고려대학교 구로병원 강 대 현 Goal of Ultrasound QA/QC Program To make sure a system is set up correctly and performs to specified standards. (

Slides:



Advertisements
Similar presentations
Chapter3 Pulse-Echo Ultrasound Instrumentation
Advertisements

Foundations of Medical Ultrasonic Imaging
Quality Control in Diagnostic Radiology
Quality assurance and Quality control in medical radiography
RAD 354 Chapt. 30 Digital QC Performance Assessment Standards Luminance Meter Digital Display QC QC by Technologist.
Quality Assurance and Digital Radiography
Seeram Chapter 13: Single Slice Spiral - Helical CT
ACR Ultrasound PM / Standards
1 CORTEX TECHNOLOGY DENMARK. 2  To obtain objective data  To visualize invisible changes and conditions  To learn more - faster  To help improve.
Resident Physics Lectures
James R. Halama PhD, Medical Physicist David A. Urani MBA, CNMT, Manager Rebecca A. Sajdak BA, CNMT, Sr. Technologist Robert H. Wagner MD A Program for.
Hospital Physics Group
Ultrasound Modes A Mode presents reflected ultrasound energy on a single line display. The strength of the reflected energy at nay particular depth is.
- Doppler Ultrasound & Obstetrical Ultrasound
Ultrasound Imaging Atam Dhawan.
Basic quality control tests in medical radiography
Sound In Medicine. Sound waves Sound is a mechanical waves travel in air as longitudinal waves in which vibrations in air causes increases (compressions)
Ultrasound – Physics & Advances
Quality Assurance/Quality Control Chapter 10. Quality Assurance Definition A system of activities whose purpose is to provide assurance that the overall.
DOPPLER SONOGRAPHY By Dr/ Dina Metwaly.
Unit IV Analyzing the Image. Unit IV Analyzing the Image.
THE ULTRASOUND IMAGE: GENERATION AND DISPLAY
Ultrasound Physics Sound is a mechanical, longitudinal wave that travels in a straight line Sound requires a medium through which to travel Ultrasound.
Ultrasonography.
Quality Control Rad T 110.
Digital Image Characteristic
CT Quality Control for CT Scanners. Quality Control in CT A good idea? Yes Required for accreditation? Sometimes Improves image quality? Sometimes Depends.
Pegasus Lectures, Inc. Volume II Companion Presentation Frank Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition.
George David Associate Professor Ultrasound Physics 04: Scanner ‘97.
Resident Categorical Course
ACR Full-Field Digital Mammography Accreditation Update
 Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy.
Dr Mohamed el Safwany, MD..  The student should learn at the end of this lecture medical ultrasound imging.
US FUNDAMENTALS: B-MODE AND DOPPLER MACHINE SETTINGS.
The Leeds Teaching Hospitals NHS Trust Is Grey the New Black? Image Display Optimisation & Consistency for Softcopy Reporting Mark Barnfield St James’s.
Display of Motion & Doppler Ultrasound
Review of Ultrasonic Imaging
Quality Assurance, Quality Control & Equipment Bushong Ch. 19 & 20
Julianna Ianni Meher Juttukonda David Morris Advisor: Dr. Jadrien Young, M.D.
Sarah Gillies Ultrasound Sarah Gillies
Doppler Ultrasound Dr Mohamed El Safwany, MD.. Introduction The Doppler Effect refers to the change in frequency that results when either the detector/observer.
Saudi Board of Radiology: Physics Refresher Course Kostas Chantziantoniou, MSc 2, DABR Head, Imaging Physics Section King Faisal Specialist Hospital &
Ultrasound Physics Reflections & Attenuation ‘97.
Preliminary Ultrasound Imager Quality Test “UNIFORMITY” J. Satrapa, TCC.
Autonomous Robots Vision © Manfred Huber 2014.
Part No...., Module No....Lesson No
Quality Assurance.
Piezoelectric Effect  Sound waves striking a PZ material produce an electrical signal  Can be used to detect sound (and echoes)!
Quality Control Bushong Chapter 10.
Artifacts Ultrasound Physics George David, M.S.
By Prof. Stelmark. Digital Imaging In digital imaging, the latent image is stored as digital data and must be processed by the computer for viewing on.
Ultrasound Learning Objectives: Describe the properties of ultrasound;
 Ultrasound waves are longitudinal with high frequencies ( ≈ > 20,000 Hz, though medical Ultrasound is between 1 to 15 MHz.)  When an ultrasound reaches.
Dept. of Radiology SHIN HEUNG College & University Ph. D. Sang-Chull Ma 장소 : 대전 컨벤션센터 일자 : 주최 : 한국표준과학연구원.
National Highway Institute 5-1 REV-2, JAN 2006 EQUIPMENT FACTORS AFFECTING INERTIAL PROFILER MEASUREMENTS BLOCK 5.
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
Ultrasound Physics Image Formation ‘97. Real-time Scanning Each pulse generates one line Except for multiple focal zones frame one frame consists of many.
Chapter 10 Digital Signal and Image Processing
Ultrasound Physics Image Formation ‘97.
Lecture on Ultrasonography
Reflections & Attenuation
Impact of Obesity on Medical Imaging and Image-Guided Intervention
Ultrasound Physics Image Formation ‘97.
Quality Assurance in Diagnostic Medical Sonography HHHoldorf
Presented by Harry C. Elinsky, Jr. Filtech, Inc.
MCQs (1) – Questions Ultrasound propagates through the medium as a transverse wave Ultrasound velocity is equal to the product of frequency and wavelength.
Factors Effecting the Production
Medical Imaging Imagining Modalities.
Review of Ultrasonic Imaging
Chapter 12 Image Artifacts
Presentation transcript:

고려대학교 구로병원 강 대 현

Goal of Ultrasound QA/QC Program To make sure a system is set up correctly and performs to specified standards. ( 장비의 올바른 설치와 기준에 맞는 작동을 보장 ) To maintain the consistency of the performance. ( 장비 성능의 일관성 유지 ) To reveal problems at its earliest stage before it severely interferes with the clinical practices. ( 임상업무를 심각하게 방해 하기 전에 초기에 문제점을 밝혀낼 수 있다 ) 장비의 정확하고 지속적인 성능 유지 초음파 영상 의 진단능력을 항상 최고로 유지시키기 위함

Good Quality Assurance will ensure proper equipment operation detect gradual changes in performance minimize equipment down time reduce the number of repeat scans optimize operator and patient safety

When should quality assurance be performed? Routine preventative maintenance ( 기본예방정비 ) Performed by Sonographer on daily or weekly basis Routine performance testing ( 기본성능시험 ) Performed by Sonographer on weekly or monthly basis Acceptance testing ( 승인검사 ) Performed when new equipment arrives and is installed.

Three Levels of Ultrasound QC Testing Level 1 – Frequent QC testingLevel 1 – Frequent QC testing : Operator’s QC Level 2 – Semi-annual or annual QC testingLevel 2 – Semi-annual or annual QC testing : Physicist’s QC Level 3 – Acceptance testingLevel 3 – Acceptance testing / baseline set- up

Guideline By The Institute of Physical Sciences in Medicine (IPSM, 미국의학물리과학협회 ) Report No. 70 “Testing of Doppler Ultrasound Equipment” Edited by PR Hoskins, SB Sherriff and JA Evans York: ISPM, 1994

Guideline By The Institute of Physical Science in Medicine (IPSM, 미국 의학물리과학협회 ) Report No. 71 “Routine Quality Assurance of Ultrasound Imaging Systems” Edited by R. Price York: ISPM, 1995

Guideline By American Institute of Ultrasound in Medicine (AIUM, 미국 의학초음파 학회 ) “Quality Assurance Manual for Gray Scale Ultrasound Scanners (Stage 2)” Edited by E. Madsen AIUM, Laurel, MD, 1995

American Association of Physicists in Medicine (AAPM, 미국 의학물리협회 ) Ultrasound Task Group No. 1 “ Real-time B-mode ultrasound quality control test procedures” By MM Goodsitt et al Medical Physics. 25(8): , 1998 Aug. Guideline By

ACR Accreditation ( 인가 ) Breast Ultrasound Accreditation Program (including Ultrasound-guided Breast Biopsy) General Ultrasound Accreditation Program Obstetrical Gynecological General Vascular Combination of the above

AIUM Accreditation Ultrasound practices in various fields: Breast Obstetrical and Trimester-Specific Obstetrical Gynecological Abdominal/General

ACR Accreditation QC Requirements A QA program should be in place; The QA program must be directed by a medical physicist or the supervising physician ; Routine QC testing must be done at least semi-annually ; Do the same tests, monitor changes over time and take corrective actions when necessary ; Testing results, corrective action, and the effects of corrective action must be documented and maintained on site. There is currently no ACR-preferred single phantom.

ACR Recommended Semi-Annual QC Tests for Breast Ultrasound Accreditation Maximum depth of visualization Distance accuracy (vertical and horizontal) Uniformity Electrical-mechanical cleanliness condition Anechoic void perception Ring down Lateral resolution Quality Control Checklist

ACR Recommended Technologist’s QC Tests for Breast Ultrasound Accreditation Adherence to universal infection control procedures for each biopsy All transducers should be cleaned between patients Distance calibration – quarterly Gray scale photography - quarterly

ACR Required semi-Annual QC Tests for General Ultrasound Accreditation System sensitivity and/or penetration capability Image uniformity Photography and other hard copy recording Low contrast object detectability (optional) Assurance of electrical and mechanical safety Vertical and horizontal distance accuracy (recommended only when the program is initiated for a scanner)

Daily or weekly check lists General Machine Cleanliness Keyboard and knobs clean Monitors clean Air Filters cleans Mechanical and Electrical Wheels fastened securely and rotate easily? Wheels locks work well? Accessories fixed securely? Cords attached securely? ALI Workstation Contrast and Brightness between scanner and workstation Transducer Inspection (for each probe) Cables Transducer Cracks Matching layer delaminations

Assurance of Electrical and Mechanical Safety The equipment related physical checks are also worth attention. These include: Checking for any cracks or delamination in the transducers Noting for loose and frayed electric cables, loose handles or control arms Checking the working condition of the wheels and the wheel locks Making sure all the accessories (such as VCR, printer, etc.) are fastened securely to the system Making sure the image monitors are clean and the air filter are not dusty

General Machine Cleanliness Philips General Machine Cleanliness / Philips System Power Switch Filter Main Power Switch ( 장비 뒷 편에 위치 ) Monitor Control Filter Power Switch IU 22 System Envisor System

HDI ® 5000 System System Power Switch Main Power Switch ( 장비 뒷 편에 위치 ) Filter Monitor Control Philips

ACCUVIX System Track ball Medison

ACCUVIX Filer Medison ACCUVIX System Filter

제품 뒤쪽 화살표 부위를 청소한다. Medison SA8000 System Fan Accuvix

Acuson Sequoia 512 System Filter : 장비의 후면에 위치 System Power switch Main Power switch Key board

Acuson Antares System System Power switch Main Power switch Filter : 장비의 후면에 위치

Acuson G50/ G60S Both Filter : 본체 전면 하단 다리측 2 개 System Power switch : 장비 좌측하단

Mechanical and Electrical check

Transducer Inspection check Transducer Cleaner

Annually or semi-annually check lists Performance test Dead zone test Vertical Measurement accuracy test Horizontal Measurement Calibration System sensitivity test Uniformity test Grayscale photography test Axial resolution test Lateral resolution test Functional resolution, Definition and Fill-in Doppler QC tests Doppler signal sensitivity Doppler angle accuracy Color display and Gray-scale image congruency Range-gate accuracy Flow readout accuracy

Ultrasound QC Phantoms B-mode: 1.Multi-purpose or general purpose tissue/cyst phantom 2.Low contrast resolution phantom 3.Spherical lesion phantom 4.Beam profile and slice thickness phantom 5.Prostate QC phantom Doppler-mode: 1.Flow phantom 2.String phantom

Multipurpose Phantom Model 539

Specifications of Multipurpose Phantom Model Tissue-mimicking material - Speed of sound: 1,540m/sec - Attenuation : 0.7 ㏈ / ㎝ / ㎒ - Freezing point: 0 ℃ (32 ℉ ) - Melting point: 78 ℃ (174 ℉ ) 2. Containers - Wall : 3/8"Acrylic - Scanning surface : 0.4mm (0.015")polyurethane(saran) 3. Targets - Cyst speed of sound : 1,540m/sec - Cyst attenuation : 0.05 ㏈ / ㎝ / ㎒ - Nylon lines: 0.3mm (0.012") diameter - Phantom size and weight: 15×18×5Cm, 2.5Kg

Multipurpose Phantom Model 539 ScanningTargetDiagram

측정항목 Dead zone test System sensitivity and/ or penetration capability test Image Uniformity test Grayscale photography and other hard copy recording test Axial and lateral resolution test Functional resolution, Definition and Fill-in Vertical and horizontal distance accuracy test (recommended only when the program is initiated for a scanner)

측정 시 주의사항 Scanning : vertical to phantom surface

Dead Zone (Ring down) A group of 9 reflectors consisting of fibers are placed at different separations from the top of the phantom(~2-10mm). As the transducer scan across the top, the distance from the transducer to the first reflector completely imaged is equal to the dead zone (ring down) distance.

Dead Zone (Ring down) Reduced near field resolution by Transducer-skin interface

Dead Zone (Ring down) Dead Zone (Ring down) test Dead zone test 2mm ~ 10mm, 9 targets

Distance Accuracy test Scan the phantom with vertical column and a horizontal row of reflectors; The digital caliper readout on screen is checked against the known distance between reflectors (10cm); Criteria Vertical : 1,5% of the actual distance or 2mm, whichever is greater. Horizontal : 3% of the actual distance or 3mm, whichever is greater.

Vertical Distance accuracy test

Horizontal Distance accuracy test

System Sensitivity/Penetration This test should be done with following setting: Maximum transmit power Proper receiver gain and TGC that allows echo texture to be visible in the deep region Transmit focus at the deepest depth The maximum depth of visualization is determined by comparing the gradually weakening echo texture to electronic noises near the bottom of the image. Do this test with the same setting and monitor the changes over time.

System Sensitivity/Penetration 3.5 MHz 12 MHz 7 MHz

Axial and Lateral Resolution Spatial resolution may be evaluated by either of the following ways: Count the number of the pins distinguished without overlap in lateral and axial direction; thus the minimum target spacing is documented Measure the size of the pin in lateral and axial direction

Axial resolution test Axial Resolution Array 3.0 mm 2.0 mm 1.0 mm 0.5 mm 0.25 mm 1.0 mm 0.25mm

Lateral resolution test Lateral Resolution Array 3.0 mm 2.0 mm 1.0 mm 0.5 mm 0.25 mm 3.0 mm 3 mm

Image Uniformity Adjust the TGC controls and other sensitivity controls to obtain an image as uniform as possible Inspect the image to detect any kinds of Vertical or radially oriented streaks( 수직 또 는 방사형태로 출현하는 줄무늬 ) Dropouts Reduction of brightness near edges of the scan( 주사면 끝부분의 밝기 저하 ) Brightness transitions between focal zones ( 초점 영역 사이의 밝기 변화 )

Image Uniformity

Functional resolution, Definition and Fill-in QA Phantom 에서 anechoic structure 의 spherical targets 를 명확하게 detection 할 수 있는 능력 2, 3, 4, 6, 8mm diameter spherical targets

Functional resolution, Definition and Fill-in 1 1/2D (Matrix) Transducer Conventional array Transducer 초음파 영상은 단면 data 가 아니라 volume data 이기 때문임. 초음파 영상은 단면 data 가 아니라 volume data 이기 때문임.

Soft and/or Hard Copy Recording Ⅰ The shades of gray, weak, and strong echo texture should be optimized and consistent between the image display on the ultrasound scanner and the photographic hard copies or soft copy displays on the workstation in the reading room. For quick follow-up testing, the Gray-scale bar pattern on the clinical image display can be used

Soft and/or Hard Copy Recording Ⅱ Use the SMPTE test pattern and other patterns if they are available on the ultrasound scanner. Workstation monitor display should be included in QC tests.

Soft and/or Hard Copy Recording Ⅲ Film processor QC needs to be done daily. Darkroom fog test needs to be done at least semi-annually. Sensitometry

Low Contrast Object Detectability Scan of a low contrast resolution phantom can reveal the low contrast object detectability which is an optional test on the ACR semi-annual QC test list for general ultrasound accreditation.

Ultrasound Doppler QC Testing Doppler QC tests include Doppler signal sensitivity Doppler angle accuracy Color display and Gray-scale image congruency Range-gate accuracy Flow readout accuracy

Doppler String Phantom

Doppler Velocity Accuracy: Variations among 6 clinical units, each with transducers of 3.5MHz, 5MHz and 7 MHz. Flow Rate (ml/min) Frequency AVG 3.5MHz 5.7%10.2%6.7%8.0%10.2%13.0%9.0% 5 MHz 17.4%7.7%10.6%6.9%13.9%13.0%11.6% 7 MHz 8.2% 4.1%9.2%7.1%12.5%8.2% Average 10.4%8.7%7.1%8.0%10.4%12.9%9.6%

Discussion The guidelines for ultrasound QA/QC programs are descriptive. Maintaining an ultrasound QA/QC program is straightforward and testing consistency. However, we need to know that the tissue mimicking material can only mimic some of the properties of tissue in real clinical situations. Assessment of the efficacy and relevance of ultrasound QA procedures is rare. The guidelines need to be updated periodically as ultrasound technology develops.