh Dr Sarvesh Natani.
EPOCH 1000 Series Procedure Linear Scan and C-scan.
Saudi Board of Radiology: Physics Refresher Course Kostas Chantziantoniou, MSc 2, DABR Head, Imaging Physics Section King Faisal Specialist Hospital &
Dr Mohamed el Safwany, MD..  The student should learn at the end of this lecture medical ultrasound imging.
ULTRASONIC TECHNOLOGY Introduction to Ultrasonic Properties of Ultrasonic waves Ultrasonic Production- Magnetostriction Method.
US FUNDAMENTALS: B-MODE AND DOPPLER MACHINE SETTINGS.
Medical Imaging X-rays CT scans MRIs Ultrasounds PET scans.
GENERAL ULTRASONIC SYSTEM
Display of Motion & Doppler Ultrasound
Review of Ultrasonic Imaging
Sarah Gillies Ultrasound Sarah Gillies
Introduction to the Physics Echocardiography Introduction to the Physics Echocardiography Jose L. Rivera, M.D. January 9, 2010.
Doppler Ultrasound Dr Mohamed El Safwany, MD.. Introduction The Doppler Effect refers to the change in frequency that results when either the detector/observer.
Saudi Board of Radiology: Physics Refresher Course Kostas Chantziantoniou, MSc 2, DABR Head, Imaging Physics Section King Faisal Specialist Hospital &
Ultrasound Physics Reflections & Attenuation ‘97.
IMAGE DATA ACQUISITION
BIOE 220/RAD 220 REVIEW SESSION 3 February 6, 2011.
Piezoelectric Effect  Sound waves striking a PZ material produce an electrical signal  Can be used to detect sound (and echoes)!
ULTRASOUND – THE BASICS
Artifacts Ultrasound Physics George David, M.S.
Performance Issues in Doppler Ultrasound 1. 2 Fundamental Tradeoffs In pulsed modes (PW and color), maximum velocity without aliasing is In pulsed modes,
Ultrasound Learning Objectives: Describe the properties of ultrasound;
 Ultrasound waves are longitudinal with high frequencies ( ≈ > 20,000 Hz, though medical Ultrasound is between 1 to 15 MHz.)  When an ultrasound reaches.
Ultrasound Physics Image Formation ‘97. Real-time Scanning Each pulse generates one line Except for multiple focal zones frame one frame consists of many.
Chapter 10 Digital Signal and Image Processing
Ultrasound Physics Image Formation ‘97.
Lecture on Ultrasonography
Sound in medicine Lect.10.
Reflections & Attenuation
Single-element transducers properties
Ultrasound Physics Image Formation ‘97.
Ultrasound.
LESSON 2 IMAGING METHODS OF ACOUSTIC FIELDS
MCQs (1) – Questions Ultrasound propagates through the medium as a transverse wave Ultrasound velocity is equal to the product of frequency and wavelength.
Physics Of Ultrasound Dr. Saidatul Ardeena,
Chapter 12 Image Artifacts
Presentation transcript:

Ultrasound Modes

A Mode presents reflected ultrasound energy on a single line display. The strength of the reflected energy at nay particular depth is visualized as the amplitude of the waveform. B Mode converts A Mode information into a two dimensional grayscale display. C Mode is a color representation of blood flow velocity and direction. D Mode is a spectral representation of blood flow velocity and direction. P Mode is used to visualize very low blood flows in color. Unlike C Mode, this mode does not show the operator flow direction. Triplex is the simultaneous operation of B Mode, C Mode and D Mode. M Mode is a scrolling display allowing the operator to view and record organ motion.

Axial Resolution Another concern is Resolution. Axial resolution is corresponds directly to the wave length characteristics of the Ultrasound wave. As frequency increases wave length shortens allowing for greater resolution. What we loose is penetration. Again as frequency increases penetration decreases. Higher frequencies also provide finer tissue grain or smoothness. A less grainy look.

Lateral resolution In simple ultrasound systems Lateral resolution is attributed to physical focus characteristics of the crystal element. The concaved shape of the element provides focus to the beam and the width of the beam at any particular point effects the ability of the ultrasound system to resolve small objects that are side by side.

Transverse resolution Transverse resolution is unique to the phase array probe. It is the ability of the probe to resolve objects side by side, as in lateral resolution, but in this case it is measured transverse to what would be considered the normal imaging plane. Again this is assuming simplest probe construction.

Contrast Resolution The ability of the system to resolve adjacent bright reflectors is called contrast resolution. This is in small part due to the cumulative effects of axial and lateral resolution. The systems scan converter plays a large role is this characteristic.

Diagnostic Ultrasonography Displaying Monitor Transducer / Probe Keyboard Probe Connector Printer (B/W & Color)

Device that can change one form of energy into another. The necessary element for generating acoustic waves. Electrical excitation into motion and pressure.

(A) (B) (C) -

Transducer Design

Echoes from Two Interfaces

Echoes from Internal Organ

A- mode B- mode M- mode Doppler Pulsed Continuos

Amp H. V CRT Transducer Organ Body T/R switch Trigger Pulser Time variable gain Saw tooth voltage sweep Block diagram of an A-scan instrument. A pulser circuit triggers the transducer, and the saw-tooth generator. The T/R switch isolates the receiver amplifier during transmission. Amplitudes of the received echo signals are presented on the vertical axis of the CRT.

Brightness modulation Horiz. Vert. CRT Amp T/R switch Pulser circuit Time variable gain Saw tooth voltage sweep Beam steering control unit A pulser circuit triggers the transducer, and the saw tooth generator. The T/R switch isolates the receiver amplifier during transmission. For each scanning line, the amplitudes of the received echo signals are modulated to brightness. Steering unit is controlling the synchronization process.

Transducer Fixed organ Horiz. Vert. CRT Amp Moving Organ Body T/R switch Trigge r Pulser circuit Time variable gain Sawtooth voltage sweep Slow voltage ramp A B B

TRANSMITTERRECEIVER PULSED CONTINUES

Linear Probe Image

Sector (Phased array) Probe

Convex Probe Image

Real Time 3D

Fetal Spine

Reconstructional 3-D Obstetrics

Ultrasound Machines

Function Diagnostic ultrasound machines are used to give images of structures within the body. This chapter does not deal with other kinds of machine (e.g. therapeutic and lithotripsy). The diagnostic machine probes, which produce the ultrasound, come in a variety of sizes and styles, each type being produced for a particular special use. Some require a large trolley for all the parts of the unit, while the smallest come in a small box with only a audio loudspeaker as output. They may be found in cardiology, maternity, outpatients and radiology departments and will often have a printer attached for recording images. Unlike X-rays, ultrasound poses no danger to the human body. How it works The ultrasound probe contains a crystal that sends out bursts of high frequency vibrations that pass through gel and on through the body. Soft tissue and bone reflect echoes back to the probe, while pockets of liquid pass the ultrasound straight through. The echoes are picked up and arranged into an image displayed on a screen. The machine offers a number of processing options for the signal and image and also allows the user to measure physical features displayed on the screen. This requires the machine to incorporate a computer.

WHAT TYPES OF EXAMINATION ARE TO BE CARRIED OUT ?

1. TRANSDUCER * Curvilinear or combination of linear and sector. 2. FREQUENCY * Standard transducer should have central frequency of 3.5 MHz. 3. ANGLE for Sector probe should be 40 or more, linear array should be cm long. 4. FRAM RATE… Hz for linear array, Hz for sector array. 5. FRAM FREEZE DENSITY… at least 512*512*4 bits to provide 16 gray levels

6. ELECTRONIC CALIPERS… one pair at least, with Quantitative readout. 7. ADD DATA IS POSSIBLE… patient identification, hospital name, date of examination… etc. 8. HARD COPY… should be possible. 9. MONITOR… at least 13 cm * 10 cm (preferably larger) 10. STABILIZING… should be able to stabilize voltage variation of +/- 10%. 11. Biometric tables… (it may not be universal and should be adjusted for local standards.

WHAT WE HAVE TO CHECK WHEN WE RECEIVE THE SCANER USER’S MANUAL SERVICE MANUAL

1. Voltage setting should be compatible with the electrical supply. 2. Interference on the screen/ whit sparks. 3. Transducer and cables test. 4. Check the cursor / measuring length, … 5. Check the accessibility of the biometrics or measurement tables.

Any missing or distorted image that does not match the real image of the part being examined

Acoustic characteristics of the tissues. Scanner’s settings. Lack of user’s experience. Defected part within the scanner.

COMMON ARTIFACTS: Cyst’s artifact (strong back-wall effect). Abdominal wall artifact. Gas artifact. Reverberation artifact. Incomplete imaging artifact. Gain artifact. Shadows artifact.

Visually inspects all transducers. “Cable, cracked surface, punctured, discolored casing” Visually inspect the power cords. Verify that the trackball and DGC controls appears clean and free from gel or other contaminants. Once the system is powered on: Verify that the monitor displays CORRECTLY the connected transducer’s identification, current date, time.

FOCUS. DEPTH GAIN COMPENSATION. OVERALL GAIN. ZOOM. MONITOR (B/C).

MAIN UNIT Surface of the system. DGC slides control. Trackball. Unit filters. TRANSDUCERS Linear, convex, ….. Endocavity, interoperative, …